| ID |
Date |
Author |
Subject |
Text |
 |
|
44
|
Wed Aug 18 11:58:49 2021 |
NTD RSS | Misc fittings and fasteners plus gasket inventory | Attachments 1-3 fasteners, fittings found
amongst earlier shipments. I assume that
the STFC DL cleaned items are for the internal |
22x |
|
373
|
Wed Aug 10 10:33:03 2022 |
TD, ML, JM, NP | Meeting at YR09 to discuss options for CARME operations to February 2022 |
Initial discussions of proposals to provide
open CARME chamber and provide access to |
   |
|
403
|
Wed May 24 09:54:56 2023 |
JM, CB | Maximum detector positions | The detectors were moved into the beam
axis using the motors to check the maximum
distance they can be moved in without collision |
9x |
|
459
|
Sun Jan 7 15:38:57 2024 |
CB, JM | Manifolds mostly assembled + bias overnight check | Assembled most of the connectors on the
new (green) manifolds. Two blanking pulgs
stuck when mounting - cannot remove them |
|
|
461
|
Mon Jan 8 19:29:57 2024 |
CB, JM | Manifold repaired + fee assembly in bench | Davide extracted the stuck blind plugs
and re-threaded all missing ports. Completed
manifold assembly, except for main inlet/outlet |
   |
|
463
|
Wed Jan 10 19:27:08 2024 |
CB, JM | Manifold and more bias tests | Mounted water manifold on CARME with help
from GSI plumber. Connected all the hoses
with Swagelok self-locking connectors. Tested |
|
|
129
|
Fri Oct 22 13:11:26 2021 |
TD | MSL type TTT14 assembly drawings & pin assignments | |
 |
|
264
|
Thu Feb 24 04:17:35 2022 |
CB, TD (remotely) | MIDAS DAQ no longer working | DAQ server crashed around 24:00
Rebooted server but could not could
not rescue remotely - could not contact raspberry |
 |
|
729
|
Tue Jun 10 09:44:20 2025 |
JG | MBS not running for a week | The MBS daq was not writing data since June
2. 22:39 (run_0035.lmd)
The reason is that
the rfio server was running on atppc025 which |
|
|
730
|
Tue Jun 10 09:44:30 2025 |
JG | MBS not running for a week | The MBS daq was not writing data since June
2. 22:39 (run_0035.lmd)
The reason is that
the rfio server was running on atppc025 which |
|
|
731
|
Tue Jun 10 09:44:33 2025 |
JG | MBS not running for a week | The MBS daq was not writing data since
June 2. 22:39 (run_0035.lmd)
This means, no detector and target |
|
|
517
|
Thu Feb 15 13:18:30 2024 |
JG | MBS DAQ: target & ring values | We are running an MBS DAQ to record the
attached list of slow control values of the
ring and the target. In most cases this is |
|
|
682
|
Mon May 12 13:25:09 2025 |
JG | MBS DAQ - overview | An MBS DAQ is running to collect two types
of data:
|
|
|
385
|
Wed Jan 18 13:32:36 2023 |
PJCS | MACB settings with either Emulator or VETAR | When using the VITAR connected
to the HDMI port of the root MACB the settings
for all the MACB in the system 0x3. |
   |
|
56
|
Fri Sep 3 14:14:40 2021 |
TD | Leybold IE514 manual & calibration data | |
 |
|
392
|
Sun Apr 30 10:01:39 2023 |
CB, JM | Left hand detectors unmounted | (29 April)
Inspected CARME after opening.
See attached. Some key takeaways. |
26x |
|
393
|
Mon May 1 08:26:09 2023 |
CB, JM | Left hand detectors remounted | (31 April)
Remounted 3335-12 on top and new
3335-1 on bottom of left hand plate. Used |
|
|
466
|
Sat Jan 13 14:32:21 2024 |
CB, JM | Leaky manifold and cable management | Found two leaks in the welding of the manifolds,
marked them in tape. Manifold to be repaired
on Monday. Remember to re-install missing |
  |
|
673
|
Tue Apr 15 12:23:10 2025 |
JM | Leakage current / CARME issues | 14/04/25
Arrived at GSI in the afternoon.
Found that the coolant loop on |
18x |
|
415
|
Thu Jul 6 15:26:39 2023 |
JM | Leak update | The leak on the actuator arm bellows has
been attempted to be sealed using UHV vacuum
seal spray. The spray had the opposite to |
|