363
|
Thu Oct 18 00:31:55 2018 |
AT & AT | 252Cf long measure, z=0cm, threshold 5% lower |
18/10/18 START: 08:31
18/10/18 STOP: 10:50
** Use of other configuration file:
181016_Conf_BrikenFull_2153_test_thr.xlsx
ROOT file: 181018_1050_252Cf_z0cm_lowThr.root
DLT:
191017_0831_252Cf_z0cm_lowThr_014.dlt
191017_0831_252Cf_z0cm_lowThr_015.dlt
191017_0831_252Cf_z0cm_lowThr_016.dlt
Comment:
- 252Cf source placed at the center using cardboard piece, z=0cm, orientation of the source with its axis parallel to the beam axis, i.e. axis of the source upstream to the beam.
- 3He counters with a threshold 5% lower compared to previous measurement.
- Clovers inside the moderator, surface position 11.6/2 cm from the each one.
- AIDA not inside the moderator.
- Without in beam PE shielding. |
365
|
Thu Oct 18 08:21:18 2018 |
AT2, JLT, SN | 252Cf long measure, z=0cm, no clovers |
18/10/18 START: 17:22
18/10/18 STOP: 20:09
** Use of other configuration file:
181016_Conf_BrikenFull_2153_test_thr.xlsx
ROOT file: 181018_1722_252Cf_z0cm_2010.root
DLT:
181018_1722_252Cf_z0cm_021.dlt
181018_1722_252Cf_z0cm_022.dlt
181018_1722_252Cf_z0cm_023.dlt
Comment:
- 252Cf source placed at the center using cardboard piece, z=0cm, source axis perpendicular to
the beam axis.
- Clovers WITHOUT the moderator.
- AIDA not inside the moderator.
- Without In beam PE shielding. |
361
|
Wed Oct 17 09:20:20 2018 |
AT & AT | 252Cf long measure |
18/10/17 START: 18:16
18/10/17 STOP: 22:00
ROOT file: 181017_1816_252Cf_z0cm_2200.root
DLT:
181017_1816_252Cf_z0cm_000.dlt
181017_1816_252Cf_z0cm_001.dlt
181017_1816_252Cf_z0cm_002.dlt
181017_1816_252Cf_z0cm_003.dlt
Comment:
- 252Cf source placed at the center using cardboard piece, z=0cm, orientation of the source similar to past year
measurements, i.e. axis of the source perpendicular to the beam axis.
- Clovers inside the moderator, surface position 11.6/2 cm from the each one.
- AIDA not inside the moderator.
- Without in beam PE shielding. |
338
|
Sat Oct 21 17:14:55 2017 |
JA, ATD, JLT | 252Cf efficiency calibration |
Measurement of BRIKEN neutron counter efficiency
252Cf source in the (approximately) center
Wasabi and D4 clover inserted. G7 on the floor.
YSO on the box with light pulser.
The source was plug in a Styrofoam piece at the right height
so that it could be centered longitudinally with the D4 Clover
and transversely with the Wasabi nose.
Possible position difference with previous efficiency
measurements using the cardboard source holder
should not impact the efficiency obtained.
Start: 00:41
Stop: 01:08
Run number: 36
List mode: 171022_0041_252Cf_036.dlt
Root file: 171022_0041_0108_252Cf.root
|
Attachment 1: Screenshot_from_2017-10-22_01-21-51.png
|
|
48
|
Wed Jul 20 11:40:00 2016 |
J.L. Tain, J. Agramunt, R. Baeturia, A. Tarifeņo, A. Tolosa, ... | 252Cf |
First measurement with full 140 tube version
of the BRIKEN neutron detector
Directory: /data/160716Test01/
Configuration: 160718Conf_BrikenFull.xlsx
File: 160718_FullBrikenFirstData_252Cf_002.dlt
Online: 160718_FullBrikenFirstData_252Cf.root
All tubes and channels working!
Sample of screenshots included. |
Attachment 1: RIKEN-252Cf-V1A1.png
|
|
Attachment 2: UPC-252Cf-V1A3.png
|
|
Attachment 3: ORNL1-252Cf-V1A5.png
|
|
Attachment 4: ORNL2-252Cf-V1A6.png
|
|
50
|
Wed Jul 20 12:00:15 2016 |
J.L. Tain, J. Agramunt, A. Tarifeņo, A. Tolosa, ... | 252Cf |
Measurement with the source at the center
Listmode: 160719_BRIKEN_252Cf_009.dlt
Online: 160719_BRIKEN_252Cf.root
BRIKEN: 1.03697E7
Pulser: 5985
Rate: 17326 n/s |
51
|
Wed Jul 20 16:07:58 2016 |
A.Tarifeņo, J.L. Tain, J. Agramunt, A. Tolosa, ... | 252Cf |
19/Jul/2016
Scanning of the efficiency as a function of longitudinal position
File:160719_BRIKEN_252Cf_XXmm_*.dlt
Online:160719_BRIKEN_252Cf_XXmm.root
Run: 011-025
Distances: -275,-175,-125,-75,-50.-25,+25,+50,+75,+125,+175,+225,+275,+375mm |
54
|
Wed Jul 20 16:50:12 2016 |
J.L. Tain, J. Agramunt, A. Tarifeņo, A. Tolosa, S. Go... | 252Cf |
Measurement of the 252Cf source at the central position
Efficiency and neutron-gamma correlations
Listmode: 160719_2233_252Cf_XX.dlt, Run: 029-030
Online: 160719_2233_252Cf_029_030.root |
65
|
Thu Oct 6 14:50:30 2016 |
JL Tain, A Tolosa, A Tarifeno,... | 252Cf |
First measurement with 252Cf source to check the calibration.
AIDA was in place so the source was at the right height
but displaced about 10cm from the center.
We use the last configuation file used in July:
160718Conf_BrikenFull_cal5.xlsx
tmeas=25min
file: 161006_2235_252Cf.root
The noise is rather small.
The calibration of July is nearly OK! |
Attachment 1: 252Cf-1st.pdf
|
|
Attachment 2: 161006_2235_252Cf.pdf
|
|
84
|
Thu Oct 13 23:41:25 2016 |
JL Tain et al | 252Cf |
New BRIKEN energy calibration
We check the 3He tubes energy calibration with 252Cf source
We Found that only one tube He006 was slightly shifted and correct it.
Check measurement:
File: 161014_0113_252Cf.root
See figure |
Attachment 1: 252Cf-161014.pdf
|
|
102
|
Thu Nov 3 04:46:24 2016 |
Alvaro Tolosa | 252Cf |
Check of BRIKEN calibration and thresholds
Source more or less at the center
Online File: 161103_1319_252Cf.root |
Attachment 1: 161103_1319_252Cf.root
|
155
|
Tue Nov 22 10:17:57 2016 |
ATS, PV, LIU | 252Cf |
Measurement with 252Cf at Z=22.5cm from the moderator border, opposite side from AIDA.
Clovers and Veto detectors are turned off.
Beam is off.
Files:
161122_1820_001.dlt
161122_1820_002.dlt
161122_1820_001.root
Obs: There is a problem in the online data for this run. DAQ for VME1 do not stored data. This problems seems to
be a crash on the DAQ. I restarted the system and run again the measurement. The display for voltages has lost
connection. The channels on VME1 have been checked, no problems is observed in the waveforms.
Again, this time without observing problems:
File:
161122_2208_003.dlt
161122_2208_003.root
BRIKEN spectrum is attached for this run. Counts in the pulser are 1.123e4 |
Attachment 1: 161122_2208_003_BRIKEN.svg
|
|
158
|
Thu Nov 24 14:36:27 2016 |
AT | 252Cf |
Check of calibrations and rates for neutron detectors with Conf File 161107Conf_BrikenFull.xlsx
Measurement 161122_2208_003.root has been used to check the calibrations and counting rates in the BRIKEN
detector. Condition of the measurement:
252Cf at 22.5cm from moderator, opposite sede from AIDA
RESULTS
Calibrations factors:
Calibrations of the neutron detectors are reasonable in the current conditions. A check detector by detector
(see attached file) shows that the full peak for most of the counters is around 764keV. Only 3 counters are
shifted more than 10keV:
He134 741.901
He135 750.312
He140 748.587
The impact on the counting rates would not be serious.
Counting rates (n/pulser):
#########################################
Total neutron rate: 1516.28
#########################################
BY RING:
#########################################
R1 359.057
R2 222.997
R3 386.299
R4 253.833
R5 143.434
R6 90.6357
R7 60.025
The total current counting rate is lower than rate obtained in Oct 12th (total 1550.91 n/pulser). The difference
is due mainly to a decrease in the counting rates for rings 1 to 4. |
Attachment 1: CalibrationCheck_161122.xlsx
|
197
|
Thu Mar 23 10:03:51 2017 |
JL Tain, J Agramunt, V. Phong, C. Griffin, O. Hall, S. Nishimura | 252Cf |
252Cf source at the center for calibration
23/03/2017
Start: 19:02
Stop: 19:38
Root file: 170323_1902_1938_252Cf.root |
Attachment 1: 170323_1902_1938_252Cf.pdf
|
|
202
|
Mon Mar 27 09:32:55 2017 |
JLT JA CG OH DK TD RC VP GK LJ SN | 252Cf |
252Cf at the center for BRIKEN calibration
Start: 17:31
Stop: 18:00
File: 170327_1800_252Cf.root
3He tube calibration done with script BrikenCal.cpp
Result stored in 170327_1800_252Cf_Cal.ods (attached)
After visual superposition of all spectra we decided
to change only the parameters of He003 and He133
in the new calibration file. |
Attachment 1: 170327_1800_252Cf.pdf
|
|
Attachment 2: 170327_1800_252Cf.root
|
Attachment 3: 170327_1800_252Cf_Cal.ods
|
301
|
Thu May 25 13:42:45 2017 |
AT, PC,KR,JLT,AA | 252Cf |
252Cf source in Z=0cm position.
Start:21:41
Stop:21:52
Files
170525_2141_252Cf_z0cm_006.dlt
170525_2141_2152_252Cf_z0cm.root
We realized that the time stamp was not reset. |
302
|
Thu May 25 13:56:00 2017 |
AT, JLT, AA | 252Cf |
252Cf source in Z=0cm position.
Start:21:57
Stop: 22:26
Files
170525_2157_252Cf_z0cm_008.dlt
170525_2157_2226_252Cf_z0cm.root
The time stamp was reset finally (at the third try !!!)
There is a file 170525_2154_252Cf_z0cm_007.dlt which was taken for a short time, until we noticed again that the
time stamp was not reset. |
93
|
Tue Oct 18 09:36:07 2016 |
Alvaro Tolosa | 2016-Oct, Logbook |
Paper logbook for October-2016
For correct visualization,
please download it. |
Attachment 1: 16Oct_logbook.pdf
|
|
94
|
Tue Oct 18 09:41:43 2016 |
Alvaro Tolosa | 2016 - July, Logbook |
Paper logbook for July-2016
For correct visualization,
please download it. |
Attachment 1: 16Jul_logbook.pdf
|
|
324
|
Tue Jun 13 06:50:12 2017 |
Briken Colaboration | 152Eu calibration |
After experiment 152Eu calibration
Working directory: /data/1706FinalCalibrations
Configuration file: 170604_Conf_BrikenFull_AfterCrisis.xlsx
DLT file name: 170613_1451_152Eu_000.dlt
Start: 14:51
Stop: 15:24
Online rootfile:152Eu_cal_z0_1451_1524_170613.root
COMMENTS:
1. Source centered in both BRIKEN and clovers (faced towards clover D4)
2. No AIDA in
3. D4 clover pulsers are at 253, 260, 265, and 271 keV
4. G7 clover pulsers are beyond 5 MeV. No pulser for crystal G7Green |
Attachment 1: 170604_Conf_BrikenFull_AfterCrisis.xlsx
|