Sun Sep 30 01:13:47 2018, OH, TD, Sunday 30 September 8x
|
09.10 DAQ continues OK - current file Sep18/R2_6
09.15 all system wide checks OK *except* nnaida7 fails ADC calibration
|
Sat Sep 29 06:25:23 2018, OH, TD, Saturday 29 September 18x
|
14:25 Moved from 8x DSSD setup to 6x DSSD
Top two detectors removed.
Detector alignments in snout unchanged so can use the 6xDSSD mark from June 2018
|
Fri Sep 28 02:08:07 2018, OH, TD, Friday 28 September 6x
|
10:08 Normal system restart ok
System currently in 8DSSD configuration as was left in June 2018
|
Wed Aug 22 11:38:11 2018, OH, TD, AIDA GSI Setup Noise Tests 13x
|
Currently seems to be a problem with the memory check in the system wide checks that produces the message shown
- attachment 1
|
Wed Aug 22 13:50:30 2018, OH, TD, Reply by PCS, AIDA GSI Setup Noise Tests
|
> Currently seems to be a problem with the memory check in the system wide checks that produces the message shown
> - attachment 1
>
|
Tue Aug 21 12:29:44 2018, OH, TD, AIDA@DESPEC dhcpd.conf, dhcpd startup and log messages  
|
Attachment 1 - DHCP configuration file /etc/dhcpd.conf
|
Wed Aug 22 09:10:28 2018, OH, TD, AIDA@DESPEC dhcpd.conf, dhcpd startup and log messages
|
>
> Attachment 1 - DHCP configuration file /etc/dhcpd.conf
>
|
Tue Jul 3 02:24:45 2018, OH, Tuesday 3 July   
|
10:23 Bias and leakage currents ok - attachment 1
Fee temperatures ok - attachment 2
System wide checks ok
|
Mon Jul 2 02:36:41 2018, OH, Monday 2 July
|
10:35 Bias and leakage currents ok - attachment 1
FEE Temps ok - attachment 2
System wide checks all ok
|
Sun Jul 1 08:25:24 2018, OH, Inventory
|
Attached is an updated inventory of the cables and adapter boards we have in RIKEN. |
Sun Jul 1 02:26:50 2018, OH, Sunday 1st July
|
10:25 Bias and leakage currents ok - attachment 1
FEE Temperatures ok
System wide checks ok
|
Sat Jun 30 03:05:36 2018, OH, Saturday 30 June    
|
11:05 Bias and leakage currents ok - attachment 1
FEE temperatures ok
System wide checks ok
|
Fri Jun 29 02:26:34 2018, OH, Friday 29 June   
|
10:27 Bias and leakage currents ok - attachment 1
Fee temperatures ok - attachment 2
Temperature 24.7C, Humdidity 44.5%, D.P +11.8C
|
Thu Jun 28 05:22:31 2018, OH, Analysis of R7 -> R9    
|
R7_0 -> R7_1 High Energy pulser run
R8_0 -> R8_1 Low energy puler run
R9_52 -> R9_62 Alpha run
|
Thu Jun 28 02:19:30 2018, OH, Thursday 28th June  
|
10:18 Bias and leakage currents ok - attachment 1
FEE Temperatures ok
System wide checks all ok
|
Wed Jun 27 02:23:03 2018, OH, Wednesday 26 June  
|
10:20 Bias and leakage currents ok - attachment 1
FEE temperatures ok
System wide checks all ok
|
Tue Jun 26 05:44:44 2018, OH, Tuesday 26 June 
|
04:12 TD found the nnaida14 appeared to have become noisy at some point. Biases/;eakages/temps ok
ASIC check load appears to have fixed problem
Noise appears to affect R9_0 to R9_51
|
Mon Jun 25 01:56:08 2018, OH, TD, Monday 25th June 21x
|
09:55 Alpha run stopped
10:45 Investigating the large noise in FEE28
|
Sun Jun 24 09:08:03 2018, 3OH, TD, June 2018 New Stack Configuration 
|
Stack has been changed in configuration from a 6 DSSD stack to 8 DSSD stack.
What was detectors 1-6 are now 3-8 with 2 new detectors added to the front.
|
Sun Jun 24 01:40:08 2018, OH, TD, Sunday 24 June 2018 15x
|
09.38 DAQ continues OK
|
Sat Jun 23 02:44:36 2018, OH, TD, Saturday 23 June 2018 16x
|
DSSSD stack dismounted yesterday to change configuration from 6x to 8x DSSSDs.
|
Sat Jun 23 02:28:13 2018, TD, OH, New fuses fitted to USB-controlled ac mains relay
|
The input fuses of the USB-controlled ac mains relay have been changed from 2x 5x20mm 250V/16A to 2x 5x20mm
250V/20A cartridge, time-lag Littelfuse 0215020.MXP |