Sun Jun 2 04:00:50 2019, TD, Sunday 2 June 21x
|
12.00 DAQ continues file May19/R15_169
ASIC settings 2019May19-17.54.03
|
Sat Apr 1 16:16:05 2017, OH, CG, Sunday 2 April   
|
00:12 BigRips started run 0223 (Started at 00:12)
AIDA file R20_344
System wide checks all ok
|
Sun Jul 1 02:26:50 2018, OH, Sunday 1st July
|
10:25 Bias and leakage currents ok - attachment 1
FEE Temperatures ok
System wide checks ok
|
Sun Nov 19 01:46:10 2017, CB, TD, Sunday 19 November 14x
|
10.45 Detector biases & leakage currents OK - see attachments 1
FEE64 Temperatures OK - see attachment 2
DAQ statistics OK - note *all* slow comparators 0xa - see attachment 3
|
Sun May 19 02:28:07 2019, OH, TD, CA, Sunday 19 May 11x
|
10:28 DAQ Continues ok
Alpha run continues - Currently on file R7_5
|
Sun Jun 18 02:56:43 2017, TD, Sunday 18 June 11x
|
10.56 Detector biases & leakage currents OK - see attachments 1 & 2
USB-controlled ac mains relay input #1 fuse blown - replaced
|
Sun Oct 16 03:14:41 2016, TD, Sunday 16 October 16x
|
11.12 DAQ start (R47)
Detector bias and leakage currents OK - see attachments 1 & 2
|
Sun Jun 16 04:09:40 2019, CA, CB, ML, TD, CG, Sunday 16 June noon-midnight 16x
|
12.10 DAQ continues OK file RIBF147/R2_145
100Sn setting, no degrader
|
Sat Jun 15 16:53:22 2019, CB, OH, DK, ML, Sunday 16 June 37x
|
00.30 AIDA file R2_26
DTAS run 007
|
Sun Jul 16 14:20:31 2023, TD, Sunday 16 July contd. 27x
|
15.20 All ERNI pins of aida01 adaptor PCB OK => FEE64 #1 ASIC #2 channel #5 issue faulty test C? faulty ASIC channel?
Fitted drain wire from aida06 & aida08 adaptor PCBs to FEE64 cooling plates
|
Sun May 14 01:52:49 2017, TD, OH, Sunday 14 May 16x
|
09.54 DAQ continues OK
Current file May2017/R7_292
|
Mon May 13 03:43:08 2019, CA, TD, OH, Sunday 12 May 14x
|
19.12 all system wide checks OK *except* nnaida7 fails ADC calibration
|
Sun Jun 12 03:05:37 2016, TD, AE, CG, MA, Sunday 12 June 2016 part 2 11x
|
BigRIPS switched from 60Ti to 64V
|
Sat Jun 11 17:24:11 2016, DK, TD, Sunday 12 June 2016 14x
|
Runs 41 and 42 are quite short. Run 41 was interrupted by the beam dropping (operators called), and run 42 we are just trying to get EURICA to
synchronize with AIDA DAQ.
1:08 Started AIDA Run 43 for 60Ti beam (EURICA run 2027; BigRIPS run 2132; CAEN Run 3018) |
Sat Nov 10 21:53:29 2018, CA, CB, TD, OH, Sunday 11 November 21x
|
06.53 DAQ continues OK - data transfer disabled
06.55 all system wide checks OK *except* nnaida7 fails ADC calibration
|
Sun Jun 11 08:23:23 2017, PV, Sunday 11 June 16:00-00:00 21x
|
16:40 beam lost - stripper foil problem
AIDA file RIBF128/R9_55
|
Sun Jun 11 00:07:24 2017, OH, Sunday 11 June 08:00-16:00 25x
|
08:10 BRIKEN and BigRIPS stops DAQs
AIDA on file R7_1287
|
Sun Dec 10 13:06:27 2023, TD, Sunday 10 December 14x
|
14.05 S4 closed
Water flow and temperature ( gauges outside S4 entrance ) OK
|
Sun Oct 1 12:38:56 2023, TD, Sunday 1 October 13x
|
12.50 Power up
All system wide checks OK
|
Sun Nov 6 01:34:15 2016, TD, OH, CG, Sun 6th November 18x
|
10:34 Detector Bias a leakage currents all ok. (Attachments 1 and 2)
Temperatures all ok (Attachment 3)
All system wide checks ok other than clock status see attachment 4
|