AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  CARME, Page 31 of 34  ELOG logo
ID Date Author Subjectdown
  270   Thu Feb 24 14:31:25 2022 JM, ML, CNBeam Tuning Run 76

15:30

1.5 MeV/u beam currently being tuned

Target is ON, setpoint = 135 K

System checks OK - white rabbit fail diff ~ 9

Temperatures OK

Merger OK

Data rate ~ 1400 kb/sec

Leakage current ~ 3.41 uA

Aida 1 and 2 missing channels from previous days appear working on stat layout 2 and xy pixel total plot (attachment 3)

17:12

Beam tuning complete. See attached Ex rate histogram.

The cryo compressor is being turned OFF for around 1 hour to remove any frozen nitrogen which may be the cause of the increased pressures observed.

Beam OFF to do so.

System checks OK - white rabbit fail diff ~ 9

Temperatures OK

Merger OK

Data rate ~ 200 kb/sec

Compressor OFF, setpoint on controller is 273 K

18:51

Pressure in the ring recovered quickly after closing valves to gas jet (attachment 5)

Increasing temperature caused a large spike in pressure in the inlets. E1 turbo briefly shut down due to the pressure increase, since restarted and pressure back in the -4 mbar range. (attachment 6)

Temperature being decreased

Beam back ON

data rate ~ 1500 kb/sec

merger OK

Temperatures OK

System checks OK - white rabbit diff ~ 12

21:58

After cooling the jet target, expected density is not achieved. E4 pressure is ~4E-10 mbar compared to ~2E-8mbar previously. Temperature reduced to 127K E4 pressure increases however stilll within -10 mbar range.

With beam on target we see rutherford peak at ~3 MeV, however it is much reduced.

Beam OFF to investigate inside the cave.

Decreasing temperature of nozzle in 0.5 k steps to see effect on E4 and S1 dump.

 

Attachment 1: 2022-02-24_15-35-53-Ex-rate-histogram.png
2022-02-24_15-35-53-Ex-rate-histogram.png
Attachment 2: 2022-02-24_15-39-08-leakage-current.png
2022-02-24_15-39-08-leakage-current.png
Attachment 3: 2022-02-24_15-44-06-aida1-5-stat.png
2022-02-24_15-44-06-aida1-5-stat.png
Attachment 4: 2022-02-24_17-07-52-xyTotal.png
2022-02-24_17-07-52-xyTotal.png
Attachment 5: vacplot24.png
vacplot24.png
Attachment 6: gasjet-vacplot.png
gasjet-vacplot.png
  226   Tue Feb 22 04:09:28 2022 JM, CB, LV, PM, AKBeam ON, Target ON Run 54 - first event at 10MeV !

05:00

Beam ON ~2E6 pps, still fluctates between 2e6 and 6e6 (like a step function)

Target ON and open to the ring

YR09xxx ETarget ~1.8e-10

Target E4 inlet pressure 8.3E-9 mbar

Dump S3 ~ 6.8E-10 mbar

Nozzle temperature 150 K

Run 54 started, monitoring code restarted.

Attachment 1: Screenshot_from_2022-02-22_05-07-47.png
Screenshot_from_2022-02-22_05-07-47.png
Attachment 2: Screenshot_from_2022-02-22_05-07-33.png
Screenshot_from_2022-02-22_05-07-33.png
Attachment 3: Screenshot_from_2022-02-22_05-07-19.png
Screenshot_from_2022-02-22_05-07-19.png
Attachment 4: Screenshot_from_2022-02-22_05-06-58.png
Screenshot_from_2022-02-22_05-06-58.png
Attachment 5: Screenshot_from_2022-02-22_05-06-30.png
Screenshot_from_2022-02-22_05-06-30.png
Attachment 6: Screenshot_from_2022-02-22_05-06-19.png
Screenshot_from_2022-02-22_05-06-19.png
Attachment 7: Screenshot_from_2022-02-22_05-06-06.png
Screenshot_from_2022-02-22_05-06-06.png
Attachment 8: Screenshot_from_2022-02-22_05-05-40.png
Screenshot_from_2022-02-22_05-05-40.png
Attachment 9: Screenshot_from_2022-02-22_05-05-30.png
Screenshot_from_2022-02-22_05-05-30.png
Attachment 10: Screenshot_from_2022-02-22_05-05-11.png
Screenshot_from_2022-02-22_05-05-11.png
Attachment 11: Screenshot_from_2022-02-22_05-04-25.png
Screenshot_from_2022-02-22_05-04-25.png
  228   Tue Feb 22 05:47:18 2022 JM, CB, LV, PM, AKBeam ON, Target ON Run 54 - 2 events at 10MeV!

06:47

Beam ON ~2.0E6 pps, still fluctates between 1e6 and 6e6 (like a step function)

Target ON and open to the ring

YR09xxx ETarget ~1.8e-10 (saturated)

Target E4 inlet pressure 8.0E-9 mbar

Dump S3 ~ 6.6E-10 mbar

Nozzle temperature 150 K

Run 54 started, monitoring code restarted.

Attachment 1: Screenshot_from_2022-02-22_06-43-36.png
Screenshot_from_2022-02-22_06-43-36.png
Attachment 2: Screenshot_from_2022-02-22_06-43-20.png
Screenshot_from_2022-02-22_06-43-20.png
Attachment 3: Screenshot_from_2022-02-22_06-42-48.png
Screenshot_from_2022-02-22_06-42-48.png
Attachment 4: Screenshot_from_2022-02-22_06-42-27.png
Screenshot_from_2022-02-22_06-42-27.png
Attachment 5: Screenshot_from_2022-02-22_06-42-00.png
Screenshot_from_2022-02-22_06-42-00.png
Attachment 6: Screenshot_from_2022-02-22_06-46-02.png
Screenshot_from_2022-02-22_06-46-02.png
Attachment 7: Screenshot_from_2022-02-22_06-45-48.png
Screenshot_from_2022-02-22_06-45-48.png
Attachment 8: Screenshot_from_2022-02-22_06-45-30.png
Screenshot_from_2022-02-22_06-45-30.png
Attachment 9: Screenshot_from_2022-02-22_06-44-28.png
Screenshot_from_2022-02-22_06-44-28.png
Attachment 10: Screenshot_from_2022-02-22_06-43-54.png
Screenshot_from_2022-02-22_06-43-54.png
Attachment 11: Screenshot_from_2022-02-22_06-46-32.png
Screenshot_from_2022-02-22_06-46-32.png
  223   Tue Feb 22 00:32:35 2022 JM, CB, LV, PM, AKBeam ON, Target ON Run 54

01:30

Beam ON ~6E6 pps

Target ON and open to the ring

Target E4 inlet pressure 8.12E-9 mbar

Dump S3 ~ 7E-10 mbar

Nozzle temperature 150 K

1:40

Run 54 started, monitoring code restarted.

Attachment 1: 2022-02-22_01-38-53-Ex-histogram.png
2022-02-22_01-38-53-Ex-histogram.png
Attachment 2: 2022-02-22_01-38-17-XYtotal.png
2022-02-22_01-38-17-XYtotal.png
  224   Tue Feb 22 02:02:11 2022 JM, CB, LV, PM, AKBeam ON, Target ON Run 54

03:00

Beam ON ~7E6 pps

Target ON and open to the ring

Target E4 inlet pressure 8.5E-9 mbar

Dump S3 ~ 7.3E-10 mbar

Nozzle temperature 150 K

Run 54 started, monitoring code restarted.

Attachment 1: TapeServer2022-02-22_02-49-29.png
TapeServer2022-02-22_02-49-29.png
Attachment 2: Leakage_current2022-02-22_02-50-39.png
Leakage_current2022-02-22_02-50-39.png
Attachment 3: Temperature2022-02-22_02-52-02.png
Temperature2022-02-22_02-52-02.png
Attachment 4: Screenshot_from_2022-02-22_02-54-34.png
Screenshot_from_2022-02-22_02-54-34.png
Attachment 5: Good_events2022-02-22_02-55-14.png
Good_events2022-02-22_02-55-14.png
Attachment 6: AIDA_CorrelationScreenshot_from_2022-02-22_02-56-04.png
AIDA_CorrelationScreenshot_from_2022-02-22_02-56-04.png
Attachment 7: Spectrum_Browser2022-02-22_02-59-28.png
Spectrum_Browser2022-02-22_02-59-28.png
Attachment 8: LowEx2022-02-22_03-00-02.png
LowEx2022-02-22_03-00-02.png
Attachment 9: LowEy2022-02-22_03-00-28.png
LowEy2022-02-22_03-00-28.png
Attachment 10: LowEXYTotal2022-02-22_03-01-16.png
LowEXYTotal2022-02-22_03-01-16.png
  225   Tue Feb 22 03:05:06 2022 JM, CB, LV, PM, AKBeam ON, Target ON Run 54

04:00

Beam ON ~5.5E6 pps, but it fluctates between 2e6 and 6e6 (like a step function) in about each 5-10min

Target ON and open to the ring

YR09xxx ETarget ~1.8e-10

Target E4 inlet pressure 8.4E-9 mbar

Dump S3 ~ 7.0E-10 mbar

Nozzle temperature 150 K

Run 54 started, monitoring code restarted.

Attachment 1: Screenshot_from_2022-02-22_04-02-25.png
Screenshot_from_2022-02-22_04-02-25.png
Attachment 2: Screenshot_from_2022-02-22_04-03-08.png
Screenshot_from_2022-02-22_04-03-08.png
Attachment 3: Screenshot_from_2022-02-22_04-03-44.png
Screenshot_from_2022-02-22_04-03-44.png
Attachment 4: Screenshot_from_2022-02-22_04-03-24.png
Screenshot_from_2022-02-22_04-03-24.png
Attachment 5: Screenshot_from_2022-02-22_04-04-28.png
Screenshot_from_2022-02-22_04-04-28.png
Attachment 6: Screenshot_from_2022-02-22_04-04-10.png
Screenshot_from_2022-02-22_04-04-10.png
Attachment 7: Screenshot_from_2022-02-22_04-03-55.png
Screenshot_from_2022-02-22_04-03-55.png
Attachment 8: Screenshot_from_2022-02-22_04-10-57.png
Screenshot_from_2022-02-22_04-10-57.png
Attachment 9: Screenshot_from_2022-02-22_04-10-37.png
Screenshot_from_2022-02-22_04-10-37.png
  227   Tue Feb 22 05:12:38 2022 JM, CB, LV, PM, AKBeam ON, Target ON Run 54

06:00

Beam ON ~1.5E6 pps, still fluctates between 1e6 and 6e6 (like a step function)

Target ON and open to the ring

YR09xxx ETarget ~1.8e-10

Target E4 inlet pressure 8.2E-9 mbar

Dump S3 ~ 6.5E-10 mbar

Nozzle temperature 150 K

Run 54 started, monitoring code restarted.

Attachment 1: Screenshot_from_2022-02-22_06-11-18.png
Screenshot_from_2022-02-22_06-11-18.png
Attachment 2: Screenshot_from_2022-02-22_06-10-59.png
Screenshot_from_2022-02-22_06-10-59.png
Attachment 3: Screenshot_from_2022-02-22_06-10-22.png
Screenshot_from_2022-02-22_06-10-22.png
Attachment 4: Screenshot_from_2022-02-22_06-10-09.png
Screenshot_from_2022-02-22_06-10-09.png
Attachment 5: Screenshot_from_2022-02-22_06-09-54.png
Screenshot_from_2022-02-22_06-09-54.png
Attachment 6: Screenshot_from_2022-02-22_06-09-31.png
Screenshot_from_2022-02-22_06-09-31.png
Attachment 7: Screenshot_from_2022-02-22_06-09-12.png
Screenshot_from_2022-02-22_06-09-12.png
Attachment 8: Screenshot_from_2022-02-22_06-08-58.png
Screenshot_from_2022-02-22_06-08-58.png
Attachment 9: Screenshot_from_2022-02-22_06-08-22.png
Screenshot_from_2022-02-22_06-08-22.png
Attachment 10: Screenshot_from_2022-02-22_06-07-57.png
Screenshot_from_2022-02-22_06-07-57.png
Attachment 11: Screenshot_from_2022-02-22_06-07-46.png
Screenshot_from_2022-02-22_06-07-46.png
  219   Mon Feb 21 17:26:19 2022 JM, CB, SHB, OH (remotely)Beam ON run 53

21/2/22 16:00

Call from main control room, beam is ON, probably has been for several hours judging by rate on xy total plot.

Changes made to online monitoring code by OH, include setting to only 1 DSSD, correct offsets from the pulser walkthrough, offline analysis capabilities and X/Y position correction. 

Monitoring code restarted in CARME_SORT/build_no_output

17:31 Run 50 ended, run 51 started now beam is definitely in the ring. Tried to insert span zero schottky into aida2 correlation scalar, however rate was far too high.

18:03 Removed span zero schottky signal, started new run 52.

With beam ON we noticed some missing channels in the xy total plot. Aida 1 and 2 stat totals on spectrum browser show missing channels. Asic check on both to no effect. Fees power cycled.

19:57 Started run 53. Currently waiting for beam

21:20 Beam ON

1*L plots checked for aida 1, several channels showed no data (noise)

Slow comparator LEC threshold lowered to 0x32  to aida 1, checked 1*L plots and channels are now showing lots of noise. So those channels appear to be quiet using 0x64 threshold. (attachment 3)

Threshold increased back to 0x64

22:33

System checks OK -white rabbit fails on all. ~30 seperation

Temperatures OK

Statistics within range - new scalars added for aida02 and aida05

Data being written ~300 kb/sec

Leakage current ~3.6 uA

 

00:39

System checks checks OK -white rabbit fails on all. ~42 seperation

Temperatures OK

Statistics within range with new scalars

Data being written 256 kb/sec

Leakage current ~3.615 uA

XY counts dominated by hot channels - most blue pixels are ~10 counts

 

Attachment 1: 2022-02-21_18-50-06-xyTotal.png
2022-02-21_18-50-06-xyTotal.png
Attachment 2: 2022-02-21_19-05-27-aida1-4-rate.png
2022-02-21_19-05-27-aida1-4-rate.png
Attachment 3: 2022-02-21_22-17-54-aida1-1*L-lowerThreshold.png
2022-02-21_22-17-54-aida1-1*L-lowerThreshold.png
Attachment 4: 2022-02-21_22-35-19-goodEvents.png
2022-02-21_22-35-19-goodEvents.png
Attachment 5: 2022-02-21_22-35-53-ADC-dataitems.png
2022-02-21_22-35-53-ADC-dataitems.png
Attachment 6: 2022-02-21_22-36-28-aida-correlation.png
2022-02-21_22-36-28-aida-correlation.png
Attachment 7: 2022-02-21_22-58-18-Ex-histogram.png
2022-02-21_22-58-18-Ex-histogram.png
Attachment 8: 2022-02-21_22-59-22-XY-total.png
2022-02-21_22-59-22-XY-total.png
Attachment 9: 2022-02-22_00-42-14-XY-total.png
2022-02-22_00-42-14-XY-total.png
  213   Sat Feb 19 20:08:43 2022 JM, CBBeam ON run 49

Run started 19/2/22 21:05. Beam ON

Temperatures OK

System wide checks OK - white rabbit fails on all.

Statistics are within normal range (10-20K goodevents)

Leakage current 3.6 uA

potentiometer = 20.28 mm

 

20/2/22

Leakage current ~3.675 uA. Looks stable now motors can not move (attachment 1)

potentiometer = 20.28 mm

Temperatures OK

System wide checks OK - white rabbit fails all

Statistic within normal range (see attachment 2 - ADC data items) Counter for aida 1 and aida 2 much higher due to noise bursts (elog 212)

 

XY total plot (attachment 3)

Low energy Ex histogram (attachment 4)

Low energy Ey histogram (attachment 5)

Tape Server data written (attachment 6)

 

Run end 20/2/22 15:08

Attachment 1: 2022-02-20_14-16-16-leakage-current.png
2022-02-20_14-16-16-leakage-current.png
Attachment 2: 2022-02-20_14-18-07-statistic-ADCdata.png
2022-02-20_14-18-07-statistic-ADCdata.png
Attachment 3: 2022-02-20_14-20-43-xyTotal.png
2022-02-20_14-20-43-xyTotal.png
Attachment 4: 2022-02-20_14-24-19-LowEnergy-ExTotal.png
2022-02-20_14-24-19-LowEnergy-ExTotal.png
Attachment 5: 2022-02-20_14-28-02-LowEnergy-EyTotal.png
2022-02-20_14-28-02-LowEnergy-EyTotal.png
Attachment 6: 2022-02-20_14-25-45-Tape-server-data-written.png
2022-02-20_14-25-45-Tape-server-data-written.png
  359   Tue Mar 1 23:43:15 2022 David, GeorgeBeam Intensity Drop

00:38
Intensity in the ring dropped by 2 orders of magnitude to 5e5 on injection

Lost rutherford rate (as seen in last e-log post)
All pressures seem fine

00:48
Called main control room, said they will take a look

00:59
Servo motor control restarted and homed

01:11
Ion source is confirmed to be off
Zoran will try and fix it, should take 1hr or so and not guaranteed to work.

 

  610   Mon Jun 10 07:40:47 2024 RSS, DBBeam

08:26: There is no beam and Gleb is informed.

08:46: Beam is back.

  167   Mon Jan 10 20:52:53 2022 JM, TD, OHBaking tent removed, VI tests

Baking tent, frame and trays fully removed from CARME chamber. Removal of baking frame required removing air lines from upper gate valve and repositioning of vacuum hose connecting turbo pump to backing line and turbo power cable. Some difficulty encountered removing upper baking frame now the target support frame is assembled. Small modification seperating section in two would aid future installation and disammebly. 

Baking trays removed with difficulty. Main baking tray bent to extract from the frame -> updated design required for future bakeout cycle. 

VI tests performed on the upper and lower detectors. Upper detector shows characteristic curve (attached plot), lower detector shows no response to increased voltage. Multimeter used on adapter boards with resistance of ~18.5 kΩ observed between bias pins for the upper detector. Open line seen between bias pins on the lower detector, some connections observed for other pins on the board -> requires further investigation.  

NOTE - Channel 0 on bias supply is unstable, should not be used. 

 Most equipment for installation of FEE's and coolant manifolds found, assembley of FEE holders started. Full list of current swagelock inventory as follows:

SS-QTM2-D-8M0 MALE QUICK DISCONNECT 15
SS-8M0-1-6RS MNIFOLD ADAPTOR 40
NY-8M4-1 ferrule fronts (nylon) 120
NY-8M3-1 ferrule rears (nylon) 120
SS-QTM2A-B-8M0 FEMALE QUICK DISCONNECT 15
Attachment 1: 20220110_175547.jpg
20220110_175547.jpg
Attachment 2: 20220110_175811.jpg
20220110_175811.jpg
Attachment 3: vi-curve.png
vi-curve.png
  424   Fri Jul 21 08:21:54 2023 TD, RSSBaking tent manual
09.20 Pirani 1.00e-3 mbar IE514 3.90e-8 mbar

      Today - install TCs, connect to Pico controller and test

10.00 Bakeout tent setup guide (draft) - attachment 1

14:30 Tested all TCs and are fine

05.25 Pirani 1.00e-3 mbar IE514 3.71e-8 mbar
Attachment 1: CARME-baking-guide_(1).pdf
CARME-baking-guide_(1).pdf CARME-baking-guide_(1).pdf CARME-baking-guide_(1).pdf CARME-baking-guide_(1).pdf CARME-baking-guide_(1).pdf CARME-baking-guide_(1).pdf
  431   Thu Aug 10 07:51:43 2023 JMBaking tent and heater testing

09/08/23

Attached is the current pumpdown plot taken from grafana. 

I have started to re-work the baking tent in order to close gaps from its initial installation. The heaters work and were tested to a temperature of 50C. Testing is required to gauge the heater temperature required to heat the baking volume, we suspect the heaters will be several hundred degrees. 

11/08/23

Baking tent has been fully constructed again, see photos attached.  Some thermocouples have been re-positioned, new thermocouples positions are given in elog 433. GSI ion pump cable has been connected. Roughing pump on the outside of the berylium window has been connected and berylium window is at < 10 mbar. 

Attachment 1: current-pump-down.png
current-pump-down.png
Attachment 2: 20230812_120951.jpg
20230812_120951.jpg
Attachment 3: 20230812_120925.jpg
20230812_120925.jpg
Attachment 4: 20230811_152942.jpg
20230811_152942.jpg
Attachment 5: 20230811_152926.jpg
20230811_152926.jpg
Attachment 6: 20230811_152923.jpg
20230811_152923.jpg
Attachment 7: 20230811_152918.jpg
20230811_152918.jpg
Attachment 8: 20230811_152913.jpg
20230811_152913.jpg
  13   Wed Sep 16 08:52:03 2020 CB (remotely)Baking status 16-17 September
16 September

08:50
Baking continues OK. Temperatures stable, did not increase with respect to yesterday evening.
See attached.

15:00
Baking continues OK.
See attached.

20:22
Baking continues OK.
See attached.

17 September

08:20
Baking continues OK. Temperatures slightly decreasing?
See attached.
Attachment 1: 20200916_0850_temps.png
20200916_0850_temps.png
Attachment 2: 20200916_0850_ctrl.png
20200916_0850_ctrl.png
Attachment 3: 20200916_1500_temps.png
20200916_1500_temps.png
Attachment 4: 20200916_2022_temps.png
20200916_2022_temps.png
Attachment 5: 20200917_0820_temps.png
20200917_0820_temps.png
  11   Mon Sep 14 12:07:40 2020 JM, CB (remotely)Baking status
Mon 14 Sep 09:50

Jordan moved the TC08 picolog 2 away from the chamber and built a makeshit heatshield. 
Picolog 2 was facing a slit in the cloth thorugh which the scroll pump flexible hose is connected.
Baking restarted in automatic mode from 80 C

12:09
Baking progressed OK - see attached.

~13:00
CJC 2 getting too high again. Baking went off. Jordan caught it and temporarily set in manual mode.
Moved picolog 2 further away. Increased amount of heatshield. Restarted in auto

17:30
Baking continues OK - see attached.
Will leave overnight.

19:30
Baking continues OK, T=124C , see attached.
However rates shown on Baking.x program appear nonsensical. Too high, too low, sometimes they all go to zero.
Clearly a bug in the code. Should not affect baking for now.
Will try to debug.
Attachment 1: 20200914_1207_temps.png
20200914_1207_temps.png
Attachment 2: 20200914_1207_ctrl.png
20200914_1207_ctrl.png
Attachment 3: 20200914_1727_temps.png
20200914_1727_temps.png
Attachment 4: 20200914_2000_temps.png
20200914_2000_temps.png
  16   Fri Sep 18 10:28:57 2020 JM, CB(remotely)Baking status
10:28
Reducing temperature in steps. System cooling down. See attached.

Pressure on the leak cart: 1.6e-6 mbar
Attachment 1: 20200918_1029_temps.png
20200918_1029_temps.png
  22   Sat Sep 26 09:31:46 2020 JM, CB (both remotely)Baking over weekend

Sat 26 Sep

09:15 Baking started automatic mode

13:30 Baking continues, heating is slower than expected due to high chamber bottom temperatures

00:00 Baking stopped. Temperatures won't increase due to high temperature of the bottom of the chamber producing thermal gradients that are too high. I believe this is due to heaters being closer to the chamber as tray sits higher up than in previous test (see heaters.png). A shield above the heaters, similar to those we utilised in early tests in Edinburgh, may help to disperse the heat more evenly around the baking volume and not directly at the bottom of the chamber. I can improvise a shield similar to those using pieces of the aluminium outer layer of the baking setup I have spare on Monday and restart baking. This will also allow me to replace the two damaged thermocouples

 

Sun 27 Sep

11:50 Baking restarted. Manually set temperature to 50 C. Started automatic mode for additional safety checks. Will check again / increase temperature manually in a few hours

13:45 Found baking had stopped due to automatic mode detecting negative temperature gradient in TC7 Floating (bottom). See plot attached. Restarted baking at 50 C in manual mode

17:30 Found temperatures had stabilised. Increased temperature to 60 C. Issues seems to be that bottom thermocouples run hotter than the reference thermocouples of the RE72. The temperature the RE72 are set to maintain is not the temperature of the chamber. This is the cause of the strong gradients. Swapping the position of the thermocouples may be sufficient. Thermocouples presumed faulty on Friday may be working OK - could just be positioned in hot spot immediately above heaters.

18:00 Baking off to allow cooling down overnight.

Attachment 1: temps.png
temps.png
Attachment 2: temps2.png
temps2.png
Attachment 3: Heaters.jpg
Heaters.jpg
Attachment 4: 20200927_1732_temps.png
20200927_1732_temps.png
  10   Sun Sep 13 08:06:25 2020 CB (remotely)Baking over the weekend -2
Sun 13 Sep 08:06

Baking reached 93 C yesterday afternoon at 16:18 and paused as a result of a percieved temperature gradient in
the octagon. 
See attached.

Left system overnight to check whether octagon gradient was going to equilibrate via convection. 
No changes this morning.
Thermocouple 1-8 labeled "Octagon (side)" likely is not in thermal contact with the octagon. 
It appears to have a temperature lower than room temperature. Possibly sits outside the tent? Gradient will not
decrease with time.

Temporarily overrode automatic mode, set temperature to 100 C, restarted automatic mode. Will keep monitoring.


Sun 13 Sep 10:00

Vessel reached 100 C, no gradients. All OK, except octagon gradient (ignoring).
Manually increased temperature to 110 C. Restored automatic mode.

Sun 13 Sep 12:00

Baking stopped automatically at 10:30 due to CJC temperature (aka room temperature 2) in TC-08 Picolog 2
exceeding 40 C.
Max operating temperature of picolog TC08 is 50 C, and accuracy is lost above 30 C.
https://www.picotech.com/data-logger/tc-08/usb-tc-08-specifications

Stopped baking. Removed thermocouple 1-8 "Octagon (side)" from Octagon gradient calculation and recompiled.
Restarted manually at T = 80 C

Please move picolog away from the tent tomorrow, and we can then increase temperature. Tent does not need to be
opened.

Sun 13 Sep 19:00
Vessel temperature at 80 C stable. 
Other temperatures stable. 
Leaving at 80 C overnight.
Attachment 1: 20201309_0806_ctrl.png
20201309_0806_ctrl.png
Attachment 2: 20201309_0806_plot.png
20201309_0806_plot.png
  9   Sat Sep 12 08:38:52 2020 CB (remotely)Baking over the weekend
Saturday 12 Sep 08:43

Remotely checked status of baking. See attached.

Baking appears to be proceeding OK. No thermocouples disconnected.

Speed is slower than previously observed, limited by thermal gradients observed across both octagon and CARME
vessel as a result of new thermocouple placement.
Octagon gradient may not be real, and could be a result of placement of Octagon side thermocouple?

Main concern is temperature recorded by CJC in Picolog2 (aka Room temperature 2) at 31 C, increasing. 
The second picolog may be too close to the tent. If the temperature keeps increasing it may stop working. Will
keep monitoring.
Attachment 1: 20200912_0839_ctrl.png
20200912_0839_ctrl.png
Attachment 2: 20200912_0839_plot.png
20200912_0839_plot.png
ELOG V3.1.4-unknown