Sat Jun 30 03:05:36 2018, OH, Saturday 30 June    
|
11:05 Bias and leakage currents ok - attachment 1
FEE temperatures ok
System wide checks ok
Currently on file R9_74
Temperature 25.3C, Humidity 42.9%, D.P +12.1C
13:27 Bias and leakage currents ok - attachment 2
FEE Temperatures ok
System wide checks ok
Analysis of the pulser run R9_52 -> R9_74 has shown no major losses in terms of apparent broken bond wires.
The large gap in FEE64 29 can be explained by an ASIC dropping out: See https://elog.ph.ed.ac.uk/AIDA/730
Another odd trend appears in the channels relating to DSSD 8 on the x side. - attachment 5
The pattern shows one strip on one strip off for the first 32 strips.
nnaida18 is also the FEE which is running at a much higher rate during this run. - attachment 3
I'm not sure that I believe that these channels are actually relate to broken bond wires.
Here are the hit patterns for the 8 DSSD - attachment 4
Attached is also the x and y histograms for DSSD 8 which shows a weird pattern - attachment 5 |
Sat Apr 30 16:27:50 2016, TD, Saturday 30 April 2016 photographs 6x
|
Attachments 1-4 MSL type 2977-2- adaptor PCB configurations (looking downstream) L, D, R, U
Attachments 5-6 MSL type 2977-20 illustrating polyethyelene 'bagging' of 3M 1345-wrapped
ribbon cables |
Sat Apr 30 07:46:47 2016, CG,PV,TD, Saturday 30 April 2016 11x
|
MSL type BB18 serial # 2977-20 bias +100V I_L +12.45uA T=23 deg C
(biased by nnaida # 3 & 5 )
DSSSD - AIDA adaptor PCB cabling
4 off LH Coupler (Kapton PCB, 5cm), 2x 34-way Samtec ribbon cable (62cm p+n, 53cm n+n),
RH coupler(Kapton PCB, 10cm)
+ 3M 1245 1.4mil copper foil screen ribbon cables (i.e. not RH & LH couplers)
+ drain wires -> gold-plated Lemo-00 test input connectors
to FEE64s in 4x FEE module crates
3M 1245-wrapped cables enclosed by polyethyelene bags (supplied to deliver 90cm Printech Kapton PCBs)
to ensure no electrical contact with snout, DSSSD mounting tubes, snout base etc
Snout drain wire to FEE module Cu block
nnaida 3 & 8 AIDA adaptor PCB rev C
nnaida 5 & 9 AIDA adaptor PCB rev C (LK1 fitted)
ground links LK3 & LK7 *not* fitted to nnaida3, 5, 8, & 9 AIDA adaptor PCBs
LK1 fitted to nnaida5 ( cf. https://elog.ph.ed.ac.uk/AIDA/190 )
Heavy duty copper cable connects copper front end frames of FEE modules
Nitto 5011N conductive gasket between FEE module and front end frames
Standard ASIC settings
nnaida 3 & 8 - negative input
nnaida 5 & 9 - positive input
Pulser BNC PB-4
Fall time 1.0ms
Rate c.100Hz
Delay min
Amplitude 50,000
Polarity -
Pulse top Tail
Atten 10x
+ polarity via EG&G Ortec 433A Sum & Invert Amplifier
Standard ASIC settings
Shaping time 8us
LEC/MEC Slow com 32 (dec)
Fast com 32 (dec)
HEC Fast com 2(dec)
Attachments 1-4 ASIC #1 waveforms from each FEE64
5-8 ASIC #2 spectra + 2.0.L pulser peak width
Pulser peak widths (ch FWHM) versus shaping time ( spectrum 2.4.L )
tau nnaida3 nnaida5 nnaida8 nnaida9
8 68 60 70 62
4 83 49 88 48
2 100 47 109 43
1 172 60 165 42
0.5 380 86 309 48
Note: n+n ohmic strips spectra > 100 ch FWHM display non-gaussian peaks/double peaking
15.45 Attempted to increase detector bias from +100V to +150V
Leakage current increased ~0.5uA/V for bias > 115V
Detector bias supply trips (20uA) at +120V
Remain at detector bias +100V
16.17 Switch polarity of detector bias from + to -
Move LK1 from nnaida5 to nnaida3
tau nnaida3 nnaida5 nnaida8 nnaida9
8 61 64 65 61
n+n ohmic strips ~10% improvement, p+n junction strips no significant change
16.30 Fit LK3/LK7 jumpers to connect adaptor PCB ground to detector PCB ground
tau nnaida3 nnaida5 nnaida8 nnaida9
8 61 64 62 60
No significant change to cf. previous test
18.30 Install second DSSSD
Upstream 3058-6 bias +100V I_L 4.15uA
Downstream 2977-20 bias +100V I_L 11.8uA
+ 207Bi source ~3cm upstream of 3058-6
Standard ASIC settings
shaping time 6us
slow comp 8(dec)
HEC fast comp 2(dec)
LEC/MEC fast comp 16(dec)
Pulser peak widths ~60 ch FWHM per previous tests today
Discriminator masks TBD
22.27 R2 to disk (directory /data20/TapeData/May2016)
MERGE data links 2-9
Run data to disk overnight. Add two more DSSSDs tomorrow.
09.21 R2 stopped |
Sat Sep 3 05:54:14 2016, TD, Saturday 3 September 16x
|
14.20 R13 continues
DSSSD bias/leakage current OK - see attachments 1-2
FEE64 temperatures OK - see attachment 3
System-wide checks OK *except* nnaida19 fails ASIC timestamp check (no change)
Merger errors messages for Q 19 - see attachment 4
AIDA SYNC, PAUSE, RESUME, Correlation L, SYNC time warp, Good Events, AIDA disc, nnaida6, nnaida5 &
nnaida20 statistics - see attachments 5-15
Analysis of R13_317 - FEE64 data synchronised - see attachment 16
aidas1> ./analyser2 /TapeData/Aug2016/R13_317
*** GREAT format 3.2.0 analyser - TD - May 2014
*** ERROR: READ I/O error: 5002
blocks: 32000
ADC data format: 107206849 ( 490913.0 Hz)
Other data format: 153881151 ( 704640.2 Hz)
Sample trace data format: 0 ( 0.0 Hz)
Undefined format: 0 ( 0.0 Hz)
Other data format type: PAUSE: 134 ( 0.6 Hz)
RESUME: 134 ( 0.6 Hz)
SYNC100: 1997059 ( 9144.8 Hz)
FEE64 disc: 151882150 ( 695486.5 Hz)
MBS info: 1674 ( 7.7 Hz)
Other info: 0 ( 0.0 Hz)
ADC data range bit set: 1095 ( 5.0 Hz)
Timewarps: ADC: 0 ( 0.0 Hz)
PAUSE: 0 ( 0.0 Hz)
RESUME: 0 ( 0.0 Hz)
SYNC100: 0 ( 0.0 Hz)
FEE64 disc: 0 ( 0.0 Hz)
MBS info: 0 ( 0.0 Hz)
Undefined: 0 ( 0.0 Hz)
Sample trace: 0 ( 0.0 Hz)
Timestamp elapsed time: 218.383 s
FEE module #: 1 elapsed dead time 0.437 s
FEE module #: 2 elapsed dead time 0.298 s
FEE module #: 3 elapsed dead time 0.000 s
FEE module #: 4 elapsed dead time 0.000 s
FEE module #: 5 elapsed dead time 2.671 s
FEE module #: 6 elapsed dead time 1.429 s
FEE module #: 7 elapsed dead time 0.000 s
FEE module #: 8 elapsed dead time 0.000 s
FEE module #: 9 elapsed dead time 0.000 s
FEE module #: 10 elapsed dead time 0.000 s
FEE module #: 11 elapsed dead time 0.000 s
FEE module #: 12 elapsed dead time 0.000 s
FEE module #: 13 elapsed dead time 0.000 s
FEE module #: 14 elapsed dead time 0.129 s
FEE module #: 15 elapsed dead time 0.000 s
FEE module #: 16 elapsed dead time 0.000 s
FEE module #: 17 elapsed dead time 0.010 s
FEE module #: 18 elapsed dead time 0.000 s
FEE module #: 19 elapsed dead time 0.392 s
FEE module #: 20 elapsed dead time 0.000 s
FEE module #: 21 elapsed dead time 0.000 s
FEE module #: 22 elapsed dead time 0.021 s
FEE module #: 23 elapsed dead time 0.000 s
FEE module #: 24 elapsed dead time 0.000 s
FEE module #: 25 elapsed dead time 0.000 s
FEE module #: 26 elapsed dead time 0.000 s
FEE module #: 27 elapsed dead time 0.000 s
FEE module #: 28 elapsed dead time 0.000 s
FEE module #: 29 elapsed dead time 0.000 s
FEE module #: 30 elapsed dead time 0.000 s
FEE module #: 31 elapsed dead time 0.000 s
FEE module #: 32 elapsed dead time 0.000 s
*** Program elapsed time: 19.699s ( 1624.430 blocks/s, 101.527 Mb/s)
18.30 HDMI cabling to NIM MACB modules mechanically secured by 1U 'cable organiser' installed above MACB NIM bin
and by AIDA stand frame
ALL system-wide checks OK *except* nnaida19 fails ASIC timestamp check (no change)
18.55 R16 starts
ASIC settings 2016Sep02-18.17.01
Discriminator masks -> 0x0
Correlation scaler trigger delay 0x180 -> 0x5
Analysis of R16_0 - FEE64 data synchronised |
Sat Nov 3 00:37:29 2018, CA, CB, TD, OH, Saturday 3 November 9x
|
09.37 DAQ continues OK - data transfer disabled
FEE64 LEC/MEC slow comparator threshold 0xa
09.39 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK *except* nnaida17 c. 400k - see attachment 4
nnaida2 rate due to one hot channel as before
1.8.L spectra - see attachments 5 & 6
N.B. evidence of minimally ionising particles (presumably protons) from beam to F11 E detector
~10-15cm Pb bricks were placed within BRIKEN (CH2)n stack to stop heavy particles
1.8.W spectra 20us FSR - see attachments 7 & 8
10.10 LEC/MEC slow comparator from 0xa to 0x64
3x ASIC check load
DAQ stop
all histograms zero'd
enable data transfer
DAQ start - file Oct18/R16
beam background
good event statistics OK *except* nnaida2 20kHz - see attachment 9 |
Sat Jun 3 08:07:48 2017, OH, CB, AE, Saturday 3 June 16:00-24:00    
|
16:26 DAQ continues OK
all system wide checks ok, *except* nnaida1 clock status (6), nnaida1, 21 ADC calibration
Leakage currents OK - attachment 1
FEE64 temperatures OK - attachment 2
Good event statistics OK - attachment 3
No new merger errors - attachment 4
Sorted run R1_1540. DSSSD rates as follows
*** data items: 8958582 ( 115623.74 Hz)
*** ADC events: 954672 ( 12321.45 Hz)
*** time warps: 0 ( 0.00 Hz)
*** DSSSD # 1 count: 392147 old count: 361510 dt: 8.78 s LEC rate: 44646.94 Hz
*** DSSSD # 2 count: 557051 old count: 514085 dt: 8.78 s LEC rate: 63421.69 Hz
*** DSSSD # 3 count: 185346 old count: 172449 dt: 8.78 s LEC rate: 21102.12 Hz
*** DSSSD # 4 count: 154380 old count: 142180 dt: 8.78 s LEC rate: 17576.56 Hz
*** DSSSD # 5 count: 67195 old count: 62386 dt: 8.78 s LEC rate: 7650.32 Hz
*** DSSSD # 6 count: 131951 old count: 122163 dt: 8.78 s LEC rate: 15022.96 Hz
*** DSSSD # 1 count: 185 old count: 171 dt: 8.78 s HEC rate: 21.06 Hz
*** DSSSD # 2 count: 450 old count: 410 dt: 8.78 s HEC rate: 51.23 Hz
*** DSSSD # 3 count: 379 old count: 351 dt: 8.78 s HEC rate: 43.15 Hz
*** DSSSD # 4 count: 297 old count: 269 dt: 8.78 s HEC rate: 33.81 Hz
*** DSSSD # 5 count: 132 old count: 127 dt: 8.78 s HEC rate: 15.03 Hz
*** DSSSD # 6 count: 20 old count: 19 dt: 8.78 s HEC rate: 2.28 Hz
Pulser event still synchronised - see attachment 5
16:34 BigRIPS & BRIKEN stop runs
AIDA on file R1_1542
16:35 BigRIPS starts run 1120
BRIKEN starts run 109
AIDA on file R1_1543
Zero'd all SortSas
17:33 bigRIPS and BRIKEN stop daqs
AIDA on file R1_1556
17:34 bigRIPS starts run 1121
BRIKEN starts run 110
AIDA on file R1_1556
17:54 System wide checks all ok
19:50 bigRIPS starts run 1123
BRIKEN starts run 112
AIDA on file R1_1588
21:00 End of 78Ni run
last AIDA file R1_1606
Stopped AIDA DAQ
Refilled chiller
21:18 Performing pulser walkthrough.
Mistakenly started Tape server before starting DAQ. There will be no R2 run.
Starting R3 - pulser walkthrough.
Pulser frequency 250 Hz
Pulser amplitudes 90 000, 80 000, 70 000 ... 10 000
Can't see pulser peak in spectra. Issue unclear
21:40 Restarted high-amplitude pulser walkthrough run. Run RIBF127R1/R4_0
Pulser frequency 250 Hz
Pulser amplitudes 90 000, 80 000, 70 000 ... 10 000
Pulser peak now visible in histograms
Started low-amplitude pulser walkthrough run. Run RIBF127R1/R5_0
Pulser frequency 250 Hz
Pulser amplitudes 12 000, 10 000, ... 2 000
Set pulser amplitude back to 90 000, frequency to 25 Hz
23:17 Merger and Tape Server stopped
Links to FEEs disabled
All Browser tabs closed
Terminal sessions closed
23:25 Started R1 in /TapeData/RIBF128/
23:43 Start ASIC data readout
System wide checks ok
ASIC load check ok |
Sat Jun 3 00:00:11 2017, LJB, PW, TD, OH, Saturday 3 June 08.00-16.00 19x
|
08.29 BRIKEN 100; BigRIPS 1111 stopping run.
AIDA on run R1_1429
08.30 BigRIPS starts run 1112
BRIKEN starts run 101
AIDA on run R1_1430
09.08 System wide checks OK
FEE64 temperatures OK - attachment 1
Good events statistics OK - attachment 2
Leakage currents OK - attachment 3
No new merger errors - attachment 4
HEC x vs y plot OK - attachment 6
Zero SortSas
09.30 BRIKEN 101; BigRIPS 1112 stopping run. AIDA on run R1_1442
09.30 BigRIPS starts run 1113
BRIKEN starts run 102
AIDA on run R1_1443
09.55 Beam off
09.57 Beam on
10.32 BRIKEN 102; BigRIPS 1113 stopping run. AIDA on run R1_1457
10.33 BigRIPS starts run 1114
BRIKEN starts run 103
AIDA on run R1_1457
10.37 Online analysis of R1_1456, dead time below 0.3%, see attachment 7
11.08 System wide checks OK
FEE64 temperatures OK - attachment 8
Good events statistics OK - attachment 9
Leakage currents OK - attachment 10
No new merger errors - attachment 11
HEC x vs y plot OK - attachment 12
Zero SortSas
11.30 BRIKEN 103; BigRIPS 1114 stopping run. AIDA on run R1_1470
11.35 BigRIPS starts run 1115
BRIKEN starts run 104
AIDA on run R1_1472
12.25 BRIKEN 104; BigRIPS 1115 stopping run. AIDA on run R1_1483, BRIKEN timestamp reset
12.29 DAQ starts
AIDA file RIBF127R1/R1_1485
BigRIPS file 1116
BRIKEN file 105
DAQs synchronised
13.11 Analysis of R1_1484 - deadtime <0.5%, no timewarps, no MBS timewarps - see attachment 13
13.24 Beam lost AIDA file RIBF127R1/R1_1497
BRIKEN, BigRIPS DAQs stop
13.28 Beam returns AIDA file RIBF127R1/R1_1497
13.30 DAQ starts
AIDA file RIBF127R1/R1_1497
BigRIPS file 1117
BRIKEN file 106
13.54 Beam lost AIDA file RIBF127R1/R1_1504
13.57 Beam returns AIDA file RIBF127R1/R1_1505
14:30 BigRips and BRIKEN stop their daqs
AIDA on file R1_1513
14:31 BRIKEN starts file 107
BigRIPS starts file 1118
AIDA on file R1_1513
14:36 System wide checks ok
15.29 System wide checks OK
FEE64 temperatures OK - attachment 14
Good events statistics OK - attachment 15
Leakage currents OK - attachment 16
No new merger errors - attachment 17
HEC x vs y plot OK - attachment 18
Zero SortSas
15.31 BigRips 1118 and BRIKEN 107 stop their daqs
AIDA on file R1_1527
15:31 BRIKEN starts file 108
BigRIPS starts file 1119
AIDA on file R1_1527
14.57 Analysis of R1_1532 - deadtime <0.5%, no timewarps, no MBS timewarps - see attachment 19 |
Fri Jun 2 16:02:37 2017, CB, Saturday 3 June 00:00-08:00 28x
|
00:02 Noise in nnaida 1 and 2 appears to be reduced.
However, some channels in nnaida1, ASIC4 and nnaida2, ASIC 1 have been dropped. See attachment 1
00:05 Analysed RIBF127R1/R1_1314 - see attachment 2
No timewarps, SYNC100 OK, dead time negligible
No mismatches between dead time and idle time
00:07 Sorted RIBF127R1/R1_1314 - rates OK
*** DSSSD # 1 count: 35064 old count: 0 dt: 1.00 s LEC rate: 35055.03 Hz
*** DSSSD # 2 count: 49092 old count: 0 dt: 1.00 s LEC rate: 49079.44 Hz
*** DSSSD # 3 count: 16520 old count: 0 dt: 1.00 s LEC rate: 16515.77 Hz
*** DSSSD # 4 count: 14753 old count: 0 dt: 1.00 s LEC rate: 14749.23 Hz
*** DSSSD # 5 count: 7128 old count: 0 dt: 1.00 s LEC rate: 7126.18 Hz
*** DSSSD # 6 count: 13937 old count: 0 dt: 1.00 s LEC rate: 13933.43 Hz
*** DSSSD # 1 count: 19 old count: 0 dt: 1.00 s HEC rate: 19.00 Hz
*** DSSSD # 2 count: 36 old count: 0 dt: 1.00 s HEC rate: 35.99 Hz
*** DSSSD # 3 count: 44 old count: 0 dt: 1.00 s HEC rate: 43.99 Hz
*** DSSSD # 4 count: 39 old count: 0 dt: 1.00 s HEC rate: 38.99 Hz
*** DSSSD # 5 count: 17 old count: 0 dt: 1.00 s HEC rate: 17.00 Hz
*** DSSSD # 6 count: 2 old count: 0 dt: 1.00 s HEC rate: 2.00 Hz
Zero'd all histogram and SortSas
00:10 System console log ttyUSB7 (nnaida1) attached - 3
00:22 BigRIPS & BRIKEN stop runs
AIDA on file R1_1318
00:26 BigRIPS starts run 1103
BRIKEN starts run 92
AIDA on file R1_1319
Zero'd all SortSas
01:18 DAQ continues OK
FEE64 temperatures - OK
Good event statistics - OK
System wide checks - OK except clock status for nnaida1 (status=6), and ADC calibration for nnaida 1, 21
No new Merger error messages
HEC x vs y plot OK - attachment 4
01:26 BigRIPS & BRIKEN stop runs
AIDA on file R1_1333
01:30 BigRIPS starts run 1104
BRIKEN starts run 93
AIDA on file R1_1334
02:06 Finished compressing RIBF127R1/R1_6* and R1_7*
Beginning compression of RIBF127R1/R1_8* and R1_9*
Filesystem Size Used Avail Use% Mounted on
/dev/mapper/vg_aidas1-lv_root
50G 43G 3.9G 92% /
tmpfs 3.9G 1.1G 2.9G 27% /dev/shm
/dev/sda1 477M 77M 376M 17% /boot
/dev/mapper/vg_aidas1-lv_home
860G 580G 236G 72% /home
/dev/mapper/Data1-data10
7.7T 6.6T 824G 89% /data10
/dev/sdg2 7.3T 3.7T 3.7T 50% /ntfs
02:25 BigRIPS & BRIKEN stop runs
AIDA on file R1_1346
02:27 BigRIPS starts run 1105
BRIKEN starts run 94
AIDA on file R1_1347
02:31 DAQ continues OK
Sorted R1_1347. DSSSD rates are
*** data items: 1044352 ( 89746.26 Hz)
*** ADC events: 125937 ( 10822.38 Hz)
*** time warps: 0 ( 0.00 Hz)
*** DSSSD # 1 count: 36971 old count: 33324 dt: 1.11 s LEC rate: 33252.59 Hz
*** DSSSD # 2 count: 49630 old count: 45531 dt: 1.11 s LEC rate: 44638.40 Hz
*** DSSSD # 3 count: 16611 old count: 15180 dt: 1.11 s LEC rate: 14940.33 Hz
*** DSSSD # 4 count: 16456 old count: 15001 dt: 1.11 s LEC rate: 14800.92 Hz
*** DSSSD # 5 count: 7463 old count: 6781 dt: 1.11 s LEC rate: 6712.40 Hz
*** DSSSD # 6 count: 15199 old count: 13737 dt: 1.11 s LEC rate: 13670.34 Hz
*** DSSSD # 1 count: 23 old count: 19 dt: 1.11 s HEC rate: 20.69 Hz
*** DSSSD # 2 count: 53 old count: 45 dt: 1.11 s HEC rate: 47.67 Hz
*** DSSSD # 3 count: 45 old count: 43 dt: 1.11 s HEC rate: 40.47 Hz
*** DSSSD # 4 count: 41 old count: 36 dt: 1.11 s HEC rate: 36.88 Hz
*** DSSSD # 5 count: 20 old count: 18 dt: 1.11 s HEC rate: 17.99 Hz
*** DSSSD # 6 count: 3 old count: 1 dt: 1.11 s HEC rate: 2.70 Hz
HEC x vs y plot OK - cleared SortSas
02:40 Leakage currents OK - attachment 5
FEE64 temperatures OK - attachment 6
Good events statistics OK - attachment 7
No new merger errors - attachment 8
03:28 Analysed RIBF127R1/R1_1359 - see attachment 9
No timewarps, SYNC100 OK
Dead time OK
03:29 BigRIPS & BRIKEN stop runs
AIDA on file R1_1360
03:31 BigRIPS starts run 1106
BRIKEN starts run 95
AIDA on file R1_1361
03:48 HEC x vs y plot OK. Cleared SortSas.
04:04 DAQ continues OK
All system-wide checks OK, *except* nnaida1 clock status (=6) and nnaida1, 21 ADC time calibration
04:30 BigRIPS & BRIKEN stop runs
AIDA on file R1_1374
04:32 BigRIPS starts run 1107
BRIKEN starts run 96
AIDA on file R1_1375
04:34 Sorted R1_1375. DSSSD rates are as follows
*** data items: 1101465 ( 93446.58 Hz)
*** ADC events: 125547 ( 10651.21 Hz)
*** time warps: 0 ( 0.00 Hz)
*** DSSSD # 1 count: 44101 old count: 39716 dt: 1.12 s LEC rate: 39211.50 Hz
*** DSSSD # 2 count: 64123 old count: 57044 dt: 1.12 s LEC rate: 57013.65 Hz
*** DSSSD # 3 count: 18735 old count: 16851 dt: 1.12 s LEC rate: 16657.84 Hz
*** DSSSD # 4 count: 18739 old count: 16870 dt: 1.12 s LEC rate: 16661.40 Hz
*** DSSSD # 5 count: 7490 old count: 6811 dt: 1.12 s LEC rate: 6659.58 Hz
*** DSSSD # 6 count: 15970 old count: 14515 dt: 1.12 s LEC rate: 14199.40 Hz
*** DSSSD # 1 count: 26 old count: 24 dt: 1.12 s HEC rate: 23.12 Hz
*** DSSSD # 2 count: 43 old count: 39 dt: 1.12 s HEC rate: 38.23 Hz
*** DSSSD # 3 count: 50 old count: 46 dt: 1.12 s HEC rate: 44.46 Hz
*** DSSSD # 4 count: 40 old count: 36 dt: 1.12 s HEC rate: 35.57 Hz
*** DSSSD # 5 count: 17 old count: 16 dt: 1.12 s HEC rate: 15.12 Hz
*** DSSSD # 6 count: 7 old count: 6 dt: 1.12 s HEC rate: 6.22 Hz
04:36 Leakage currents OK - attachment 10
FEE64 temperatures OK - attachment 11
Good events statistics OK - attachment 12
No new merger errors - attachment 13
05:09 DAQ continues OK. Analysed RIBF127R1/R1_1383 - see attachment 14
No timewarps, SYNC100 OK
Dead time OK
Hit pattern OK - see attachment 15
05:28 Beam lost. BRIKEN & BigRIPS stopping run.
AIDA on run RIBF127R1/R1_1387
05:31 Beam is back. Issue unclear.
BigRIPS starts run 1108
BRIKEN starts run 97
AIDA on run R1_1388
05:37 Lost beam again. It appears to be an issue with "power"
AIDA on run RIBF127R1/R1_1389
BRIKEN & BigRIPS still running.
05:39 BigRIPS no longer sending signal: lost synchronisation.
05:43 Beam is back. Synchronisation is back.
BRIKEN & BigRIPS stopping run.
05:47 BigRIPS starts run 1109
BRIKEN starts run 98
AIDA on run R1_1391
05:53 Analysed R1_1391 - see attachment 16
No timewarps, SYNC100 OK
Dead time OK
DAQ continues OK
06:05 Sorted R1_1395. DSSSD rates as follows
*** data items: 1158578 ( 105681.80 Hz)
*** ADC events: 120781 ( 11017.26 Hz)
*** time warps: 0 ( 0.00 Hz)
*** DSSSD # 1 count: 54590 old count: 48655 dt: 1.12 s LEC rate: 48746.42 Hz
*** DSSSD # 2 count: 79085 old count: 70259 dt: 1.12 s LEC rate: 70619.36 Hz
*** DSSSD # 3 count: 24651 old count: 21385 dt: 1.12 s LEC rate: 22012.24 Hz
*** DSSSD # 4 count: 20753 old count: 18595 dt: 1.12 s LEC rate: 18531.50 Hz
*** DSSSD # 5 count: 8376 old count: 7452 dt: 1.12 s LEC rate: 7479.39 Hz
*** DSSSD # 6 count: 17321 old count: 15715 dt: 1.12 s LEC rate: 15466.88 Hz
*** DSSSD # 1 count: 27 old count: 24 dt: 1.12 s HEC rate: 24.11 Hz
*** DSSSD # 2 count: 43 old count: 38 dt: 1.12 s HEC rate: 38.40 Hz
*** DSSSD # 3 count: 41 old count: 35 dt: 1.12 s HEC rate: 36.61 Hz
*** DSSSD # 4 count: 38 old count: 33 dt: 1.12 s HEC rate: 33.93 Hz
*** DSSSD # 5 count: 14 old count: 14 dt: 1.12 s HEC rate: 12.50 Hz
*** DSSSD # 6 count: 1 old count: 1 dt: 1.12 s HEC rate: 0.89 Hz
06:19 Leakage currents OK - attachment 17
FEE64 temperatures OK - attachment 18
Good events statistics OK - attachment 19
No new merger errors - attachment 20
06:23 HEC x vs y plot OK - attachment 21
Zero'd all SortSas
06:27 Lost the beam. AIDA on run R1_1401.
Same "power-related" issue as before. Expected 2-3 minutes shutdown.
Lost synchronisation.
06:32 Beam is back. Synchronisation is back.
BRIKEN & BigRIPS stopping run.
06:33 BigRIPS starts run 1110
BRIKEN starts run 99
AIDA on run R1_1402
06:58 DAQ continues OK
HEC x vs y plot OK - attachment 22
Zero'd all SortSas
Analysed R1_1407 - attachment 23
No timewarps, SYNC100 OK
Dead time OK
07:06 Sorted R1_1407. Pulser events are still synchronised - attachment 24
07:30 BRIKEN & BigRIPS stopping run.
AIDA on run R1_1415
07:33 BigRIPS starts run 1111
BRIKEN starts run 100
AIDA on run R1_1416
07:35 Leakage currents OK - attachment 25
FEE64 temperatures OK - attachment 26
Good events statistics OK - attachment 27
No new merger errors - attachment 28
HEC x vs y plot OK
Zero'd SortSas |
Sat Dec 3 10:49:36 2016, TD, Saturday 3 December  
|
19.48 DSSSD bias & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
Ambient temperature +25.1 deg C, d.p. -3.0 deg C, RH 15.6% |
Sat Sep 29 06:25:23 2018, OH, TD, Saturday 29 September 18x
|
14:25 Moved from 8x DSSD setup to 6x DSSD
Top two detectors removed.
Detector alignments in snout unchanged so can use the 6xDSSD mark from June 2018
Rev F adapter cards have been put in place for all backside FEEs
All adapter boards are now Rev E or F
All adapter boards have an additional ground cable from the screw on the copper block to the lemo ground
connection
Power on and detectors biased (Cabling changed to reflect new 1-6)
Leakage and bias ok - attachment 1
System wide checks all ok
FEE temperatures ok - attachment 2
DAQ Started
Stats are looking very good - attachment 3
Rates look ok (Lin and Log) - attachments 4 & 5
Front side waveforms look ok - attachment 6
Backside waveforms look ok - attachment 7
Pulser widths are showing double peaking. - attachments 8 & 9
we are not sure if this is down to the pulser or not
Expanded timescales of the waveform, 0, 10, 100 and 1000 cycles
Layout 7 - attachments 10 - 14
Layout 8 - attachments 15 - 18
15:24 Pulser turned off in prep for alpha run
Slow comp set to 0x64
15.42 DAQ start file Sep18/R2
Test and calibration using background alpha particles
ASIC settings 2018Jun22-11.28.18
*except* slow comparator 0xa -> 0x64
BNC PB-4 pulser OFF
Merger c. 30k items/s TapeServer 230kb/s |
Sat Oct 29 03:31:03 2016, TD, Saturday 29 October 11x
|
11.30 DAQ continues (R87)
System-wide checks OK *except* SYNC errors & SYNC pulses - see attachments 1-4
11.36 DSSSD bias & leakage currents OK - see attachments 5-8
FEE64 temperatures OK - see attachment 9
16.46 DAQ continues (R87)
System-wide checks OK *except* SYNC errors & SYNC pulses - see attachments 5-8
18.55 DAQ stalled
Most recent file R87_274 last accessed 18.47
Restart merger
DAQ STOP
System wide checks
DAQ RESET & SETUP
System-wide checks
Disable H & data transfer #1
Restart merger & tapeserver
Enable H & data transfer #1
19.10 DAQ starts (R89) |
Sat Oct 28 03:33:10 2017, TD, OH, Saturday 28 October 13x
|
11.33 DAQ running OK - date transfer not enabled
BNC PB-4
Amplitude 90,000
x5 attenuator IN
t_r 100ns tau_d 50us
- polarity
frequency range 25Hz
delay MIN, tail pulse, INT ref
ASIC settings 2017Oct23-14.38.03
*all* FEE64s slow comparators 0xa *except* nnaida13 0xe, nnaida19-22 0xc
Detector biases & leakage currents OK - see attachments 1 & 2
FEE64 Temperatures OK - see attachment 3
DAQ statistics OK - see attachment 4
*all* system wide checks OK *except* nnaida21 fails ADC calibration
Julabo FL11006 recirculating chiller set point +20 deg C, actual +20.0 deg C
Ambient temperature +26.2 deg C, d.p. +6.2 deg C, RH 28.3%
14.40 Detector biases & leakage currents OK - see attachments 5 & 6
FEE64 Temperatures OK - see attachment 7
DAQ statistics OK - see attachment 8
*all* system wide checks OK *except* nnaida21 fails ADC calibration
14.50 DAQ starts
AIDA data file Oct17/R18
14.53 per FEE64 rates (20us/channel) - see attachment 9
No indication of c. 400us period noise as indicated by DAQ statistics attachment 8
15.46 DAQ stopped
DAQ start - data tranfser disabled
19.40 Detector biases & leakage currents OK - see attachments 10 & 11
FEE64 Temperatures OK - see attachment 12
DAQ statistics OK - see attachment 13
*all* system wide checks OK *except* nnaida21 fails ADC calibration |
Sat Oct 27 01:22:43 2018, CA, OH, Saturday 27 October   
|
09:22 DAQ found crashed. Appears to have crashed at 20:23 last night whilst writing R3_0 to file.
Run Control is non responsive
Leakage currents okay - attachment 1
Unable to ping any of the nnaida FEEs
Restarted merger in an attempt to clear links still no response.
Have powered down FEE64
Replaced blown fuse
Checked temperatures of FEEs, all okay - attachment 2
10:19 Restarted alpha run, write to file Oct18/R4.
System wide checks - all ok.
Slow comparator threshold changed to 100 from 10
14:12 FEE temperatures ok (attachment 3)
Bias/leakage currents ok (attachment 4)
System wide checks ok
Alpha run on R4_1
17.45 DAQ continues OK, file Oct18/R4_3 |
Sat May 27 05:38:39 2017, CB, TD, OH, Saturday 27 May 9x
|
13.38 DAQ continues OK
Current file May2017/R27_49
Detector biases; leakage currents OK - see attachment 1
FEE64 Temperatures OK - see attachment 2
good events; statistics OK - see attachment 3
No warnings/errors reported by merger since DAQ GO - see attachment 4
*all* system wide checks OK *except* ADC calibration of nnaida1, 21 and nnaida1 clock status (=6)
10.30 E17/F11 ambient temperature +23.8 deg C, RH 40.9%, d.p. +9.6 deg C
Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 75%)
13:40 Stopped DAQ running on alpha background calibration run
Current file May2017/R49
tyUSB7 log (nnaida1) attached
13:43 Detector bias OFF
14.16 DAQ starts
File May2017/R28
Pulser walkthrough
BNC PB-4
Amplitude 90,000-10,000 @ 10,000 step
x5 attenuator IN
t_r 100ns tau_d 50us
- polarity
frequency 250Hz range
delay MIN, tail pulse, INT ref
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator 0x0a
LEC/MEC fast comparator 0x20
See attachment 6
14.25 DAQ stop OK
14.27 DAQ starts
File May2017/R29
Pulser walkthrough
BNC PB-4
Amplitude 12,000-2,000 @ 2,000 step
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator 0x0a
LEC/MEC fast comparator 0x20
See attachment 7
14.34 DAQ stop OK
14.38 DAQ starts
File May2017/R30
Background, pulser OFF
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator 0x0a
LEC/MEC fast comparator 0x20
15.16 DAQ stop OK
May2017/R30_8
16:17 DAQ starts okay
current run May2017/R31_0
leakage current and temperature fine
see attachment 8 and 9
|
Sat Nov 26 12:18:25 2016, AE, Saturday 26 November, Part III 6x
|
19:30 Power cycled of AIDA reuired after Ge filling
20:04 Started new run: R10_0
BRIKEN run: 20161126_2005
BigRIPS run: 40Mg1011
ASIC settings: 2016Nov17-10.50.10
Degraders:
3mm Pb (in opening of PE shielding wall)
variable= 2 and 3 IN; 1.3mm W + 2 mm Al (positions confirmed when entering F11 to move Pb degrader downstream)
Pulser widths (ch 8 of ASIC 1 - see attachments):
P-N strips:
Integration of /tmp/nnaida1_1.8.L.msf title
Peak width = 63.64+/- 1.4850, Peak area = 4675.00+/- 70.59
Integration of /tmp/nnaida2_1.8.L.msf title
Peak width = 189.85+/- 2.4288, Peak area = 4585.00+/- 67.71
Integration of /tmp/nnaida5_1.8.L.msf title
Peak width = 70.83+/- 1.5868, Peak area = 5889.00+/- 80.70
Integration of /tmp/nnaida6_1.8.L.msf title
Peak width = 83.75+/- 0.7634, Peak area = 6194.00+/- 78.70
Integration of /tmp/nnaida9_1.8.L.msf title
Peak width = 73.87+/- 0.6702, Peak area = 6206.00+/- 78.78
Integration of /tmp/nnaida10_1.8.L.msf title
Peak width = 79.73+/- 0.9011, Peak area = 6215.00+/- 78.84
Integration of /tmp/nnaida13_1.8.L.msf title
Peak width = 96.67+/- 0.8988, Peak area = 6166.00+/- 78.52
Integration of /tmp/nnaida14_1.8.L.msf title
Peak width = 71.27+/- 1.3208, Peak area = 5701.50+/- 80.27
Integration of /tmp/nnaida17_1.8.L.msf title
Peak width = 265.98+/- 2.8845, Peak area = 5625.00+/- 75.00
Integration of /tmp/nnaida18_1.8.L.msf title
Peak width = 188.75+/- 2.3814, Peak area = 5920.00+/- 76.94
Integration of /tmp/nnaida21_1.8.L.msf title
Peak width = 80.82+/- 0.9310, Peak area = 5925.00+/- 80.03
Integration of /tmp/nnaida22_1.8.L.msf title
Peak width = 79.79+/- 1.1019, Peak area = 5748.50+/- 80.41
N-N strips
Integration of /tmp/nnaida3_1.8.L.msf title
Peak width = 77.05+/- 0.6356, Peak area = 7049.00+/- 83.96
Integration of /tmp/nnaida4_1.8.L.msf title
Peak width = 80.16+/- 0.6881, Peak area = 7048.00+/- 83.95
Integration of /tmp/nnaida7_1.8.L.msf title
Peak width = 81.78+/- 0.7073, Peak area = 7046.00+/- 83.94
Integration of /tmp/nnaida8_1.8.L.msf title
Peak width = 81.94+/- 0.8639, Peak area = 6921.00+/- 84.69
Integration of /tmp/nnaida11_1.8.L.msf title
Peak width = 84.16+/- 0.7746, Peak area = 6924.00+/- 84.66
Integration of /tmp/nnaida12_1.8.L.msf title
Peak width = 90.02+/- 0.7720, Peak area = 7059.00+/- 84.02
Integration of /tmp/nnaida15_1.8.L.msf title
Peak width = 90.08+/- 0.9683, Peak area = 6707.00+/- 84.93
Integration of /tmp/nnaida16_1.8.L.msf title
Peak width = 98.99+/- 0.9699, Peak area = 6807.50+/- 84.25
Integration of /tmp/nnaida19_1.8.L.msf title
Peak width = 138.33+/- 1.2900, Peak area = 6747.00+/- 82.14
Integration of /tmp/nnaida20_1.8.L.msf title
Peak width = 111.96+/- 1.3730, Peak area = 6848.00+/- 84.99
Integration of /tmp/nnaida23_1.8.L.msf title
Peak width = 106.07+/- 0.9108, Peak area = 6955.00+/- 83.40
Integration of /tmp/nnaida24_1.8.L.msf title
Peak width = 63.04+/- 0.5770, Peak area = 6978.50+/- 84.64 |
Sat Nov 26 07:11:34 2016, TD, DK, AE, Saturday 26 November, Part II
|
16.11 The cyclotron and BigRIPS condition are now restored and we have good beam
Count rate at F11 plastic is ~8 cps
At F8, they are focued on 41Al (D5 Brho = 7.5235 Tm)
At F11, they are focused on 40Mg (D8 Brho = 6.3208 Tm)
Presently around Run R9_19
3mm Pb is on the exit window of the first MUSIC at F11
All variable degraders were out (that is not what we intended to have...see next note below)
16.35 The variable degraders were not in the positions we thought. Now we make it as follows:
Variable degraders 1.0 mm W (+1 mm Al) and 0.3 mm W (+1 mm Al)
R9_24 just began
18:00 Beam stopped for stripper foil change
Moved Pb degrader downstream, from window of MUSIC1 detector to hole of polyethylene shield in front of BRIKEN.
18:12 Stopped AIDA DAQ at run R9_43 (Ge filling will start soon)
|
Fri Nov 25 15:56:36 2016, CG, DK, Saturday 26 November 11x
|
00.55 Bias and leakage currents ok (attachments 1+2).
Temps ok (attachment 3).
01.40 DAQ starts (RIBF03R1 - Run R1)
Pulser walkthrough
BNC PB_4 amplituides 90,000 - 10,000 @ 10,000 step
x5 attenuator IN
frequency 250Hz range
t_r 50ns tau_d 50us
polarity -, tail pulse, delay MIN
01.55 Ambient temperature +24.8 deg C, d.p. -2.2 deg C, RH 16.6%
02.05 DAQ starts (RIBF03R1 - Run R2)
Pulser walkthrough - same settings as above.
BNC PB_4 amplitudes 14,000 - 2,000 @ 2,000 step
02.48 Run R3 started.
Pulser off. Background run.
15cm Pb wall removed
6mm Pb degrader installed between MUSIC 1 & 2
variable degrader 1mm, 0.5mm & 0.3mm W + 1mm Al attached to each W sheet.
04.14 Noticed R3 stopped
05.40 Difficulty starting DAQ
'want first SYNC' from merger
lost contact with nnaida19 - restarting merger
TclHttpd server on nnrpi1 crashed - restarted remotely
unable to power cycle - need to check fuses when next obtain access
whilst waiting boot aidas1
F11 rate c. 450Hz
AIDA veto plastic c. 3200Hz
07.30 R7 started with beam.
07.38 Beam stopped.
R7 continues.
08.12 Beam returned to F11.
File R7_9
08.50 Beam stopped.
Start of R7_17.
Using implantdecay.f saw no HEC events in any DSSD. Removed 2mm of Pb degrader.
We re-insert all the variable Al degraders (for now) for safety. We can remove them again when the beam
comes back, as necessary.
Secondary beam current expected to remain at ~50cps.
Now no beam for ~1hr.
10.15 Beam comes back
There was a minor delay to get the beam back from some accelerator issue (SRQ issue I am told)
F11 plastic rate is about 10 cps, still I can see nothing in AIDA low gain
Main exp. team confirms several 40Mg events within minutes, and makes jokes about old Nature paper with 10 days of
data to do that.
10.25 Trying to find AIDA's implants, using MIDAS Sort in online mode
Removed the variable degraders, one-by-one. Still nothing? (this is now circa file R7_36)
Is there a possibility that the "low gain" setting of the ASIC is too high energy for something like 40Mg?
We try to check the implant rate with the AIDA veto plastic
During this time, we moved degraders in and out, several times, watching the oscilloscope and visual scalers by
video link in F11
10.39 We check some scalers of the dE before AIDA (including F11 x plastic behind AIDA)
We can see a reasonable effect as expected for these detector count rates
The count rate never exceeds about 50 to 60 cps at maximum for dE
I don't trust MIDAS Sort in online mode
it is known to show factor 10 or less of the real rate, previously)
Check the data file R7_37 -> may show 1 to 2 Hz
We should decrease the low gain (heavy ion) threshold or dynamic range for the ASICs?
But DK is not an expert of that, and doesn't want to cause AIDA to cease functionality...
11.03 Checking runs later than R7_37 (i.e. _38 to _42)
Never see anything. R7_37 must have been taken with some amount of degraders in
We surmise that the ions are mostly punching through AIDA, owing to low Z.
11.05 We put in the 0.5 W + 1 mm Al degrader back, to see if we can get back the condition of _37...
R7_44 after a short delay should have this condition.
There can be 1 cps or something, though with small statistics it is hard to say...
It seems to be definitely non-zero though, compared to some runs like _38 to _42
11.10 Add the 1.0 W + 1mm Al (with the 0.5 W and 1 mm Al) degrader
R7_45 begins ~400 MB before this time.
We can see some kinds of definite hit events **with low stats** (1 to 2 Hz in several detectors in some data blocks)
It seems to be, if all degraders are out, that the beam punches through AIDA, and dE is below the thresholds
Of course, the count rate at F11 is only 10 cps, so we cannot conclude much with strong confidence...
11.17 Remove all variable degraders again (near halfway point of R7_46)
We need a decent LISE++ file to get a clue of where these ions are stopping...!
11.42 Beam is stopped because one of the cyclotron RF systems is down.
This occurs around the middle of file R7_50
12.40 Still waiting for beam. I try a more definitive analysis of R7_45
Partial disk analysis of file R7_45:
First part of data look like this:
*** DSSSD # 1 count: 4 old count: 2 dt: 1.06 s HEC rate: 1.89 Hz
*** DSSSD # 2 count: 1 old count: 0 dt: 1.06 s HEC rate: 0.94 Hz
*** DSSSD # 3 count: 0 old count: 0 dt: 1.06 s HEC rate: 0.00 Hz
*** DSSSD # 4 count: 0 old count: 0 dt: 1.06 s HEC rate: 0.00 Hz
*** DSSSD # 5 count: 0 old count: 0 dt: 1.06 s HEC rate: 0.00 Hz
*** DSSSD # 6 count: 10 old count: 10 dt: 1.06 s HEC rate: 0.00 Hz
Notice that we get 10 total counts in DSSD#6, and practically 0 in any other layer (2 in DSSD#1)
Now, as the analysis progresses, we do not increase the counts in DSSD#6, but we start to see events in other layers
*** DSSSD # 1 count: 128 old count: 128 dt: 343.81 s HEC rate: 0.37 Hz
*** DSSSD # 2 count: 115 old count: 115 dt: 343.81 s HEC rate: 0.33 Hz
*** DSSSD # 3 count: 91 old count: 90 dt: 343.81 s HEC rate: 0.26 Hz
*** DSSSD # 4 count: 41 old count: 40 dt: 343.81 s HEC rate: 0.12 Hz
*** DSSSD # 5 count: 13 old count: 13 dt: 343.81 s HEC rate: 0.04 Hz
*** DSSSD # 6 count: 12 old count: 12 dt: 343.81 s HEC rate: 0.03 Hz
*** ENTRY finish ends
Notice how only 12 counts in DSSD#6 (two new counts) but 128 counts in DSSD#1 (126 new counts).
We can conclude that with 3 mm Pb (fixed), and 0.5 mm W and 1.0 mm Al, some of the beam is stopping in DSSD #6, and mostly
punches through.
Then, when 1.0 mm W and 1.0 mm Al are added, we shift the implantation to the beginning of AIDA (or some in the air
before AIDA).
So, we should try the 1 mm W (+1 mm Al) and 0.3 mm W (+1 mm Al) once the beam returns.
13.34 Stop run 7.
Change the medium/low energy selection on ASIC control from 0x0 to 0x1 (from low to medium energy mode)
(Already, the high energy setting is 0x2 and we cannot decrease it)
Start run 8. It looks okay, though of course now data rate becomes much smaller.
13.36 Beam is now on again. F11 rate has decreased 1 order of magnitude (now it is 0 to 2 Hz)
14.00 As expected, data rate becomes quite low, as the low energy mode has typically many events (10kHz or so)
But now we don't see anything in the high energy mode as before
3 mm Pb remains in, and variable degraders are all out (mistake...we meant for some to be in)
Strange! But the beam at F11 does not seem to be the same rate as before the cyclotron issues...
Realized that someone was checking the beam at an earlier focal plane and that F11 has no beam...
(Post-dated note: forgot to hit "Start ASIC Readout" --- duh oops.
RUN 8 IS JUNK, AS NO ASIC READOUT BY USER ERROR (DK)
14.12 Stop run 8. Change the medium energy mode back to low energy mode of all ASICs. Restart as run 9.
But the merger data links don't alternate colors when I reload the merger screen. Is it bad?
Data rate is still very low, but that doesn't make much sense.
Okay, now I start ASIC Readout! Woohoo!
But it seems some magnetic dipole of BigRIPS must have an improper setting, explaining the 1 cps rate at F11...
15.33 Bias and leakage currents OK (attachments 9 & 10).
FEE64 temperatures OK (attachment 11).
15.40 SRIM 2013
145MeV/u (5800MeV) 40Mg + 3mm Pb
sigma 0.55 deg
ang X (str) 0.90 deg
assume Pb - DSSSD distance c. 2m => 2000 x tan(0.55 deg) ~ 17mm rms
consider reducing Pb - DSSSD distance |
Sat May 25 06:27:12 2019, TD, Saturday 25 May 12x
|
14.32 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C
Water level c. 60% M-H markers
Temperature +25.6 deg C dew point +3.1 deg C RH 23.1%
DAQ continues OK to file May19/R11_3
alpha background
ASIC settings 2019May19-17.54.03
slow comparator 0x64
BNC PB-4 Pulser OFF
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachments 4
Merger OK - see attachment 5
TapeServer OK - see attachment 6
1.8.L spectra - see attachments 7 & 8
p+n FEE64s 1.8.W spectra 20us FSR - attachment 9
n+n FEE64s 1.8.W spectra 20us FSR - attachment 10
14.37 analysis of May19/R11_2 & R11_3 - see attachments 11 & 12
deadtime OK *except* nnaida1
ADC data for FEE64s nnaida5 & nnaida6 hot channels (Stat spectrum channel # 104)
Also observed at a much lower rate for nnaida1 & nnaida2
Statistics imply nnaida1 discriminators running at high rate for some part of R11_2 & R11_3
14.39 Check ASIC Control *all* ASICs & FEE64s |
Sat Apr 25 03:56:44 2015, JA, LHB, TD, AE & CG, Saturday 25 April 2015   
|
1130: Rotated forward water inlet connection to point towards rear of AIDA stand to remedy mechanical
conflict with EURICA support frame - see attachments. During drain and re-fill of Julabo FL1106
recirculating chiller no evidence of sediment in, or discoloration of, the water. Coolant loop closed
and tested OK.
CG installing (new, production batch) second and third MSL type BB18(DS)-1000 detectors to stack using
the new FCT reduced capacitance 0.9m, 68-way to 68-way, flexible Kapton PCBs - see attachments.
1200: Installed two new BB18 detectors (3131-11 and 3058-09) to nose mount.
Modified isolated kaptons labelled and number 1 and 5.
Source (207Bi) mounted upstream of first detector approx 5-10mm away.
Configuration - 3131-11 (#1) furthest upstream, 3058-09 (#2) in middle, 2977-20 (#3) at rear, followed by
10mm of Al.
1800: Cabled everything up.
3131-11 connected to nnaida10+6 (p-side) and nnaida11+16 (n-side). HV from single supply.
3058-09 connected to nnaida2+5 (p-side) and nnaida4+7 (n-side). HV from ch0 of quad supply.
2977-20 connected to nnaida1+14 (p-side) and nnaida3+8 (n-side). HV from ch1 of quad supply.
Typical resistance of new kaptons ~6.5 ohm.
Typical resistance of old kaptons ~3.2 ohm.
Makes sense if track width has been roughly halved, doubling resistivity.
At +100V bias:
- 3131-11 has leakage current 5.88 uA.
- 3058-09 has leakage current 3.82 uA.
- 2977-20 has leakage current 10.99 uA.
2000: Carried out brief pulser test of all connected modules.
Many spectra exhibit double-peaking of the pulser peak or otherwise non-Gaussian peaks.
Of the ones the have single Gaussian shaped peaks, many now have much larger FWHM than previously observed. |
Sat Jun 23 02:44:36 2018, OH, TD, Saturday 23 June 2018 16x
|
DSSSD stack dismounted yesterday to change configuration from 6x to 8x DSSSDs.
10.15 Overnight, the LHS input fuse (FEE64s 1-16) of the USB-controlled ac mains relay failed.
New fuses installed - see Elog https://elog.ph.ed.ac.uk/AIDA/720
10.25 Successful power-up
System-wide checks OK *except*
nnaida25 failed global clock check code 6 - see attachment 1
failed ADC calibration
Histogramming enabled for all FEE64s
Following DAQ GO *and* before START ASIC Readout only FEE64s 1-24 *and* 25 remain enabled
10.55 *all* FEE64 temperatures OK - see attachment 2
11.00 Stop DAQ
Start MERGE and TapeServer
Configure MERGE for 32x FEE64s
Histogramming *and data transfer #1 enabled for all FEE64s
Following DAQ GO *and* before START ASIC Readout only FEE64s 1-24 *and* 25 remain histogram enabled, *all*
FEE64s enabled for data transfer #1 - see attachment 3
11.44 Observe low leakage current (as expected) with detector bias 100V - see attachments 4 & 5
HV#2 reads zero which may be be too low - to be checked with DSSSD
18.15 June18/R4
Histogramming not available nnaida26-32
Pulser data
ASIC settings 2018Jun22-11.28.18
*all* slow comparators 0xa
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
t_r 100ns, tau_d 1ms
frequency c. 2Hz
polarity +
delay MIN, ref INT, tail pulse
detector biases & leakage currents OK - see attachments 6 & 7
ambient temperature +25.4 deg C, d.p. +10.1 deg C, RH 37.6%
FEE64 temperatures OK - see attachment 8
DAQ statistics - nnaida22, 25, 28, 29 & 30 high - see attachment 9
MERGE/Tapeserver - see attachments 10 & 11
18.28 DAQ stop (final file R4_4)
Analysis of R4_0 - note lots of HEC data for nnaida25
*** TDR format 3.3.0 analyser - TD - June 2017
*** ERROR: READ I/O error: 5002
blocks: 32000
ADC data format: 258702070 ( 1412508.5 Hz)
Other data format: 2377771 ( 12982.6 Hz)
Sample trace data format: 0 ( 0.0 Hz)
Undefined format: 0 ( 0.0 Hz)
Other data format type: PAUSE: 66 ( 0.4 Hz)
RESUME: 66 ( 0.4 Hz)
SYNC100: 2234733 ( 12201.6 Hz)
FEE64 disc: 142906 ( 780.3 Hz)
MBS info: 0 ( 0.0 Hz)
Other info: 0 ( 0.0 Hz)
ADC data range bit set: 4319574 ( 23584.8 Hz)
Timewarps: ADC: 0 ( 0.0 Hz)
PAUSE: 0 ( 0.0 Hz)
RESUME: 0 ( 0.0 Hz)
SYNC100: 0 ( 0.0 Hz)
FEE64 disc: 0 ( 0.0 Hz)
MBS info: 0 ( 0.0 Hz)
Undefined: 0 ( 0.0 Hz)
Sample trace: 0 ( 0.0 Hz)
*** Timestamp elapsed time: 183.151 s
FEE elapsed dead time(s) elapsed idle time(s)
1 0.000 0.000
2 0.000 0.000
3 0.000 0.000
4 0.000 0.000
5 0.000 0.000
6 0.000 0.000
7 0.000 0.000
8 0.000 0.000
9 0.000 0.000
10 0.000 0.000
11 0.000 0.000
12 0.000 0.000
13 0.000 0.000
14 0.016 0.000
15 0.000 0.000
16 0.000 0.000
17 0.000 0.000
18 0.000 0.000
19 0.000 0.000
20 0.000 0.000
21 0.000 0.000
22 0.245 0.000
23 0.022 0.000
24 0.000 0.000
25 0.738 0.771
26 0.000 0.000
27 0.028 0.000
28 0.023 0.000
29 0.000 0.000
30 0.186 0.194
31 0.364 0.000
32 0.246 0.000
*** Statistics
FEE ADC Data Other Data Sample Undefined Pause Resume SYNC100 Disc MBS Other
HEC Data
0 0 0 0 0 0 0 0 0 0 0
0
1 3222303 76527 0 0 0 0 69859 6668 0 0
569
2 2941948 76748 0 0 0 0 69857 6891 0 0
0
3 5670829 75859 0 0 0 0 69850 6009 0 0
0
4 2125453 74279 0 0 0 0 69858 4421 0 0
0
5 2136921 77632 0 0 0 0 69862 7770 0 0
0
6 2391640 73715 0 0 0 0 69858 3857 0 0
0
7 5890950 74230 0 0 1 1 69858 4370 0 0
4
8 3513119 75980 0 0 0 0 69864 6116 0 0
11
9 2263483 74984 0 0 0 0 69859 5125 0 0
0
10 2586263 77464 0 0 0 0 69857 7607 0 0
0
11 1010024 75142 0 0 0 0 69860 5282 0 0
0
12 1589585 74212 0 0 0 0 69857 4355 0 0
0
13 1683394 76604 0 0 0 0 69859 6745 0 0
0
14 6156870 74846 0 0 4 4 69856 4982 0 0
25
15 4365448 74530 0 0 0 0 69856 4674 0 0
0
16 4573069 75434 0 0 0 0 69864 5570 0 0
0
17 2440883 73800 0 0 0 0 69858 3942 0 0
5
18 3555638 78294 0 0 0 0 69853 8441 0 0
0
19 1011281 77659 0 0 0 0 69859 7800 0 0
0
20 1348970 74278 0 0 0 0 69856 4422 0 0
123
21 3718102 79600 0 0 0 0 69862 9738 0 0
0
22 31353026 76209 0 0 10 10 69765 6424 0 0
10
23 7467427 75889 0 0 3 3 69851 6032 0 0
15
24 5627550 75520 0 0 0 0 69855 5665 0 0
0
25 18997744 69607 0 0 13 13 69581 0 0 0
4318786
26 4444974 69852 0 0 0 0 69852 0 0 0
0
27 7927708 69857 0 0 3 3 69851 0 0 0
6
28 24389685 69867 0 0 9 9 69849 0 0 0
9
29 52059614 69858 0 0 0 0 69858 0 0 0
0
30 17131235 69796 0 0 4 4 69788 0 0 0
3
31 12650414 69742 0 0 11 11 69720 0 0 0
8
32 12456520 69757 0 0 8 8 69741 0 0 0
0
*** Timewarps
FEE ADC Pause Resume SYNC100 Disc MBS Undefined Samples
0 0 0 0 0 0 0 0 0
1 0 0 0 0 0 0 0 0
2 0 0 0 0 0 0 0 0
3 0 0 0 0 0 0 0 0
4 0 0 0 0 0 0 0 0
5 0 0 0 0 0 0 0 0
6 0 0 0 0 0 0 0 0
7 0 0 0 0 0 0 0 0
8 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0
10 0 0 0 0 0 0 0 0
11 0 0 0 0 0 0 0 0
12 0 0 0 0 0 0 0 0
13 0 0 0 0 0 0 0 0
14 0 0 0 0 0 0 0 0
15 0 0 0 0 0 0 0 0
16 0 0 0 0 0 0 0 0
17 0 0 0 0 0 0 0 0
18 0 0 0 0 0 0 0 0
19 0 0 0 0 0 0 0 0
20 0 0 0 0 0 0 0 0
21 0 0 0 0 0 0 0 0
22 0 0 0 0 0 0 0 0
23 0 0 0 0 0 0 0 0
24 0 0 0 0 0 0 0 0
25 0 0 0 0 0 0 0 0
26 0 0 0 0 0 0 0 0
27 0 0 0 0 0 0 0 0
28 0 0 0 0 0 0 0 0
29 0 0 0 0 0 0 0 0
30 0 0 0 0 0 0 0 0
31 0 0 0 0 0 0 0 0
32 0 0 0 0 0 0 0 0
*** MBS Timewarps
FEE
0 0
1 0
2 0
3 0
4 0
5 0
6 0
7 0
8 0
9 0
10 0
11 0
12 0
13 0
14 0
15 0
16 0
17 0
18 0
19 0
20 0
21 0
22 0
23 0
24 0
25 0
26 0
27 0
28 0
29 0
30 0
31 0
32 0
*** Program elapsed time: 21.492s ( 1488.913 blocks/s, 93.057 Mb/s)
19.10 June18/R5
Histogramming not available nnaida26-32
alpha background
ASIC settings 2018Jun22-11.28.18
*all* slow comparators 0x64
BNC PB-4 Pulser OFF |
|