ID |
Date |
Author |
Subject |
319
|
Mon Feb 28 02:16:00 2022 |
JM, OF | Current status |
03:21
Temperature setpoint 110 K
Rutherford rate stable. Min peak height ~ 10^2 counts, max ~ 3*10^2 counts.
Temperatures OK
System checks OK - wr attached
Merger OK
Data rate ~ 1500 kb/sec
Leakage current ~ 3.228 uA
Ring vacuum stable, Avg = 3.1E-10 mbar
E1 pump current = 2.5 A |
Attachment 1: 2022-02-28_03-11-52-Rates_1.png
|
|
Attachment 2: 2022-02-28_03-12-28-ExTotal.png
|
|
Attachment 3: 2022-02-28_03-15-07-xyTotal.png
|
|
Attachment 4: 2022-02-28_03-17-43-wr.png
|
|
Attachment 5: 2022-02-28_03-19-41-leakage-current.png
|
|
Attachment 6: 20220228_032156.jpg
|
|
320
|
Mon Feb 28 03:16:37 2022 |
JM, OF | Current status |
04:24
Temperature setpoint 110 K
Rates appeared stable throughout the hour ~10^2 peak height. Last 5 minutes rutherford peak height appears reduced, cycling from 10^1 to 10^2. If this continues with sustained low density, temperature will be cycled to 125 K and back down.
Merger OK
Data rate = 1300 kb/sec
Ring avg pressure = 3.1E-10 mbar
Temperatures OK
System checks OK - wr fail attached
Leakage current ~ 3.228 uA
04:33
Rutherford rate appears to have increased and stabilised again around 10^2 peak height. |
Attachment 1: 2022-02-28_04-23-06-rates.png
|
|
Attachment 2: 2022-02-28_04-22-28-goodEvents.png
|
|
Attachment 3: 2022-02-28_04-22-05-correlation8.png
|
|
Attachment 4: 2022-02-28_04-21-39-ADCitems.png
|
|
Attachment 5: 2022-02-28_04-21-17-wr.png
|
|
Attachment 6: 20220228_042449.jpg
|
|
321
|
Mon Feb 28 04:14:26 2022 |
JM, OF | Current status |
05:20
Temperature setpoint = 110 k
Rates are stable however are slightly lower than previous. Min peak height ~40 counts, max ~ 10^2.
Merger OK
Data rate ~ 1200 kb/sec
Ring avg pressure ~ 2.9E-10 mbar
Temperatures OK
System checks OK - wr fail attached
Leakage current ~ 3.222 uA |
Attachment 1: 2022-02-28_05-20-11-correlation8.png
|
|
Attachment 2: 2022-02-28_05-19-44-ADCitems.png
|
|
Attachment 3: 2022-02-28_05-19-19-goodEvents.png
|
|
Attachment 4: 2022-02-28_05-18-55-wr.png
|
|
Attachment 5: 2022-02-28_05-18-13-Exrate.png
|
|
Attachment 6: 2022-02-28_05-21-31-ExTotal.png
|
|
Attachment 7: 2022-02-28_05-22-26-xyTotal.png
|
|
Attachment 8: 20220228_052350.jpg
|
|
Attachment 9: 20220228_052326.jpg
|
|
323
|
Mon Feb 28 06:12:58 2022 |
JG, AK, US | Current status |
07:20
Temperature setpoint = 110 k
Rutherford rates shaky, peak max at ~ 3x10^2
Merger OK
Data rate ~ 1200 kb/sec
Ring avg pressure ~ 2.5E-10 mbar
Temperatures OK
System checks OK - wr fail attached
Leakage current ~ 3.222 uA |
Attachment 1: Screenshot_from_2022-02-28_07-16-18.png
|
|
Attachment 2: Screenshot_from_2022-02-28_07-16-02.png
|
|
Attachment 3: Screenshot_from_2022-02-28_07-15-49.png
|
|
Attachment 4: Screenshot_from_2022-02-28_07-15-28.png
|
|
Attachment 5: Screenshot_from_2022-02-28_07-12-06.png
|
|
Attachment 6: Screenshot_from_2022-02-28_07-11-59.png
|
|
Attachment 7: Screenshot_from_2022-02-28_07-11-48.png
|
|
Attachment 8: Screenshot_from_2022-02-28_07-17-27.png
|
|
324
|
Mon Feb 28 07:31:25 2022 |
JG, AK | Current status |
08:30
Temperature setpoint = 110 k
Rutherford rates shaky, peak max at ~ 3x10^2
Merger OK
Data rate ~ 1500 kb/sec
Ring avg pressure ~ 2.7E-10 mbar
Temperatures OK
System checks OK - wr fail attached
Leakage current ~ 3.22 uA |
Attachment 1: Screenshot_from_2022-02-28_08-30-34.png
|
|
Attachment 2: Screenshot_from_2022-02-28_08-30-13.png
|
|
Attachment 3: Screenshot_from_2022-02-28_08-29-59.png
|
|
Attachment 4: Screenshot_from_2022-02-28_08-29-44.png
|
|
Attachment 5: Screenshot_from_2022-02-28_08-29-28.png
|
|
Attachment 6: Screenshot_from_2022-02-28_08-29-17.png
|
|
Attachment 7: Screenshot_from_2022-02-28_08-29-10.png
|
|
Attachment 8: Screenshot_from_2022-02-28_08-29-04.png
|
|
326
|
Mon Feb 28 08:36:35 2022 |
JG, AK, CB | Current status |
09:35
beam intensity at injection: 8.2e7 ions
Temperature setpoint = 110 k
Rutherford rates shaky, peak max at ~ 3x10^2
Merger OK
Data rate ~ 1400 kb/sec
Ring avg pressure ~ 2.7E-10 mbar
Temperatures OK
System checks OK - wr fail attached
Leakage current ~ 3.22 uA |
Attachment 1: Screenshot_from_2022-02-28_09-36-05.png
|
|
Attachment 2: Screenshot_from_2022-02-28_09-35-56.png
|
|
Attachment 3: Screenshot_from_2022-02-28_09-35-35.png
|
|
Attachment 4: Screenshot_from_2022-02-28_09-35-18.png
|
|
Attachment 5: Screenshot_from_2022-02-28_09-35-07.png
|
|
Attachment 6: Screenshot_from_2022-02-28_09-34-21.png
|
|
Attachment 7: Screenshot_from_2022-02-28_09-34-06.png
|
|
Attachment 8: Screenshot_from_2022-02-28_09-33-48.png
|
|
327
|
Mon Feb 28 09:31:48 2022 |
JG, AK, CB | Current status |
10:35
beam intensity at injection: 1e8 ions
Temperature setpoint = 110 k
Rutherford rates shaky, peak max at ~ 3x10^2
Merger OK
Data rate ~ 1200 kb/sec
Ring avg pressure ~ 2.7E-10 mbar
Temperatures OK
System checks OK - wr fail attached
Leakage current ~ 3.22 uA |
Attachment 1: Screenshot_from_2022-02-28_10-31-27.png
|
|
Attachment 2: Screenshot_from_2022-02-28_10-31-16.png
|
|
Attachment 3: Screenshot_from_2022-02-28_10-31-06.png
|
|
Attachment 4: Screenshot_from_2022-02-28_10-30-49.png
|
|
Attachment 5: Screenshot_from_2022-02-28_10-30-30.png
|
|
Attachment 6: Screenshot_from_2022-02-28_10-30-24.png
|
|
Attachment 7: Screenshot_from_2022-02-28_10-30-00.png
|
|
Attachment 8: Screenshot_from_2022-02-28_10-29-26.png
|
|
328
|
Mon Feb 28 10:55:48 2022 |
JG, US | Current status |
11:55
beam intensity at injection: 8e7 ions
Temperature setpoint = 110 k
Rutherford rates shaky, peak max at ~ 3x10^2
Merger OK
Data rate ~ 1600 kb/sec
Ring avg pressure ~ 2.7E-10 mbar
Temperatures OK
System checks OK - wr fail attached
Leakage current ~ 3.22 uA |
Attachment 1: Screenshot_from_2022-02-28_11-57-17.png
|
|
Attachment 2: Screenshot_from_2022-02-28_11-57-04.png
|
|
Attachment 3: Screenshot_from_2022-02-28_11-56-48.png
|
|
Attachment 4: Screenshot_from_2022-02-28_11-54-34.png
|
|
Attachment 5: Screenshot_from_2022-02-28_11-53-25.png
|
|
Attachment 6: Screenshot_from_2022-02-28_11-52-21.png
|
|
Attachment 7: Screenshot_from_2022-02-28_11-52-14.png
|
|
Attachment 8: Screenshot_from_2022-02-28_11-51-06.png
|
|
329
|
Mon Feb 28 11:57:22 2022 |
JG, AK, CB | Current status |
12:55
beam intensity at injection: 9e7 ions
Temperature setpoint = 110 k
Rutherford rates shaky, peak max at ~ 3x10^2
Merger OK
Data rate ~ 1400 kb/sec
Ring avg pressure ~ 2.8E-10 mbar
Temperatures OK
System checks OK (wr fail)
Leakage current ~ 3.22 uA |
Attachment 1: Screenshot_from_2022-02-28_12-56-50.png
|
|
Attachment 2: Screenshot_from_2022-02-28_12-56-40.png
|
|
Attachment 3: Screenshot_from_2022-02-28_12-56-29.png
|
|
Attachment 4: Screenshot_from_2022-02-28_12-56-16.png
|
|
Attachment 5: Screenshot_from_2022-02-28_12-55-36.png
|
|
Attachment 6: Screenshot_from_2022-02-28_12-55-08.png
|
|
Attachment 7: Screenshot_from_2022-02-28_12-54-54.png
|
|
Attachment 8: Screenshot_from_2022-02-28_12-54-33.png
|
|
330
|
Mon Feb 28 13:16:39 2022 |
JG, AK, CB | Current status |
14:16
beam intensity at injection: 7e7 ions
Temperature setpoint = 110 k
Rutherford rates shaky, peak max at ~ 1x10^2
Merger OK
Data rate ~ 1500 kb/sec
Ring avg pressure ~ 2.3E-10 mbar
Temperatures OK
System checks OK (wr fail)
Leakage current ~ 3.23 uA |
Attachment 1: Screenshot_from_2022-02-28_14-16-00.png
|
|
Attachment 2: Screenshot_from_2022-02-28_14-15-43.png
|
|
Attachment 3: Screenshot_from_2022-02-28_14-15-30.png
|
|
Attachment 4: Screenshot_from_2022-02-28_14-15-10.png
|
|
Attachment 5: Screenshot_from_2022-02-28_14-14-29.png
|
|
Attachment 6: Screenshot_from_2022-02-28_14-14-22.png
|
|
Attachment 7: Screenshot_from_2022-02-28_14-14-14.png
|
|
Attachment 8: Screenshot_from_2022-02-28_14-13-24.png
|
|
335
|
Mon Feb 28 22:34:51 2022 |
JM, EM | Current status |
23:35
Temperature setpoint = 116 K
Detector position IN, potentiometer reads 34.89 mm
Rutherford rate fluctuates however is consistently aboe 100 counts peak height
Merger OK
Data rate = 1600 kb/sec
Leakage current = 3.23 uA
Temperatures OK
System checks OK - wr attached
Base Current Difference
aida01 fault 0xab60 : 0xab8c : 44
aida02 fault 0x88d3 : 0x88fe : 43
aida03 fault 0x23aa : 0x23d6 : 44
aida04 fault 0x15d7 : 0x1603 : 44
aida05 fault 0x2f03 : 0x2f0f : 12
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
|
Attachment 1: 2022-02-28_23-40-14-xyTotal.png
|
|
Attachment 2: 2022-02-28_23-39-56-Exrate.png
|
|
Attachment 3: 2022-02-28_23-39-26-goodEvents.png
|
|
Attachment 4: 2022-02-28_23-38-58-correlation8.png
|
|
Attachment 5: 2022-02-28_23-38-31-ADCitems.png
|
|
337
|
Tue Mar 1 02:15:29 2022 |
JM, EM | Current status |
03:15
The nozzle motor control at the target crashed, but operation has been restored.
Temperature setpoint = 116 K
Detector position IN potentionmeter = 34.89mm
Rutherford rate fluctuates, around 100 counts peak height
Merger OK
Data rate = 1356 kb/sec
Leakage current = 3.234 uA
Temperatures OK
System checks OK - wr fails
Base Current Difference
aida01 fault 0xab60 : 0xab8c : 44
aida02 fault 0x88d3 : 0x88fe : 43
aida03 fault 0x23aa : 0x23d6 : 44
aida04 fault 0x15d7 : 0x1603 : 44
aida05 fault 0x2f03 : 0x2f0f : 12
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring pressure = 4.4 E-10 mbar |
Attachment 1: 2022-03-01_03-21-28_Exrate.png
|
|
Attachment 2: 2022-03-01_03-22-30_ExTotal.png
|
|
Attachment 3: 2022-03-01_03-23-06_EyTotal.png
|
|
Attachment 4: 2022-03-01_03-25-54_GoodEvents.png
|
|
Attachment 5: 2022-03-01_03-26-48_Correlations.png
|
|
Attachment 6: 2022-03-01_03-27-23_ADCdataitems.png
|
|
339
|
Tue Mar 1 03:34:28 2022 |
JM, EM | Current status |
Temperature setpoint = 116 k
Detector IN potentiometer = 34.89 mm
Rutherford rate is reduced from previous entries. Avergae peak height appears to be ~50 counts. Fluctuations in peak height and E4 pressure are quite high. Temperature has been cycled, nozzle position has been optimised and different temperatures have been trialled however density cannot be maintained at a decent level.
Merger OK
Data rate ~ 1500 kb/sec
Leakage current = 3.2295 uA
Temperatures OK
System checks OK - wr fails
Base Current Difference
aida01 fault 0xab60 : 0xab8c : 44
aida02 fault 0x88d3 : 0x88fe : 43
aida03 fault 0x23aa : 0x23d6 : 44
aida04 fault 0x15d7 : 0x1603 : 44
aida05 fault 0x2f03 : 0x2f0f : 12
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring pressure = 4.4 E-10 mbar
|
Attachment 1: 2022-03-01_04-37-03-Ex-rate.png
|
|
Attachment 2: 2022-03-01_04-41-46-correlation8.png
|
|
Attachment 3: 2022-03-01_04-41-15-goodEvents.png
|
|
Attachment 4: 2022-03-01_04-40-49-ADCitems.png
|
|
Attachment 5: 2022-03-01_04-43-53-XYtotal.png
|
|
Attachment 6: 2022-03-01_04-44-33-EyTotal.png
|
|
Attachment 7: 2022-03-01_04-45-07-ExTotal.png
|
|
357
|
Tue Mar 1 22:10:26 2022 |
David, George | Current status |
Ring current 6.67e7 particles/s
pressure at S1: 2.3-2.8e-8
Merger, Tape server (WS3): 1574 kB/s, writing ok
Leakage current (WS4): 3.20 uA
DAQ (WS2): temperature checked, clock status: ok, adc calibration: ok, white rabbit check: failed
Online monitoring (WS 5): done |
Attachment 1: 2022-03-01_23-14-43aida.png
|
|
Attachment 2: 2022-03-01_23-13-57exrate.png
|
|
Attachment 3: 2022-03-01_23-13-40xyto.png
|
|
Attachment 4: 2022-03-01_23-13-10extot.png
|
|
358
|
Tue Mar 1 23:30:04 2022 |
David, George | Current status |
00:30
Pressure in S1 slightly decreased from before but still high enough to measure data.
One new count at 20MeV
Target density at S1: 2e-8 - 2.7e-8
Pressures OK (one small spike just before 00:00 in YR08)
System Checks OK
White rabbit fail diff: 49
Data write speed: ~1.4Mb/s
Beam intensity at injection: ~7.7e7
Cryo Temp Setpoint: 110K
(Beam intensity in ring dropped as I was writing which is why there is no rutherford peak on exRate image)
|
Attachment 1: 2022-03-02_00-39-47xytotal.png
|
|
Attachment 2: 2022-03-02_00-39-20exRate.png
|
|
Attachment 3: 2022-03-02_00-38-44extotal.png
|
|
364
|
Wed Mar 2 14:15:16 2022 |
JM, PM | Current status |
15:15
Temperature setpoint = 113 K
Target density is OK for now. S1 vaies between 2E-8 and 3.5E-8 mbar. E4 between 1E-9 and 1E-8 mbar.
Detector motors functioning well. Moving between 23.1 mm and 67.43mm. Move in and move out counters the same. Correlation scalar 8 for detectors moving is seen on aida1
Rutherford rate peak varies between 10^3 peak height when fully in and ~10^2 when fully out. No peak from halo seen in rate.
Beam intensity ~ 1E8
Merger OK
Data rate ~ 1700 kb/sec
Run is Run 87 - started 10:34
Data spy restarted at 15:00 so only include Run87 data.
Leakage current = 3.205 uA
Fee Temperatures OK
System checks OK - wr fails
Base Current Difference
aida01 fault 0xab60 : 0xaba3 : 67
aida02 fault 0x88d3 : 0x8914 : 65
aida03 fault 0x23aa : 0x23ed : 67
aida04 fault 0x15d7 : 0x161a : 67
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring Pressure = 5.5E-10 mbar
16:45
Temperature setpoint = 113 K
Target density is OK for now. S1 vaies between 2E-8 and 3E-8 mbar. E4 between 1E-9 and 1E-8 mbar.
Detector motors functioning well. Moving between 23.1 mm and 67.43mm. Move in and move out counters the same. Correlation scalar 8 for detectors moving is seen on aida1
Rutherford rate peak varies between 10^3 peak height when fully in and ~10^2 when fully out. No peak from halo seen in rate.
Beam intensity ~ 1E8
Merger OK
Data rate ~ 1300 kb/sec
Temperatures OK
Leakage current ~3.205 uA
System chacks OK - wr fail
Base Current Difference
aida01 fault 0xab60 : 0xaba3 : 67
aida02 fault 0x88d3 : 0x8914 : 65
aida03 fault 0x23aa : 0x23ed : 67
aida04 fault 0x15d7 : 0x161a : 67
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring pressure ~ 4.4 E-10 mbar
Statistics similar to previous check.
|
Attachment 1: 2022-03-02_15-23-02-ExRate.png
|
|
Attachment 2: 2022-03-02_15-24-14-XYtotal.png
|
|
Attachment 3: 2022-03-02_15-28-04-correlation8.png
|
|
Attachment 4: 2022-03-02_15-27-13-ADCitems.png
|
|
Attachment 5: 2022-03-02_15-26-34-GoodEvents.png
|
|
Attachment 6: 2022-03-02_16-51-19-aida1-4-stats.png
|
|
365
|
Wed Mar 2 17:21:35 2022 |
JM, PM | Current status |
18:21
Temperature setpoint = 113k
Target density OK S1 varies between 2E-8 and 2.5E-8 mbar, E4 between 1E-9 and 5E-9 mbar. (slightly lower than previous check)
Detectors moving well between 23.1mm and 67.43 mm, Move in and move out counters the same. Correlation scalar is on aida1
Rutherford peak varies between 10^3 and 10^2 depending on detector position. No halo peak in rate. Noise peak varies from ~100 peak height when fully in to ~400 peak height when fully out.
Beam intensity ~ 1E8
Merger OK
Data rate ~ 1800 kb/sec
Leakage current ~3.21 uA
Temperature OK
System checks OK - wr fails
Base Current Difference
aida01 fault 0xab60 : 0xaba3 : 67
aida02 fault 0x88d3 : 0x8914 : 65
aida03 fault 0x23aa : 0x23ed : 67
aida04 fault 0x15d7 : 0x161a : 67
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring pressure = 4.4 E-10 mbar
19:20
Temperature setpoint = 113k
Target density OK S1 varies between 2E-8 and 3E-8 mbar, E4 between 1E-9 and 1E-8 mbar. (slightly higher than previous check)
Detectors moving well between 23.1mm and 67.43 mm, Move in and move out counters the same. Correlation scalar is on aida1
Rutherford peak varies between 10^3 and 10^2 depending on detector position. No halo peak in rate.
Beam intensity ~ 1E8
Merger OK
Data rate ~ 1800 kb/sec
Leakage current ~3.20 uA
Temperature OK
System checks OK - wr fails
Base Current Difference
aida01 fault 0xab60 : 0xaba4 : 68
aida02 fault 0x88d3 : 0x8915 : 66
aida03 fault 0x23aa : 0x23ee : 68
aida04 fault 0x15d7 : 0x161b : 68
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring pressure = 4.3 E-10 mbar
Statistics similar to previous check. |
Attachment 1: 2022-03-02_18-28-37-Exrate.png
|
|
Attachment 2: 2022-03-02_18-29-38-Xytotal.png
|
|
Attachment 3: 2022-03-02_18-30-33-ExTotal.png
|
|
Attachment 4: 2022-03-02_18-32-51-correlation8.png
|
|
Attachment 5: 2022-03-02_18-31-50-GoodEvents.png
|
|
Attachment 6: 2022-03-02_18-31-24-ADCitems.png
|
|
Attachment 7: 2022-03-02_19-48-56-rateIN.png
|
|
Attachment 8: 2022-03-02_19-50-51-rateOUT.png
|
|
366
|
Wed Mar 2 19:25:25 2022 |
JM, PM | Current status |
20:25
Observed short burst of low energy noise on aida1, lasted <30s (attachment 8).
Temperature setpoint = 113 K
Target density OK S1 vaies between 2E-8 and 3E-8 mbar, E4 between 1E-9 and 1E-8 mbar.
Detector moving well between 23.1 mm and 67.43mm, move in and move out counters the same. Correlation scalar on aida1.
Rutherford peak varies between 10^3 and 10^2 depending on detector position, noise peak varies with detector position -> lower further in.
Beam intensity ~1E8
Merger OK
Data rate ~ 1900 Kb/sec
Leakage current = 3.208 uA
Temperatures OK
System checks - wr fails
Base Current Difference
aida01 fault 0xab60 : 0xaba4 : 68
aida02 fault 0x88d3 : 0x8915 : 66
aida03 fault 0x23aa : 0x23ee : 68
aida04 fault 0x15d7 : 0x161b : 68
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring pressure = 5.5 E-10 mbar. Ring pressure tends to oosciilate between ~4 and 5.5 E-10 mbar
21:21
Burst of noise not observed again
Temperature setpoint = 113 K
Target density OK S1 vaies between 2E-8 and 3E-8 mbar, E4 between 1E-9 and 1E-8 mbar.
Detector moving well between 23.1 mm and 67.43mm, move in and move out counters the same. Correlation scalar on aida1.
Rutherford peak varies between 10^3 and 10^2 depending on detector position, noise peak varies with detector position -> lower further in.
Beam intensity ~1E8
Merger OK
Data rate ~ 1900 Kb/sec
Leakage current = 3.206 uA
Temperatures OK
System checks - wr fails
Base Current Difference
aida01 fault 0xab60 : 0xaba4 : 68
aida02 fault 0x88d3 : 0x8915 : 66
aida03 fault 0x23aa : 0x23ee : 68
aida04 fault 0x15d7 : 0x161b : 68
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Avg ring pressure 5.4E-10 mbar |
Attachment 1: 2022-03-02_20-37-20-rateIN.png
|
|
Attachment 2: 2022-03-02_20-36-42-rateOUT.png
|
|
Attachment 3: 2022-03-02_20-38-16-XYtotal.png
|
|
Attachment 4: 2022-03-02_20-39-02-Extotal.png
|
|
Attachment 5: 2022-03-02_20-42-28-ADCitems.png
|
|
Attachment 6: 2022-03-02_20-42-02-goodEvents.png
|
|
Attachment 7: 2022-03-02_20-41-36-correlation8.png
|
|
Attachment 8: 2022-03-02_20-48-27-noise.png
|
|
Attachment 9: 2022-03-02_21-24-40-ExTotalZoom.png
|
|
Attachment 10: 2022-03-02_21-24-07-EYTotalZoom.png
|
|
367
|
Wed Mar 2 21:16:06 2022 |
JM, PM | Current status |
22:16
Temperature setpoint = 113k
Not observed noise burst again
Target density OK, S1 between 2E-8 and 3E-8mbar, E4 between 1E-9 and 1E-8 mbar.
Detector moving well, potentionmer between 23.1 mm and 67.43 mm, move in/move out counters the same. Correlation scalar on aida1
Ruthrford peak varies between 10^3 and 10^2 peak height depending on detector position, noise rate also varies with position.
Beam intensity ~1E8
Merger OK
Data rate ~ 1200 kb/sec
Temperatures OK
System checks OK - wr fails
Base Current Difference
aida01 fault 0xab60 : 0xaba4 : 68
aida02 fault 0x88d3 : 0x8915 : 66
aida03 fault 0x23aa : 0x23ee : 68
aida04 fault 0x15d7 : 0x161b : 68
aida05 fault 0x2f03 : 0x2f20 : 29
White Rabbit error counter test result: Passed 0, Failed 5
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
Ring pressure oscillation is 30 minute cycle on all areas of the ring. |
671
|
Thu Apr 10 07:02:12 2025 |
CB (remotely) | Current logging enabled - sort of |
There appear to be issues with Shahab's caenlogging program. It crashes occasionally, with an error indicating the number of columns in the terminal is incorrect. We will have to ask Shahab.
For now
1. Start the caen_screen.sh script on nnrp1 (aka CARMEMotorsPi - why was it renamed? It should stay as CARMEMotorsPi to avoid duplicate names).
This starts connect to the caen power module via TTY inside a screen called caenhv. Currently on desktop 1
2. On nnrpi1, start the caenlogging program --settings caenlogger_settings_gsi.toml
3. On carme-gsi, in ~/Programs/caenlogger start the gnuplot script
If the caenlogger script crashes, restart it on the second terminal
More investigations required on the crash. |