Mon Jun 6 13:46:45 2016, TD, ML & YS, /var/log/messages
|
aidas1 log |
Mon Jun 6 11:17:30 2016, TD, ML & YS, Re-configure merger to ignore nnaida19 
|
19.00 Power-cycled and rebooted AIDA FEE64s
|
Mon Jun 6 08:57:34 2016, GL,PW, YS, TD, ML, AIDA run 1112 - Sync problem
|
After starting run 1112, we noticed that the merge program was waiting for a first sync. It then turned out that all syncs had disappeared. This required
an entire reboot of the AIDA DAQ. Consequently, AIDA run 1112 should not be used. |
Mon Jun 6 08:51:50 2016, TD, Change of set point for Julabo FL11006 recirculating chiller 
|
The Julabo FL11006 recirculating chiller set point has been changed from +20.0 deg C to +15.0 deg C.
|
Sun Jun 5 17:09:49 2016, CG, DK, AE, Monday 6th June 20x
|
00.45 Run R1100 started on n-rich setting (EURICA run_1100, BigRIPS run 1165)
Degraders as noted in previous entry
Check clock status failed for nnaida5. ADC calib failed for nnaida8. nnaida19 failed |
Sun Jun 5 02:14:10 2016, TD, AE, CG & DK, Sunday 5 June 2016 8x
|
10.10 Detector biases & leakage currents - OK - see attachments 1-2
10.40 Detector biases & leakage currents - OK - see attachments 3-4
|
Sun Jun 5 01:32:44 2016, TD, DK, GL, AE, PJW, First RIB implantation: Run 67
|
We restarted the merger (Stop, Setup & Configure, Go, Pause) and then we could restart the DAQ as normal. Rate is ~500 Hz. |
Sun Jun 5 01:20:39 2016, DK, AE, GL, PJW, Merger wants first SYNC
|
Trying to start the AIDA DAQ now that the RI beam is (nearly) ready. However, we cannot get merging to work. But the merger status has "Merge
State = GOing : paused : xfer enabled :
want first SYNC" but the last part ought to show "output paused" or after merge begins, "ready for SYNC". We restarted |
Sat Jun 4 08:46:20 2016, TD, AIDA noise - state of play 2 27x
|
Attachments 1-23 AIDA FEE64s nnaida1-nnaida23 ASIC #1 waveforms
|
Sat Jun 4 04:53:25 2016, TD, AE, CG, DK, GL YL & PJW (Team AIDA), Saturday 4 June 2016 8x
|
12.27 DAQ run R61 started 01.30
Appears to have stopped abruptly at ~06.55
All FEE64 AIDA SYNCs zero
|
Fri Jun 3 03:10:55 2016, TD, AE, CG & DK, Friday 3 June 2016 12x
|
11.00 Detector biases and leakage currents - see attachments 1-2
FEE64 temperatures - see attachment 3
|
Thu Jun 2 09:04:57 2016, TD, [How to] Monitor AIDA
|
Every 4h
1) Record DSSSD #1 = #6 detector bias voltages and leakage currents in written logbook
|
Thu Jun 2 01:33:13 2016, TD, AE, CG & DK, Thursday 2 June 2016    
|
09.30 DAQ left running overnight
AIDA SYNCs on all FEE64s this morning
|
Wed Jun 1 01:33:58 2016, TD, Wednesday 1 June 2016 9x
|
09.30 Detector biases and leakage currents - ambient temp +24.9 deg C - attachments 1-2
10.08 Connection lost to nnaida27 and nnaida28
|
Wed Jun 1 09:00:13 2016, TD - Reply PJCS, Wednesday 1 June 2016
|
> 09.30 Detector biases and leakage currents - ambient temp +24.9 deg C - attachments 1-2
>
> 10.08 Connection lost to nnaida27 and nnaida28
|
Wed Jun 1 13:58:48 2016, TD - Reply PJCS, Wednesday 1 June 2016   
|
> Check the Local Controls for the setting of the LMK3200 control rgister.
> If you recall from the problem with nnaida5 the value should be 0xB.
|
Wed Jun 1 12:14:08 2016, Patrick Coleman-Smith, [HowTo] React to "Check Clock Status => FEE64 Failure"
|
If there are more than one failures then check how they map to the MACB HDMI connections.
Are they all connected to the just one MACB ? => Check the HDMI cable at the MACB Layer Port.
Are they all connected to several MACBs which have a common "Parent" ? => Check the HDMI cable at the "Parent"
|
Wed Jun 1 07:49:45 2016, TD, Detector noise - state of play 10x
|
Screenshots of ASIC#1 waveforms for each side of each of the DSSSD #1 - #6 are attached.
Standard ASIC settings - tau = 8us
|
Wed Jun 1 01:32:04 2016, TD, Co/Bi source run (R19) 
|
09.18 R19 (AIDA) stopped
AIDA SYNC lost for most (but not all) FEE64s - see attachments 1-2
|
Tue May 31 16:03:24 2016, AE. VP, Co/Bi source run (R17) 8x
|
Set discriminator mask for DSSD2 through DSSD6 (see attachments). Fast comparator threshold set to 9 for all DSSDs.
New source run for beta-gamma correlation, now with all DSSDs included in EURICA trigger.
|
Tue May 31 08:06:43 2016, TD, AE & CG, Tuesday 31 May 2016 7x
|
Attachment 3 - AIDA discriminator with pulser ~5Hz
standard ASIC settings (tau=8us, slow comparator 0x10, LEC/MEC comparator 0x9)
|
Mon May 30 18:29:54 2016, AE, Co/Bi source run (R7)    
|
Some comments regarding source run for beta-gamma calibration, to complement EURICA16 elog entry:
http://ribf-exp.riken.jp/elog/EURICA+U-campaign+2016/
|
Mon May 30 22:57:58 2016, AE, Co/Bi source run (R7)
|
> Some comments regarding source run for beta-gamma calibration, to complement EURICA16 elog entry:
> http://ribf-exp.riken.jp/elog/EURICA+U-campaign+2016/
>
|