ID Datedown Author Subject
  105   Wed Sep 24 21:35:27 2025 Emmanuel26AL New processing routine with FFT
  104   Thu Sep 18 14:11:29 2025 CLW, EONew processing

Comparison strip 6 amplitude spectra anew (red) vs (old) processing (Screenshot). Slight shift in amplitude and slightly less counts with new processing in t+alpha peaks (1% level). This could be either due to less noise background, shift in neutron energy calibration (amplitude spectra shown with a cut of 100 keV). pdf plot shows new vs old for neutron energy vs counts. only minor differences at low neutron energy, new processing makes spectra cleaner at higher energy.

Shape of the Li 240 keV resonance is shifted in new processing - either due to different offset in tof-ene calibration, or time stamp for signals in one of the processing versions is wrong. Need a new comparison to simulated spectra. 

  103   Tue Sep 9 13:59:29 2025 SelinUserInputs for Processing

This is the UserINputs for the processing of Ca41. The dead time used for the processing has been 2us. Further processing of a B run with dead time set to 0 (as seen in the inputs) showed that the dead time should be around 1 us.

  102   Mon Aug 25 10:47:44 2025 SelinPulser Test - Week 1 and 2

This is the pulser test between runs 221376 and 221377 (06/08/2025) which is in pink and 221439 and 221440 (13/08/2025) which is in cyan.

The back strips seem to be affected by gain as the pulser amplitudes are shifted.

  101   Thu Aug 21 16:19:43 2025 EmmanuelChecking DEED 11 with the Pulser

We checked DEED 11 with the Pulser and it was still unresponsive (See screenshot).

Next, we will change the LEMO cables and do another Pulser Test.

  100   Mon Aug 18 09:19:27 2025 EmmanuelDEED 11 NO Gamma Flash Signal

We observed that DEED11 no longer have Gamma flash signals.

Before the drop, it had some weird flunctuations.

Attached are the plots.

We will check with the pulser tomorrow.

 

 

  99   Wed Aug 6 11:20:52 2025 SelinPulser Test

The pulser test was performed with a frequency of 500 Hz and with voltage settings of: 2V, 1.6V, 1.2V, 0.8V, 0.6V, 0.4V, 0.3V, 0.2V, 0.1V, 0.05V, 0.025V.

The run was started with calibration settings and then the voltage was changed every 2.5-3 minutes.

From the attached plots, it can be seen that everything is as expected. This means that the pulser multipolarity is consistent with front and back strips, pulser signals are more frequent and the pulser amplitude decreases with decreasing V.

The front and back pulser tests have been done with separate runs and the settings have been reverted to their original values (100 Hz, 2V, pulser is off).

  98   Mon Aug 4 16:45:43 2025 SelinBoron-3 Consistency Check

These are the consistency checks done for Boron-3 sample.

Please note that the back strips for the first run are mismatched due to the settings in DAQ.

  97   Mon Aug 4 16:44:06 2025 SelinBoron-1 Consistency Check

These are the consistency checks done for Boron-1 sample.

  96   Mon Aug 4 16:32:56 2025 SelinLithium Consistency Check

These are the consistency checks done for Lithium sample.

Please note that in the earlier runs, the cables 22 and 23 were switched.

  95   Mon Jul 21 10:30:01 2025 SelinThe 2D Energy Spectrum of Ca41

The 2D energy spectrum of Ca41 (36 runs). The expected alpha energies are 4.8 MeV (alpha_0) and 2.7 MeV (alpha_1).

  94   Fri Jul 18 12:04:36 2025 CLWPresence at CERN

Excel sheet for presence at CERN during 41Ca run. Please update if required. 

https://docs.google.com/spreadsheets/d/19OzH7bpnHb-YeOskNBGlrhEkPZmDsPDw9y0tleU0Soo/edit?gid=0#gid=0

  93   Wed Jul 16 13:56:00 2025 CLWLinks to photos and proton counting excel sheet

Photos setup

Excel Sheet 

  92   Wed Jul 16 11:45:39 2025 SelinBoron Spectra Update After Setting Change + Pulser Removed

Uncalibrated amplitude spectra of Boron (run221188) after the settings were changed, the pulser was removed, and the threshold in the UserInput has been lowered. Both alphas can be seen in the front and back strips. The ratios between the alphas are similar for front (0.066) and back (0.063) strips.

1- Uncalibrated amplitude spectra of strip 2.

2- 1- Uncalibrated amplitude spectra of strip 19.

  90   Wed Jul 16 08:55:22 2025 SelinRun221141 Boron Uncalibrated Data

Plots of Boron #1 run (221141) from the Ca41 campaign with the right voltage applied (11.07.2025).

1- Uncalibrated amplitude spectra of strip 1 (dedicated).

2- Uncalibrated 2d energy plot of Boron dedicated data.

3- Uncalibrated amplitude spectra of strip 22 (dedicated). The alpha-0 on the back strips has lower counts.

  89   Wed Jul 16 08:51:06 2025 SelinRun221136 Lithium Uncalibrated Data

Plots of Lithium #3 run (221136) from the Ca41 campaign with the right voltage applied (11.07.2025).

1- Uncalibrated amplitude spectra of strip 1 (dedicated).

2- Uncalibrated amplitude spectra of strip 1 (parasitic).

3- Uncalibrated 2d energy plot of Lithium dedicated data.

  88   Sat Jul 12 11:34:31 2025 SelinDetector-Cable Connections

List of the detectors and the connections.

  87   Sat Jul 12 11:25:35 2025 Selin The Detector Bias-Leakage Current Data

This Excel sheet has the detector bias-leakage current data and plot for both detectors.

Detector 2 has lower leakage current and a flatter plateau region.

  86   Sat Jul 12 10:30:06 2025 Francisco García InfantesReport Experimental SetUp Ca41

I'm attaching a report for the experimental setup so far.

There is a main problem which haven't been solved yet and I don't know if that can cause problem in the future analysis.

Channels #23, #24, #31 and #32 are not working well. The signal os these channels look attenuated from the pulser, make impossible to see then in the Boron run.

As these channels are in the back side of the detector, it could be possible to run like? Unless we have more ideas about to solve the problem.

The front channels work well all of them, and the rest of the back channels work well as well.

I attach the current for the +15 and -15 volt applied to the PreAmp (low voltage) and the leakage current from the detector is 0.1 uA.

  85   Fri Jul 11 09:16:18 2025 TDSilena 7710 Quad Bias Supply manual and internal configuration
Manual - see attachment 1


Internal configuration

Voltage full scale - 400V/10uA

JP1/1-4 fitted
JP4 fitted *see note below
DVM card JP1 AB
HV jumpers AB

Current readout resolution - 10nA

DVM card JP2 LM
JP5/1-5 GH


Polarity - channels 1-4 negative

Note with channel 1 voltage set to -180V the output voltage measured by DMM is -40V.
The output impedance of the Silena 7710 outputs may be comparable to the input impedance of the DMM (DMM typically c. 20-40M) so the measured voltage will be a *lower* 
limit.

With JP4 removed (per manual instructions for voltage full scale 400V/10uA) the front panel reads -20V with an output voltage measured by DMM of -40V.
Again, the measured voltage will be a lower limit.

By observation of the behaviour of the electronic noise of p+n junction and n+n Ohmic strips as a function of bias for MSL type W1 DSSSD 3353-4 (75um) the voltage 
applied to the DSSSD is > depletion.
MSL QA tests report depletion voltage 28V and test/operating voltage 28+10V = 38V.