Entry  Thu Jan 27 15:19:18 2022, OH, JM, Enabling internet access on the carme-gsi server 
To use both anydesk and firefox on the carme server require an internet connection.

As the server is on the cryring local network an ssh tunnel needs to be opened to the GSI proxy server.
Entry  Mon Jun 10 18:22:47 2024, JM, Electron cooler calculations space-charge-2.pngspace-charge-1.pngBeamCalc.png
Ion beam energy is calculated using Beamcalc (https://web-docs.gsi.de/~lestinsk/beamcalc/electron.php)

Electron current is set to 4.3 mA and can be taken from Grafana

Alpha factor is 50
Entry  Tue Sep 26 10:13:14 2023, TD, ELOTECH PID Temperature Controller R4000 manual ( German and English ) BA-Isoheat_R4000-XX-X-000-X-X_DE_V1.03.pdfR4000-XX-X-000-X-X_EN_V1.30.pdf
 
Entry  Wed Feb 23 20:06:43 2022, TD CN, E141 kinematics E141_kinematics.txt

Assuming DSSSD is 1000um thick we can expect to observe protons for E_lab=11MeV as follows
Entry  Sun Oct 31 07:29:17 2021, CB JM + RSS OG, Detectors mounted on plate in CARME 8x
(Wed 27 October)

Dismounted plate from rod. Dismounted broken detectors from plate and placed back in box. Dismounted "dog leg" struts and replaced
with fully un-threaded struts.
Entry  Sun Oct 31 07:21:09 2021, CB JM RSS, Detectors arrived in GSI 6x
(Tue 26 October)

Two new CARME DSSD arrived in GSI and collected. Visually inspected both sides. No obvious damage. Photos attached. Second detector (3335-2)
was shipped in upside-down box. Does not seem to have made any difference.
Entry  Fri May 26 09:08:06 2023, JM, CB, Detector movement  20230525_135756.jpgdetector-movement-in.mp4
Hopefully we will not be opening the chamber for quite some time so find attached images of all four installed detectors and a video of them both moving
into the beam axis for future reference
Entry  Mon Feb 5 20:15:31 2024, CB, TD, Detector move in/out signals 20240205_174307.jpg
17.43 setup signals from detector movement controller.

See attachment 1
Entry  Wed Sep 29 13:13:51 2021, JM RSS PB(remotely), Detector frame assembley new_carrier_simple_holes.PDFScreenshot_20210929-110343_Telegram.jpg20210930_164625.jpg20210930_164650.jpg
29/9/21

We have opened the cleaned parts for assembley of the moving detector frames. We appear to have everything we require. 

Standoffs - 12 M3x35 bolts, Many M3 washers, Many M3 nuts, 16 standoff tubes. 
Entry  Tue Feb 6 09:39:54 2024, CB, Detector bias tests 13x
10:30 Found ADC#4 on aida 1 and 4 are suddenly noisy. See attach 1-2. Checkloaded multiple times - no effect.

Leakage current overnight was stable, but a clear transient can be observed yesterday around the times of the DSSD motion test. We did not observe
this in the cave looking at the CAEN module - it might have been too fast.
Entry  Tue Feb 1 14:43:20 2022, JM, TD, Detector Bias, FEEs 22-02-01_15-03-56_aida03pulserADCspectra.png2022-02-01_15-14-07_aida03_pulser_DAQserver.png2022-02-01_15-22-23aida03_waveforms_pulser.png2022-02-01_15-28-45aida03_waveforms_pulser2.png2022-02-01_15-54-57rates_aidaALL.png
Detector is not biasing as it has in previous tests. Last bias 15/1/22 see previous elogs, max leakage current ~10 uA at 150V

Detector bias increases well above 10uA for voltages as low as 30-40V. Optical windows covered, Y09 valves closed, large ion pump OFF and both
ion gauges OFF produces similar results but with a slightly lower leakage current.
Entry  Wed Feb 2 21:53:32 2022, JM, TD, Detector Bias 20220202_110757.jpg20220202_193636.jpg20220202_193651.jpgmotion.png
FEE mounts for 5,6,7,8 disassembled to access adaptor boards. No sign of debris which could cause short on adaptor boards. 

Adaptor boards removed and checked against spares with multimeter. Both show same characteristics. Pins on feedthrough checked using multimeter.
No longer see -18.5 KOhm between bias pins, no see 140 kOhm which is similar to bias pin to non bias pin from prevous tests.
Entry  Sun Apr 20 07:30:29 2025, CB, DSSD leakage currents over Easter Screenshot_2025-04-20_083007.pngScreenshot_2025-04-20_193226.pngScreenshot_2025-04-21_085130.png
Just after midnight on 20/04 it appears DSSD #4 leakage current has perfectly stabilised. See attached.

DSSDs remain biased.

 
    Reply  Fri Jun 7 15:53:35 2024, TD, DSSD leakage current increase - temperature increase in CRYRING hall Capture.PNG
Re-increase in DSSSD leakage current c. 14-15.00 6.6.24 

There was a significant temperature spike in the CRYRING hall on the afternoon of 6.6.24 which has, partly, recovered - see attachment 1




Quote:


Entry  Tue Feb 20 07:50:10 2024, JM, DAQ restart 
DAQ restarted. Run46. All appears OK, aida01 shows high rates across ASIC 4 even with no beam which could not be resolved via ASIC control. May suggest
some damge.

Merger is running and tape server is writing data to disk. Data rate is about 1300 kB/sec from 800 kb/sec due to increase in aida01 rate.
Entry  Tue May 27 05:03:41 2025, JM, DAQ restart 7x
06:00

FEE modules power cycled.

No issues on power cycle restart.
Entry  Mon Feb 21 20:56:51 2022, SHB, JM, CB, OH (remotely), DAQ connection for ion current cry.jpg
signals from ICT and CRYRADIO are now connected to AIDA2 and AIDA5 respectively. ICT is set to factor 200 and CRYRADIO to 40 for now.

 

(AIDA3 --> electron cooler)
Entry  Mon Jun 16 08:17:41 2025, JG, D2 consumption 
The current bottle of D2 was installed on friday 06.06. at around 12:00. I assume we started at 200 bar.

Right now we have 50 bar remaining in this bottle (16.06. at 8:00).

This is an average consumption of ~15 bar/day
Entry  Fri Apr 18 16:18:33 2025, JM, D+D BBN energies test 16x
D+D minimum energy Ecm = 150 keV which corresponds to Ed=150 keV/u or Ed=300 keV. Therefore we need a threshold below this.

Test 1 - Current status

Current slow comparator threshold is 0x64 (1 MeV). ADC data items rate in all FEEs is essentially zero with this threshold. (except aida03 and
Entry  Tue Mar 1 05:23:02 2022, JM, EM, Current status  2022-03-01_06-42-53-ExRate.png
06:20

Servo motors moved to fully out position as some work on the beam was neccessary. Beam injection had dropped from ~9e7 to around 1e7. Potentiometer
reads 19.10 mm.   
ELOG V3.1.3-7933898