Entry  Sun Sep 30 01:13:47 2018, OH, TD, Sunday 30 September 8x
09.10 DAQ continues OK - current file Sep18/R2_6

09.15 all system wide checks OK *except* nnaida7 fails ADC calibration
Entry  Sat Sep 29 06:25:23 2018, OH, TD, Saturday 29 September 18x
14:25 Moved from 8x DSSD setup to 6x DSSD
      Top two detectors removed.
      Detector alignments in snout unchanged so can use the 6xDSSD mark from June 2018
Entry  Fri Sep 28 02:08:07 2018, OH, TD, Friday 28 September 6x
10:08 Normal system restart ok
      System currently in 8DSSD configuration as was left in June 2018
Entry  Wed Aug 22 11:38:11 2018, OH, TD, AIDA GSI Setup Noise Tests 13x
Currently seems to be a problem with the memory check in the system wide checks that produces the message shown
- attachment 1
    Reply  Wed Aug 22 13:50:30 2018, OH, TD, Reply by PCS, AIDA GSI Setup Noise Tests 
> Currently seems to be a problem with the memory check in the system wide checks that produces the message shown
> - attachment 1
Entry  Tue Aug 21 12:29:44 2018, OH, TD, AIDA@DESPEC dhcpd.conf, dhcpd startup and log messages dhcpd.confdhcdConfStartup.txtmessagesLog.txt

Attachment 1 - DHCP configuration file /etc/dhcpd.conf
    Reply  Wed Aug 22 09:10:28 2018, OH, TD, AIDA@DESPEC dhcpd.conf, dhcpd startup and log messages 

> Attachment 1 - DHCP configuration file /etc/dhcpd.conf
Entry  Tue Jul 3 02:24:45 2018, OH, Tuesday 3 July 180703_1023_bias.png180703_1025_FEETemp.png180703_1406_bias.png180703_1406_stats.png
10:23 Bias and leakage currents ok - attachment 1
      Fee temperatures ok - attachment 2
      System wide checks ok
Entry  Mon Jul 2 02:36:41 2018, OH, Monday 2 July 180702_1036_bias.png
10:35 Bias and leakage currents ok - attachment 1
      FEE Temps ok - attachment 2
      System wide checks all ok
Entry  Sun Jul 1 08:25:24 2018, OH, Inventory 180701_Cable_Inventory.xlsx
Attached is an updated inventory of the cables and adapter boards we have in RIKEN.
Entry  Sun Jul 1 02:26:50 2018, OH, Sunday 1st July 180701_1025_bias.png
10:25 Bias and leakage currents ok - attachment 1
      FEE Temperatures ok
      System wide checks ok
Entry  Sat Jun 30 03:05:36 2018, OH, Saturday 30 June 180630_1104_bias.png180630_1327_BIAS.png180630_1335_Stats.png180630_8dssd_map.png180630_xandydssd8.png
11:05 Bias and leakage currents ok - attachment 1
      FEE temperatures ok
      System wide checks ok
Entry  Fri Jun 29 02:26:34 2018, OH, Friday 29 June 180629_1026_bias.png180629_1028_feetemp.png180629_1445_bias.png180629_1954_bias.png
10:27 Bias and leakage currents ok - attachment 1
      Fee temperatures ok - attachment 2
      Temperature 24.7C, Humdidity 44.5%, D.P +11.8C
Entry  Thu Jun 28 05:22:31 2018, OH, Analysis of R7 -> R9 lowPulserADC.pnghighPulserADC.pngalphaADC.pngalphaOld.pngExEyPulser.PNG
R7_0 -> R7_1 High Energy pulser run
R8_0 -> R8_1 Low energy puler run
R9_52 -> R9_62 Alpha run
Entry  Thu Jun 28 02:19:30 2018, OH, Thursday 28th June 180628_1018_bias.png180628_1249_bias.png180628_1922_FEE29ASIC2.png
10:18 Bias and leakage currents ok - attachment 1
      FEE Temperatures ok
      System wide checks all ok
Entry  Wed Jun 27 02:23:03 2018, OH, Wednesday 26 June 180628_1023_Bias.png180628_1023_Bias.png180626_1527_feetemps.png
10:20 Bias and leakage currents ok - attachment 1
      FEE temperatures ok
      System wide checks all ok
Entry  Tue Jun 26 05:44:44 2018, OH, Tuesday 26 June 180526_1345_Bias.png180626_2152_bias.png
04:12 TD found the nnaida14 appeared to have become noisy at some point. Biases/;eakages/temps ok
      ASIC check load appears to have fixed problem
      Noise appears to affect R9_0 to R9_51
Entry  Mon Jun 25 01:56:08 2018, OH, TD, Monday 25th June 21x
09:55 Alpha run stopped

10:45 Investigating the large noise in FEE28
Entry  Sun Jun 24 09:08:03 2018, 3OH, TD, June 2018 New Stack Configuration IMG_20180623_104349.jpgIMG_20180623_113315.jpg
Stack has been changed in configuration from a 6 DSSD stack to 8 DSSD stack.
What was detectors 1-6 are now 3-8 with 2 new detectors added to the front.
Entry  Sun Jun 24 01:40:08 2018, OH, TD, Sunday 24 June 2018 15x

09.38 DAQ continues OK
Entry  Sat Jun 23 02:44:36 2018, OH, TD, Saturday 23 June 2018 16x

      DSSSD stack dismounted yesterday to change configuration from 6x to 8x DSSSDs.
Entry  Sat Jun 23 02:28:13 2018, TD, OH, New fuses fitted to USB-controlled ac mains relay 2290195.pdf

The input fuses of the USB-controlled ac mains relay have been changed from 2x 5x20mm 250V/16A to 2x 5x20mm
250V/20A cartridge, time-lag Littelfuse 0215020.MXP
ELOG V3.1.4-unknown