Entry  Wed Sep 24 21:35:27 2025, Emmanuel, 26AL New processing routine with FFT 29x
 
Entry  Thu Sep 18 14:11:29 2025, CLW, EO, New processing  Screenshot_2025-09-18_at_12.27.35.pngnewprocessing.pdf
Comparison strip 6 amplitude spectra anew (red) vs (old) processing (Screenshot). Slight shift in amplitude and slightly less counts with new processing
in t+alpha peaks (1% level). This could be either due to less noise background, shift in neutron energy calibration (amplitude spectra shown with a cut
of 100 keV). pdf plot shows new vs old for neutron energy vs counts. only minor differences at low neutron energy, new processing makes spectra cleaner
Entry  Tue Sep 9 13:59:29 2025, Selin, UserInputs for Processing UserInput_Silicons_Ca41.h
This is the UserINputs for the processing of Ca41. The dead time used for the processing has been 2us. Further processing of a B run with dead time set
to 0 (as seen in the inputs) showed that the dead time should be around 1 us.
Entry  Mon Aug 25 10:47:44 2025, Selin, Pulser Test - Week 1 and 2 pulserfront6.pdfpulserfront14.pdfpulserback27.pdfpulserback30.pdf
This is the pulser test between runs 221376 and 221377 (06/08/2025) which is in pink and 221439 and 221440 (13/08/2025) which is in cyan.

The back strips seem to be affected by gain as the pulser amplitudes are shifted.
Entry  Thu Aug 21 16:19:43 2025, Emmanuel, Checking DEED 11 with the Pulser DEED11_Pulser.png
We checked DEED 11 with the Pulser and it was still unresponsive (See screenshot).

Next, we will change the LEMO cables and do another Pulser Test.
Entry  Mon Aug 18 09:19:27 2025, Emmanuel, DEED 11 NO Gamma Flash Signal DEED11_18AUG2025_C.pngDEED11_17AUG2025_B.pngDEED11_17AUG2025_D.pngDEED11_17AUG2025.pngDEED11_18AUG2025.png
We observed that DEED11 no longer have Gamma flash signals.

Before the drop, it had some weird flunctuations.

Attached are the plots.
Entry  Wed Aug 6 11:20:52 2025, Selin, Pulser Test front_2V.pngfront_less2V.pngfront_less1V.pngback_2V.png
The pulser test was performed with a frequency of 500 Hz and with voltage settings of: 2V, 1.6V, 1.2V, 0.8V, 0.6V, 0.4V, 0.3V, 0.2V, 0.1V, 0.05V, 0.025V.

The run was started with calibration settings and then the voltage was changed every 2.5-3 minutes.

From the attached plots, it can be seen that everything is as expected. This means that the pulser multipolarity is consistent with front and
Entry  Mon Aug 4 16:45:43 2025, Selin, Boron-3 Consistency Check B3-1-8.pngB3-9-16.pngB3-17-24.pngB3-25-32.png
These are the consistency checks done for Boron-3 sample.

Please note that the back strips for the first run are mismatched due to the settings in DAQ.
Entry  Mon Aug 4 16:44:06 2025, Selin, Boron-1 Consistency Check B1-1-8.pngB1-9-16.pngB1-17-24.pngB1-25-32.png
These are the consistency checks done for Boron-1 sample.
Entry  Mon Aug 4 16:32:56 2025, Selin, Lithium Consistency Check Li1-8.pngLi9-16.pngLi17-24.pngLi25-32.png
These are the consistency checks done for Lithium sample.

Please note that in the earlier runs, the cables 22 and 23 were switched.
Entry  Mon Jul 21 10:30:01 2025, Selin, The 2D Energy Spectrum of Ca41 Screenshot_from_2025-07-21_08-51-09.png
The 2D energy spectrum of Ca41 (36 runs). The expected alpha energies are 4.8 MeV (alpha_0) and 2.7 MeV (alpha_1).
Entry  Fri Jul 18 12:04:36 2025, CLW, Presence at CERN 
Excel sheet for presence at CERN during 41Ca run. Please update if required. 

https://docs.google.com/spreadsheets/d/19OzH7bpnHb-YeOskNBGlrhEkPZmDsPDw9y0tleU0Soo/edit?gid=0#gid=0
Entry  Wed Jul 16 13:56:00 2025, CLW, Links to photos and proton counting excel sheet 
Photos
setup

Excel
Entry  Wed Jul 16 11:45:39 2025, Selin, Boron Spectra Update After Setting Change + Pulser Removed B_Strip2_ded.pngB_Strip19_ded.png
Uncalibrated amplitude spectra of Boron (run221188) after the settings were changed, the pulser was removed, and the threshold in the UserInput has been
lowered. Both alphas can be seen in the front and back strips. The ratios between the alphas are similar for front (0.066) and back (0.063) strips.

1- Uncalibrated amplitude spectra of strip 2.
Entry  Wed Jul 16 08:55:22 2025, Selin, Run221141 Boron Uncalibrated Data B_Strip1.pngB_ded_2d.pngB_Strip22.png
Plots of Boron #1 run (221141) from the Ca41 campaign with the right voltage applied (11.07.2025).

1- Uncalibrated amplitude spectra of strip 1 (dedicated).

2- Uncalibrated 2d energy plot of Boron dedicated data.
Entry  Wed Jul 16 08:51:06 2025, Selin, Run221136 Lithium Uncalibrated Data Li_Strip1.pngLi_Strip1_Par.pngLi_ded_2d.png
Plots of Lithium #3 run (221136) from the Ca41 campaign with the right voltage applied (11.07.2025).

1- Uncalibrated amplitude spectra of strip 1 (dedicated).

2- Uncalibrated amplitude spectra of strip 1 (parasitic).
Entry  Sat Jul 12 11:34:31 2025, Selin, Detector-Cable Connections Detector-Cable.pdf
List of the detectors and the connections.
Entry  Sat Jul 12 11:25:35 2025, Selin, The Detector Bias-Leakage Current Data Ca-VoltageCurrent.xlsx
This Excel sheet has the detector bias-leakage current data and plot for both detectors.

Detector 2 has lower leakage current and a flatter plateau region.
Entry  Sat Jul 12 10:30:06 2025, Francisco García Infantes, Report Experimental SetUp Ca41 Experimental_Setup_Ca41.pdfCurrent_Value_Voltage_-15.jpegCurrent_Value_Voltage_15.jpegLeakage_Current_Detector.jpeg
I'm attaching a report for the experimental setup so far.

There is a main problem which haven't been solved yet and I don't know if that can cause problem in the future analysis.

Channels #23, #24, #31 and #32 are not working well. The signal os these channels look attenuated from the pulser, make impossible to see
Entry  Fri Jul 11 09:16:18 2025, TD, Silena 7710 Quad Bias Supply manual and internal configuration Silena_7710_quad_bias_supply.pdf
Manual - see attachment 1