ID |
Date |
Author |
Subject |
|
366
|
Wed Mar 2 19:25:25 2022 |
JM, PM | Current status | 20:25
Observed short burst of low energy noise on aida1, lasted <30s (attachment 8).
Temperature setpoint = 113 K
Target density OK S1 vaies between 2E-8 and 3E-8 mbar, E4 between 1E-9 and 1E-8 mbar.
Detector moving well between 23.1 mm and 67.43mm, move in and move out counters the same. Correlation scalar on aida1.
Rutherford peak varies between 10^3 and 10^2 depending on detector position, noise peak varies with detector position -> lower further in.
Beam intensity ~1E8
Merger OK
Data rate ~ 1900 Kb/sec
Leakage current = 3.208 uA
Temperatures OK
System checks - wr fails
Base Current Difference
aida01 fault 0xab60 : 0xaba4 : 68
aida02 fault 0x88d3 : 0x8915 : 66
aida03 fault 0x23aa : 0x23ee : 68
aida04 fault 0x15d7 : 0x161b : 68
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring pressure = 5.5 E-10 mbar. Ring pressure tends to oosciilate between ~4 and 5.5 E-10 mbar
21:21
Burst of noise not observed again
Temperature setpoint = 113 K
Target density OK S1 vaies between 2E-8 and 3E-8 mbar, E4 between 1E-9 and 1E-8 mbar.
Detector moving well between 23.1 mm and 67.43mm, move in and move out counters the same. Correlation scalar on aida1.
Rutherford peak varies between 10^3 and 10^2 depending on detector position, noise peak varies with detector position -> lower further in.
Beam intensity ~1E8
Merger OK
Data rate ~ 1900 Kb/sec
Leakage current = 3.206 uA
Temperatures OK
System checks - wr fails
Base Current Difference
aida01 fault 0xab60 : 0xaba4 : 68
aida02 fault 0x88d3 : 0x8915 : 66
aida03 fault 0x23aa : 0x23ee : 68
aida04 fault 0x15d7 : 0x161b : 68
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring pressure 5.4E-10 mbar |
|
365
|
Wed Mar 2 17:21:35 2022 |
JM, PM | Current status | 18:21
Temperature setpoint = 113k
Target density OK S1 varies between 2E-8 and 2.5E-8 mbar, E4 between 1E-9 and 5E-9 mbar. (slightly lower than previous check)
Detectors moving well between 23.1mm and 67.43 mm, Move in and move out counters the same. Correlation scalar is on aida1
Rutherford peak varies between 10^3 and 10^2 depending on detector position. No halo peak in rate. Noise peak varies from ~100 peak height when fully in to ~400 peak height when fully out.
Beam intensity ~ 1E8
Merger OK
Data rate ~ 1800 kb/sec
Leakage current ~3.21 uA
Temperature OK
System checks OK - wr fails
Base Current Difference
aida01 fault 0xab60 : 0xaba3 : 67
aida02 fault 0x88d3 : 0x8914 : 65
aida03 fault 0x23aa : 0x23ed : 67
aida04 fault 0x15d7 : 0x161a : 67
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring pressure = 4.4 E-10 mbar
19:20
Temperature setpoint = 113k
Target density OK S1 varies between 2E-8 and 3E-8 mbar, E4 between 1E-9 and 1E-8 mbar. (slightly higher than previous check)
Detectors moving well between 23.1mm and 67.43 mm, Move in and move out counters the same. Correlation scalar is on aida1
Rutherford peak varies between 10^3 and 10^2 depending on detector position. No halo peak in rate.
Beam intensity ~ 1E8
Merger OK
Data rate ~ 1800 kb/sec
Leakage current ~3.20 uA
Temperature OK
System checks OK - wr fails
Base Current Difference
aida01 fault 0xab60 : 0xaba4 : 68
aida02 fault 0x88d3 : 0x8915 : 66
aida03 fault 0x23aa : 0x23ee : 68
aida04 fault 0x15d7 : 0x161b : 68
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring pressure = 4.3 E-10 mbar
Statistics similar to previous check. |
|
364
|
Wed Mar 2 14:15:16 2022 |
JM, PM | Current status | 15:15
Temperature setpoint = 113 K
Target density is OK for now. S1 vaies between 2E-8 and 3.5E-8 mbar. E4 between 1E-9 and 1E-8 mbar.
Detector motors functioning well. Moving between 23.1 mm and 67.43mm. Move in and move out counters the same. Correlation scalar 8 for detectors moving is seen on aida1
Rutherford rate peak varies between 10^3 peak height when fully in and ~10^2 when fully out. No peak from halo seen in rate.
Beam intensity ~ 1E8
Merger OK
Data rate ~ 1700 kb/sec
Run is Run 87 - started 10:34
Data spy restarted at 15:00 so only include Run87 data.
Leakage current = 3.205 uA
Fee Temperatures OK
System checks OK - wr fails
Base Current Difference
aida01 fault 0xab60 : 0xaba3 : 67
aida02 fault 0x88d3 : 0x8914 : 65
aida03 fault 0x23aa : 0x23ed : 67
aida04 fault 0x15d7 : 0x161a : 67
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring Pressure = 5.5E-10 mbar
16:45
Temperature setpoint = 113 K
Target density is OK for now. S1 vaies between 2E-8 and 3E-8 mbar. E4 between 1E-9 and 1E-8 mbar.
Detector motors functioning well. Moving between 23.1 mm and 67.43mm. Move in and move out counters the same. Correlation scalar 8 for detectors moving is seen on aida1
Rutherford rate peak varies between 10^3 peak height when fully in and ~10^2 when fully out. No peak from halo seen in rate.
Beam intensity ~ 1E8
Merger OK
Data rate ~ 1300 kb/sec
Temperatures OK
Leakage current ~3.205 uA
System chacks OK - wr fail
Base Current Difference
aida01 fault 0xab60 : 0xaba3 : 67
aida02 fault 0x88d3 : 0x8914 : 65
aida03 fault 0x23aa : 0x23ed : 67
aida04 fault 0x15d7 : 0x161a : 67
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring pressure ~ 4.4 E-10 mbar
Statistics similar to previous check.
|
|
363
|
Wed Mar 2 09:38:50 2022 |
Spillmann, Uwe | ion source restarted |
detector position
out: 23.1mm
in: 67.4mm
tape: next file
gas jet S1: 2.0-2.2 ** -8
|
|
362
|
Wed Mar 2 06:00:49 2022 |
Spillmann, Uwe | waiting | 07:00
Target density is hovering between 2.0e-8 and 2.3e-8 |
|
361
|
Wed Mar 2 03:40:40 2022 |
David, George | Temperature Cycle | 04:40
Target density is hovering between 2.0e-8 and 2.3e-8, sometimes dipping below 2.0e-8
We will conduct a temperature walkthrough while there is no beam,
04:40: setpoint 125K
04:46 setpoint 95K
04:56 setpoint 108K
04:59 setpoint 110K
05:00 setpoint 112K (here improved from 110K)
05:03 setpoint 114K
05:08 setpoint 113K
05:11
113K seems good, fluctuating between 2.3e-8 and 3.1e-8
will leave temperature here |
|
360
|
Wed Mar 2 01:24:48 2022 |
David, George | Ion Source Update | 02:25
Zoran was unable to ignite the ion source.
He has given some parameters to the control room so they can wiggle stuff around and try and ignite it.
He also said it may at some point ignite itself.
In worse case scenario, the person who usually manages the ion source will come in from 7am-8am, read Zoran's notes and try and get it working again.
Also target density seems a little worse but still hanging above 2e-8 to 2e-4 |
|
359
|
Tue Mar 1 23:43:15 2022 |
David, George | Beam Intensity Drop | 00:38
Intensity in the ring dropped by 2 orders of magnitude to 5e5 on injection
Lost rutherford rate (as seen in last e-log post)
All pressures seem fine
00:48
Called main control room, said they will take a look
00:59
Servo motor control restarted and homed
01:11
Ion source is confirmed to be off
Zoran will try and fix it, should take 1hr or so and not guaranteed to work.
|
|
358
|
Tue Mar 1 23:30:04 2022 |
David, George | Current status | 00:30
Pressure in S1 slightly decreased from before but still high enough to measure data.
One new count at 20MeV
Target density at S1: 2e-8 - 2.7e-8
Pressures OK (one small spike just before 00:00 in YR08)
System Checks OK
White rabbit fail diff: 49
Data write speed: ~1.4Mb/s
Beam intensity at injection: ~7.7e7
Cryo Temp Setpoint: 110K
(Beam intensity in ring dropped as I was writing which is why there is no rutherford peak on exRate image)
|
|
357
|
Tue Mar 1 22:10:26 2022 |
David, George | Current status | Ring current 6.67e7 particles/s
pressure at S1: 2.3-2.8e-8
Merger, Tape server (WS3): 1574 kB/s, writing ok
Leakage current (WS4): 3.20 uA
DAQ (WS2): temperature checked, clock status: ok, adc calibration: ok, white rabbit check: failed
Online monitoring (WS 5): done |
|
356
|
Tue Mar 1 21:34:06 2022 |
PM, LV | check | no change:
Check Clock Status': Passed 5, Failed 0
Check ADC Calibration: Passed 5, Failed 0
'Check the White rabbit decoder status: Passed 0, Failed 5
target density at S1: 2.2-2.7e-8
temp: 110K
beam at injection: 7e7 |
|
355
|
Tue Mar 1 20:41:28 2022 |
PM, LV | check | no change:
Check Clock Status': Passed 5, Failed 0
Check ADC Calibration: Passed 5, Failed 0
'Check the White rabbit decoder status: Passed 0, Failed 5
target density at S1: 2.3-2.7e-8
temp: 110K
beam at injection: 7e7 |
|
354
|
Tue Mar 1 19:41:24 2022 |
PM, LV | target temperature walkthrough | A target temperature walkthrough has been initiated at 20:41 since the target density at S1 had slightly decreased (with ~0.1-0.2)
Current temprature 116K
set to 126K at 20:41
set to 95K at 20:47:30
set to 108.5K at 21:01:30, at this temperature was a sweetspot observerd for target density.
set to 109.7 at 21:08:30
set to 110K at 21:13:30 - we keep at this value for a while. the values at S1 are changing between 2.2-2.9e-8 |
|
353
|
Tue Mar 1 19:40:28 2022 |
PM, LV | check | no change:
Check Clock Status': Passed 5, Failed 0
Check ADC Calibration: Passed 5, Failed 0
'Check the White rabbit decoder status: Passed 0, Failed 5
target density at S1: 2.1-2.3e-8
temp: 116K
beam at injection: 7e7 |
|
352
|
Tue Mar 1 18:43:41 2022 |
PM, LV | check | no change:
Check Clock Status': Passed 5, Failed 0
Check ADC Calibration: Passed 5, Failed 0
'Check the White rabbit decoder status: Passed 0, Failed 5
target density at S1: 2.1-2.4e-8
temp: 116K |
|
351
|
Tue Mar 1 17:36:04 2022 |
PM, LV | check | no change:
Check Clock Status': Passed 5, Failed 0
Check ADC Calibration: Passed 5, Failed 0
'Check the White rabbit decoder status: Passed 0, Failed 5
target density at S1: 2.1-2.4e-8
temp: 116K |
|
350
|
Tue Mar 1 16:57:34 2022 |
CB, LV, PMH | New run | Beam issues due to too long storage time in the last hours. Motors moved out and not moving.
Decreased storage time to 70 s. Seems OK now. Restarted motor movement as before
Started new run R86 |
|
349
|
Tue Mar 1 14:01:01 2022 |
MS, SHB | check | 'Check Clock Status', 'Check ADC Calibration' and 'Check the White rabbit decoder status' : no change |
|
348
|
Tue Mar 1 12:58:32 2022 |
CB, MS, SS, ML | Synced pneumatic motion | New run R84
Attempting to sync pneumatic movement with ring. Still see something on t out -> p out
Command to move out was being given relative to the wrong time.
Fixed it. Still see one-two events every time we move out around 4-6 MeV. Attempted to greatly increase time before next injection - no difference. Microphonics from moving the detectors? It's just one-two counts on entire detector, acceptable.
New run R85
Now moving in/out
Potentiometer: 23.10 mm to 67.19 mm |
|
347
|
Tue Mar 1 12:33:21 2022 |
MS, SHB | check | 'Check Clock Status', 'Check ADC Calibration' and 'Check the White rabbit decoder status': no change |
|