Entry  Mon Feb 28 02:16:00 2022, JM, OF, Current status 6x

03:21

Temperature setpoint 110 K

Rutherford rate stable. Min peak height ~ 10^2 counts, max ~ 3*10^2 counts.

Temperatures OK

System checks OK - wr attached

Merger OK

Data rate ~ 1500 kb/sec

Leakage current ~ 3.228 uA

Ring vacuum stable, Avg = 3.1E-10 mbar

E1 pump current = 2.5 A

Entry  Mon Feb 28 03:16:37 2022, JM, OF, Current status 6x

04:24

Temperature setpoint 110 K

Rates appeared stable throughout the hour ~10^2 peak height. Last 5 minutes rutherford peak height appears reduced, cycling from 10^1 to 10^2. If this continues with sustained low density, temperature will be cycled to 125 K and back down.

Merger OK

Data rate = 1300 kb/sec

Ring avg pressure = 3.1E-10 mbar

Temperatures OK

System checks OK - wr fail attached

Leakage current ~ 3.228 uA

04:33

Rutherford rate appears to have increased and stabilised again around 10^2 peak height.

Entry  Mon Feb 28 04:14:26 2022, JM, OF, Current status 9x

05:20

Temperature setpoint = 110 k

Rates are stable however are slightly lower than previous. Min peak height ~40 counts, max ~ 10^2.

Merger OK

Data rate ~ 1200 kb/sec

Ring avg pressure ~ 2.9E-10 mbar

Temperatures OK

System checks OK - wr fail attached

Leakage current ~ 3.222 uA

Entry  Mon Feb 28 06:12:58 2022, JG, AK, US, Current status 8x

07:20

Temperature setpoint = 110 k

Rutherford rates shaky, peak max at ~ 3x10^2

Merger OK

Data rate ~ 1200 kb/sec

Ring avg pressure ~ 2.5E-10 mbar

Temperatures OK

System checks OK - wr fail attached

Leakage current ~ 3.222 uA

Entry  Mon Feb 28 07:31:25 2022, JG, AK, Current status 8x

08:30

Temperature setpoint = 110 k

Rutherford rates shaky, peak max at ~ 3x10^2

Merger OK

Data rate ~ 1500 kb/sec

Ring avg pressure ~ 2.7E-10 mbar

Temperatures OK

System checks OK - wr fail attached

Leakage current ~ 3.22 uA

Entry  Mon Feb 28 08:36:35 2022, JG, AK, CB, Current status 8x

09:35

beam intensity at injection: 8.2e7 ions

Temperature setpoint = 110 k

Rutherford rates shaky, peak max at ~ 3x10^2

Merger OK

Data rate ~ 1400 kb/sec

Ring avg pressure ~ 2.7E-10 mbar

Temperatures OK

System checks OK - wr fail attached

Leakage current ~ 3.22 uA

Entry  Mon Feb 28 09:31:48 2022, JG, AK, CB, Current status 8x

10:35

beam intensity at injection: 1e8 ions

Temperature setpoint = 110 k

Rutherford rates shaky, peak max at ~ 3x10^2

Merger OK

Data rate ~ 1200 kb/sec

Ring avg pressure ~ 2.7E-10 mbar

Temperatures OK

System checks OK - wr fail attached

Leakage current ~ 3.22 uA

Entry  Mon Feb 28 10:55:48 2022, JG, US, Current status 8x

11:55

beam intensity at injection: 8e7 ions

Temperature setpoint = 110 k

Rutherford rates shaky, peak max at ~ 3x10^2

Merger OK

Data rate ~ 1600 kb/sec

Ring avg pressure ~ 2.7E-10 mbar

Temperatures OK

System checks OK - wr fail attached

Leakage current ~ 3.22 uA

Entry  Mon Feb 28 11:57:22 2022, JG, AK, CB, Current status 8x

12:55

beam intensity at injection: 9e7 ions

Temperature setpoint = 110 k

Rutherford rates shaky, peak max at ~ 3x10^2

Merger OK

Data rate ~ 1400 kb/sec

Ring avg pressure ~ 2.8E-10 mbar

Temperatures OK

System checks OK (wr fail)

Leakage current ~ 3.22 uA

Entry  Mon Feb 28 13:16:39 2022, JG, AK, CB, Current status 8x

14:16

beam intensity at injection: 7e7 ions

Temperature setpoint = 110 k

Rutherford rates shaky, peak max at ~ 1x10^2

Merger OK

Data rate ~ 1500 kb/sec

Ring avg pressure ~ 2.3E-10 mbar

Temperatures OK

System checks OK (wr fail)

Leakage current ~ 3.23 uA

Entry  Mon Feb 28 22:34:51 2022, JM, EM, Current status 2022-02-28_23-40-14-xyTotal.png2022-02-28_23-39-56-Exrate.png2022-02-28_23-39-26-goodEvents.png2022-02-28_23-38-58-correlation8.png2022-02-28_23-38-31-ADCitems.png

23:35

Temperature setpoint = 116 K

Detector position IN, potentiometer reads 34.89 mm

Rutherford rate fluctuates however is consistently aboe 100 counts peak height

Merger OK

Data rate = 1600 kb/sec

Leakage current = 3.23 uA

Temperatures OK

System checks OK - wr attached

         Base         Current     Difference
aida01 fault      0xab60 :      0xab8c :      44  
aida02 fault      0x88d3 :      0x88fe :      43  
aida03 fault      0x23aa :      0x23d6 :      44  
aida04 fault      0x15d7 :      0x1603 :      44  
aida05 fault      0x2f03 :      0x2f0f :      12  
White Rabbit error counter test result: Passed 0, Failed 5

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

 

Entry  Tue Mar 1 02:15:29 2022, JM, EM, Current status 6x

03:15

The nozzle motor control at the target crashed, but operation has been restored.

Temperature setpoint = 116 K

Detector position IN potentionmeter = 34.89mm

Rutherford rate fluctuates, around 100 counts peak height

Merger OK

Data rate = 1356 kb/sec

Leakage current = 3.234 uA

Temperatures OK

System checks OK - wr fails


         Base         Current     Difference
aida01 fault      0xab60 :      0xab8c :      44  
aida02 fault      0x88d3 :      0x88fe :      43  
aida03 fault      0x23aa :      0x23d6 :      44  
aida04 fault      0x15d7 :      0x1603 :      44  
aida05 fault      0x2f03 :      0x2f0f :      12  
White Rabbit error counter test result: Passed 0, Failed 5

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

Avg ring pressure = 4.4 E-10 mbar

Entry  Tue Mar 1 03:34:28 2022, JM, EM, Current status 7x

Temperature setpoint = 116 k

Detector IN potentiometer = 34.89 mm

Rutherford rate is reduced from previous entries. Avergae peak height appears to be ~50 counts. Fluctuations in peak height and E4 pressure are quite high. Temperature has been cycled, nozzle position has been optimised and different temperatures have been trialled however density cannot be maintained at a decent level. 

Merger OK

Data rate ~ 1500 kb/sec

Leakage current = 3.2295 uA

Temperatures OK

System checks OK - wr fails

         Base         Current     Difference
aida01 fault      0xab60 :      0xab8c :      44  
aida02 fault      0x88d3 :      0x88fe :      43  
aida03 fault      0x23aa :      0x23d6 :      44  
aida04 fault      0x15d7 :      0x1603 :      44  
aida05 fault      0x2f03 :      0x2f0f :      12  
White Rabbit error counter test result: Passed 0, Failed 5

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR


Avg ring pressure = 4.4 E-10 mbar

Entry  Tue Mar 1 22:10:26 2022, David, George, Current status 2022-03-01_23-14-43aida.png2022-03-01_23-13-57exrate.png2022-03-01_23-13-40xyto.png2022-03-01_23-13-10extot.png

Ring current 6.67e7 particles/s

pressure at S1: 2.3-2.8e-8

Merger, Tape server (WS3): 1574 kB/s, writing ok

Leakage current (WS4): 3.20 uA

DAQ (WS2): temperature checked, clock status: ok, adc calibration: ok, white rabbit check: failed

Online monitoring (WS 5): done

Entry  Tue Mar 1 23:30:04 2022, David, George, Current status 2022-03-02_00-39-47xytotal.png2022-03-02_00-39-20exRate.png2022-03-02_00-38-44extotal.png

00:30
Pressure in S1 slightly decreased from before but still high enough to measure data.
One new count at 20MeV

Target density at S1: 2e-8 - 2.7e-8
Pressures OK (one small spike just before 00:00 in YR08)
System Checks OK
White rabbit fail diff: 49
Data write speed: ~1.4Mb/s
Beam intensity at injection: ~7.7e7
Cryo Temp Setpoint: 110K

(Beam intensity in ring dropped as I was writing which is why there is no rutherford peak on exRate image)
 

Entry  Wed Mar 2 14:15:16 2022, JM, PM, Current status 6x

15:15

Temperature setpoint = 113 K

Target density is OK for now. S1 vaies between 2E-8 and 3.5E-8 mbar. E4 between 1E-9 and 1E-8 mbar.

Detector motors functioning well. Moving between 23.1 mm and 67.43mm. Move in and move out counters the same. Correlation scalar 8 for detectors moving is seen on aida1

Rutherford rate peak varies between 10^3 peak height when fully in and ~10^2 when fully out. No peak from halo seen in rate.

Beam intensity ~ 1E8

Merger OK

Data rate ~ 1700 kb/sec

Run is Run 87 - started 10:34

Data spy restarted at 15:00 so only include Run87 data.

Leakage current = 3.205 uA

Fee Temperatures OK

System checks OK - wr fails

         Base         Current     Difference
aida01 fault      0xab60 :      0xaba3 :      67  
aida02 fault      0x88d3 :      0x8914 :      65  
aida03 fault      0x23aa :      0x23ed :      67  
aida04 fault      0x15d7 :      0x161a :      67  
aida05 fault      0x2f03 :      0x2f20 :      29  
White Rabbit error counter test result: Passed 0, Failed 5

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

Avg ring Pressure = 5.5E-10 mbar

16:45

Temperature setpoint = 113 K

Target density is OK for now. S1 vaies between 2E-8 and 3E-8 mbar. E4 between 1E-9 and 1E-8 mbar.

Detector motors functioning well. Moving between 23.1 mm and 67.43mm. Move in and move out counters the same. Correlation scalar 8 for detectors moving is seen on aida1

Rutherford rate peak varies between 10^3 peak height when fully in and ~10^2 when fully out. No peak from halo seen in rate.

Beam intensity ~ 1E8

Merger OK

Data rate ~ 1300 kb/sec

Temperatures OK

Leakage current ~3.205 uA

System chacks OK - wr fail

         Base         Current     Difference
aida01 fault      0xab60 :      0xaba3 :      67  
aida02 fault      0x88d3 :      0x8914 :      65  
aida03 fault      0x23aa :      0x23ed :      67  
aida04 fault      0x15d7 :      0x161a :      67  
aida05 fault      0x2f03 :      0x2f20 :      29  
White Rabbit error counter test result: Passed 0, Failed 5

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

Avg ring pressure ~ 4.4 E-10 mbar

Statistics similar to previous check.

 

Entry  Wed Mar 2 17:21:35 2022, JM, PM, Current status 8x

18:21

Temperature setpoint = 113k

Target density OK S1 varies between 2E-8 and 2.5E-8 mbar, E4 between 1E-9 and 5E-9 mbar. (slightly lower than previous check)

Detectors moving well between 23.1mm and 67.43 mm, Move in and move out counters the same. Correlation scalar is on aida1

Rutherford peak varies between 10^3 and 10^2 depending on detector position. No halo peak in rate. Noise peak varies from ~100 peak height when fully in to ~400 peak height when fully out.

Beam intensity ~ 1E8

Merger OK

Data rate ~ 1800 kb/sec

Leakage current ~3.21 uA

Temperature OK
System checks  OK - wr fails

         Base         Current     Difference
aida01 fault      0xab60 :      0xaba3 :      67  
aida02 fault      0x88d3 :      0x8914 :      65  
aida03 fault      0x23aa :      0x23ed :      67  
aida04 fault      0x15d7 :      0x161a :      67  
aida05 fault      0x2f03 :      0x2f20 :      29  
White Rabbit error counter test result: Passed 0, Failed 5

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

Avg ring pressure = 4.4 E-10 mbar

19:20

Temperature setpoint = 113k

Target density OK S1 varies between 2E-8 and 3E-8 mbar, E4 between 1E-9 and 1E-8 mbar. (slightly higher than previous check)

Detectors moving well between 23.1mm and 67.43 mm, Move in and move out counters the same. Correlation scalar is on aida1

Rutherford peak varies between 10^3 and 10^2 depending on detector position. No halo peak in rate.

Beam intensity ~ 1E8

Merger OK

Data rate ~ 1800 kb/sec

Leakage current ~3.20 uA

Temperature OK
System checks  OK - wr fails

         Base         Current     Difference
aida01 fault      0xab60 :      0xaba4 :      68  
aida02 fault      0x88d3 :      0x8915 :      66  
aida03 fault      0x23aa :      0x23ee :      68  
aida04 fault      0x15d7 :      0x161b :      68  
aida05 fault      0x2f03 :      0x2f20 :      29  
White Rabbit error counter test result: Passed 0, Failed 5

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

Avg ring pressure = 4.3 E-10 mbar

Statistics similar to previous check.

Entry  Wed Mar 2 19:25:25 2022, JM, PM, Current status 10x

20:25

Observed short burst of low energy noise on aida1, lasted <30s (attachment 8).

Temperature setpoint = 113 K

Target density OK S1 vaies between 2E-8 and 3E-8 mbar, E4 between 1E-9 and 1E-8 mbar.

Detector moving well between 23.1 mm and 67.43mm, move in and move out counters the same. Correlation scalar on aida1.

Rutherford peak varies between 10^3 and 10^2 depending on detector position, noise peak varies with detector position -> lower further in.

Beam intensity ~1E8

Merger OK

Data rate ~ 1900 Kb/sec

Leakage current = 3.208 uA

Temperatures OK

System checks - wr fails

         Base         Current     Difference
aida01 fault      0xab60 :      0xaba4 :      68  
aida02 fault      0x88d3 :      0x8915 :      66  
aida03 fault      0x23aa :      0x23ee :      68  
aida04 fault      0x15d7 :      0x161b :      68  
aida05 fault      0x2f03 :      0x2f20 :      29  
White Rabbit error counter test result: Passed 0, Failed 5

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

Avg ring pressure = 5.5 E-10 mbar. Ring pressure tends to oosciilate between ~4 and 5.5 E-10 mbar

21:21

Burst of noise not observed again

Temperature setpoint = 113 K

Target density OK S1 vaies between 2E-8 and 3E-8 mbar, E4 between 1E-9 and 1E-8 mbar.

Detector moving well between 23.1 mm and 67.43mm, move in and move out counters the same. Correlation scalar on aida1.

Rutherford peak varies between 10^3 and 10^2 depending on detector position, noise peak varies with detector position -> lower further in.

Beam intensity ~1E8

Merger OK

Data rate ~ 1900 Kb/sec

Leakage current = 3.206 uA

Temperatures OK

System checks - wr fails

         Base         Current     Difference
aida01 fault      0xab60 :      0xaba4 :      68  
aida02 fault      0x88d3 :      0x8915 :      66  
aida03 fault      0x23aa :      0x23ee :      68  
aida04 fault      0x15d7 :      0x161b :      68  
aida05 fault      0x2f03 :      0x2f20 :      29  
White Rabbit error counter test result: Passed 0, Failed 5

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

Avg ring pressure 5.4E-10 mbar

Entry  Wed Mar 2 21:16:06 2022, JM, PM, Current status 

22:16

Temperature setpoint = 113k

Not observed noise burst again

Target density OK, S1 between 2E-8 and 3E-8mbar, E4 between 1E-9 and 1E-8 mbar.

Detector moving well, potentionmer between 23.1 mm and 67.43 mm, move in/move out counters the same. Correlation scalar on aida1

Ruthrford peak varies between 10^3 and 10^2 peak height depending on detector position, noise rate also varies with position.

Beam intensity ~1E8

Merger OK

Data rate ~ 1200 kb/sec

Temperatures OK

System checks OK - wr fails

         Base         Current     Difference
aida01 fault      0xab60 :      0xaba4 :      68  
aida02 fault      0x88d3 :      0x8915 :      66  
aida03 fault      0x23aa :      0x23ee :      68  
aida04 fault      0x15d7 :      0x161b :      68  
aida05 fault      0x2f03 :      0x2f20 :      29  
White Rabbit error counter test result: Passed 0, Failed 5

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

Ring pressure oscillation is 30 minute cycle on all areas of the ring.

Entry  Thu Apr 10 07:02:12 2025, CB (remotely), Current logging enabled - sort of 

There appear to be issues with Shahab's caenlogging program. It crashes occasionally, with an error indicating the number of columns in the terminal is incorrect. We will have to ask Shahab.

For now

1. Start the caen_screen.sh script on nnrp1 (aka CARMEMotorsPi -  why was it renamed? It should stay as CARMEMotorsPi to avoid duplicate names).

This starts connect to the caen power module via TTY inside a screen called caenhv. Currently on desktop 1

 

2. On nnrpi1, start the caenlogging program --settings caenlogger_settings_gsi.toml

 

3. On carme-gsi, in ~/Programs/caenlogger start the gnuplot script

 

If the caenlogger script crashes, restart it on the second terminal

More investigations required on the crash.