Entry  Mon Mar 25 12:57:30 2024, Clw, Adapters for ntof setup 20240325_125536.jpg20240325_125528.jpg

10 new Adapters for lemo to bnc on my desk

Entry  Thu Sep 12 11:07:29 2024, Nick, 40K Borrowed Items 

We borrowed from n_TOF:

1x O-ring for circular side-flanges on the chamber (the borrowed part has a small green marker cross on it)

10x LIMO-to-BNC connectors (BNC female) from electronics lab outside EAR1

Entry  Mon Sep 16 12:36:45 2024, TD, 34-way IDC to 37-way D connector ribbon cables 20240916_121057_(1).jpg20240916_121052_(1).jpg20240916_121047_(1).jpg20240916_121003_(1).jpg
The 2x sets of 34-way IDC to 37-way D connctor ribbon cables used at GSI for the two-alpha decay measurement have been located - see attachments 1-4.

They have been packed for shipment to CERN and delivered to stores for shipment. Awaiting quotation and P&M PO.
Entry  Sat Aug 12 09:24:19 2023, Nikolay Sosnin, 26Al Protons 

Here is a Google spreadsheet for counting protons for the runs:

https://docs.google.com/spreadsheets/d/1INX8G9GAu-M70SdZdz55qXnMZNXKXLW7xfFH-eaLuQw/edit?usp=sharing

Entry  Wed Sep 27 09:54:06 2023, Emmanuel, 10B sample in 20230926_113136.jpg20230926_113811.jpg20230926_115036.jpg20230926_115053.jpg20230926_111737.jpg

We switched from Aluminium target to Boron Target (number 1) on Tuesday (26th of Sept.). The Aluminium target was placed in the material room by the RP. We will change to LiF (number 3) on Thursday (28th of Sept.)

Entry  Fri Oct 16 16:11:41 2015, Sarah Lonsdale, 09/10/15 Photos of detectors (reversed dE wiring) and upper level frame for future set up, leakage current 2015-10-09_09.16.16.jpg2015-10-09_09.16.22.jpg2015-10-09_09.29.23.jpg2015-10-09_09.29.35.jpg2015-10-09_09.30.02.jpg

 Leakage before take down:
de 0.07 micro A
E 0.47 micro A

Entry  Fri Oct 16 16:01:41 2015, Sarah Lonsdale, 08/10/15 Removed source and sample, leakage current check 

Removed source and removed Li sample for blank frame test

Leakage
DE 0.06micro A
E 0.41 micro A

 

DAQ 
56.25 sample rate all except EBED 5,6,7

Entry  Fri Oct 16 15:56:01 2015, Sarah Lonsdale, 07/10/15 Alpha source in, change to pulser delay 6x

Alpha source in - containing 148Gd, 241Am, 239Pu, 244Cm (4x 1kBq)

Leakage on arrival
dE 0.08microA , 19V

E 0.48microA 132V

At leaving
dE 0.08 micro A
E 0.41 micro A

Pulser to 20ms (2015-10-07_13.58.08.jpg)


DAQ:

EBED 5,6,7 have post sample 16382
Increased further at 4pm - 16000+ post sample for 225 samples

 

 

Entry  Fri Oct 16 15:36:13 2015, Sarah Lonsdale, 06/10/15 Leakage, dE signal problem solving and external pulser trigger 6x

 Leakage tues am
0.03 microA dE
0.52 microA E

Before leaving at 11.30am or so-
Bias dE up to 20V, leakage to 0.07microA
E has 0.51 microA

Problem solving dE signals...
SIG l6d6 a has signal of 600mV
SIG 24 has signal of 500mV
White noise is comparable, but transient ringing on l6d6 could indicate higher capacitance

Testing dE for signal with alpha source- but removed before continuing run.
dE sees particles after rotating ribbon by 180 degrees (abnormal detector)

External pulser triggering (2015-10-06_11.30.jpg )
Pulsar trigger, +2.5V external from proton pulse (10ms before beam), lasts 1ms (width).

Convert pulse input from TTL to NIM to go into delay (15ms).
Start delay in dual timer, then output to trigger pulser (TTL)

Entry  Fri Oct 16 15:41:14 2015, Sarah Lonsdale, 06/10/15 Changes to DAQ and pulser settings 2015-10-06_delay.jpg2015-10-06_trig.jpg

 Pulser out of 5,6,7 EFED- extended to 32ms

Changed full scale on ADC to 0.2V,
Offset changed by factor of 10.
Zero suppression on, de -3mV thresh
E +-5mV thresh
Post sample 4096

5pm increased post sample for spd-02886 to 8192 as higher sample rate due to pick up.

Corrected EBED 4 offset.

(Again, all DAQ settings also recorded to the nToF erunbook)

Changed pulser to 13ms.

Entry  Mon Oct 12 11:50:52 2015, Sarah Lonsdale, 05/10/15 Pulser, Wiring information and DAQ initial set up 2015-10-05_15.19.54.jpg2015-10-05_15.20.24.jpg2015-10-05_15.22.39.jpg2015-10-05_15.22.46.jpg

600mV in 50ohm coax with 25 hz pulser (positive signal) - see 2015-10-05_15.19.54.jpg and 2015-10-05_15.22.46.jpg



Leakage currents on detectors:

In air
10v: 0.03 micro amp det 1 leakage -dE
132v: 0.47micro amp det 2 leakage- E det

In vacuum
DE 0.04 micro amp
E 0.46 micro amp

Same size of signal at ADC hence 50 ohms coaxial.

The following channel explanations refer to cables at control area (not vault) see (
2015-10-05_15.19.54.jpg) for full wiring information.


dE det= DEED 0 to -2V full scale, Lower limit 1800mV, Threshold 0V, thresh neg, pre 1024, post 2048, 20ms window, sample rate 225 MS/s
- L6D6 A = de 4
- L6D6 B = de 5
- L6D6 C = de 6
- L6D6 D = de 7
- LaBr 3 = de 8
- 21 = de 9

E det front= EFED 0 to -2V full scale, Lower limit 1800mV, Threshold 0V, thresh neg, pre 1024, post 2048, 20ms window, sample rate 225 MS/s
- 10 = Ef 3
- 12 = Ef 4
- 13 = Ef 5
- 14 = Ef 6
- 15 = Ef 7

E det back= EBED 0 to +2V full scale, Lower limit -200mV, Threshold 0V, thresh pos, pre 1024, post 2048, 20ms window, sample rate 225 MS/s (except EBED 5,6,7 which MUST be 900MS/s due to beam pick up signal)
- 16 = Eb 3
- 18 = Eb 4
- 20 = Eb 5
- 38 = Eb 6
- 40 = Eb 7

Run number 203385 on https://project-ntof.web.cern.ch/project-ntof/eRunbookEAR2.php

Entry  Mon Oct 12 12:04:51 2015, Sarah Lonsdale, 05/10/15 Photo of detector distances and layout 8x

Fairly self explanatory, note that images with time-stamps before 13.08.29 were taken before fitting the o-ring on the flange and therefore lateral positions will be different to photos after 13.08.29. Individual detector heights were not changed.

Entry  Mon Oct 12 12:07:24 2015, Sarah Lonsdale, 05/10/15 Changes to DAQ settings 

 Note that detailed DAQ information available from project-ntof.cern.ch erunbook ear2.

As a rough guide: 6/7 mV per MeV

5mV thresh back side E, -5mV front E
-5mV dE
Implemented from Run 203388

From Run 203389
-7mV in Deed 8&9, EFED
+7mV in EBED

Run 203390
Thresholds back to -6mV for deed 8&9

Gamma flash ends at 70micro sec, corresponds to neutron energy of ~200keV...

Entry  Mon Oct 12 11:27:02 2015, Sarah Lonsdale, 05/10/15 - Set up notes 8x

 20 micron (see 2015-10-05_12.46.59.jpg for detailed information) and 500 micron detectors

Testing preamps using pulser through common wire to both sets of preamps (see 2015-10-05_13.12.01.jpg).

64 x 14mA pos signals 15V
64 x 4mA neg signals 15V (see 2015-10-05_12.02.00.jpfor example on oscilloscope)

Sample has 105 microgram Li-6 (LiF - 95% enriched) 105 micron thick (2015-10-05_12.12.15.jpis Li-6 side up, 2015-10-05_12.12.26.jpis Li-6 side down... do not mix up!)

dE uses 22, 23 on ribbon connector, E 24, 25
dE from bias supply 2, E from bias 1

de and E pos signals needed to be inverted using jumpers. de negs not used, only 16ch (see 2015-10-05_14.59.12.jpg)


de first and last (0,15) are noisy- don't use. E detector ch 0 doesn't work.
Channels used dE 4-9, E(pos) 3-7 and E(neg) 3-7 (see
2015-10-05_15.11.37.jpg and 2015-10-05_15.11.44.jpg)

Entry  Sat Jul 12 11:25:35 2025, Selin, The Detector Bias-Leakage Current Data Ca-VoltageCurrent.xlsx

This Excel sheet has the detector bias-leakage current data and plot for both detectors.

Detector 2 has lower leakage current and a flatter plateau region.