AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40 MONNET
  DESPEC, Page 34 of 37  ELOG logo
ID Date Author Subject Text Attachments
  75   Thu Oct 31 15:24:35 2019 TDwaveform spectra issues?   12x 
  74   Thu Oct 31 14:27:06 2019 NHWR TimestampsAll 12 FEEs have valid WR Timestamps
Had to powercycle aida09 once as before raw
readout was displaying upper 12 bits of WR
 WRTimes.png 
  73   Thu Oct 31 14:23:37 2019 TDaida09 asic 1 no data readout?   10.png11.png12.png13.png 
  72   Thu Oct 31 10:33:34 2019 TD[How To] Start TclHttpd server on Raspberry Pi1) Login to Raspberry Pi 

ssh pi@nnrpi2
  
  71   Thu Oct 31 09:36:37 2019 TDRIKEN LayOut directory   LayOut.tar.gz 
  70   Thu Oct 31 09:00:58 2019 CA, TD, NHThursday 31 October10.05 DSSSD # 1-3 detector biases & leakage
currents OK - see attachment 1
      Ambient temperature +21.2 deg C, d.p.
 14x 
  69   Wed Oct 30 11:51:08 2019 CA, TD, NHDetector biasing12.51 - detector bias and leakage currents
- detector 3 has no current
                                        
 bias30Oct.png30Octbias2.pnghvsketch.pdf 
  68   Wed Oct 30 08:06:49 2019 CA, TD, NHDSSSD stackRibbon cables fitted with kapton PCB couplers
-> 3 isolated + 1 non-isolated for each cable
length/DSSSD
 6x 
  67   Mon Oct 28 17:02:11 2019 TDAIDA @ DESPEC cable length estimatesAIDA snout length 37cm (from end of AIDA snout
support assembly)
  
  66   Mon Sep 16 12:24:46 2019 NHAIDA Monitoringhttps://docs.google.com/spreadsheets/d/1fZBybHodSvacUdKpDM_VKznqTrh49CKnqKVQVMEDWGs/edit?usp=sharing

A spreadsheet of ambient conditions
and water temperature in S4 for AIDA.
  
  65   Fri Aug 16 15:09:05 2019 NHAida Waves (All FEEs)Included waves for all 12 FEEs

FEES 1, 5, 9 all show this HF noise
on it, and are all located at the top of
 12x 
  64   Thu Aug 15 14:09:39 2019 NHAIDA WaveformsNoise seemed less strong today, unsure
of cause. I had re-taped snout as tape was
not sticking well.
 10x 
  63   Wed Aug 14 14:40:18 2019 NHHigh Frequency Noise in AIDAFigures of intense high frequency noise
pickup in AIDA.

Will investigate source in S4.
 HF.pngHFZ.png 
  61   Tue Aug 13 13:55:45 2019 NHNew Merger/MBS Test RunsAIDA Finally working again so can test
VP's added WAVE histogramming mode.

Fig 1. Waveforms for FEE12 ASIC
 6x 
  60   Tue Aug 13 09:17:02 2019 NHAIDA 12/08/2019The other 6 FEEs were powered on one at
a time to check no issues. None found.
All 12 FEEs plugged back into the
 AIDATemps-1208191457.png 
  59   Fri Aug 9 12:33:14 2019 NHAida Test 9/8/19 - FEES 1-6The water issue has been resolved - merely
incorrect temperature guages (replaced).
Water is 20 C as required.
  
  58   Mon Aug 5 12:57:48 2019 NHS4 ConditionsTemp: 26 C
Humidity: 50%
Dew Point: ~15 C
  
  57   Wed Jul 10 10:46:05 2019 NHAIDA/Water 10.07.2019S4 Conditions:

25.2 C, 35.3% RH, Td = 8.8 C

Water temperature: 20 C
  
  56   Mon Jul 8 16:02:52 2019 NH, PJCSAida 28/6/2019 - WATER OFFIt occurs to to wonder if there is now
water inside the modules. If the metal of
the cooling plates is damp on the outside
  
  55   Mon Jul 1 09:42:16 2019 NHAIDA - 01/07/2019Water is still off.

Checked dewpoint sensor still flashing
red light (indicating an error). Pipes warmed.
  
ELOG V3.1.3-7933898