Sat Nov 11 23:30:25 2017, RG MB, Run 101
|
Header: CH2 #30+25, Beam energy 4.0 MeV/u
Ender:
====================================================
Run number : 101
Date : 12/11/2017
Start
time : 00:27
Stop time : 02:29
Trigger rate (MIDAS) : 6
Blocks written : 484
——————————————————————————————
Run
Scalars (approx rates)
0- Ungated : 4
1- Gated : 4
2- dE : 146151
3- E |
Sun Nov 12 01:28:26 2017, RG MB, Run 102 - Junk
|
Header: CH2 #30+25, Beam energy 4.0 MeV/u
Ender: Run stopped short as the RF cavity tripped
====================================================
|
Sun Nov 12 01:46:23 2017, RG MB, Run 103
|
Header: CH2 #30+25, beam energy 4.0 MeV/u
Ender: Run stopped as the beam went off. Noticed on the HI det scalar
====================================================
|
Sun Nov 12 04:22:57 2017, RG MB, Run 104
|
Header: CH2 #30+25, beam energy 4.0 MeV/u
Ender:
====================================================
|
Sun Nov 12 06:23:36 2017, RG MB, Run 105
|
Header: CH2 #30+25, beam energy 4.0 MeV/u
Ender: run ends OK to change beam energy from 4.0MeV/u to 4.29MeV/u
====================================================
|
Thu Nov 16 11:20:51 2017, RG CLW MD, Run 162 -HI det calibration
|
Header: 4alpha source - Heavy Ion detector calibration
Ender:
====================================================
|
Thu Nov 16 15:14:39 2017, RG CLW MD, Run 165 Pulse Walkthrough dE p-side
|
Header: Pulser walkthrough, dE p-side
Ender: positive pulser, fall time 1ms, delay 250 ns, 1.6 V in steps of 0.2, 250 Hz
====================================================
|
Thu Nov 16 15:24:42 2017, RG CLW MD, Run 166 Pulse Walkthrough dE n-side
|
Header: Pulser walkthrough, dE n-side
Ender: negative pulser, rise time 0.05 us, fall time 1ms, delay .5 us,ref INT, width 0.5 us, Amplitude: 7 on
dial, att. 5, in steps of 1 , 250 Hz
|
Thu Nov 16 15:33:24 2017, RG CLW MD, Run 168 Pulse Walkthrough E p-side
|
Header: Pulser walkthrough, E p-side
Ender: positive pulser, fall time 1ms, delay 250 ns, 6 V in steps of 0.5, 250 Hz
====================================================
|
Thu Nov 16 15:43:48 2017, RG CLW MD, Run 169 Pulse Walkthrough E n-side
|
Header: Pulser walkthrough, E n-side
Ender: negative pulser, rise time 0.05 us, fall time 1ms, delay .5 us,ref INT, width 0.5 us, Amplitude: 6 on
dial, no attenuation, in steps of 1 , 250 Hz
|
Thu Nov 16 15:57:29 2017, RG CLW MD, Run 170 Pulse Walkthrough HI p-side
|
Header: Pulser walkthrough, HI p-side
Ender: positive pulser, fall time 1ms, delay 250 ns, 10 V in steps of 1, 250 Hz
====================================================
|
Thu Nov 16 16:06:07 2017, RG CLW MD, Run 171 Pulse Walkthrough HI n-side
|
Header: Pulser walkthrough, HI n-side
Ender: negative pulser, rise time 0.05 us, fall time 1ms, delay .5 us,ref INT, width 0.5 us, Amplitude: 9 on
dial, no attenuation, in steps of 1 , 250 Hz
|
Thu Nov 16 16:14:04 2017, RG CLW MD, Run 172 Pulse Walkthrough HI n-side
|
Header: Pulser walkthrough, HI n-side
Ender: Change to digital pulser to check as non-linearity was observed at low amplitudes.
====================================================
|
Thu Nov 16 16:24:29 2017, RG CLW MD, Run 173 Pulse Walkthrough HI n-side
|
Header: Pulser walkthrough, HI n-side, trigger on HI detector this time
Ender: Change to digital pulser to check as non-linearity was observed at low amplitudes.
====================================================
|
Wed Nov 15 11:33:31 2017, RG CLW, Run 155 - calibration del E
|
Header: 4alpha source run for delE detector calibration
Ender:
====================================================
|
Wed Nov 15 15:41:41 2017, RG CLW, Run 157 - EJunk
|
Header: 4alpha source
Ender:
====================================================
|
Wed Nov 15 16:05:25 2017, RG CLW, Run 159 - E detector calibration
|
Header: 4alpha source run for E det calubration
Ender:
====================================================
|
Wed Nov 15 16:24:35 2017, RG CLW, Run 160 - strip numbering E detector
|
Header: 4alpha source off center (to the top)
Ender:
====================================================
|
Thu Nov 9 17:35:32 2017, RG, Run 67
|
Header: 4.7 MeV/u and CH2_30+25.
Ender:
====================================================
Run number : 67
Date : 09/11
Start time :
18:30
Stop time : 19:57
Trigger rate (MIDAS) : 10
Blocks written : 463
——————————————————————————————
Run
Scalars (approx rates)
0- Ungated : 10
1- Gated : 9
2- dE : 48
3- E |
Tue Oct 24 09:27:24 2017, Peter Black, DK, Safety words
|
Plate 1. Contents of chamber
|