Entry  Thu Feb 24 08:30:36 2022, JG, MS, AIDA5 calibration test failed - solved Screenshot_from_2022-02-24_09-09-36.pngScreenshot_from_2022-02-24_09-29-24.png
UPDATE: THE PROBLEM IS SOLVED FOR NOW:

Doing the calib for the 5th time eventually worked.

------------
Entry  Thu Feb 24 09:43:16 2022, JG, MS, check 13x
Beam is still beeing tuned. Target is off.

 

- Motor on manual
Entry  Mon Jan 15 18:51:28 2024, JG, DR, TD, Monday 15 January 

Cooling water manifolds
Entry  Tue Jan 16 08:57:44 2024, JG, DR, TD, Tuesday 16 January 8x
10.00 Outflow manifold returned from workshop with leaking attachment point welding re-worked.

11.30 Outflow hoses re-attached to outflow manifold. Faulty male Swagelok replaced.
Entry  Tue May 27 15:18:46 2025, JG, CN, JM, run 24 7x
Run 24 running

27/05/2025

We start a new run with real data taking condititon: beam is ok, target density is 5e11 cm-2, detectors are moving in-out periodically.
Entry  Mon Feb 28 13:56:02 2022, JG, CB, Temperature cycle to 125K 
14:45

Rutherford rate was decreased significantly for longer period.

Temperature increased to 125 K
Entry  Wed Oct 18 16:47:36 2023, JG, CB, Power outage 
Local power outage at the CRYRING today, while JG was trying to activate the ion pumps. Unlcear what caused it.

Valve on scroll worked as intended, but CARME reacehd 1E-7 mbar from turbo (?) venting.

Power restored. Pressure back in the 1E-11 mbar range, decreasing.
Entry  Mon Feb 28 06:12:58 2022, JG, AK, US, Current status 8x
07:20

Temperature setpoint = 110 k

Rutherford rates shaky, peak max at ~ 3x10^2
Entry  Mon Feb 28 08:36:35 2022, JG, AK, CB, Current status 8x
09:35

beam intensity at injection: 8.2e7 ions

Temperature setpoint = 110 k
Entry  Mon Feb 28 09:31:48 2022, JG, AK, CB, Current status 8x
10:35

beam intensity at injection: 1e8 ions

Temperature setpoint = 110 k
Entry  Mon Feb 28 11:57:22 2022, JG, AK, CB, Current status 8x
12:55

beam intensity at injection: 9e7 ions

Temperature setpoint = 110 k
Entry  Mon Feb 28 13:16:39 2022, JG, AK, CB, Current status 8x
14:16

beam intensity at injection: 7e7 ions

Temperature setpoint = 110 k
Entry  Mon Jun 16 11:12:22 2025, JG, AK, AR, EH, run 83 8x
12:15 restarted measurement with target - Run 83

same settings as for runs 76, 77 etc

ROOT histograms NOT zeroed
Entry  Mon Feb 28 07:31:25 2022, JG, AK, Current status 8x
08:30

Temperature setpoint = 110 k

Rutherford rates shaky, peak max at ~ 3x10^2
Entry  Mon Jun 2 20:03:43 2025, JG, AG, CB, EW, EM, Run 45/46, 0.5 MeV/A 13x
Run 45 started.

D+ beam deuterium target. D+ beam energy is 0.5 MeV/u. (beam calc calculation is 0.5143 MeV/u). See attachment for beamCalc

Target density is about 2e12 at/cm2
Entry  Mon Jun 2 17:25:31 2025, JG, AG, Run 43 scraper tests 
Same settings are Run 42 but bunched beam and scrapers set to:  30 / - 30 mm.

 
Entry  Mon Jun 2 17:48:07 2025, JG, AG, Run 44 scraper tests Screenshot_from_2025-06-02_18-51-24.pngScreenshot_from_2025-06-02_19-27-31.png
Same settings are Run 43 but with coasting beam and scrapers set to:  30 / - 30 mm.

 

all checks ok, but WR timestamp error, see attached
Entry  Mon Feb 28 07:35:51 2022, JG, current logger crashed Screenshot_from_2022-02-28_08-38-35.png
The current logger did crash in the evening of 25.02. and stopped logging since then, it seems.

I restarted it now (28.02. 8:30).

Seems to run stable for now.
Entry  Thu Feb 15 13:18:30 2024, JG, MBS DAQ: target & ring values 2024_jet_daq.txt
We are running an MBS DAQ to record the attached list of slow control values of the ring and the target. In most cases this is achieved by converting
analog monitor voltages to frequencies and recording them with a VME scaler (VULOM4B with scaler-firmware). The complete list of signals is attached including
a channel mapping.
Entry  Mon May 12 13:25:09 2025, JG, MBS DAQ - overview 
An MBS DAQ is running to collect two types of data:

1. WR timestamps for machine events (injection, target ON/OFF, ...)