408
|
Sat Jun 24 10:38:42 2023 |
JM, TD | Section 9 closed, begin pumping |
23/06/23
The CARME chamber and interaction chamber were connected and section 9 has been fully closed. New valves have been installed on the interaction chamber, in addition to new diamond windows. A leak cart has been connected to the backing line of CARME. Before begining the pumpdown, the manual valve between the interaction chamber and the electron target was closed.
The pumpdown was started at 17:15, the valve on the scroll pump was very slowly opened to not damage the berylium window. By 18:00 a pressure of 1 mbar was achieved and the turbopump was activated. By 18:20 a pressure of 2E-3 mbar was achieved, the ion gauge was activated and briefly flashed 5E-4 mbar but this is over its limit so the gauge shutdown. At 19:20, 1.3E-3 mbar was achieved, the ion gauge was activated but the reading was 1.7E-4 mbar which is still over its limit. The turbo fan was not working and temperatures on the turbo were ~50C and increasing, so it was decided to close the main gate valve and turno off the turbo overnight in order to fix the fan.
A bias test of the detectors was also performed while in vacuum, T=20.5 C. All detectors show expected V-I curve and the leakage current is ~1-2 uA lower than the previous test in air, probably due to reduced light leaks. The top right detector did initially produce >10 uA at 20 V but the adaptor card was not properly seated on the p-n feed-through. Seating the adaptor card correctly yielded the expected V-I curve. Further inspection of the pins on this feed-through will be performed tomorrow.
24/06/23
Overnight the pressure had increased to 0.8 mbar. Opening the gate valve decreased the pressure by an order of magnitude. The turbo fan was fixed and the turbo was re-activated, the pressure dropped to 1E-3 mbar level within 20 minutes. Ion gauge was turned on after 30 minutes with a pressure of 3.7e-5 mbar. Pump down plot and comparison to previous pump downs will be uploaded in a seperate elog. Turbo pump maximum temperature reads 46C and has been consistent for several hours.
V-I tests completed for the right hand side detectors. Leakage currents ~0.4 uA higher than yesterday which is consistent with the increase in temperature of ~ 0.5-0.7C. |
Attachment 1: 23-06-23-VI.png
|
|
Attachment 2: 24-06-23-VI.png
|
|
222
|
Mon Feb 21 21:04:08 2022 |
SHB, JM, CB, OH (remotely) | Schottky RF view from the beam inside |
|
Attachment 1: pio-2022.02.21.21.54.46.895.PNG
|
|
662
|
Sat Apr 5 16:13:32 2025 |
TD | Saturday 5 April |
12.00 CARME IE514 ion gauge 4.2e-11 mbar
CARME IE514 OFF
aida08 red coolant line and HDMI cable found disconnected - blue coolant line, network cable and FEE^$ power are still connected
Now re-connected
13.57 DSSSD bias & leakage current - attachment 1
leakage currents high (especially DSSSD#1 and DSSSD#3) cf. most recent values from June 2024 values https://elog.ph.ed.ac.uk/CARME/654
another ion gauge on with line of sight to DSSSDs?
FEE64 power ON
FEE64 temps OK - attachment 2
All system wide checks OK - attachments 3-6
WR timestamps OK - attachment 7
18.00 DSSSD bias OFF
FEE64 power OFF |
Attachment 1: Screenshot_from_2025-04-05_13-57-06.png
|
|
Attachment 2: Screenshot_from_2025-04-05_16-00-19.png
|
|
Attachment 3: Screenshot_from_2025-04-05_16-01-31.png
|
|
Attachment 4: Screenshot_from_2025-04-05_16-04-22.png
|
|
Attachment 5: Screenshot_from_2025-04-05_16-04-36.png
|
|
Attachment 6: Screenshot_from_2025-04-05_16-04-47.png
|
|
Attachment 7: Screenshot_from_2025-04-05_16-06-41.png
|
|
449
|
Sat Sep 30 10:58:14 2023 |
TD | Saturday 30 September |
10.50 CARME IE514 7.14e-10 mbar, ambient temperature 20.8 deg C
CARME bakeout monitor, ELOTECH R4000 temperature controller front panel and Grafana bakeout dashboard- attachments 1-3
Change ELOTECH R4000 set points for CARME heaters #1 and #3 to 125 deg C
14.20 CARME IE514 8.25e-10 mbar, ambient temperature 20.8 deg C
CARME bakeout monitor, ELOTECH R4000 temperature controller front panel and Grafana bakeout dashboard- attachments 4-6
Change ELOTECH R4000 set points for CARME heaters #1 and #3 to 150 deg C
18.30 CARME IE514 1.29e-9 mbar, ambient temperature 20.7 deg C
CARME bakeout monitor, ELOTECH R4000 temperature controller front panel and Grafana bakeout dashboard- attachments 7-9
Change ELOTECH R4000 set points for CARME heaters #1 and #3 to 175 deg C and heater #2 to 125 deg C |
Attachment 1: 1000007236.jpg
|
|
Attachment 2: 1000007239.jpg
|
|
Attachment 3: 1000007237.jpg
|
|
Attachment 4: 1000007251.jpg
|
|
Attachment 5: 1000007250.jpg
|
|
Attachment 6: 1000007252.jpg
|
|
Attachment 7: 1000007256.jpg
|
|
Attachment 8: 1000007257.jpg
|
|
Attachment 9: 1000007255.jpg
|
|
548
|
Fri Apr 26 23:41:16 2024 |
TD | Saturday 27 April |
00.34 DAQ continues data file R3_9
DSSSD bias & leakage current OK - attachment 1
FEE64 temps OK - attachment 2
ADC data item stats OK - attachment 3
per FEE64 Rate spectra - attachment 4
Merger, TapeServer etc - attachments 5-7
06.41 DAQ continues data file R3_10
DSSSD bias & leakage current OK - attachment 8
FEE64 temps OK - attachment 9
ADC data item stats OK - attachment 10
per FEE64 Rate spectra - attachment 11
Merger, TapeServer etc - attachments 12-14
10.56 DAQ continues data file R3_11
DSSSD bias & leakage current OK - attachment 15
FEE64 temps OK - attachment 16
ADC data item stats OK - attachment 17
per FEE64 Rate spectra - attachment 18
Merger, TapeServer etc - attachments 19-21
14.33 DAQ continues data file R3_12
DSSSD bias & leakage current OK - attachment 22
FEE64 temps OK - attachment 23
ADC data item stats OK - attachment 24
per FEE64 Rate spectra - attachment 25
Merger, TapeServer etc - attachments 26-28
18.31 DAQ continues data file R3_13
DSSSD bias & leakage current OK - attachment 29
FEE64 temps OK - attachment 30
ADC data item stats OK - attachment 31
per FEE64 Rate spectra - attachment 32
per FEE64 1.8.L spectra - attachments 33-34
Merger, TapeServer etc - attachments 35-37 |
Attachment 1: Screenshot_from_2024-04-27_00-34-52.png
|
|
Attachment 2: Screenshot_from_2024-04-27_00-35-25.png
|
|
Attachment 3: Screenshot_from_2024-04-27_00-36-07.png
|
|
Attachment 4: Screenshot_from_2024-04-27_00-36-34.png
|
|
Attachment 5: Screenshot_from_2024-04-27_00-36-49.png
|
|
Attachment 6: Screenshot_from_2024-04-27_00-37-14.png
|
|
Attachment 7: Screenshot_from_2024-04-27_00-37-37.png
|
|
Attachment 8: Screenshot_from_2024-04-27_06-41-34.png
|
|
Attachment 9: Screenshot_from_2024-04-27_06-42-06.png
|
|
Attachment 10: Screenshot_from_2024-04-27_06-42-48.png
|
|
Attachment 11: Screenshot_from_2024-04-27_06-43-08.png
|
|
Attachment 12: Screenshot_from_2024-04-27_06-43-19.png
|
|
Attachment 13: Screenshot_from_2024-04-27_06-43-36.png
|
|
Attachment 14: Screenshot_from_2024-04-27_06-43-50.png
|
|
Attachment 15: Screenshot_from_2024-04-27_10-55-54.png
|
|
Attachment 16: Screenshot_from_2024-04-27_10-56-21.png
|
|
Attachment 17: Screenshot_from_2024-04-27_10-56-38.png
|
|
Attachment 18: Screenshot_from_2024-04-27_10-57-26.png
|
|
Attachment 19: Screenshot_from_2024-04-27_10-57-45.png
|
|
Attachment 20: Screenshot_from_2024-04-27_10-58-04.png
|
|
Attachment 21: Screenshot_from_2024-04-27_10-58-16.png
|
|
Attachment 22: Screenshot_from_2024-04-27_14-33-33.png
|
|
Attachment 23: Screenshot_from_2024-04-27_14-34-04.png
|
|
Attachment 24: Screenshot_from_2024-04-27_14-34-52.png
|
|
Attachment 25: Screenshot_from_2024-04-27_14-35-08.png
|
|
Attachment 26: Screenshot_from_2024-04-27_14-35-19.png
|
|
Attachment 27: Screenshot_from_2024-04-27_14-35-50.png
|
|
Attachment 28: Screenshot_from_2024-04-27_14-35-34.png
|
|
Attachment 29: Screenshot_from_2024-04-27_18-31-01.png
|
|
Attachment 30: Screenshot_from_2024-04-27_18-31-22.png
|
|
Attachment 31: Screenshot_from_2024-04-27_18-31-40.png
|
|
Attachment 32: Screenshot_from_2024-04-27_18-32-27.png
|
|
Attachment 33: Screenshot_from_2024-04-27_18-33-04.png
|
|
Attachment 34: Screenshot_from_2024-04-27_18-33-47.png
|
|
Attachment 35: Screenshot_from_2024-04-27_18-34-04.png
|
|
Attachment 36: Screenshot_from_2024-04-27_18-34-22.png
|
|
Attachment 37: Screenshot_from_2024-04-27_18-34-44.png
|
|
287
|
Sat Feb 26 04:36:38 2022 |
RSS TD | Saturday 26 February 07:00-15:00 |
05:36 E4 pressure c. 1.7e-9mbar *not* 3e-10mbar as reported earlier
See attachments 1 & 2
E4 pressure not < 5e-10mbar - wait to startup gas jet target
06:47 RSS & TD on-site
07.00 E4 1.70e-9mbar
S3 9.42e-10mbar
07:00 Checks: all good except for the white rabbit decoder status.
07:15 All histograms cleared in the run control tab.
07.32 E4 1.60e-9mbar
S3 9.00e-10mbar
08:06 E4 1.49e-9mbar
S3 8.59e-10mbar
08:30 E4 1.43e-9mbar
S3 8.34e-10mbar
09:00 E4 1.35e-9mbar
S3 8.03e-10mbar
09:30 E4 1.29e-9mbar
S3 7.70e-10mbar
10:00 E4 1.23e-9mbar
S3 7.49e-10mbar
10:00 Checks: all good except for the white rabbit decoder status.
10:30 E4 1.18e-9mbar
S3 7.22e-10mbar
11:00 E4 1.14e-9mbar
S3 7.02e-10mbar
11:20 Compressor was switched on.
11:30 E4 1.10e-9mbar
S3 6.82e-10mbar
12:00 E4 1.35e-9mbar
S3 6.64e-10mbar
12:30 E4 1.23e-9mbar
S3 6.46e-10mbar
14:45 Checks: all good except for the white rabbit decoder status.
17.20 System wide checks OK *except* WR decoder status
Base Current Difference
aida01 fault 0xab60 : 0xab85 : 37
aida02 fault 0x88d3 : 0x88f7 : 36
aida03 fault 0x23aa : 0x23cf : 37
aida04 fault 0x15d7 : 0x15fc : 37
aida05 fault 0x2f03 : 0x2f0e : 11
White Rabbit error counter test result: Passed 0, Failed 5
FEE64 temperatures OK - attachment 26
stats (adc data items) OK - attachment 27
DSSSD bias -100V leakage current -3.31uA |
Attachment 1: Capture_1.PNG
|
|
Attachment 2: Capture_2.PNG
|
|
Attachment 3: 7-spectrum-id8-2022-02-26_07-14-49.png
|
|
Attachment 4: 7-spectrum-id7-2022-02-26_07-14-12.png
|
|
Attachment 5: 7-spectrum-42022-02-26_07-13-21.png
|
|
Attachment 6: 7-spectrum-3-2022-02-26_07-12-50.png
|
|
Attachment 7: 7-spectrum2-2022-02-26_07-12-02.png
|
|
Attachment 8: 7-spectrum1-2022-02-26_07-10-55.png
|
|
Attachment 9: 7-spectrum-2022-02-26_07-10-10.png
|
|
Attachment 10: 7-correlation-2022-02-26_07-09-11.png
|
|
Attachment 11: 7-adc-2022-02-26_07-08-53.png
|
|
Attachment 12: 7-good-events-2022-02-26_07-08-27.png
|
|
Attachment 13: 10-id2-2022-02-26_10-06-08.png
|
|
Attachment 14: 10-id1-2022-02-26_10-04-01.png
|
|
Attachment 15: 10-correlation-2022-02-26_10-03-08.png
|
|
Attachment 16: 10-adc-2022-02-26_10-02-47.png
|
|
Attachment 17: 10-good-2022-02-26_10-02-27.png
|
|
Attachment 18: IMG_7113.JPG
|
|
Attachment 19: IMG_7112.JPG
|
|
Attachment 20: IMG_7111.JPG
|
|
Attachment 21: IMG_7114.JPG
|
|
Attachment 22: 2-adc-2022-02-26_14-45-37.png
|
|
Attachment 23: 2-good-2022-02-26_14-45-16.png
|
|
Attachment 24: 2-correlation-2022-02-26_14-44-56.png
|
|
Attachment 25: 2-id1-2022-02-26_14-48-13.png
|
|
Attachment 26: Screenshot_from_2022-02-26_17-18-11.png
|
|
Attachment 27: Screenshot_from_2022-02-26_17-18-25.png
|
|
477
|
Mon Jan 22 16:29:31 2024 |
TD | Saturday 20 January |
11.00 No WLAN/WiFi
came-gsi private LAN appears to be OK
AIDA FEE64 power ON
11.09 FEE64 temps - attachment 1
aida06 low, otherwise OK
11:10 CARME IE514 4.0e-11 mbar, temp unknown
IE514 OFF
11.22 Detector bias ON - attachment 2
found ch 2 up ramp set to 63V/s! changed to 1V/s
11.30 system wide checks OK except
aida02 & aida06 fail WR decoder status
WR ts control - attachments 3-4
12.02 ASIC settings updated for aida09-aida16
ASIC settings DB key 2024Jan20-12:02:08
Unable to save DB key to Options - use ASIC settings restore as workaround
12.30 Layouts for 16x FEE64s created - directory ~/LayOut/16xFEE64
Note layouts include spectra from aida07, aida08 and aida16 so an error message will be generated but it can be ignored
12.40 per FEE64 Rate spectra - attachments 5-6
per FEE64 p+n & nn 1.8.W spectra - attachments 7-8
12.57 per FEE64 p+n & n+n 1.8.W spectra 20us FSR - attachments 9-10
Test signals
EG&G Ortec 448 Pulser
Amplitude 90,000
attenuation x10
Normalise 5.00
t_r 100ns
tau_d 50us
Frequency 50Hz ( relay line freq )
Polarity +
test + direct from 448, test - from Cooknell SA1 Sum & Invert
Test + daisy chain -> aida01 - aida02 - aida05 - aida06 - 50 Ohm
Test - daisy chain -> aida03 - aida04 - aida07 - aida08 - 50 Ohm
aida08 ( DSSSD #2 n+n Ohmic bias ) LK2 removed, LK1 added
aida04 ( DSSSD #1 n+n Ohmic bias ) no LK2 fitted, LK1 added
DSSSD#1 SHV - 2x Lemo 00.250 cable replaced by standard SHV - BNC/BNC Lemo-00.250 coaxial cable
Adaptor PCB ground daisy chain aida01 - aida02 - aida03 - aida04
DSSSD#2 - DSSSD#4 HV cabling unchanged
15.30 Power ON
System wide checks OK except
aida02 fail WR decoder status - attachment 11
FEE64 temperatures - attachment 12
aida06 ASIC temp low
DSSSD bias - attachment 13
ADC data item stats - attachment 14
slow comparator 0x64
per FEE64 Rate and 1.8.W 20us FSR spectra - attachments 15-17
16.00 per FEE64 rate spectra - attachment 18
16.15 per FEE64 Rate and 1.8.W 20us FSR spectra - attachments 19-21
17.36 Switch from VETAR2 ( all MACB modules mode 0x3 ) to NIM WR emulator ( root MACB module 0xd, all others 0x3 )
All FEE64s fail Global clock and WR decoder - attachments 22-24
WR timetsamp control - attachment 25
18.10 Noted MACB #4 mode actually set to 0x2 *not* 0x3 - changed to 0x3
Re-test with VETAR2 WR - same behaviour as observed this morning - see attachments 26-30
Summary
n+n sides appear to be OK but noisy
no data observed from aida06 & aida11
aida11 known to be disconnected from adaptor PCB
aida15 asic #2 may be faulty
aida01, aida05, aida12 - data suggest FEE64/adaptor PCB alignment issue ?
aida02, aida03, aida09, aida14 - data suggest adaptor PCB damaged pins ?
aida07 - network config issue ?
aida08 - not completing boot
need to attach sys console to aida07 & aida08 to diagnose problem. |
Attachment 1: Screenshot_from_2024-01-20_11-10-03.png
|
|
Attachment 2: Screenshot_from_2024-01-20_11-22-20.png
|
|
Attachment 3: Screenshot_from_2024-01-20_11-31-31.png
|
|
Attachment 4: Screenshot_from_2024-01-20_11-31-39.png
|
|
Attachment 5: Screenshot_from_2024-01-20_12-18-27.png
|
|
Attachment 6: Screenshot_from_2024-01-20_12-18-37.png
|
|
Attachment 7: Screenshot_from_2024-01-20_12-44-07.png
|
|
Attachment 8: Screenshot_from_2024-01-20_12-42-58.png
|
|
Attachment 9: Screenshot_from_2024-01-20_12-59-12.png
|
|
Attachment 10: Screenshot_from_2024-01-20_13-00-18.png
|
|
Attachment 11: Screenshot_from_2024-01-20_15-30-48.png
|
|
Attachment 12: Screenshot_from_2024-01-20_15-40-44.png
|
|
Attachment 13: Screenshot_from_2024-01-20_15-40-58.png
|
|
Attachment 14: Screenshot_from_2024-01-20_15-43-06.png
|
|
Attachment 15: Screenshot_from_2024-01-20_15-44-18.png
|
|
Attachment 16: Screenshot_from_2024-01-20_15-45-14.png
|
|
Attachment 17: Screenshot_from_2024-01-20_15-45-47.png
|
|
Attachment 18: Screenshot_from_2024-01-20_16-03-18.png
|
|
Attachment 19: Screenshot_from_2024-01-20_16-15-24.png
|
|
Attachment 20: Screenshot_from_2024-01-20_16-15-52.png
|
|
Attachment 21: Screenshot_from_2024-01-20_16-16-37.png
|
|
Attachment 22: Screenshot_from_2024-01-20_17-36-22.png
|
|
Attachment 23: Screenshot_from_2024-01-20_17-36-28.png
|
|
Attachment 24: Screenshot_from_2024-01-20_17-36-43.png
|
|
Attachment 25: Screenshot_from_2024-01-20_17-38-22.png
|
|
Attachment 26: 1000009752.jpg
|
|
Attachment 27: 1000009753.jpg
|
|
Attachment 28: 1000009754.jpg
|
|
Attachment 29: 1000009755.jpg
|
|
Attachment 30: 1000009756.jpg
|
|
399
|
Sat May 13 11:34:25 2023 |
TD, RSS | Saturday 13 May |
12.34 DSSSD#3 bias -149.95V leakage current -6.763uA temperature +21.2 deg C ( https://web-docs.gsi.de/~lestinsk/tempplot.php )
CRYRING cave lights ON - leakage current measurements @ 60s intervals
6.763
6.764
6.764
6.765
6.767
6.767
6.767
6.767
6.768
6.769
12.48 CRYRING cave lights OFF
13.03 leakage current -6.771uA - no obvious change due to the CRYRING cave lighting
17.31 DSSSD#3 bias -149.95V leakage current -6.658uA temperature +21.0 deg C ( https://web-docs.gsi.de/~lestinsk/tempplot.php )
leakage current measurements @ 60s intervals
6.656
6.655
6.655
6.655
6.654
6.652
6.652
6.651
6.650
6.649
CRYRING cave temperature variations/cycles ~0.1 deg C at +21 deg C ambient temperature can produce leakage current variations ~60nA for a leakage current of ~6.6uA
td@winder:~/f77$ ./jgen
*** Enter reference temperature (K)
294
*** Enter minimum temperature, maximum temperature
*** and temperature increment (K)
290 298 0.1
Temperature (K) Ratio (J(T)/J(294.0))
290.0 0.702
290.1 0.708
290.2 0.714
290.3 0.721
290.4 0.727
290.5 0.734
290.6 0.740
290.7 0.747
290.8 0.754
290.9 0.761
291.0 0.767
291.1 0.774
291.2 0.781
291.3 0.788
291.4 0.795
291.5 0.802
291.6 0.809
291.7 0.817
291.8 0.824
291.9 0.831
292.0 0.839
292.1 0.846
292.2 0.854
292.3 0.861
292.4 0.869
292.5 0.877
292.6 0.884
292.7 0.892
292.8 0.900
292.9 0.908
293.0 0.916
293.1 0.924
293.2 0.932
293.3 0.941
293.4 0.949
293.5 0.957
293.6 0.966
293.7 0.974
293.8 0.983
293.9 0.991
294.0 1.000
294.1 1.009
294.2 1.018
294.3 1.027
294.4 1.036
294.5 1.045
294.6 1.054
294.7 1.063
294.8 1.072
294.9 1.082
295.0 1.091
295.1 1.101
295.2 1.110
295.3 1.120
295.4 1.130
295.5 1.139
295.6 1.149
295.7 1.159
295.8 1.169
295.9 1.179
296.0 1.190
296.1 1.200
296.2 1.210
296.3 1.221
296.4 1.231
296.5 1.242
296.6 1.253
296.7 1.263
296.8 1.274
296.9 1.285
297.0 1.296
297.1 1.308
297.2 1.319
297.3 1.330
297.4 1.342
297.5 1.353
297.6 1.365
297.7 1.376
297.8 1.388
297.9 1.400
STOP *** Program ends |
107
|
Tue Oct 5 11:44:11 2021 |
CB, TD | SAES manuals & cable drawings |
NEG power
SIP power
UHV1400 operating instructions |
Attachment 1: NEG_POWER_user_manual.pdf
|
|
Attachment 2: operating_instructions_SIP_POWER.pdf
|
|
Attachment 3: operating_instructions_UHV1400-UHV700_wafer_modules.pdf
|
|
Attachment 4: RFION9901221.pdf
|
|
Attachment 5: RFKIT9903220.pdf
|
|
Attachment 6: RFKIT9904220_(2).pdf
|
|
Attachment 7: RFKIT9905220_(2).pdf
|
|
Attachment 8: RFNEG9904221.pdf
|
|
3
|
Tue Aug 18 17:35:20 2020 |
TD | SAES UHV1400 pin assignment |
Using a digital multimeter confirmed continuity for SAES-supplied
power cable (Amphenol (circular) 6-pin plug to Han 05 plug) for
the following pins
pin A - pin 5
pin F - pin 3
which correspond to the type K chromel/alumel thermocouple contacts
and Neg PSU output connector pins for the thermocouple (e.g. see p.34
of Neg PSU manual). |
Attachment 1: NEG_POWER_user_manual.pdf
|
|
Attachment 2: 20200813_121944_(1).jpg
|
|
Attachment 3: 20200813_134548.jpg
|
|
Attachment 4: RFKIT9904220_(2).pdf
|
|
Attachment 5: RFKIT9905220_(2).pdf
|
|
558
|
Wed May 29 16:46:21 2024 |
NP, AB, PM | Run7 - Measurements only with ion beam. |
18:19 Start or run 7: In beam has been optimzed up to now by Zoran and Claude. ~7E7 particles at injection. From now on, only the injection beamline is optimized by Operators.
Detectors IN (left arm pot =18.86 mm, right arm pot = 19.80 mm)
No Target
Beam is in the ring for ~6 seconds and total cycle time between injections is ~8 seconds
Injection beamline is still in process of optimization by operators
21:30
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK
Leakage current stable
23:30
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK
Leakage current stable
01:30
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK
Leakage current stable
Beam and Electron Coller failure (~00:35); Beam back (~01:08); Cooler back (~01:20)
03:00
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK
Leakage current stable
Beam and Electron Coller failure (~01:35); Beam back (~01:55); Cooler back (~01:55)
04:40
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK
Leakage current stable
06:00
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK
Leakage current stable
07:00
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK
Leakage current stable
|
Attachment 1: Screenshot_from_2024-05-29_21-49-57.png
|
|
Attachment 2: Screenshot_from_2024-05-29_21-50-14.png
|
|
556
|
Wed May 29 08:10:32 2024 |
JM | Run4 - alpha background |
09:00
R4 alpha background continues, currently R4_12
Temps OK, system checks unchanged from previous status (OK, aida16 clock and ADC issue) |
518
|
Thu Feb 15 21:07:00 2024 |
JM, AB, AR, OF | Run33 - cooled beam on target |
22:27
Run 33 started cooled beam on target.
Detectors IN (left arm pot =36.5 mm, right arm pot = 70.49 mm)
Electron cooler voltage 754.945 this corresponds to 1.3499 MeV/u. This is above the resonance of interest but cooler settings cannot be altered without major beam tuning.
Target density is 4.1e11 at/cm2, ring pressure looks OK YR09 1.2e-10 mbar.
Horizontal beam position is -7.5 mm, vertical beam position is now 0.0 mm. This results in no beam halo in bottom left detector which should be good for seeing Rutherford.
Beam is in the ring for 8 seconds and total cycle time between injections is 12 seconds
FEE temperatures are OK. 3 FEEs are slightly too. (attachment 1)
System checks are OK.
Statistics look OK. High in the top left detector which is expected as majority of halo is there. (attachment 2+3)
Merger OK Tape server OK - 851 kB/s
Merger statistics (attachment 4)
Leakage current is increasing but is likely temperature related. ML temp plot has same pattern and it is a WARM day. (attachment 5)
Monitoring code - xy plot and Ey energy histograms (attachment 6+7). Rutherford rate looks higher -> looking peaky
00:14
System checks all OK
FEE temperatures unchanged. Tom has now set up plotting of aida09 fee temperatures in the second workspace.
Merger OK, tape server OK.
Target density remains about 4.1e11 at/cm2. Ring pressure ~1e-10 mbar. YR08 read close to 1e-9 but beam is able to circulate in the ring so will not change for now.
Rate statistics in all FEEs remain unchanged.
02:10
System checks OK
FEE temperatures unchanged
Merger OK
Target density OK. Ring pressure not great but cannot change remotely due to script not in screen and poor coding.
Beam stil on.
Leakage current trending up.
04.00 FEE64 temperatures - attachments 10-11
aida09 ( and other FEE64s ) on downward trend
CRYRING hall temperature - attachment 12
05:00
Ring pressure worsening and target getting thinner (4E11). Tweak nozzle when in the lab physically.
Beam still on around 3E7
Temperatures & Leakage currents OK. Same trends
Merger continues OK
Stats OK
System wide checks OK
07:00
Fee temperatures have reduced - all green
System check OK
ADC data items statistics unchanged.
Merger and tape server OK
Leakage current plot (attachment 15)
monitoring plots - xy plot and EY energy histogram (attachment 16+17 )
Pressure has dropped slightly over the last hour, target density remains stable 4.2e11. Beam transformer has dropped off in the last hour -> less beam target interactions -> lower pressure?
We still see count rates in the detectors.
08:18
Inconsistent beam into the ring continues. HKR are aware.
Temperatures OK
System checks OK
10:00
Beam is much more consistent now. We observe vast majority of shots injected into the ring and corresponding rate increase in the detectors via the monitoring code.
FEE temperatures OK
System wide checks OK
Merger and tape server OK
New merger stats (attachment 18)
ADC data item statistics OK
Leakage current plot (attachment 19)
Attachments 20+21: Cooler settings before and after changing to the new beam energy of 1.3435 MeV/u (copy from GSI olog)
10:24
Run33 stopped. Run 34 started. Changing beam energy to be on resonance.
Detectors moved fully out. Target moved out
Final spectrum browser stat for all FEEs before run ended (attachment 22) |
Attachment 1: Screenshot_from_2024-02-15_22-15-04.png
|
|
Attachment 2: Screenshot_from_2024-02-15_22-17-29.png
|
|
Attachment 3: Screenshot_from_2024-02-15_22-18-21.png
|
|
Attachment 4: Screenshot_from_2024-02-15_22-20-52.png
|
|
Attachment 5: Screenshot_from_2024-02-15_22-21-43.png
|
|
Attachment 6: Screenshot_from_2024-02-15_22-24-12.png
|
|
Attachment 7: Screenshot_from_2024-02-15_22-24-21.png
|
|
Attachment 8: Screenshot_from_2024-02-16_02-07-38.png
|
|
Attachment 9: Screenshot_from_2024-02-16_02-10-16.png
|
|
Attachment 10: Screenshot_from_2024-02-16_04-03-13.png
|
|
Attachment 11: Screenshot_from_2024-02-16_04-04-08.png
|
|
Attachment 12: Capture.PNG
|
|
Attachment 13: Screenshot_from_2024-02-16_05-04-19.png
|
|
Attachment 14: Screenshot_from_2024-02-16_05-04-47.png
|
|
Attachment 15: Screenshot_from_2024-02-16_07-06-22.png
|
|
Attachment 16: Screenshot_from_2024-02-16_07-06-40.png
|
|
Attachment 17: Screenshot_from_2024-02-16_07-07-40.png
|
|
Attachment 18: Screenshot_from_2024-02-16_10-07-21.png
|
|
Attachment 19: Screenshot_from_2024-02-16_10-08-02.png
|
|
Attachment 20: olog-before.pdf
|
|
Attachment 21: olog-after.pdf
|
|
Attachment 22: Screenshot_from_2024-02-16_10-13-24.png
|
|
516
|
Thu Feb 15 12:40:52 2024 |
JM. CB, AR, AB | Run30 - uncooled beam on target |
Run started 13:40
Target density 4.3e11, Beam injection is ~1e7, Pressure in the ring is ~1e-10 mbar.
Detectors have been moved IN with servo motors only (left arm pot = 36.5 mm, right arm pot =37.7 mm
Time between injections is 12 seconds. Beam is in the ring for 3 seconds of the cycle.
System wide checks OK
Temperatures OK
Leakage current plot attached -> temperature fluctuations (attachment 1)
ADC data items stats (attachment 2)
Merger and tape server OK
All histograms and statistics zero'd
14:17
Cave access required. No beam
15:24
No beam. System wide checks ok. Merger ok.
Attatchment 3: stats
Attatchment 4: Temperatures generally okay, 3 were red before reload, one red after reload.
Attatchment 5-8: Waveforms. Layout 7 and 8 look different from yesterday, consistnely different after several reloads.
Attatchment 9: Leakage current
Attatchment 10: LOWENERGYEXEY PLOTS
17:12
Tunning beam. System wide checks ok. Merger ok.
Attatchment 11: stats
Attatchment 12: Temperatures generally okay, 3 were red reloaed and stayed red but not so high.
Attatchment 13-16: Waveforms. Were the same as previous check.
Attatchment 17: Leakage current. They are tunning the beam so its a little bit changed.
19:28
Still hvae unstable noncooled beam, currently being worked on.
Tunning beam. System wide checks ok. Merger ok.
Attatchment 18: Temperatures generally okay, 3 were red reloaed and stayed red but not so high and were slighlty dicrease.
Attatchment 19-22: Waveforms. Were the same as previous check.
Attatchment 23: Leakage current. Slowly increasing from last check in, highest its been but conditions are not great at the moment.
21:10
We did not have any update about the beam. They still work on it.
Tunning beam. System wide checks ok. Merger ok. Stats were ok.
Attatchment 24: Temperatures generally okay, Same 3 were red and they increased a little bit. Nothing too concerning.
Attatchment 25-28: Waveforms. Were the same as previous check.
Attatchment 29: Leakage current. Contining to increase by small amount from last check.
|
Attachment 1: Screenshot_from_2024-02-15_13-45-42.png
|
|
Attachment 2: Screenshot_from_2024-02-15_13-48-57.png
|
|
Attachment 3: Screenshot_from_2024-02-15_15-19-31.png
|
|
Attachment 4: Screenshot_from_2024-02-15_15-19-55.png
|
|
Attachment 5: Screenshot_from_2024-02-15_15-21-25.png
|
|
Attachment 6: Screenshot_from_2024-02-15_15-21-53.png
|
|
Attachment 7: Screenshot_from_2024-02-15_15-22-32.png
|
|
Attachment 8: Screenshot_from_2024-02-15_15-23-13.png
|
|
Attachment 9: Screenshot_from_2024-02-15_15-23-38.png
|
|
Attachment 10: Screenshot_from_2024-02-15_15-24-02.png
|
|
Attachment 11: Screenshot_from_2024-02-15_17-00-51.png
|
|
Attachment 12: Screenshot_from_2024-02-15_17-01-28.png
|
|
Attachment 13: Screenshot_from_2024-02-15_17-03-31.png
|
|
Attachment 14: Screenshot_from_2024-02-15_17-05-14.png
|
|
Attachment 15: Screenshot_from_2024-02-15_17-07-20.png
|
|
Attachment 16: Screenshot_from_2024-02-15_17-09-17.png
|
|
Attachment 17: Screenshot_from_2024-02-15_17-11-12.png
|
|
Attachment 18: Screenshot_from_2024-02-15_19-22-45.png
|
|
Attachment 19: Screenshot_from_2024-02-15_19-23-53.png
|
|
Attachment 20: Screenshot_from_2024-02-15_19-24-43.png
|
|
Attachment 21: Screenshot_from_2024-02-15_19-25-59.png
|
|
Attachment 22: Screenshot_from_2024-02-15_19-26-38.png
|
|
Attachment 23: Screenshot_from_2024-02-15_19-28-03.png
|
|
Attachment 24: Screenshot_from_2024-02-15_21-05-57.png
|
|
Attachment 25: Screenshot_from_2024-02-15_21-07-39.png
|
|
Attachment 26: Screenshot_from_2024-02-15_21-08-10.png
|
|
Attachment 27: Screenshot_from_2024-02-15_21-08-48.png
|
|
Attachment 28: Screenshot_from_2024-02-15_21-09-29.png
|
|
Attachment 29: Screenshot_from_2024-02-15_21-10-01.png
|
|
583
|
Mon Jun 3 21:37:20 2024 |
CN, EM, CB | Run30 |
Beam energy = 3 MeV
Horizontal beam position = 7.0 mm, vertical beam position = 0.0 mm.
Beam intensity at inj ~ 2.5e7
Detectors are positioned IN. Left arm = 36.54, right arm = 64.60 mm
22:52 (Attachments 1 -3)
Beam is very unstable
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK
00:47
Beam is still very unstable.
We checked for beam-target overlap position and found a clear increase in counts at beam position 4.0 mm. We have left the beam at horizontal = 4.0 mm, vertical = 0.0
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK |
Attachment 1: Screenshot_from_2024-06-03_23-15-37.png
|
|
Attachment 2: Screenshot_from_2024-06-03_23-16-18.png
|
|
Attachment 3: Screenshot_from_2024-06-03_23-17-16.png
|
|
491
|
Mon Feb 5 05:45:26 2024 |
CB TD | Run3 - end |
06.40 DAQ continues file G22-88/R3_42
DSSSD bias OK - attachment 1
ambient temperature 16.8 deg C
FEE64 temperatures OK - attachment 2
ADC data item stats - attachments 3
aida15 rate c. 34k ( 2x 'hot' channels, 1x channel missing, all other channels OK )
aida01 rate c. 75k ( 2x 'hot' channels )
per FEE64 Stat spectra - attachments 4-5
Merger and TapeServer - attachments 6-7
All system wide checks OK except
aida02 & aida08 FPGA timestamp error - attachment 8
WR timestamps OK - attachment 9
09.47 DAQ continues file G22-88/R3_46
DSSSD bias OK - attachment 10
ambient temperature 16.8 deg C
FEE64 temperatures OK - attachment 11
ADC data item stats - attachments 12
aida15 rate c. 34k ( 2x 'hot' channels, 1x channel missing, all other channels OK )
aida01 rate c. 75k ( 2x 'hot' channels )
per FEE64 Stat spectra - attachments 13-14
Merger and TapeServer - attachments 15-16
09.52 per FEE64 Rate spectra - attachments 17-18
per FEE64 1.8.L spectra - attachments 19-20
per FEE64 1.8.H spectra - attachments 21-22
per p+n FEE64 1.8.W spectra - 20us, 200us, 2ms & 20ms FSR - attachments 23-26
per n+n FEE64 1.8.W spectra - 20us, 200us, 2ms & 20ms FSR - attachments 27-30
10.11 Merger stats blocks 0-4, 14-16 - attachments 31-38
10.14 aida01-aida04, aida14-aida16 stats- attachments 39-45
10.21 Overnight V-I attach 46
10.23 Save current ASIC settings 2024Feb05-10.22.28
changes to disc mask and LED waveform threshold for some FEE64s
10.32 Analysis of data file G22-88/R3_45 - attachments 47
deadtime aida01 c. 0.5%, all other FEE64s deadtime c. 0%
13.07 DAQ stop file G22-88/R3_50
Stop merger
Stop tapeserver
Afternoon: Tested detector motion. Channels on p+n strips drop out with high threshold (1 MeV), but not with low threshold (100 keV). No issues on n+n strips. Reasons unclear. |
Attachment 1: Screenshot_from_2024-02-05_06-40-57.png
|
|
Attachment 2: Screenshot_from_2024-02-05_06-41-22.png
|
|
Attachment 3: Screenshot_from_2024-02-05_06-41-47.png
|
|
Attachment 4: Screenshot_from_2024-02-05_06-42-04.png
|
|
Attachment 5: Screenshot_from_2024-02-05_06-42-29.png
|
|
Attachment 6: Screenshot_from_2024-02-05_06-42-55.png
|
|
Attachment 7: Screenshot_from_2024-02-05_06-43-16.png
|
|
Attachment 8: Screenshot_from_2024-02-05_06-44-10.png
|
|
Attachment 9: Screenshot_from_2024-02-05_06-44-55.png
|
|
Attachment 10: Screenshot_from_2024-02-05_09-46-59.png
|
|
Attachment 11: Screenshot_from_2024-02-05_09-47-39.png
|
|
Attachment 12: Screenshot_from_2024-02-05_09-47-52.png
|
|
Attachment 13: Screenshot_from_2024-02-05_09-48-22.png
|
|
Attachment 14: Screenshot_from_2024-02-05_09-48-29.png
|
|
Attachment 15: Screenshot_from_2024-02-05_09-48-51.png
|
|
Attachment 16: Screenshot_from_2024-02-05_09-49-05.png
|
|
Attachment 17: Screenshot_from_2024-02-05_09-52-04.png
|
|
Attachment 18: Screenshot_from_2024-02-05_09-52-14.png
|
|
Attachment 19: Screenshot_from_2024-02-05_09-52-57.png
|
|
Attachment 20: Screenshot_from_2024-02-05_09-53-33.png
|
|
Attachment 21: Screenshot_from_2024-02-05_09-55-33.png
|
|
Attachment 22: Screenshot_from_2024-02-05_09-56-09.png
|
|
Attachment 23: Screenshot_from_2024-02-05_09-57-05.png
|
|
Attachment 24: Screenshot_from_2024-02-05_09-58-51.png
|
|
Attachment 25: Screenshot_from_2024-02-05_09-59-23.png
|
|
Attachment 26: Screenshot_from_2024-02-05_09-59-56.png
|
|
Attachment 27: Screenshot_from_2024-02-05_10-03-26.png
|
|
Attachment 28: Screenshot_from_2024-02-05_10-02-25.png
|
|
Attachment 29: Screenshot_from_2024-02-05_10-01-57.png
|
|
Attachment 30: Screenshot_from_2024-02-05_10-01-25.png
|
|
Attachment 31: Screenshot_from_2024-02-05_10-11-41.png
|
|
Attachment 32: Screenshot_from_2024-02-05_10-11-46.png
|
|
Attachment 33: Screenshot_from_2024-02-05_10-11-51.png
|
|
Attachment 34: Screenshot_from_2024-02-05_10-11-57.png
|
|
Attachment 35: Screenshot_from_2024-02-05_10-12-04.png
|
|
Attachment 36: Screenshot_from_2024-02-05_10-12-39.png
|
|
Attachment 37: Screenshot_from_2024-02-05_10-12-44.png
|
|
Attachment 38: Screenshot_from_2024-02-05_10-13-11.png
|
|
Attachment 39: Screenshot_from_2024-02-05_10-14-43.png
|
|
Attachment 40: Screenshot_from_2024-02-05_10-14-50.png
|
|
Attachment 41: Screenshot_from_2024-02-05_10-15-43.png
|
|
Attachment 42: Screenshot_from_2024-02-05_10-15-48.png
|
|
Attachment 43: Screenshot_from_2024-02-05_10-15-54.png
|
|
Attachment 44: Screenshot_from_2024-02-05_10-16-00.png
|
|
Attachment 45: Screenshot_from_2024-02-05_10-16-06.png
|
|
Attachment 46: Screenshot_from_2024-02-05_10-21-03.png
|
|
Attachment 47: R3_45
|
*** TDR format 3.3.0 analyser - TD - May 2021
*** ERROR: READ I/O error: 5002
blocks: 32000
ADC data format: 261748120 ( 83699.2 Hz)
Other data format: 171882 ( 55.0 Hz)
Sample trace data format: 0 ( 0.0 Hz)
Undefined format: 0 ( 0.0 Hz)
Other data format type: PAUSE: 194 ( 0.1 Hz)
RESUME: 194 ( 0.1 Hz)
SYNC100: 43648 ( 14.0 Hz)
WR48-63: 43648 ( 14.0 Hz)
FEE64 disc: 84198 ( 26.9 Hz)
MBS info: 0 ( 0.0 Hz)
Other info: 0 ( 0.0 Hz)
ADC data range bit set: 3 ( 0.0 Hz)
Timewarps: ADC: 0 ( 0.0 Hz)
PAUSE: 0 ( 0.0 Hz)
RESUME: 0 ( 0.0 Hz)
SYNC100: 0 ( 0.0 Hz)
WR48-63: 0 ( 0.0 Hz)
FEE64 disc: 0 ( 0.0 Hz)
MBS info: 0 ( 0.0 Hz)
Undefined: 0 ( 0.0 Hz)
Sample trace: 0 ( 0.0 Hz)
*** Timestamp elapsed time: 3127.246 s
FEE elapsed dead time(s) elapsed idle time(s)
0 17.539 0.000
1 0.000 0.000
2 0.000 0.000
3 0.000 991.338
4 0.000 0.000
5 0.000 0.000
6 0.000 0.000
7 0.000 0.000
8 0.000 0.000
9 0.000 0.000
10 0.000 0.000
11 0.000 0.000
12 0.000 0.000
13 0.000 0.000
14 0.500 0.000
15 0.000 0.000
16 0.000 0.000
17 0.000 0.000
18 0.000 0.000
19 0.000 0.000
20 0.000 0.000
21 0.000 0.000
22 0.000 0.000
23 0.000 0.000
24 0.000 0.000
25 0.000 0.000
26 0.000 0.000
27 0.000 0.000
28 0.000 0.000
29 0.000 0.000
30 0.000 0.000
31 0.000 0.000
32 0.000 0.000
*** Statistics
FEE ADC Data Other Data Sample Undefined Pause Resume SYNC100 WR48-63 Disc MBS Other HEC Data
0 172462915 139302 0 0 170 170 27567 27567 83828 0 0 0
1 45 2 0 0 0 0 0 0 2 0 0 1
2 58 0 0 0 0 0 0 0 0 0 0 0
3 167035 79 0 0 0 0 24 24 31 0 0 0
4 332 0 0 0 0 0 0 0 0 0 0 0
5 257 0 0 0 0 0 0 0 0 0 0 0
6 254 0 0 0 0 0 0 0 0 0 0 0
7 370 0 0 0 0 0 0 0 0 0 0 0
8 336 0 0 0 0 0 0 0 0 0 0 0
9 75 0 0 0 0 0 0 0 0 0 0 0
10 99 0 0 0 0 0 0 0 0 0 0 0
11 240 164 0 0 0 0 0 0 164 0 0 0
12 1266 0 0 0 0 0 0 0 0 0 0 0
13 218 0 0 0 0 0 0 0 0 0 0 0
14 89113402 32165 0 0 24 24 16057 16057 3 0 0 2
15 1218 170 0 0 0 0 0 0 170 0 0 0
16 0 0 0 0 0 0 0 0 0 0 0 0
17 0 0 0 0 0 0 0 0 0 0 0 0
18 0 0 0 0 0 0 0 0 0 0 0 0
19 0 0 0 0 0 0 0 0 0 0 0 0
20 0 0 0 0 0 0 0 0 0 0 0 0
21 0 0 0 0 0 0 0 0 0 0 0 0
22 0 0 0 0 0 0 0 0 0 0 0 0
23 0 0 0 0 0 0 0 0 0 0 0 0
24 0 0 0 0 0 0 0 0 0 0 0 0
25 0 0 0 0 0 0 0 0 0 0 0 0
26 0 0 0 0 0 0 0 0 0 0 0 0
27 0 0 0 0 0 0 0 0 0 0 0 0
28 0 0 0 0 0 0 0 0 0 0 0 0
29 0 0 0 0 0 0 0 0 0 0 0 0
30 0 0 0 0 0 0 0 0 0 0 0 0
31 0 0 0 0 0 0 0 0 0 0 0 0
32 0 0 0 0 0 0 0 0 0 0 0 0
*** Timewarps
FEE ADC Pause Resume SYNC100 WR48-63 Disc MBS Undefined Samples
0 0 0 0 0 0 0 0 0 0
1 0 0 0 0 0 0 0 0 0
2 0 0 0 0 0 0 0 0 0
3 0 0 0 0 0 0 0 0 0
4 0 0 0 0 0 0 0 0 0
5 0 0 0 0 0 0 0 0 0
6 0 0 0 0 0 0 0 0 0
7 0 0 0 0 0 0 0 0 0
8 0 0 0 0 0 0 0 0 0
9 0 0 0 0 0 0 0 0 0
10 0 0 0 0 0 0 0 0 0
11 0 0 0 0 0 0 0 0 0
12 0 0 0 0 0 0 0 0 0
13 0 0 0 0 0 0 0 0 0
14 0 0 0 0 0 0 0 0 0
15 0 0 0 0 0 0 0 0 0
16 0 0 0 0 0 0 0 0 0
17 0 0 0 0 0 0 0 0 0
18 0 0 0 0 0 0 0 0 0
19 0 0 0 0 0 0 0 0 0
20 0 0 0 0 0 0 0 0 0
21 0 0 0 0 0 0 0 0 0
22 0 0 0 0 0 0 0 0 0
23 0 0 0 0 0 0 0 0 0
24 0 0 0 0 0 0 0 0 0
25 0 0 0 0 0 0 0 0 0
26 0 0 0 0 0 0 0 0 0
27 0 0 0 0 0 0 0 0 0
28 0 0 0 0 0 0 0 0 0
29 0 0 0 0 0 0 0 0 0
30 0 0 0 0 0 0 0 0 0
31 0 0 0 0 0 0 0 0 0
32 0 0 0 0 0 0 0 0 0
*** Program elapsed time: 24.477s ( 1307.373 blocks/s, 81.711 Mb/s)
|
490
|
Sun Feb 4 08:40:04 2024 |
CB TD | Run3 - contd |
09.30 DAQ continues file G22-88/R3_17
DSSSD bias OK - attachment 1
ambient temperature 16.8 deg C
FEE64 temperatures OK - attachment 2
ADC data item stats - attachment 3
aida15 rate c. 25k ( 2x 'hot' channels, 1x channel missing, all other channels OK )
aida01 rate c. 45k ( 2x 'hot' channels )
per FEE64 Stat spectra - attachments 4-5
per FEE64 1.8.L spectra - attachments 6-7
per FEE64 1.8.W spectra - 20us FSR - attachments 8-9
Merger and TapeServer - attachments 10-11
09.36 Check ASIC load
No change to aida01 ADC data item rate
Last access times of data files indicate that data rate changed c. 05.00 - 06.00 this am.
No obvious vacuum/temperature transients in last six hours.
09.55 All system wide checks OK except
FPGA timestamp errors - attachment 12
WR timestamps OK - attachment 13
10:06 V-I plot overnight - attach 14.
13.35 DAQ continues file G22-88/R3_22
DSSSD bias OK - attachment 15
ambient temperature 16.8 deg C
FEE64 temperatures OK - attachment 16
ADC data item stats - attachment 17
aida15 rate c. 35k ( 2x 'hot' channels, 1x channel missing, all other channels OK )
aida01 rate c. 51k ( 2x 'hot' channels )
per FEE64 Rate spectra - attachments 18
per FEE64 Stat spectra - attachments 19-20
Merger and TapeServer - attachments 21-22 |
Attachment 1: Screenshot_from_2024-02-04_09-30-41.png
|
|
Attachment 2: Screenshot_from_2024-02-04_09-31-07.png
|
|
Attachment 3: Screenshot_from_2024-02-04_09-31-40.png
|
|
Attachment 4: Screenshot_from_2024-02-04_09-32-26.png
|
|
Attachment 5: Screenshot_from_2024-02-04_09-32-42.png
|
|
Attachment 6: Screenshot_from_2024-02-04_09-33-23.png
|
|
Attachment 7: Screenshot_from_2024-02-04_09-34-00.png
|
|
Attachment 8: Screenshot_from_2024-02-04_09-34-37.png
|
|
Attachment 9: Screenshot_from_2024-02-04_09-35-21.png
|
|
Attachment 10: Screenshot_from_2024-02-04_09-37-23.png
|
|
Attachment 11: Screenshot_from_2024-02-04_09-38-28.png
|
|
Attachment 12: Screenshot_from_2024-02-04_09-55-02.png
|
|
Attachment 13: Screenshot_from_2024-02-04_09-55-18.png
|
|
Attachment 14: Screenshot_from_2024-02-04_10-05-07.png
|
|
Attachment 15: Screenshot_from_2024-02-04_13-35-51.png
|
|
Attachment 16: Screenshot_from_2024-02-04_13-36-21.png
|
|
Attachment 17: Screenshot_from_2024-02-04_13-36-41.png
|
|
Attachment 18: Screenshot_from_2024-02-04_13-37-35.png
|
|
Attachment 19: Screenshot_from_2024-02-04_13-38-27.png
|
|
Attachment 20: Screenshot_from_2024-02-04_13-38-52.png
|
|
Attachment 21: Screenshot_from_2024-02-04_13-40-02.png
|
|
Attachment 22: Screenshot_from_2024-02-04_13-40-20.png
|
|
515
|
Thu Feb 15 10:39:50 2024 |
JM, JG, ML, AG, CB | Run29 - beam optimisation |
11:27
Run 29 started. Beam energy is 1.3435 MeV/u with a cooler voltage of 751.3 V.
We observe significantly greater count rate for top right detector (DSSD#1) now that the beam energy has been lowered. ADC over range? Beam target overlap change?
Plan to check optimisation of beam-target overlap and beam settings again.
12:40
Beam lost. Detectors out. R29_11
13:37
Electron cooler is not working again. We will have un-cooled beam again.
Will start a new run 30 for un-cooled beam on target. |
582
|
Mon Jun 3 15:17:41 2024 |
CN, EM, CB | Run29 |
16:15
In the previous run while working on the target it was discovered that the valve was closed! Probaly this happened around 10 am, we are not sure.
The gas target inlet valve is open again, working on the target.
Beam energy = 3 MeV
Horizontal beam position = 7.0 mm, vertical beam position = 0.0 mm.
Beam intensity at inj ~ 2.5e7
Detectors are positioned IN pneumatic only. Left arm = 62.71, right arm = 64.60 mm
Histograms and stats zero'd, monitoring code restarted.
Beam in the ring ~ 7 minutes.
Pressure in YR09 E target ~ 4.5e-11 mbar.
18:03 ( Attchments 1 -4)
Still no target!
Merger OK, Tape server OK
ADC data item statistics looks as before. OK
FEE temperatures look OK
System checks OK
20:23
We have the target back. S1( grafana) 4.8e-8. The target thickness read by the system 5E11 atoms/cm2 but apparently the formula is wrong and the value should be almost e12 atoms/cm2.
. We have now proper beam and target and we
|
Attachment 1: Screenshot_from_2024-06-03_18-04-47.png
|
|
Attachment 2: Screenshot_from_2024-06-03_18-05-11.png
|
|
Attachment 3: Screenshot_from_2024-06-03_18-06-09.png
|
|
Attachment 4: Screenshot_from_2024-06-03_18-06-40.png
|
|
Attachment 5: Screenshot_from_2024-06-03_20-35-00.png
|
|
514
|
Wed Feb 14 16:52:46 2024 |
JM, AR, CB, JG | Run28 - beam on target |
Beam energy - 1.361 MeV/u
17:12
Run 28 Started - beam on target
Detectors IN (left arm pot = 36.5 mm, right arm pot = 70.5 mm)
Beam position (vertical offset -9mm, horizontal offset -7.5mm)
Ions circulate in the ring for 3 seconds before they are dumped. Total time between injections is 12 seconds.
Target density apporx 4.2e11 atoms/cm2, pressures in the ring are below 1e-10 mbar. (Attachment 1+2)
Leakage current plot (attachment 3) - change in leakage current is consistent with temperature variation from ML online plot.
18:18
FEE temperatures OK - aida09 is slightly hot - 65.06 C so slightly above vertex temp value of 65C and stays this way over several reloads (attachment 4)
System wide checks OK
Statistics (attachments 5+6) - aida09 and aida12 slightly high stats but still <1K. Check load does not resolve.
Tape server OK - writing ~730 kB/sec
Merger OK
Monitoring code XY plots and Ey energy histograms (attachments 7+8)
18:53 All checks OK
21:58
All system checks ok. Temperatures and leakage current ok. Merger ok.
Attatchment 9: Rates and stats - all below 1k
Attatchment 10: Temperatures
Attatchment 11-14: Waveforms layouts 1,2,7,8 respectively look okay
Attatchment 15: Leakge current
Attatchment 16: lowenergyXY plots
Vacuum pressure is ok.
23:01
system wide checks ok. Temperatures ok. Leakage current ok. Merger ok all went green. Waveforms same as 10pm check.
Statistics - aida09 and aida12 went over 1k (attach 17), but reloaded and settled down again seconds appart (attatch 18). Reloaded again at 23:14 and stats were aove 1k again (attatch 19).
Attatchment 20 is the leakage current, still ok.
Vacuum pressure ok (attatchment 21) - rises after small bump down from the regular value.
01:00
System wide checks ok. Temperatures ok. Leakage current ok (see attatchment 22). Merger ok all went green. Waveforms the same as 10pm check.
Statistics - aida09 and aida12 occasionally just over 1k when reloading a few times, but goes below mostly.
Vacuum pressure ok, stablised after small bump around 10ish (attatchment 23)
03:00
Ring vacuum suffered from long-term creep, especially in YR08. Tweaked nozzle position. Pressure improves, target density slightly affected. Attachment 24-25
Leakage current and temperature indicate we have an increase in temperature in the ring. This is confirmed by vacplot. Attachment 26-27
Merger continues OK 28
Rates and Stats show we have no beam. Going to the lab 29-30.
R28_22 NO BEAM
03:45
No beam in the ring, we are missing all the spills. Called HKR.
Detectors out in case beam manipulation is required.
Stat plot looked OK. 31
04:35
DAQ continues OK on run R28_24, still no beam
Temperatures slightly lower - OK
System wide checks - OK
Stats - OK
Merger - OK
05:44
Beam is back. It had been away since 1:30 am due to a fault in the new EC power supply. Fault reason unclear.
Detectors back in (left 36.54, right 70.49 mm)
R28_26 BEAM ON
(Beam likely off from R28_ 21-25 inclusive)
Temps as before
Stats now OK - att 32
Merger OK - att 33, 34
Stat spectra OK - att 35. Note slope in ASICS (Rutherford?)
Leakage current trending up - likely temperature related.
Ring pressure OK
Gas target 4.14E11
Beam at injection around 7.5E6. Could be higher. Asked for optimisation later today assuming current configuration holds.
06:59
Temperatures OK
System checks OK
Stats looking good in all FEEs except aida01 and aida15
Leakage current has levelled off - looks OK
Ring pressure looks OK and is trending down - target density 4.1e11
Merger and tape server OK
08:30
System checks all OK
Temperatures OK
Merger / tape server OK - data being written R28_29
Target density 4.1e11, ring pressure stable
09:00
Attachment 38 Spectrum browser stats for all FEEs (log scale)
Attachment 39-42 Spectrum browser *L histograms for select channels for each detector. MIDAS data is consistent with inline monitoring plots.
System checks OK
Beam still on - target density and ring pressure remain stable
11:28 We have access to the voltage dividers for the electron cooler. The voltage is -763.625 V. This corresponds to an electron energy of 746.795 eV and a beam energy of 1.361 MeV/u using ML ion calc.
This is above the resonance energy of 1.3435 MeV/u. We will now end the run and set the cooler voltage to this ion energy.
Run ended. Run 29 started while setting beam energy.
|
Attachment 1: 20240214_180331.jpg
|
|
Attachment 2: 20240214_180351.jpg
|
|
Attachment 3: Screenshot_from_2024-02-14_17-26-37.png
|
|
Attachment 4: Screenshot_from_2024-02-14_18-11-58.png
|
|
Attachment 5: Screenshot_from_2024-02-14_18-14-06.png
|
|
Attachment 6: Screenshot_from_2024-02-14_18-14-15.png
|
|
Attachment 7: Screenshot_from_2024-02-14_18-17-32.png
|
|
Attachment 8: Screenshot_from_2024-02-14_18-17-46.png
|
|
Attachment 9: Screenshot_2024-02-14_at_21.51.31.png
|
|
Attachment 10: Screenshot_2024-02-14_at_21.52.19.png
|
|
Attachment 11: Screenshot_2024-02-14_at_21.53.59.png
|
|
Attachment 12: Screenshot_2024-02-14_at_21.54.41.png
|
|
Attachment 13: Screenshot_2024-02-14_at_21.55.47.png
|
|
Attachment 14: Screenshot_2024-02-14_at_21.56.24.png
|
|
Attachment 15: Screenshot_2024-02-14_at_21.57.05.png
|
|
Attachment 16: Screenshot_2024-02-14_at_21.58.02.png
|
|
Attachment 17: Screenshot_2024-02-14_at_23.02.54.png
|
|
Attachment 18: Screenshot_2024-02-14_at_23.03.31.png
|
|
Attachment 19: Screenshot_2024-02-14_at_23.14.11.png
|
|
Attachment 20: Screenshot_2024-02-14_at_23.08.34.png
|
|
Attachment 21: Screenshot_2024-02-14_at_23.17.27.png
|
|
Attachment 22: Screenshot_2024-02-15_at_01.08.36.png
|
|
Attachment 23: Screenshot_2024-02-15_at_01.19.22.png
|
|
Attachment 24: Gasjet.png
|
|
Attachment 25: Pressure.png
|
|
Attachment 26: Screenshot_from_2024-02-15_03-07-02.png
|
|
Attachment 27: Screenshot_from_2024-02-15_03-07-44.png
|
|
Attachment 28: Screenshot_from_2024-02-15_03-20-37.png
|
|
Attachment 29: Screenshot_from_2024-02-15_03-14-58.png
|
|
Attachment 30: Screenshot_from_2024-02-15_03-20-18.png
|
|
Attachment 31: Screenshot_from_2024-02-15_03-59-38.png
|
|
Attachment 32: Screenshot_from_2024-02-15_05-47-18.png
|
|
Attachment 33: Screenshot_from_2024-02-15_05-47-31.png
|
|
Attachment 34: Screenshot_from_2024-02-15_05-47-45.png
|
|
Attachment 35: Screenshot_from_2024-02-15_05-48-27.png
|
|
Attachment 36: Screenshot_from_2024-02-15_05-48-41.png
|
|
Attachment 37: Screenshot_from_2024-02-15_07-01-48.png
|
|
Attachment 38: Screenshot_from_2024-02-15_08-48-58.png
|
|
Attachment 39: Screenshot_from_2024-02-15_08-52-55.png
|
|
Attachment 40: Screenshot_from_2024-02-15_08-54-09.png
|
|
Attachment 41: Screenshot_from_2024-02-15_08-56-27.png
|
|
Attachment 42: Screenshot_from_2024-02-15_08-58-01.png
|
|