12
|
Sun May 3 09:04:07 2015 |
J.L. Tain, J. Agramunt, G. Kiss | Summary of neutron measurements |
U: (Up: 2 RIKEN tubes at 40cm from AIDA)
D: (Down: 12 RIKEN tubes at 80cm from AIDA)
US: (2 ORNL 2" tubes close to the floor at 1.8m from AIDA)
Natural background:
File: 150426_Bckg_1425_1805.root
Rate per tube (n/s): U=0.0071, D=0.0045, US=0.0222
Notes:
-This rate seems quite small but we should remember
that we are underground.
-The fact that U is larger than D could be explained as
related that the amount of PE around tubes is larger.
-For comparison:
BELEN-20@JYFL=0.6n/s (no beam), 0.9n/s (beam)
BELEN-30@GSI=1.2n/s (beam)
BELEN-48@JYFL=1.5n/s
110Zr setting:
File:150501_16111810_110Zr.root
Rate per tube (n/s): U=0.483, D=0.139, US=0.292
Ratio to background: U=68.3, D=31.1, US=13.2
Individual tube distribution:
U1 0.280 D1 0.087 D7 0.084 US1 0.132
U2 0.249 D2 0.092 D8 0.084 US2 0.188
D3 0.083 D9 0.077
D4 0.075 D10 0.070
D5 0.073 D11 0.069
D6 0.065 D12 0.054
Notes:
-There is huge increase of with beam
-The drop of increase with distance is consistent
with either beam induced or decay of implants
110Zr setting:
File: 150501_1733_1900_110Zr.root (actually 02/05)
Rate per tube (n/s): U=0.869, D=0.242, US=0.516
Ratio to 1st 110Zr: U=1.80, D=1.74, US=1.77
Notes:
-We don't quite understand the reason for the near factor two
difference since apparently the settings are the same and beam was
rather stable
In order to investigate the point of origin of the neutrons
we add extra PE covering D5,D6,D11,D12 from the top and
D1,D2,D7,D8 from the front
88Ge setting:
File: 150502_21502333_88Ge.root
Rate per tube (n/s): U=1.271, D=0.349, US=0.706
Ratio to 1st 110Zr: U=2.63, D=2.51, US=2.42
Individual rates:
U1 1.340 D1 0.424 D7 0.402 US1 0.576
U2 1.202 D2 0.446 D8 0.381 US2 0.835
D3 0.408 D9 0.342
D4 0.393 D10 0.315
D5 0.359 D11 0.250
D6 0.275 D12 0.193
Notes:
-One notices a relative drop of rate for D6, D11 and D12
pointing to a source located in the vicinity of AIDA
94Se setting:
File: 150503_1219_1319_94Se.root
Rate per tube (n/s): U=1.267, D=0.349, US=0.717
Ratio to 1st 110Zr: U=2.62, D=2.51, US=2.46
Notes:
-Very similar to 88Ge measurement |
348
|
Wed Oct 25 01:02:14 2017 |
RG | Status of the YSO array | The presently used detector has 48x48x5 mm3 YSO array with 1x1mm2 segments. They are separated with ESR. We use H8500 pspmt and 2 mm quartz diffuser with rough edges. Coupling is done with sylgard. Anger logic uses a conventional resistive network with 1kohm and we have 4 position signals and dynode. The position signals (xa,xb,ya,yb) are split in Low gain branch connected directly to 5V range input in Struck and High gain branch, which uses Phillips fast amplifier to boost the gain. The use of the amplifiers enables to operate the detector at low PSPMT gain and should give us a chance to work with the strong light produced by the implant. The implant signals were simulated using blue diode and a pulse generator. We observed significant drift of the position for very large pulses, but this was for now attributed to be the pulse shape effect due to the use of the diode and reltively slow pulser.
YSO for now should run at -700 V. The maximum operating voltage is 1200V and should not be exceeded. At 1200 V the ions induced signals may saturate our electronics.
Tuesday evening:
Detector remounted in the pedestal and aligned with WASABI with 1mm+1.5mm spacers. The pedestal is 25.8 mm tall and the box is 60 mm tall. We checked also the orientation of the spectra and chose AIDA convention (look upstream). With this pedestal the front of the box is centered with the back of the clover leaf.
Wednestay morning:
I opened the detector and rewrapped it with new teflon tape. I noticed a small piece if sylgard peeled off, which may explain why the we had some edge effect yesterday. The detector was remounted on the new pedestal which will allow it to be positioned at the center of the clover. The signals looked OK on the oscilloscope.
|
210
|
Wed Mar 29 13:31:37 2017 |
J Agramunt, R. Caballero, T. Davinson, C. Griffin, O. Hall, D. Kahl, G. Kiss, J. Liu, S. Nishimura, V. Phong, JL Tain | Start of parasitic | Start of parasitic Ca beam
(main experiment "New isotope search", Ahn et al.)
New directory: /data/1703Ahn/
Configuration file: 170329_Conf_BrikenFull.xlsx |
358
|
Tue Oct 16 03:15:03 2018 |
ATD & JLT | Start of REP run preparation (RIBF148) | 16-Oct-2018
- Start DACQ computer: OK!
- Create directory: /data/1810Preparation/
- Copy last configuration RIBF139 file: 171119_1126_Conf_BrikenFullYSO.xlsx
- Return PB-5 pulser into crate and set parameters
- Star all crates: OK!
- Start DACQ: OK!
- Rename configuration file and edit to eliminate YSO related parameters
New Config file: 181016_Conf_BrikenFull.xlsx
- Start 3He tube HV: power the ethernet hub, power the pixie PC,
edit the hvmon.conf file (only tubes active), run hvmon: OK!
- We realized that current and voltage of VME crate 1 was oscillating.
We pulled out V1A7 (not used) and it became stable: seems that we are at the limit of
power supply: OK!
- We put the 252Cf source in and check gain match:
only (the gain of) three tubes moved slightly since the last run (one year ago)!
He006, He134, He135 -> new calibration: OK!
- We move SIS3820 clock from internal to external frequency:
Setup UK Tektronix clock: 25MHz, square signal, 800mVpp, -400mV offset
Plug in Lupo module (VME1 crate): OK!
ISSUE: the current placement of the table supporting the MUSIC chamber
is too close to the VME1 crate and one of the modules (V1A4)
cannot be extracted in case of failure. However V1A7 is unused
in this run and can be used as replacement. |
134
|
Tue Nov 8 02:49:19 2016 |
PV, Liu | Specification of the dE silicon detector | The specification and picture of the dE silicon detector is shown in the attached photos.
The detector has an effective area of 134 x 123 mm and 320 micrometer thick
There are 26 vertical strips. They are combined into 2 readout channels (top and bottom which respect to the orientation of the detector in BRIKEN) |
190
|
Tue Nov 29 00:30:42 2016 |
PV,SN | Sillicon dE bias voltage increased | We increase the bias voltage of the dE detector from 20 V to 100 V
Leakage current:
Before: ~0.590 uA
After 0.695 uA |
123
|
Mon Nov 7 04:37:33 2016 |
Phong,Liu,... | Si de and AIDA plastic calibration | 60Co source placed close to the detectors:
ROOT files:
161107_1221_SiDetCalibration.root
161107_1237_AIDAPlasticCalibration.root |
145
|
Tue Nov 8 17:00:27 2016 |
NTB, CG, AT | Si Delta E | At least during the timeframe in between ~ 21:30 (11/08) - 23:30 (11/09) (161108_2323_717273Ni.root) we see that the silicon dE have a very low counting rate. ~1Hz when the first AIDA detector sees ~600 Hz and a cumulative ~2kHz. The energy range in the dE stops at ~10MeV.
Edit: It appears the Si dE were removed. Some signals are still in the spectra. |
321
|
Mon Jun 12 02:45:44 2017 |
Iris Dillmann AT | Setup detectors. Id of Ancillary@BRIKEN | Attachment 1: Setup detectors
Attachment 2: example of configuration file (valid only for RIBF127). The Id of the ancillary detectors are:
DTPulser 95
V_PL_T 96
V_PL_B 97
F11_PL_R_LG 98
F11_PL_L_LG 99
V_NaI_R 100
V_NaI_L 101
F11_PL_R 102
F11_PL_L 103
Si_T 104
Si_B 105
AIDA_PL 106 |
171
|
Sat Nov 26 06:30:30 2016 |
AT | Setup | Setup for parasitic experiments
Energy 41Al @ F11 is around 145MeV/u
Degraders:
- Pb 3mm width ---> IN beam
- Tungsten:
1) 0.5 mm W + 1mm Al ---> IN/OUT beam
2) 0.3 mm W + 1mm Al ---> IN/OUT beam
3) 1.0 mm W + 1mm Al ---> IN/OUT beam |
330
|
Mon Oct 16 05:52:14 2017 |
BRIKEN collaboration | START preparation | We start to check every thing for the coming run
New directory: /data/1710Preparation/
Copy last configuration file of RIBF148 experiment
renamed:171016_Conf_BrikenFull.xlsx
Edinburgh clock parameters:
-Square signal.
-Amplitude: 1 Vpp
-Offset: 0.5V
-FreqCh1=FreqCh2
Status:
-F11 (high/low gain), AIDA_PL, DT_Pulser, Sync_Pulse, He tubes are defined and connected. Preliminary
configuration "171016_Conf_BrikenFull.xlsx" attached.
-YSO & clovers are not connected yet. |
293
|
Wed May 24 06:42:05 2017 |
JLT,AT,AM,RY,SG | START Preparation | We start to check every thing for the coming run
New directory: /data/1705Preparation/
Copy last configuration file of Ahn parasitic run
renamed:170524_Conf_BrikenFull.xlsx
252Cf source at the center.
Root file: 170524_252Cf_1st.root
Issues:
Pulsers of He006, He140 are tripled
There is a strange background above n-peak in He034
Calibration is ok except for He003, He004 and He006 !!!!
|
63
|
Thu Oct 6 02:03:32 2016 |
JL Tain | START | We started the assembly of BRIKEN and AIDA detectors in their final position
on Tuesday October 4:
Gabor Kiss, Phong Vi, Keishi Matsui, Shunji Nishimura, Shigeru Kubono, PA,
Chris Griffin, Tom Davinson, Nathan Brewster, Shintaro Go, Daniel Stracener,
Ariel Tarifeno, Jorge Agramunt, Alvaro Tolosa, Jose L Tain
Tubes were removed from BRIKEN. PE matrix lifted with crane.
Table rotated and placed in the final position (front wheels
between forward EURICA rails) by hand. Table leveled (height, angle)
with the help of laser. PE matrix placed back.
AIDA rotated and adjusted to BRIKEN by hand.
See photos |
101
|
Thu Nov 3 04:40:02 2016 |
All | START | Start of commissioning preparation
Working directory:
/data/1611Commissioning/161031Preparation/
DACQ configuration file:
161103Conf_BrikenFull.xlsx |
131
|
Mon Nov 7 13:44:26 2016 |
Alvaro, Ale, Jorge, Jose | Running Gasific and run protocol | Preliminary |
162
|
Fri Nov 25 03:41:40 2016 |
AT | Run setups | Run setups during preparation of Fallon Parasitic experiments (Kanungo experiment beam time) can be found on
http://ribfdaq/~gamma16/expdbrunsum.php
A copy of the information in this page is placed here!
According to their Setup, they are using C targets on F8 and F11. The targets can be IN/OUT. |
270
|
Sat Apr 1 02:07:49 2017 |
ID DK JA | Run 55 | Date: 1/04/2017
Start:10:04
BRIKEN run: 55
AIDA Run: R20_131
BigRIPS run: 207
DLT:170401_1002_Na37_055.dlt
BRIKEN Rate: 1.5 cps
F11 Rate: 8.5 cps
Date: 1/04/2017
Stop: 11:59
BRIKEN run: 55
AIDA Run: R20_160
BigRIPS run: 208
Root file: 170401_0157_1159_B055_Na37Ne36.root
OBSERVATIONS: 10:55 Beam stop few minutes
F11_PL_R Change the range and cleared
Two Big rips files |
252
|
Fri Mar 31 10:35:22 2017 |
PV CG JA | Run 40 | Date: 31/03/2017
Start:
BRIKEN run: 40
AIDA Run: R16_630
BigRIPS run: 184
DLT: 170331_1837_Mg40_040.dlt
BRIKEN Rate: 40/ 10= 4 cps
F11 Rate: 12 cps
Date: 31/03/2017
Stop: 19:43
BRIKEN run: 040
AIDA Run: R16_647
BigRIPS run: 185
Root file: 170331_1222_1943_B040_Mg40.root
OBSERVATIONS: BigRIPS team changed to run185 after about 11min (at 18:38) |
251
|
Fri Mar 31 09:23:45 2017 |
CG PV JA | Run 39 | Date: 31/03/2017
Start: 17:22
BRIKEN run: 039
AIDA Run: R16_611
BigRIPS run: 183
DLT: 170331_1722_Mg40_039.dlt
BRIKEN Rate: 40 cps
F11 Rate: 12 cps
Date: 31/03/2017
Stop: 18:33
BRIKEN run:39
AIDA Run: R16_628
BigRIPS run:183
Root file:170331_1222_1833_B039_Mg40.root
OBSERVATIONS: Stopped to change a stripper foil |
250
|
Fri Mar 31 08:23:32 2017 |
CG PV JA | Run 38 | Date: 31/03/2017
Start: 16:21
BRIKEN run: 038
AIDA Run: R16_581
BigRIPS run: 182
DLT: 170331_1524_Mg40_037.dlt
BRIKEN Rate: 40 cps
F11 Rate: 12 cps
Date: 31/03/2017
Stop:
BRIKEN run: 38
AIDA Run: R16_611
BigRIPS run:182
Root file:170331_1222_1721_B038_Mg40.root
OBSERVATIONS: |
|