ID |
Date |
Author |
Subject |
634
|
Fri Jun 14 22:18:18 2024 |
RSS, CB, US | Run 73 |
Beam is back.New run started.
Cooler voltage setpoint 247.3 V.
23:18
DSSSD bias & leakage current OK
FEE64 temperatures OK
ADC data item stats OK
per FEE64 Rate spectra OK
Beam intensity 6e+7 particles
Target density ~4E10 atoms/cm2
01:01
DSSSD bias & leakage current OK - attachment 1 & 5
FEE64 temperatures OK - attachment 2
ADC data item stats OK - attachment 3
per FEE64 Rate spectra OK - attachment 4
Beam intensity 7e+7 particles
Target density ~5E11 atoms/cm2
03:03
DSSSD bias & leakage current OK - attachment 6 & 7
FEE64 temperatures OK - attachment 8
ADC data item stats OK - attachment 9
per FEE64 Rate spectra OK - attachment 10
Beam intensity 7e+7 particles
Target density ~4E11 atoms/cm2
05:00
DSSSD bias & leakage current OK - attachment 11 & 15
FEE64 temperatures OK - attachment 12
ADC data item stats OK - attachment 13
per FEE64 Rate spectra OK - attachment 14
Beam intensity 7.4e+7 particles
Target density ~5E11 atoms/cm2
05:15
Turbo frequency went below 810 Hz. Set temperature increased from 38.8 K to 39.0 K.
06:50
DSSSD bias & leakage current OK - attachment 16 & 17
FEE64 temperatures OK - attachment 18
ADC data item stats OK - attachment 19
per FEE64 Rate spectra OK - attachment 20
Beam intensity 7.5e+7 particles
Target density ~2.3E11 atoms/cm2 |
Attachment 1: Screenshot_from_2024-06-15_00-59-40.png
|
|
Attachment 2: Screenshot_from_2024-06-15_00-59-53.png
|
|
Attachment 3: Screenshot_from_2024-06-15_01-00-03.png
|
|
Attachment 4: Screenshot_from_2024-06-15_01-00-12.png
|
|
Attachment 5: Screenshot_from_2024-06-15_01-00-20.png
|
|
Attachment 6: Screenshot_from_2024-06-15_03-03-21.png
|
|
Attachment 7: Screenshot_from_2024-06-15_03-03-41.png
|
|
Attachment 8: Screenshot_from_2024-06-15_03-03-49.png
|
|
Attachment 9: Screenshot_from_2024-06-15_03-04-00.png
|
|
Attachment 10: Screenshot_from_2024-06-15_03-04-08.png
|
|
Attachment 11: Screenshot_from_2024-06-15_05-00-41.png
|
|
Attachment 12: Screenshot_from_2024-06-15_05-01-00.png
|
|
Attachment 13: Screenshot_from_2024-06-15_05-01-14.png
|
|
Attachment 14: Screenshot_from_2024-06-15_05-01-23.png
|
|
Attachment 15: Screenshot_from_2024-06-15_05-01-34.png
|
|
Attachment 16: Screenshot_from_2024-06-15_06-49-51.png
|
|
Attachment 17: Screenshot_from_2024-06-15_06-49-57.png
|
|
Attachment 18: Screenshot_from_2024-06-15_06-50-13.png
|
|
Attachment 19: Screenshot_from_2024-06-15_06-50-25.png
|
|
Attachment 20: Screenshot_from_2024-06-15_06-50-33.png
|
|
633
|
Fri Jun 14 19:59:14 2024 |
US, LC, CB | New run with name of 72 is started due to target issue |
21:00
Target is lost. Background run is being collected.
New run 72 - beam on target with definite target overlap [coasting]
Cooler voltage setpoint 247.3 V.
FC20 read value 246.75 V.
Right arm at 64.60 mm
Left arm IN at 36.54 mm18:00
FakeMotion script running
horizontal beam position 1.0 mm (new beam target overlap)
vertical beam position -5mm
Beam intensity 6e+7 particles.
Target density ~3e9 at/cm2
Target temperature is 90 K. Values of E1 is 820 Hz, 2.6A, 4.2e-2 mbar
Aida03 detector T is stable.SCreenshot is attached.
22:27
Cooling fixed. Temps OK. |
Attachment 1: Screenshot_from_2024-06-14_21-08-22.png
|
|
Attachment 2: Screenshot_from_2024-06-14_21-08-22.png
|
|
632
|
Fri Jun 14 14:52:02 2024 |
JM | Beam-Target Overlap Scan |
The orbit of the beam in the ring was re-configured so that the beam position is more central.
With the target density at ~6e11 at/cm2 this allowed a target overlap scan using the beam lifetime to be performed.
Beam put into bunched mode to allow the number of particles to be observed on Grafana.
The number of particles injected at the start of the cycle was then compared to the number of particles at the end of the cycle to gauge the % of the beam lost.
A flat top region was observed where >99% of the particles in the ring were lost. The centre of this flat top corresponds to a horizontal beam position of 1.0 mm
Attachment 1 - Grafana plot of beam intensity
Attachment 2 - Plot of overlap |
Attachment 1: 20240614_153532.jpg
|
|
Attachment 2: Beam-overlap-updated.png
|
|
631
|
Fri Jun 14 14:48:02 2024 |
CB NP ML | [How To] Change target pressure. EMERGENCY ONLY! |
ONLY FOLLOW THESE INSTRUCTIONS if E1 turbo is below 820 Hz, pressure in E1 is above 2.5E-1 mbar and experts (Nikos, Andreas) cannot be reached.
Go to the red cabinet under the stairs.
Open rightmost door. See first photo. Do not press the button on the handle or the door will lock!
Turn down pressure regulator indicated in second photo by approximately 1 degree. Read regulator to know which way to turn to decrease. No immediate change should be obvious on the pressure regulator.
Check E1 pressure for 5 minutes. If no effect, turn pressure down again until an effect appears. |
Attachment 1: 20240614_142435.jpg
|
|
Attachment 2: 20240614_142440.jpg
|
|
630
|
Fri Jun 14 14:46:54 2024 |
LC, JM, CB, TD, US | Run 71 - beam on target |
15:45
New run 71 - beam on target with definite target overlap [coasting]
Cooler voltage setpoint 247.3 V.
FC20 read value 246.75 V.
Right arm at 64.60 mm
Left arm IN at 36.54 mm18:00
FakeMotion script running
horizontal beam position 1.0 mm (new beam target overlap)
vertical beam position -5mm
Beam intensity 7.0e+7 particles.
Target density ~4e11 at/cm2
Target temperature is 36.6 K. Values of E1 is 820 Hz, 2.6A, 1.4e-1 mbar
16:36
No difference observed between bunched and coasting
See attached screenshots to compare to previous run
Saw an increase in the data rate to disk ~ 2000 kB/s. On spectrum browser we saw aida15 had HEC reading hot. ASIC check on aida 15 resolved issue
18:00
Beam intensity 6e+7 particles.
Target density ~6e11 at/cm2
Target temperature is 36.6 K. Values of E1 is 820 Hz, 2.7 A, 1.5e-1 mbar
Aida03 detector temperature is 66 degree.
20:00
Beam intensity 6e+7 particles.
Target density ~6e11 at/cm2
Target temperature is 36.6 K. Values of E1 is 820 Hz, 2.7 A, 1.6e-1 mbar
Aida03 detector temperature is 63 degree. Screenshot is attached.
21:00
arget densityis dropped drastically. A temperature is ncreased signifiantly.
Beam intensity 6e+7 particles.
Target density ~2.7e9 at/cm2
Target temperature is 9 K. Values of E1 is 820 Hz, 2.7 A, 4.2e-2mbar
Aida03 detector temperature is 63 degree. Screenshot is attached.
|
Attachment 1: Screenshot_from_2024-06-14_16-35-44.png
|
|
Attachment 2: Screenshot_from_2024-06-14_16-36-06.png
|
|
Attachment 3: Screenshot_from_2024-06-14_16-36-28.png
|
|
Attachment 4: Screenshot_from_2024-06-14_19-49-02.png
|
|
629
|
Fri Jun 14 13:45:21 2024 |
JM, JS, CB, TD, EH | Run 70 - beam on target |
14:45
New run 70 - beam on target with definite target overlap
Cooler voltage setpoint 247.3 V.
FC20 read value 246.75 V.
Right arm at 64.60 mm
Left arm IN at 36.54 mm
FakeMotion script running
horizontal beam position 1.0 mm (new beam target overlap)
vertical beam position -5mm
Beam intensity 7.0e+7 particles.
Target density ~6.3e11 at/cm2
Target temperature is 36.6 K. Values of E1 is 820 Hz, 2.6A, 1.4e-1 mbar
All system checks OK, data rate is ~ 260 kB/s. Statistics attached.
xy and energy histograms following ~20 mins attached.
15:45
I realised we have been running in bunched mode not coasting as we have been previously.
See attached screenshots (4-7) of last 1 hour of running. We will move to coasting. |
Attachment 1: Screenshot_from_2024-06-14_14-51-46.png
|
|
Attachment 2: Screenshot_from_2024-06-14_14-53-57.png
|
|
Attachment 3: Screenshot_from_2024-06-14_14-55-02.png
|
|
Attachment 4: Screenshot_from_2024-06-14_15-38-58.png
|
|
Attachment 5: Screenshot_from_2024-06-14_15-39-16.png
|
|
Attachment 6: Screenshot_from_2024-06-14_15-39-36.png
|
|
Attachment 7: Screenshot_from_2024-06-14_15-40-14.png
|
|
628
|
Fri Jun 14 10:21:24 2024 |
JS, TD | Run 68 |
Frank has been tuning the beam since 08:00
detectors moved fully out
11:20
New run (R68) started at 11:20. Put into no storage mode as rates are high during beam tuning. Cut off the end of last nights run when analysing
System checks OK
13.20 DSSSD bias & leakage current OK - attachment 1
FEE64 temperatures OK - attachment 2
ADC data item stats - attachment 3
per FEE64 1.8.W spectra - attachments 4-5
per FEE64 Rate spectra - attachment 6
aida16 producing HEC data - Chck ASIC load to fix
TapeServer - no storage mode
13.35 aida01-aida04 c. 250k ADC data items/s - will await completion of current beam/target scan to fix by cycing Left DSSSD motion
14.00 CB cycles Left DSSSD motion to fix aida01-aida04 rate issue
CRYRING vacuum & hall temprature - attachments 7-8
14.28 New run data file G22-201/R69
Beam/target overlap optimised - details to follow |
Attachment 1: Screenshot_from_2024-06-14_13-22-13.png
|
|
Attachment 2: Screenshot_from_2024-06-14_13-22-18.png
|
|
Attachment 3: Screenshot_from_2024-06-14_13-22-37.png
|
|
Attachment 4: Screenshot_from_2024-06-14_13-22-58.png
|
|
Attachment 5: Screenshot_from_2024-06-14_13-23-30.png
|
|
Attachment 6: Screenshot_from_2024-06-14_13-24-08.png
|
|
Attachment 7: Screenshot_from_2024-06-14_13-25-02.png
|
|
Attachment 8: vac.PNG
|
|
Attachment 9: temp.PNG
|
|
627
|
Thu Jun 13 23:01:07 2024 |
RSS, JS, EH | Run 67 - Night shift |
Run 67
00:01
DSSSD bias & leakage current OK - attachment 1
FEE64 temperatures OK - attachment 2
ADC data item stats OK - attachment 3
per FEE64 Rate spectra OK - attachment 4
Beam intensity 5.7E7 particles
Target density - 4.98E11 atoms/cm2
02:01
DSSSD bias & leakage current OK - attachment 5
FEE64 temperatures OK - attachment 6
ADC data item stats OK - attachment 7
per FEE64 Rate spectra OK - attachment 8
Beam intensity 5.8E7 particles
Target density - 4.39E11 atoms/cm2
04:00
DSSSD bias & leakage current OK - attachment 9
FEE64 temperatures OK - attachment 10
ADC data item stats OK - attachment 11
per FEE64 Rate spectra OK - attachment 12
Beam intensity 4.1E7 particles
Target density - 3.57E11 atoms/cm2
06:00
DSSSD bias & leakage current OK - attachment 13
FEE64 temperatures OK - attachment 14
ADC data item stats OK - attachment 15
per FEE64 Rate spectra OK - attachment 16
Beam intensity 3.7E7 particles
Target density - 3.89E11 atoms/cm2
08:00
The orbit adjustments started. The detectors moved out. One of these showed rates of about 20kHz. but after moving it in and out again it returned to normal.
11:00
DSSSD bias & leakage current OK - attachment 17
FEE64 temperatures OK - attachment 18
ADC data item stats OK - attachment 19
per FEE64 Rate spectra OK - attachment 20
Beam intensity 4.9E7 particles
Target density - 5.27E11 atoms/cm2 |
Attachment 1: Screenshot_from_2024-06-14_00-02-01.png
|
|
Attachment 2: Screenshot_from_2024-06-14_00-02-20.png
|
|
Attachment 3: Screenshot_from_2024-06-14_00-02-31.png
|
|
Attachment 4: Screenshot_from_2024-06-14_00-02-40.png
|
|
Attachment 5: Screenshot_from_2024-06-14_02-00-42.png
|
|
Attachment 6: Screenshot_from_2024-06-14_02-00-58.png
|
|
Attachment 7: Screenshot_from_2024-06-14_02-01-04.png
|
|
Attachment 8: Screenshot_from_2024-06-14_02-01-13.png
|
|
Attachment 9: Screenshot_from_2024-06-14_04-00-08.png
|
|
Attachment 10: Screenshot_from_2024-06-14_04-00-26.png
|
|
Attachment 11: Screenshot_from_2024-06-14_04-00-38.png
|
|
Attachment 12: Screenshot_from_2024-06-14_04-00-52.png
|
|
Attachment 13: Screenshot_from_2024-06-14_06-00-15.png
|
|
Attachment 14: Screenshot_from_2024-06-14_06-00-31.png
|
|
Attachment 15: Screenshot_from_2024-06-14_06-00-42.png
|
|
Attachment 16: Screenshot_from_2024-06-14_06-00-51.png
|
|
Attachment 17: Screenshot_from_2024-06-14_10-53-10.png
|
|
Attachment 18: Screenshot_from_2024-06-14_10-53-54.png
|
|
Attachment 19: Screenshot_from_2024-06-14_10-54-28.png
|
|
Attachment 20: Screenshot_from_2024-06-14_10-55-05.png
|
|
626
|
Thu Jun 13 21:07:30 2024 |
LCS, US | Run 67- beam on target |
22:05
The energy is changed to 247.4 V.
Read back value is 246.849 V.
Right arm at 64.60 mm
Left arm IN at 36.54 mm
horizontal beam position 8.75 mm
vertical beam position -5mm
Beam intensity 4.0e+7 particles.
Run is R67. Target temperature is 36..6 K.
Values of E1 is 820 Hz, 2.4A, 8.69e-2 mbar, 2.55e+11/cm2. The pressure is slowly increasing since this afternoon. The recent value is E1 is 820 Hz, 2.4A, 9.0e-2 mbar, 3.28e+11/cm2
Systen checks all OK
Attachments
13.39 DSSSD bias & leakage current OK
FEE64 temps OK
ADC data item stats OK
per FEE64 Rate spectra
Merger, TapeServer et
data file R65_11
DSSSD bias & leakge current log
CARME DSSSD motion control
|
Attachment 1: Screenshot_from_2024-06-13_22-14-46.png
|
|
Attachment 2: Screenshot_from_2024-06-13_22-15-01.png
|
|
Attachment 3: Screenshot_from_2024-06-13_22-15-14.png
|
|
Attachment 4: Screenshot_from_2024-06-13_22-15-38.png
|
|
Attachment 5: Screenshot_from_2024-06-13_22-15-44.png
|
|
Attachment 6: Screenshot_from_2024-06-13_22-15-51.png
|
|
Attachment 7: Screenshot_from_2024-06-13_22-16-07.png
|
|
625
|
Thu Jun 13 14:46:17 2024 |
TD | R66 beam overlap test |
15.45 Start data file G22-201/R66
Beam/target overlap test |
624
|
Thu Jun 13 09:48:25 2024 |
TD, LCS | Run 65 - beam on target |
10:47
The energy is changed to 247.4 V.
Read back value is 246.849 V.
Right arm at 64.60 mm
Left arm IN at 36.54 mm
MBS working - hopefully
horizontal beam position 8.75 mm
vertical beam position -5mm
Beam intensity 3.0e+7 particles.
e-cooler export controls before/after energy change - attachments 1 & 2
New run is R65. Target temperature is 37.6 K.
Values of E1 is 820 Hz, 2.12A, 6.01e-2 mbar, 2.8e+11/cm2
12:34
Systen checks all OK
Target density stable at ~1.5e11 at/cm2
E1 = 6e-2 mbar, turbo is fine
13,38 CB reboots RHS CRYRING control computer (atppc017) - appears to be a now unavoidable system update/reboot
13.39 DSSSD bias & leakage current OK - attachment 3
FEE64 temps OK - attachment 4
ADC data item stats OK - attachment 5
per FEE64 Rate spectra - attachment 6
Merger, TapeServer etc - attachments 7-8
data file R65_11
DSSSD bias & leakge current log - attachment 12
CARME DSSSD motion control - attachment 13
13.42 E1 turbo 820Hz 2.2A 5.2e-2mbar
Target 3.45e+11/cm2 T1 36.6K
Beam c. 3e+7
15:20
Stats OK. Temps OK.
Merger OK. Write to disk OK.
|
Attachment 1: 20240613_104403.jpg
|
|
Attachment 2: 20240613_104840.jpg
|
|
Attachment 3: Screenshot_from_2024-06-13_13-39-26.png
|
|
Attachment 4: Screenshot_from_2024-06-13_13-39-40.png
|
|
Attachment 5: Screenshot_from_2024-06-13_13-40-09.png
|
|
Attachment 6: Screenshot_from_2024-06-13_13-40-32.png
|
|
Attachment 7: Screenshot_from_2024-06-13_13-40-52.png
|
|
Attachment 8: Screenshot_from_2024-06-13_13-41-02.png
|
|
Attachment 9: Screenshot_from_2024-06-13_13-41-16.png
|
|
Attachment 10: Screenshot_from_2024-06-13_13-41-23.png
|
|
Attachment 11: Screenshot_from_2024-06-13_15-12-19.png
|
|
Attachment 12: Screenshot_from_2024-06-13_15-13-16.png
|
|
Attachment 13: Screenshot_from_2024-06-13_15-20-28.png
|
|
Attachment 14: Screenshot_from_2024-06-13_15-20-31.png
|
|
623
|
Thu Jun 13 06:21:53 2024 |
TD, LCS | Run 64 - contd |
07.22 E1 pump 799Hz, 2.8A, 2.6e-1 mbar, 4.8e+11/cm2
07.24 DSSSD bias & leakage current OK - attachment 1
FEE64 temperatures OK - attachment 2
ADC data item stats OK - attachment 3
per FEE64 Rate spectra - attachment 4
per FEE64 1.8.L spectra - attachments 5-6
per FEE64 1.8.W spectra - 20us FSR - attachments 7-8
Merger, TapeServer etc - attachments 9-10
data file G22-201/R64_11
DSSSD bias & leakage current log - attachment 11
CRYRING hall temperature log - attachment 12
CARME DSSSD position controller - attachment 13
07.57 all histograms zero'd
07.58 E1 pump 794Hz, 2.8A, 2.6e-1 mbar, 3.1e+11/cm2
08.55 E1 pump 785Hz, 2.8A, 2.74e-1 mbar, 4.8e+11/cm2. Therefore, the temperature is changed from 38.8 K to 38.4 K. Values of E1 is changed to E1 pump 782Hz, 2.7A, 2.95e-1 mbar, 6e+11/cm2
09.15 The temperature is further changed from 38.4 K to 38.0 K.
09.25 Values of E1 is changed to E1 pump 780 Hz, 2.79 A, 3.0 e-1 mbar, 5.87 e+11/cm2
09.35 Values of E1 is changed to E1 pump 779 Hz, 2.79 A, 3.1 e-1 mbar, 2.07 e+11/cm2
09.45 Values of E1 is changed to E1 pump 778 Hz, 2.75 A, 3.0 e-1 mbar, 4.6 e+11/cm2
10.00 Values of E1 is changed to E1 pump 776 Hz, 2.78 A, 3.05 e-1 mbar, 5 e+11/cm2
Intervention by Uli. Reduced H2 regulator to c. 12 bar. Briefly opened exhaust to hasten recovery.
10.13 E1 820Hz 2.3A 7.7e-2mbar c. 2e+11/cm2
10.34 E1 820Hz 2.1A 6.2e-2mbar c. 1.2e+11/cm2 38K
Beam c. 4e+7. Cooler set 247.3V, read 246.751V. |
Attachment 1: Screenshot_from_2024-06-13_07-24-49.png
|
|
Attachment 2: Screenshot_from_2024-06-13_07-28-56.png
|
|
Attachment 3: Screenshot_from_2024-06-13_07-30-34.png
|
|
Attachment 4: Screenshot_from_2024-06-13_07-36-35.png
|
|
Attachment 5: Screenshot_from_2024-06-13_07-40-06.png
|
|
Attachment 6: Screenshot_from_2024-06-13_07-41-10.png
|
|
Attachment 7: Screenshot_from_2024-06-13_07-42-32.png
|
|
Attachment 8: Screenshot_from_2024-06-13_07-43-30.png
|
|
Attachment 9: Screenshot_from_2024-06-13_08-10-26.png
|
|
Attachment 10: Screenshot_from_2024-06-13_08-10-33.png
|
|
Attachment 11: Screenshot_from_2024-06-13_08-10-43.png
|
|
Attachment 12: Capture.PNG
|
|
Attachment 13: Screenshot_from_2024-06-13_08-16-07.png
|
|
622
|
Thu Jun 13 00:50:33 2024 |
JM, JS | Run 64 - beam on target |
01:50
Right arm at 64.60 mm
Left arm IN at 36.54 mm
MBS working - hopefully
Target ~3E11
Injection ca. 3E7
horizontal beam position 8.75 mm
vertical beam position -5mm
Changed electron cooler set point to 247.3 V
Read value 246.75 V
Collector current 4.3A
Histograms and statistics zero'd. Code reset
Rates OK. Temps OK.
NOTE. Between 23:45 and 12:45 the electron cooler was off? We called HKR and they turned it back on. Are we using the same cooler settings? Is the cooler still definitely working properly? Ask Claude in the morning
02:30
The target density started to oscillate between 2e+10 and 3.5e+10. However, the pump speed seems to be stable.
03:50
The oscillations seems to deepend between 1e+10 and 6e+10. Additionally the E1 pump slowed down to 815. As Nikos suggested during the tuning, I tried to lower the temperature from 41K to 40.8K.
04:00 - 04:30
No major change was observed, thus I increased the temperature to 42.2K in order to try to get the pump to the nominal rotation. After 10 minutes, the pump slowed down to 810 and the pressure in E1 reached 2.2e-1. Since increasing the temperature seemed to have a negative effect, I slowly decresed the temperature with steps of 0.2 K, until reaching 40K. At this point, the E1 pump returned to 819/820, the pressure of E1 lowered to 1.6e-1 and the target density seems to be more stable with a value of 2.8e+11.
05:00
The target density started decreasing to 1.5e+10 and the pump speed slowed to 815. I've tried to lower the temperature to 39.8K, but the target density started to decrease. Thus I've set the temperature to 40.2K. The pump speed raised to 818 but the target density remaind at 1e+11 level with oscillations between 8.5e+10 and 1.5e+11. After some minutes, it started slowly increasing towards the 2e+11 region. The E1 pumping speed is stable at 818 and the pressure of E1 is 2.02e-1.
05:10
Electron cooler off, restarted 10 minutes after by calling HKK.
05:30
Target density at 1.5e+11, however, the E1 raised to 2.3e-1 and the pump speed decreased to 810. Since going down seemed to solve it, I decreased the temperature to 39.6K. After 5 minutes the E1 lowered to 1.9e-1 and the pump speed is going up (currently at 814). The target density is at the same level of 1.5e+11.
06:00
Again the target density started to go down, as well the speed of the E1 pump and the E1 pressure reached 2.3e-1. I lowered the temperature to 39.2K: the target density raised to approximately 3.8e+11 (and keeps going up), the E1 pressure lowered to 1.9e-1 and the E1 pump speed started increasing from 810 to 815.
06:30
The target density starting going down again, the E1 started increasing to 2.3e-1 and the E1 pump speed decreased to 810. I lowered the temperature to 38.8K, the density raised to 3.5e+11, the E1 pressure lowered to 1.9e-1 and the pumping speed starting increasing. |
621
|
Wed Jun 12 07:16:47 2024 |
RSS, YL, JM, TD, CB | Status |
07:56 Gleb has taken the beam to try settings for the 5+ charge state. As soon as the target is ready, the 1+ charge state will be given back.
Detectors have been moved out for safety
08:10
Run 63
DSSSD bias & leakage current OK - attachment 1
FEE64 temperatures OK - attachment 2
ADC data item stats - attachment 3
per FEE64 Rate spectra - attachment 4
09:55
Run 63
DSSSD bias & leakage current OK
FEE64 temperatures OK
ADC data item stats OK
per FEE64 Rate spectra OK
10:30
Data rate appeared very high c. 2000 kB/s instead of the usual 200 kB/s.
Rate on first two ASICs of aida02 was high (~2 kHz) but rest of aida01-04 looked fine.
ASIC check load reduced rate back to normal levels. Data rate reduced back to 200 kB/sec.
New ADC offsets loaded into the CARME_Sort code
Aida03 appears to be down. Check load does not resolve issue. Moving detector arm does not resolve issue. No spike in leakage current observed. System checks appear OK. Zeroing ADC data item statistics keeps aida03 at 606023857 items and does not zero.
Issue resolved by resetting the data links for aida03
13:20
DSSSD bias & leakage current OK
FEE64 temperatures OK
ADC data item stats OK
per FEE64 Rate spectra OK
21:30
Waiting for the target.
Beam has been running, we have been running without storage.
See attachment showing beam stripe on the detectors.
|
Attachment 1: Screenshot_from_2024-06-12_08-18-27.png
|
|
Attachment 2: Screenshot_from_2024-06-12_08-18-44.png
|
|
Attachment 3: Screenshot_from_2024-06-12_08-18-55.png
|
|
Attachment 4: Screenshot_from_2024-06-12_08-19-08.png
|
|
Attachment 5: Screenshot_from_2024-06-12_11-10-34.png
|
|
Attachment 6: Screenshot_from_2024-06-12_21-39-20.png
|
|
620
|
Tue Jun 11 21:16:50 2024 |
TD | Run 63 continued |
Run 63 continued
22.06 DSSSD bias & leakage current OK - attachment 1
FEE64 temperatures OK - attachment 2
ADC data item stats - attachment 3
per FEE64 Rate spectra - attachments 4-5
aida16 producing HEC data, ASIC check load to clear
23.56 DSSSD bias & leakage current OK - attachment 6
FEE64 temperatures OK - attachment 7
ADC data item stats - attachment 8
per FEE64 Rate spectra - attachments 9
aida01-aida04 high ADC data item rates - left DSSSD motion cycle required to reset
03.28 DSSSD bias & leakage current OK - attachment 10
FEE64 temperatures OK - attachment 11
ADC data item stats - attachment 12
per FEE64 Rate spectra - attachments 13
aida01-aida04 high ADC data item rates - left DSSSD motion cycle required to reset
07.50 DSSSD bias & leakage current OK - attachment 14
FEE64 temperatures OK - attachment 15
ADC data item stats - attachment 16
per FEE64 Rate spectra - attachments 17
aida01-aida04 high ADC data item rates (see above) appears to have resolved itself or has been resolved (but not logged) |
Attachment 1: Screenshot_from_2024-06-11_22-06-38.png
|
|
Attachment 2: Screenshot_from_2024-06-11_22-07-38.png
|
|
Attachment 3: Screenshot_from_2024-06-11_22-08-37.png
|
|
Attachment 4: Screenshot_from_2024-06-11_22-12-35.png
|
|
Attachment 5: Screenshot_from_2024-06-11_22-10-38.png
|
|
Attachment 6: Screenshot_from_2024-06-11_23-56-29.png
|
|
Attachment 7: Screenshot_from_2024-06-12_00-00-16.png
|
|
Attachment 8: Screenshot_from_2024-06-12_00-02-06.png
|
|
Attachment 9: Screenshot_from_2024-06-12_00-02-35.png
|
|
Attachment 10: Screenshot_from_2024-06-12_03-25-53.png
|
|
Attachment 11: Screenshot_from_2024-06-12_03-26-33.png
|
|
Attachment 12: Screenshot_from_2024-06-12_03-27-12.png
|
|
Attachment 13: Screenshot_from_2024-06-12_03-28-02.png
|
|
Attachment 14: Screenshot_from_2024-06-12_07-50-16.png
|
|
Attachment 15: Screenshot_from_2024-06-12_07-52-53.png
|
|
Attachment 16: Screenshot_from_2024-06-12_07-53-23.png
|
|
Attachment 17: Screenshot_from_2024-06-12_07-53-53.png
|
|
619
|
Tue Jun 11 20:10:25 2024 |
JM | Run 63 - Beam only background |
21:00
Claude has achieved optimal cooling parameters for the beam
No target and internal work inside the cave tomorrow is required
Run 63 going - beam only background
Right arm at 64.60 mm
Left arm IN at 36.54 mm
MBS fakemotion script running
High rates on top left dssd again. Cycle left servo motor two times resolved issue
Rates OK. Temps OK. |
618
|
Tue Jun 11 19:44:48 2024 |
OF, SS | Target out and at He |
Target has been moved out and H2 switched off. Changed to He.
Nikos will repair nozzle tomorrow morning. Remote checks until then. |
617
|
Tue Jun 11 12:26:07 2024 |
RSS, JS, JM, CB | Run 62 - Beam optimization + Target change |
Beam optimization and target change |
616
|
Tue Jun 11 08:34:50 2024 |
JM, RSS, JS | Run 61 - beam on target |
New run 61
Right arm at 64.60 mm
Left arm IN at 36.54 mm
MBS working - hopefully
Target 4E11
Injection ca. 3E7
horizontal beam position 8.75 mm
vertical beam position -5mm
Changed electron cooler set point to 244.95 V
Read value 244.4 V
Collector current 4.31A
Histograms and statistics zero'd. Code reset
Rates OK. Temps OK.
10:24
We lost the beam for 5 minutes. No counting spike observed when we got it back.
1L10
We lost the beam again for 5 minutes, but no counting spike obsevered when we got it back.
11:50
DSSSD bias & leakage current OK - attachment 1
FEE64 temperatures OK - attachment 2
ADC data item stats - attachment 3
per FEE64 Rate spectra - attachment 4
Beam intensity - ~ 3.3E7 pps
Target density - 5.14E11 atoms/cm2
12:00 - 12:40
Large fluctuations in the target pressure, going down to 1e-9. We increased the temperature a little but it did not solve the problem.
Increased pressure 0.5 K more. Seems more stable
13:09
FEEs on det 2 stopped working. Checkloaded. Fixed.
No beam. |
Attachment 1: Screenshot_from_2024-06-11_11-47-57.png
|
|
Attachment 2: Screenshot_from_2024-06-11_11-48-20.png
|
|
Attachment 3: Screenshot_from_2024-06-11_11-48-32.png
|
|
Attachment 4: Screenshot_from_2024-06-11_11-48-44.png
|
|
615
|
Tue Jun 11 04:36:17 2024 |
CB, RSS, JS | Run 60 - as 59, left servo in |
05:36
Run started.
Conditions as before, except left servo now really in.
06:24
Run continues OK, file R60_9
Temps OK. Stats OK. Merger OK. Write to disk OK.
System wide checks as usual.
07:30
DSSSD bias & leakage current OK - attachment 4
FEE64 temperatures OK - attachment 5
ADC data item stats - attachment 6
per FEE64 Rate spectra - attachment 7
Beam intensity - ~ 4.8E7 pps
Target density - 7.25E11 atoms/cm2
9:20
DSSSD bias & leakage current OK - attachment 8
FEE64 temperatures OK - attachment 9
ADC data item stats - attachment 10
per FEE64 Rate spectra - attachment 11
Beam intensity - ~ 4E7 pps
Target density - 7E11 atoms/cm2
|
Attachment 1: Screenshot_from_2024-06-11_06-23-36.png
|
|
Attachment 2: Screenshot_from_2024-06-11_06-23-56.png
|
|
Attachment 3: Screenshot_from_2024-06-11_06-24-11.png
|
|
Attachment 4: Screenshot_from_2024-06-11_07-27-43.png
|
|
Attachment 5: Screenshot_from_2024-06-11_07-28-02.png
|
|
Attachment 6: Screenshot_from_2024-06-11_07-28-14.png
|
|
Attachment 7: Screenshot_from_2024-06-11_07-28-22.png
|
|
Attachment 8: Screenshot_from_2024-06-11_09-18-55.png
|
|
Attachment 9: Screenshot_from_2024-06-11_09-19-10.png
|
|
Attachment 10: Screenshot_from_2024-06-11_09-19-20.png
|
|
Attachment 11: Screenshot_from_2024-06-11_09-19-28.png
|
|