AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DTAS, Page 6 of 8  ELOG logo
Entry  Mon May 27 07:18:26 2019, A. Algora, AIDA Plastik 
Nishimura san found the module used for the AIDA plastik in the last BRIKEN run. We put it instead of the TFA. It was kept in the same

setting conditions as in the earlier experiment.

AIDA plastic module configuration (Ortec 671 Amplifier)
Entry  Mon May 27 07:34:58 2019, A. Algora, HV status 
EMPTY00    │   0.0 V 1500.0 uA    0.5 V    0.5 uA Off               
│00.0000
DET01        │ 599.0 V 1500.0 uA  598.5 V   90.0 uA On                
Entry  Mon May 27 08:26:20 2019, A. Algora, Co source on top of Aida 201905271703_60Co_AIDA_in.root
Start: 16:28

Stop: 17:03

Rates: 14.4 kHz (Dinodes), 12.8 kHz (Anodes)
Entry  Mon May 27 09:12:22 2019, A. Algora, Background, with AIDA Screenshot_from_2019-05-27_17-50-45.pngScreenshot_from_2019-05-27_17-51-02.png201905271803_background_AIDA_in.root
Start: 17:13

Stop: 18:03

Rates: 1.8 kHz (Dynodes), 1.8 kHz Anodes
Entry  Mon May 27 12:54:48 2019, A. Algora, Reference detector Screenshot_from_2019-05-27_20-54-21.png
The reference detector was very noisy. The threshold has been changed from 2000 to 3200. We should look at this more carefully.

The position of the fiber was also changed. It moved too close to the Cs peak. Probably it has to do with the flexibility of the plastik screw
that fixes it.
Entry  Mon May 27 13:50:54 2019, A. Algora, preparation directory moved 
The preparation directory was moved to /data/DRikenPreparation (from its first place /home/dacq/DRikenPreparation) because the lack of space.
Entry  Mon May 27 13:58:03 2019, A. Algora, Background, no lmd, AIDA in, no sources around 201905281315_background_AIDA_in.root
Start: 21:59

Stop: 13:15 (next day, 28/05)

Rates: 1.9-2.3 (dynodes), 1.8-1.9 kHz (anodes)
Entry  Tue May 28 05:30:03 2019, A. Algora, Reference detector threshold changed Screenshot_from_2019-05-28_13-29-46.png
Again the Reference detector was firing mainly in the noise. Threshold increased to 5000 !!!!

New config file was saved, rate of Ref detector now: aprox. 400 kHz
Entry  Tue May 28 05:43:41 2019, A. Algora, 133Ba source. (IR052 from the third floor) 
Start: 13:43

Stop: 14:05

Comments: The source looks also too strong. Rates: 39 kHz (Dynodes), 40 kHz (anodes).
Entry  Tue May 28 06:13:27 2019, A. Algora, Background, Aida in 
Start: 14:13

Stop: 15:47

Lmd: Background_AIDA_in_201905281413_231.dlt .. Background_AIDA_in_201905281413_233.dlt
Entry  Tue May 28 07:49:51 2019, A. Algora, Background, no lmd 
Start: 15:50

Stop: 10:09 (next day)

Root: 201905291009_background_AIDA_in.root
Entry  Wed May 29 02:07:20 2019, A. Algora, Irradiation for 24Na 
Irradiation of two pieces of aluminium of 80 mm x 40 mm x 10 mm will be performed at the second BigRips dipole.

The pieces were placed with the help of Fukuda and Nishimura san.
Entry  Wed May 29 07:04:19 2019, A. Algora, F11 plastic connected 
F11 plastic has been connected to our dacq as it was in the earlier BRIKEN setting.

The detector cables were followed and connections checked.

The outputs of the detector (L and R) go through an Ortec TFA 474 with the following settings:
Entry  Wed May 29 08:29:26 2019, A. Algora, latest conf file 20190528_Conf.xlsx
Latest configuration file from 28 of May
Entry  Wed May 29 09:08:11 2019, A. Algora, AIDA plastic setting changed 
AIDA plastic changed.

HV was at 1000 V, it was changed to 1300 V

Coarse gain was changed to x5, Fg remains as it was.
Entry  Wed May 29 09:49:25 2019, A. Algora, New configuration file 20190529_Conf.xlsx
New configuration file after a few small changes in thresholds.
Entry  Wed May 29 10:01:10 2019, A. Algora, Background, no lmd, AIDA in 201905291833_background_AIDA_in.root201905300953_background_AIDA_in.root
Start: 18:02

Stop: 18:33, partial file (201905291833_background_AIDA_in.root), Stop: 09:53 saved also next day (30/05) 201905300953_background_AIDA_in.root

Comments: AIDA in, Bigrips is doing settings (105Te to my best knowledge). There is a lead wall before our setting. No apparent change in the
Entry  Thu May 30 02:03:33 2019, A. Algora, Ref detector position changed, AIDA plastik HV changed 20190530_Conf.xlsx
The position of the reference detector in the ancilliary card was changed from channel 10 to channel 9. Remember that this is important

in relation to all files saved until now.

The HV of AIDA plastic was changed from 1300 V back to 1000 V.
Entry  Thu May 30 02:44:56 2019, A. Algora, Background, just a test 20190530_Conf.xlsx
Start: 11:01

Stop: 13:37

Root: 201905301337_background_AIDA_in.root
Entry  Thu May 30 06:41:18 2019, A. Algora, Background with the beam 
There was an small increase in the counting rate of the detectors after the beam was delivered. The rate at
F7 was of the order of 5 kHz. But remember that there is a lead wall protecting the detector. Root file saved. Look at TSumA for example

Root: 201905301442_background_AIDA_in.root
ELOG V3.1.4-unknown