ID |
Date |
Author |
Subject |
61
|
Tue Sep 7 12:40:43 2021 |
CB ML + RH | CARME survey preparation |
Rudi and survey group representative came
to CARME.
Elephant feet confirmed to be out of
position with respect to the models. Still
unclear if CARME is where it should be
with respect to the jet target and
electron target. I.e. do they have space
to mount? Do we have to move CARME again
(hopefully not!!)
Valve to downstream confirmed out of
alignment. Solution unclear but may be
needed to ensure beam goes through
detectors correctly.
Survey group will come to confirm CARME
position on Thursday 16 September. We
cannot mount baking tent until then, but
no major issue given delays due to strain
relieves and shutdown on 12-15 September. |
60
|
Tue Sep 7 11:37:05 2021 |
CB RSS + OG | Update on strain relief screw issues |
Tested every strain relief. M3 screws get stuck at different points in each hole. Only one good hole out of 16
(2x8).
They were tested in Edinburgh - unclear if this is a result of baking at high temperatures, or bad screws, or
something else.
Easiest solution may be to re-machine holes and clean the connectors here in GSI.
Now investigating options. |
59
|
Mon Sep 6 14:13:17 2021 |
CB ML | Motors mounted - first attempt |
Mounted both motors on CARME. Started
with outside motor.
M6x160 screws proved slightly too long to
fix bellows to rod and seal the vacuum.
Had to be manually shortened by 1 cm
roughly.
Mounted whole assembly but could not fix
actuator rod to moving plate due to holes
mis-alignment.
Mounted motor on CARME but could not align
holes on flange and chamber without
turning motor to the side... see attach 3 and 4
Got in contact with Peter over lunch and
understood issue.
Mounted second motor with experience
acquired. No issues. All screws go in.
Motor box level. See attach 5
Will need to dismount / remount first
motor box before pumping down. Currently
secure on CARME thanks to adjustable
support.
Note we forgot to buy DN 40 CF gaskets. ML
supplied a few. Will buy more. Needed
between bellows and rod.
Strain relief mentioned in previous ELOG
entry was taken to workshop and hole re-
drilled. Workshop reports other hole is OK
for M3 screw. Either screw used or hole
were poorly machined. M3 screws should be
correct. Re-machined part sent for
cleaning.
Will try with clean parts again tomorrow
unless we remount motors.
Activities for the day cut shorter due to
SPARC meeting this week. |
Attachment 1: 20210906_113522.jpg
|
|
Attachment 2: 20210906_113519.jpg
|
|
Attachment 3: 20210906_115601.jpg
|
|
Attachment 4: 20210906_152607.jpg
|
|
Attachment 5: 20210906_143944.jpg
|
|
58
|
Sat Sep 4 15:54:50 2021 |
CB RSS | Motor mounting preparation |
Prepared the motors to be mounted and
found all components required. No issues.
May have to partly dismount again to fix
bellows depending on ease of mounting.
Note extremely long screws mentioned in
previous Elog inventory are to be used for
the motors.
Attempted to swap Pirani gauge with ours
but realised it's CF and needs appropriate
port without risk of baking. That port
would be on the interaction chamber which
is not going to be mounted. Will probably
keep using ABD gauge for now.
Peter spotted mistake in mounting front
foot of CARME which is back to front.
Correct mount will gain us 2-3 cm but will
conflict with current CARME position fixed
by tube on the back. Will discuss options
with ML.
Motor mounting to take place on Mon with
local technical support. No activities
tomorrow unless a solution for the clamps
mentioned in previous entry is found (i.e.
what size bolts go on top). Peter will
check step file and we will consult with
local engineers on Mon pm. |
Attachment 1: 20210904_162517.jpg
|
|
Attachment 2: 20210904_162520.jpg
|
|
57
|
Fri Sep 3 15:55:00 2021 |
CB RSS | Issues with cable clamps |
Attempted to mount cable clamps for strain
relief on inner part of Allectra
feedthroughs.
SS support mounted with M3 vented screws
OK
Cannot figure out which bolt type to use
to fix Macor on SS support.
M2 falls through
M3 screws in partly and gets stuck...
Will need to figure out how to get unstuck
No M2.5 bolts available.
In contact with Peter. Not clear what the
solution was
Will try again tomorrow or move to motors
first. |
Attachment 1: 20210903_154614.jpg
|
|
Attachment 2: 20210903_162519.jpg
|
|
56
|
Fri Sep 3 14:14:40 2021 |
TD | Leybold IE514 manual & calibration data |
|
Attachment 1: 20210903_141201.jpg
|
|
Attachment 2: Ionivac_Sensors_IE414_IE514.pdf
|
|
55
|
Fri Sep 3 08:14:51 2021 |
CB RSS | Issues foreseeable splitting CARME |
In order to install the detector supports
and eventually install the detectors
themselves CARME must be split and
separated via rails.
However it does not look like we have
sufficient clearance on either side. See
pictures attached.
I suggest we mount the strain relievers
wires and motors, not split CARME, and
pump down + bake to give us time to figure
out what is different from the models and
how to fix it. Outgassing from detector
supports should be negligible wrt wires.
One possibility would be to install rail
extension on the back side of the Al frame
where the cross was meant to be supported,
temporarily remove the tube connecting
CARME to the YR09 valve and slide CARME
back. Will need to discuss with Peter. |
Attachment 1: 20210903_085801.jpg
|
|
Attachment 2: 20210903_085805.jpg
|
|
Attachment 3: 20210903_085811.jpg
|
|
Attachment 4: 20210903_085822.jpg
|
|
Attachment 5: 20210903_085958.jpg
|
|
Attachment 6: 20210903_090003.jpg
|
|
Attachment 7: 20210903_090024.jpg
|
|
54
|
Fri Sep 3 08:12:39 2021 |
CB RSS | Pressure after overnight pump down |
03.09.2021 09:00 5.27E-7 mbar
03.09.2021 13:00 4.08E-7 mbar
09;00 Will keep pumping down this morning. |
Attachment 1: 20210903_085646.jpg
|
|
Attachment 2: IMG_4487.JPG
|
|
53
|
Thu Sep 2 21:04:26 2021 |
CB RSS ML + DR OG | Gate valves mounted + pump down |
Mounted both gate valves and support. Gate
valves could not be installed one bolt
rotated from 90 degrees as per drawings
due to mechanical conflict with back foot.
Using rotating flanges some small rotation
from 90 was achieved. This may be
sufficient given the reason for the
rotation was to avoid mechanical conflicts
with ring parts that are no longer there.
Valve struts had to be modified as gate
valve is much lower and close to Al frame
than in Daresbury.
ML suggested that the reason for the
change in height mentioned in previous
entry is that the ring gate valve to YR10
is not centered. Solutions will be
considered when the other two setups in
YR09 are mounted and the alignment
situation is more clear
Noted a few blind flanges are bluish,
which indicates SS 304 instead of 316. Did
we get the wrong ones from ITL? Did we
swap them in Daresbury by mistake? Is it
just an odd colour?
Installed turbo pump on top gate valve,
Pirani gauge provided by ABD (where is
ours?) and pumped down.
Reached 10-1 mbar in 10 minutes. Turbo on.
Reached 10-5 mbar on IE514 in 17 more
minutes (where is its manual?)
Will leave pumping overnight. |
Attachment 1: 20210902_152536.jpg
|
|
Attachment 2: 20210902_153213.jpg
|
|
Attachment 3: 20210902_170002.jpg
|
|
Attachment 4: 20210902_192104.jpg
|
|
Attachment 5: 20210902_192751.jpg
|
|
Attachment 6: 20210902_193944.jpg
|
|
52
|
Thu Sep 2 10:41:16 2021 |
CB RSS + DR OG | CARME connected to the CRYRING |
Connected CARME to the CRYRING via tube
plus bellows. See attach.
Had to lift CARME slightly wrt yesterday
beam alignment to get a straight tube. |
Attachment 1: 20210902_114031.jpg
|
|
51
|
Wed Sep 1 17:14:16 2021 |
CB RSS + DR ML | CARME mounted on the CRYRING |
Lifted CARME with Al support on the
CRYRING. No issues during transportation.
Drilled new holes in the Al feet to fit
elephant feet. Unclear if the issue is
with the model, the frame or the feet
themselves. In any case this issue is
sorted.
CARME was designed to interface with the
CRYRING via the cross section with pumps
and sensors that we decided dismount. To
replace this section a length of xhv
tubing and a bellow will be employed.
Prepared today, to be mounted tomorrow.
Aligned CARME height with beam height.
This posed issues due to difficulty to
adjust bolts under the cross members used
to support the motors. See previous log
entry.
Bolts under those cross members are now
only finger tight. Unclear whether the
CARME section will drop slightly if
disconnected from main body. It's the
section housing the moving Dssds. May be
an issue for tightening it back. Will
investigate.
Gate valves and supports to be installed
tomorrow morning. |
Attachment 1: 20210901_162435.jpg
|
|
Attachment 2: 20210901_173059.jpg
|
|
Attachment 3: 20210901_174157.jpg
|
|
Attachment 4: 20210901_134635.jpg
|
|
Attachment 5: 20210901_173059.jpg
|
|
50
|
Wed Sep 1 12:05:28 2021 |
CB RSS | Preparation for CARME on CRYRING YR09 |
Mounted additional cross member for motor
support. See attach 1
Washer required to make it fit on foot on
one side.
Holes in Al support are not matching elephant feet...! Will have to re-
drill them now.
Yr09 fully dismounted downstream |
Attachment 1: 20210901_125504.jpg
|
|
Attachment 2: 20210901_125517.jpg
|
|
49
|
Tue Aug 31 18:56:35 2021 |
CB | CARME status |
(ELOG does not allow more pictures to be added to previous entry) |
Attachment 1: 20210831_171533.jpg
|
|
Attachment 2: 20210831_171458.jpg
|
|
Attachment 3: 20210831_171501.jpg
|
|
Attachment 4: 20210831_171506.jpg
|
|
Attachment 5: 20210831_171510.jpg
|
|
Attachment 6: 20210831_171514.jpg
|
|
Attachment 7: 20210831_171517.jpg
|
|
Attachment 8: 20210831_171520.jpg
|
|
Attachment 9: 20210831_171524.jpg
|
|
48
|
Tue Aug 31 18:34:41 2021 |
CB RSS + DR OG | CARME on Al stand |
Lifted CARME off the floor and mounted feet on sides. It was found impossible to put four bolts for each foot
due to alignment conflict. Three bolts were mounted per foot instead.
One of the mounting holes in CARME has been damaged and the thread has been ruined. Simple rethreading attempts
failed. Three good holes left - not an issue for now.
One of the shims between the feet and the chamber had to be modified in the workshop. Some feet were rather
difficult to mount, but in the end we succedeed.
CARME was then lifted on the Al frame and gently dropped. The crane remains in place overnight to prevent it
falling over given it is not anchored to the floor.
Front foot and back foot were also mounted.
Elbows at the back were mounted.
Currently considering whether to mount gate valves before or after lifting in place on YR09. Lift experts will
decide tomorrow what is the easiest procedure given constraints on the ring.
See other attachment and next entry for current status.
YR09 downstream will be cleared tomorrow morning. Lift will proceed tomorrow afternoon. |
Attachment 1: 20210831_144545.jpg
|
|
Attachment 2: 20210831_171442.jpg
|
|
Attachment 3: 20210831_171447.jpg
|
|
Attachment 4: 20210831_171449.jpg
|
|
Attachment 5: 20210831_171453.jpg
|
|
47
|
Tue Aug 31 11:55:11 2021 |
CB RSS | Last preparations for CARME lift |
Found strut bolts in the bag with
spherical washers. Mounted the struts and
the washers. See attach 1.
Vented CARME in preparation for lift.
Turbo off
Pressure starts increasing without
introducing leaks. Up to 10-4 mbar. IE 514
off.
Removed plastic cap from leak valve.
Pressure rate increases. Leaky valve?
Turbo motor speed reaches 0 Hz in 1h
Opened leak valve. Scroll off. Pressure to
10 mbar. Not increasing anymore. Turbo
leak aperture likely to small to vent
CARME volume.
Disconnected scroll. Carefully let air in
by tweaking flexible hose connection. Not
recommended again, but should be safe
given turbo fully stop and chamber nearly
vented.
Reached atmosphere in 10 minutes.
Now waiting for lifting support to mount
feet. |
Attachment 1: 16304077162843806956479048661374.jpg
|
|
Attachment 2: 16304077904321905824276870110392.jpg
|
|
46
|
Tue Aug 31 07:09:54 2021 |
CB RSS (rem) JM | Preparation for mounting |
(30 August 2021)
The current state of YR09 is as per attach 1. CCC has been removed, but additional pumps and gauges remain. Some
will have to be mounted on CARME. Dismounting is scheduled for this Wed.
CB and RSS have been preparing for CARME to be mounted. JM supplied pictures of mounting process in DL. Attach 2-5.
2: Studs on carriages
3: CARME feet
4: Back foot
5: Elbows support
All components required have been located in the boxes except
- Cap hex bolts (M4 x 12?) to mount studs on carriages
- Plastic rails to remove cariages (a solution with bubble-wrap was found. Extra carriages are now in a box. Do
not turn it around.)
- Baking tray does not fit with carriages. Unclear how it did before. Will try to file down edges today.
- Bolts for CARME feet are of rather poor quality, and a host of different sizes. We may have to buy some, but
unclear what length.
- Studs are of different heights. That was not clear from previous pictures (attach 6)
- New and improved valve/elbow support (attach 7) is aligned worse than old one. Will keep using old one. Should
still work somehow. |
Attachment 1: 20210830_110920.jpg
|
|
Attachment 2: Al_frame.jpg
|
|
Attachment 3: 20200814_142437.jpg
|
|
Attachment 4: 20200902_153219.jpg
|
|
Attachment 5: 20210830_160613.jpg
|
|
Attachment 6: 20210830_115903.jpg
|
|
45
|
Wed Aug 25 07:02:41 2021 |
RSS TD | 25-31 August |
Pirani IE514
mbar mbar
25/8/21 08:00 0.0030 1.86e-8
----------------------------------------------------
26/8/21 09:00 0.0030 1.77e-8
----------------------------------------------------
27/8/21 08:00 0.0030 1.76e-8
----------------------------------------------------
28/8/21 17:30 0.0029 1.63e-8
----------------------------------------------------
30/8/21 08:00 0.0028 1.43e-8
----------------------------------------------------
31/8/21 08:30 0.0027 1.40e-8 |
Attachment 1: IMG_4349.JPG
|
|
Attachment 2: IMG_4360.JPG
|
|
Attachment 3: log_scale_26.08.png
|
|
Attachment 4: turbo_pump_26.08.png
|
|
Attachment 5: scroll_pump_26.08.png
|
|
Attachment 6: IMG_4367.JPG
|
|
Attachment 7: log_scale_27.08.png
|
|
Attachment 8: IMG_4372.JPG
|
|
Attachment 9: log_scale_28.08.png
|
|
Attachment 10: IMG_4426.JPG
|
|
Attachment 11: IMG_4427.JPG
|
|
Attachment 12: log_scale_30.08.png
|
|
Attachment 13: turbo_pump_30.08.png
|
|
Attachment 14: scroll_pump_30.08.png
|
|
Attachment 15: log_scale_31.08.png
|
|
Attachment 16: turbo_pump_31.08.png
|
|
Attachment 17: scroll_pump_31.08.png
|
|
Attachment 18: IMG_4445.JPG
|
|
Attachment 19: IMG_4446.JPG
|
|
44
|
Wed Aug 18 11:58:49 2021 |
NTD RSS | Misc fittings and fasteners plus gasket inventory |
Attachments 1-3 fasteners, fittings found amongst earlier shipments. I assume that the STFC DL cleaned items are for the internal detector mount assembly.
Inventory of silver-plated copper gaskets
ID (cm)
7x DN250CF 25.4
5x DN200CF 20.0
5x DN160CF 15.3
100x DN100CF 10.0
4x DN63CF 6.3
16x DN35CF 3.5
23x DN16CF 1.6
N.B. ID is measured +/- 0.1cm and DNxxCF designation is what the Lesker catalogue says it means.
Attachments 4 & 9 2x motor rods and MACOR heat shields for cleaning
Attachment 5 - cable assemblies
Attachment 7 - what are these screws for?
|
Attachment 1: 20210818_113056.jpg
|
|
Attachment 2: 20210818_111549.jpg
|
|
Attachment 3: 20210818_111543.jpg
|
|
Attachment 4: 20210819_162634.jpg
|
|
Attachment 5: 20210819_163357.jpg
|
|
Attachment 6: 20210819_163402.jpg
|
|
Attachment 7: 20210819_163646.jpg
|
|
Attachment 8: 20210819_163651.jpg
|
|
Attachment 9: 20210819_164007.jpg
|
|
Attachment 10: 20210819_164018.jpg
|
|
Attachment 11: 20210819_164028.jpg
|
|
Attachment 12: 20210819_164043.jpg
|
|
Attachment 13: 20210819_164043.jpg
|
|
Attachment 14: 20210819_164054.jpg
|
|
Attachment 15: 20210819_164111.jpg
|
|
Attachment 16: 20210819_164234.jpg
|
|
Attachment 17: 20210819_164238.jpg
|
|
Attachment 18: 20210819_164242.jpg
|
|
Attachment 19: 20210819_164417.jpg
|
|
Attachment 20: 20210819_164421.jpg
|
|
Attachment 21: 20210819_164527.jpg
|
|
Attachment 22: 20210819_164530.jpg
|
|
43
|
Wed Aug 18 07:07:26 2021 |
RSS TD | 18-24 August |
Pirani IE514
mbar mbar
18/8/21 08:00 0.0034 3.90e-8
18/8/21 17:00 0.0032 3.72e-8
-------------------------------------------------------------
19/8/21 08:00 0.0032 3.38e-8
19/8/21 17:00 0.0032 3.31e-8
-------------------------------------------------------------
20/8/21 08:00 0.0032 3.02e-8
-------------------------------------------------------------
21/8/21 16:00 0.0032 2.59e-8
-------------------------------------------------------------
22/8/21 12:00 0.0034 2.44e-8
-------------------------------------------------------------
23/8/21 08:00 0.0030 2.23e-8
-------------------------------------------------------------
24/8/21 08:00 0.0030 2.15e-8
|
Attachment 1: IMG_4273.JPG
|
|
Attachment 2: IMG_4274.JPG
|
|
Attachment 3: log_scale_18.08.png
|
|
Attachment 4: turbo_pump_18.08.png
|
|
Attachment 5: scroll_pump_18.08.png
|
|
Attachment 6: IMG_4281.JPG
|
|
Attachment 7: log_scale_19.08.png
|
|
Attachment 8: turbo_pump_19.08.png
|
|
Attachment 9: scroll_pump_19.08.png
|
|
Attachment 10: IMG_4290.JPG
|
|
Attachment 11: IMG_4301.JPG
|
|
Attachment 12: log_scale_21.08.png
|
|
Attachment 13: turbo_pump_21.08.png
|
|
Attachment 14: scroll_pump_21.08.png
|
|
Attachment 15: IMG_4323.JPG
|
|
Attachment 16: IMG_4324.JPG
|
|
Attachment 17: log_scale_22.08.png
|
|
Attachment 18: turbo_pump_22.08.png
|
|
Attachment 19: scroll_pump_22.08.png
|
|
Attachment 20: IMG_4339.JPG
|
|
Attachment 21: IMG_4340.JPG
|
|
Attachment 22: IMG_4345.JPG
|
|
Attachment 23: log_scale_24.08.png
|
|
Attachment 24: turbo_pump_24.08.png
|
|
Attachment 25: scroll_pump_24.08.png
|
|
42
|
Fri Aug 13 07:28:37 2021 |
RSS TD | Friday 13-17 August |
Pirani IE514
mbar mbar
13/8/21 08:00 0.0075 3.53e-7
13/8/21 12:00 0.0065 2.80e-7
13/8/21 15:20 0.0065 2.40e-7
13/8/21 16:30 0.0060 2.29e-7
---------------------------------
14/8/21 12:00 0.0046 1.25e-7
14/8/21 12:00 0.0044 1.14e-7
---------------------------------
15/8/21 09:00 0.0040 8.47e-8
15/8/21 12:30 0.0040 8.00e-8
15/8/21 15:30 0.0040 7.69e-8
---------------------------------
16/8/21 08:00 0.0038 6.36e-8
16/8/21 11:30 0.0036 6.14e-8
16/8/21 17:00 0.0036 5.82e-8
---------------------------------
17/8/21 08:00 0.0034 4.82e-8
17/8/21 17:00 0.0034 4.47e-8 |
Attachment 1: IMG_4196.JPG
|
|
Attachment 2: IMG_4197.JPG
|
|
Attachment 3: IMG_4198.JPG
|
|
Attachment 4: scroll_pump_13.08.png
|
|
Attachment 5: turbo_pump_13.08.png
|
|
Attachment 6: IMG_4237.JPG
|
|
Attachment 7: IMG_4238.JPG
|
|
Attachment 8: turbo_pump_14.08.png
|
|
Attachment 9: scroll_pump_14.08.png
|
|
Attachment 10: IMG_4244.JPG
|
|
Attachment 11: IMG_4245.JPG
|
|
Attachment 12: turbo_pump_15.08.png
|
|
Attachment 13: scroll_pump_15.08.png
|
|
Attachment 14: IMG_4257.JPG
|
|
Attachment 15: IMG_4258.JPG
|
|
Attachment 16: turbo_pump_16.08.png
|
|
Attachment 17: scroll_pump_16.08.png
|
|
Attachment 18: IMG_4265.JPG
|
|
Attachment 19: IMG_4266.JPG
|
|
Attachment 20: turbo_pump_log_scale.png
|
|
Attachment 21: scroll_pump_17.08.png
|
|
Attachment 22: turbo_pump_17.08.png
|
|
Attachment 23: log_scale_17.08.png
|
|