ID |
Date |
Author |
Subject |
Text |
 |
|
210
|
Wed Mar 31 15:40:42 2021 |
PJCS | Check Options files function added | There is a new operation available in the
System Functions menu in the System Wide
Checks page. ( The page needs to be reset |
|
|
211
|
Wed Mar 31 15:46:09 2021 |
PJCS | HowTo : Synchronize the ASIC clocks. | To Synchronize the FEE64 ASIC clocks to
rise at the same timestamp time use the System
function Synchronise the ASIC clocks |
|
|
220
|
Wed Apr 14 12:14:47 2021 |
PJCS | Corrections and changes | 14/4/21 @12:00 UK time
Corrected the failure of FADC re-calibration
All modules. The problem was a missing > |
|
|
221
|
Wed Apr 14 15:14:55 2021 |
PJCS | Note Well | Please note that for the forseeable future
the system function in the menu on the System
Wide Checks browser page labelled as |
|
|
300
|
Fri May 7 19:47:54 2021 |
PJCS | HowTo mitigate excessive temperature in an FEE64 | After tests in the Daresbury T9 system.
Disabling the waveform ADCs in
an FEE64 which is running with the FPGA over |
|
|
396
|
Wed Feb 16 10:32:52 2022 |
PJCS | AIDA software changes and some suggestions | The Pi console monitor program now outputs
a full date/time to file for each line of
report. |
|
|
490
|
Wed Jun 22 10:09:12 2022 |
PJCS | INFO: FEE64 supply Voltages | Study of the FEE64 power supply distribution
has yielded the following :-
The most sensitive regulator, as |
|
|
514
|
Wed Sep 14 19:07:07 2022 |
PJCS | INFO : Three Merger Statistics explained | There are three Merger statistics that
can be used to better understand how the
data flow through the Merger system is proceeding. |
|
|
225
|
Fri Apr 16 00:47:36 2021 |
OH-ML | AIDA DAQ Reset | 1:18 (CET) No rate in AIDA 07.
Oscar did a AIDA DAQ reset and
a quick wide check. |
|
|
1
|
Thu Sep 6 15:11:17 2018 |
OH, TD | Tuesday 4 September - Thursday 6 September | Attachment 1 - FEE64 temperatures illustrating
(we assume) a faulty PSU temperature sensor
for aida07
|
19x |
|
2
|
Thu Sep 6 15:25:19 2018 |
OH, TD | AIDA@DESPEC Setup Photos | |
|
|
3
|
Thu Sep 6 15:42:53 2018 |
OH, TD | FEE64 system console logs |
Kernel panics during boot cab be observed
in files ttyUSB1 and ttyUSB5
|
|
|
5
|
Wed Sep 19 12:40:41 2018 |
OH, TD | Single DSSD Installation and Tests | On the 18th of September we installed DSSD
3208-14
|
21x |
|
214
|
Sat Apr 10 12:15:54 2021 |
OH, TD | Saturday 10 April | 12:37 DAQ found crashed. Had crashed at some
time following 7:50am following previous
statistics update
|
16x |
|
291
|
Tue May 4 13:53:22 2021 |
OH, TD | Tuesday 4 May |
NH has re-checked alignment of AIDA adaptor
PCBs cf. Elog:290
|
7x |
|
341
|
Fri May 21 07:03:50 2021 |
OH, TD | Friday 21 May 08:00-16:00 | 08:00 OH and TD take over
Current settings p+n sides at 0xc
DSSD1 n+n sides 0x20 on ASICs 1-3 and |
9x |
|
353
|
Wed Jun 2 14:53:05 2021 |
OH, TD | Merger/Tapeserver issues after reboot of aida-3 | On 01/06/21 aida-3 was rebooted as the graphical
user interface had frozen.
We were also unable to connected to MIDAS |
6x |
|
442
|
Fri May 13 07:36:18 2022 |
OH, TD | Friday 13 May | 08:36 DAQ Still running smoothly
Currently on run R1_328
At current data rates we have just |
35x |
|
486
|
Mon Jun 20 09:26:23 2022 |
OH, TD | Monday 20 June | 10:26 MACB HDMI for AIDA03 and AIDA04 reseated.
Merger and DAQ restarted - All happily
sending data
|
11x |
|
307
|
Wed May 12 06:59:55 2021 |
OH, RDP | Wed 12 May 08:00-16:00 | 08:00 While the rates appeared to have quietened
around 7am they are back to their previous
rate now.
|
   |