494
|
Thu Jun 23 15:10:17 2022 |
Philippos Papadakis | Thursday 23 June 16:00-24:00 |
16:20
System wide checks.
Clock status OK
White Rabbit Error status:
Base Current Difference
aida07 fault 0xc4fe : 0xc525 : 39
aida08 fault 0xf0e9 : 0xf166 : 125
White Rabbit error counter test result: Passed 6, Failed 2
FPGA error status:
Base Current Difference
aida07 fault 0x11 : 0x1c : 11
aida08 fault 0x1a : 0x23 : 9
Temps OK (attachment 1), Stats OK (attachment 2), Bias and leakage OK (attachment 3)
18:05
System wide checks.
Clock status OK
White Rabbit Error status:
Base Current Difference
aida07 fault 0xc4fe : 0xc527 : 41
aida08 fault 0xf0e9 : 0xf177 : 142
White Rabbit error counter test result: Passed 6, Failed 2
FPGA error status:
Base Current Difference
aida07 fault 0x11 : 0x1c : 11
aida08 fault 0x1a : 0x24 : 10
FPGA Timestamp error counter test result: Passed 6, Failed 2
Temps OK (attachment 4), Stats OK (attachment 5), Bias and leakage OK (attachment 6)
20:22
System wide checks.
Clock status OK
White Rabbit Error status:
Base Current Difference
aida07 fault 0xc4fe : 0xc529 : 43
aida08 fault 0xf0e9 : 0xf17f : 150
White Rabbit error counter test result: Passed 6, Failed 2
FPGA error status:
Base Current Difference
aida07 fault 0x11 : 0x20 : 15
aida08 fault 0x1a : 0x25 : 11
FPGA Timestamp error counter test result: Passed 6, Failed 2
Temps OK (attachment 7), Stats OK (attachment 8), Bias and leakage OK (attachment 9)
21:56
System wide checks.
Clock status OK
White Rabbit Error status:
Base Current Difference
aida07 fault 0xc4fe : 0xc529 : 43
aida08 fault 0xf0e9 : 0xf186 : 157
White Rabbit error counter test result: Passed 6, Failed 2
FPGA error status:
Base Current Difference
aida07 fault 0x11 : 0x21 : 16
aida08 fault 0x1a : 0x26 : 12
Temps OK (attachment 10), Stats OK (attachment 11), Bias and leakage OK (attachment 12)
23:50
System wide checks.
Clock status OK
White Rabbit Error status:
Base Current Difference
aida07 fault 0xc4fe : 0xc52e : 48
aida08 fault 0xf0e9 : 0xf18c : 163
White Rabbit error counter test result: Passed 6, Failed 2
FPGA error status:
Base Current Difference
aida07 fault 0x11 : 0x21 : 16
aida08 fault 0x1a : 0x27 : 13
FPGA Timestamp error counter test result: Passed 6, Failed 2
Temps OK (attachment 13), Stats OK (attachment 14), Bias and leakage OK (attachment 15) |
|
493
|
Thu Jun 23 06:56:02 2022 |
OH | Thursday 23 June 08:00-16:00 |
07:56 Taken over from AM
Current White Rabbit Error status:
Base Current Difference
aida07 fault 0xc4fe : 0xc51a : 28
aida08 fault 0xf0e9 : 0xf132 : 73
White Rabbit error counter test result: Passed 6, Failed 2
FPGA error status:
Base Current Difference
aida07 fault 0x11 : 0x18 : 7
aida08 fault 0x1a : 0x20 : 6
FPGA Timestamp error counter test result: Passed 6, Failed 2
Likely that at some point AIDA08 will drop out again. I think this occurs if there is a sudden burst of WR errors which causes the merger to drop the link
Statistics ok - attachment 1
FEE temperatures ok - attachment 2
Bias and leakage currents ok - attachment 3
Analysis of file R3_93 - attachment 4
Still deadtime in aida02 of ~15%
Patrick has said he will look into whether we are still sending info code 4 and 5 for every ADC event which could be contributing to the deadtime.
Currently 2.2 TB of space on /media/SecondDrive
Current data rate of 6.23MB/s
At current data rate this should last 4 days
08:27 Currently no beam
Problem with Alvarez 3 now
Tomorrow will have a 1 hour break due to changing of the source
09:36 Beam back
09:59 System wide checks
Base Current Difference
aida07 fault 0xc4fe : 0xc51b : 29
aida08 fault 0xf0e9 : 0xf138 : 79
White Rabbit error counter test result: Passed 6, Failed 2
Base Current Difference
aida07 fault 0x11 : 0x18 : 7
aida08 fault 0x1a : 0x20 : 6
Stats ok - attachment 5
Temperatures ok - attachment 6
Bias and leakage currents ok - attachment 7
11:14 FRS group is changing from Hg setting back to Pt
11:36 Back on Pt R3_129
14:14 Stats ok - attachment 8
Temperature ok - attachment 9
Bias and leakage currents ok - attachment 10
System wide checks:
WR:
Base Current Difference
aida07 fault 0xc4fe : 0xc523 : 37
aida08 fault 0xf0e9 : 0xf156 : 109
White Rabbit error counter test result: Passed 6, Failed 2
FPGA
Base Current Difference
aida07 fault 0x11 : 0x1b : 10
aida08 fault 0x1a : 0x20 : 6
Analysis of R3_156 - attachment 11 |
|
492
|
Wed Jun 22 23:18:45 2022 |
AM, OH, TD, MA | Thursday 23 June |
00:00 TD Noticed on the last stats uploaded that AIDA08 had stopped sending signals.
Attempted to recover restarting merger but this caused 01 to drop out.
00:10 DAQ recovered and AIDA01 recovered with a reboot
01:00 Attachments 1-4, white rabbit failures on aida07 and aida08, otherwise all good
03:00 Attachments 5-9, aida07 and aida08 failures on white rabbit and fpga timestamp, otherwise all good
05:00 Attachments 10-14, aida07 and aida08 failures on white rabbit and fpga timestamp continue, otherwise all good
07:00 Attachments 15-19, aida07 and aida08 failures on white rabbit and fpga timestamp continue, otherwise all good |
|
491
|
Wed Jun 22 15:09:37 2022 |
MA | Wednesday 22 June 16:00-00:00 |
16:00 Take over from Magda mrning shift
Beam was off of sometime for tunning to Mercurey setting.
17:00 attached 1-3
19:00 attached 4-6
21:00 attached 7-9
23:30 10-12
|
|
490
|
Wed Jun 22 10:09:12 2022 |
PJCS | INFO: FEE64 supply Voltages |
Study of the FEE64 power supply distribution has yielded the following :-
The most sensitive regulator, as regards the device voltage input, is the TPS51100 that supplies the DDR memory reference voltages. This requires +4.75 to +5.25 volts. This is supplied from the Power input connector +5v input.
The common LT3080 regulator used over much of the board is a Low Voltage Dropout regulator. This requires 0.5v difference between input and output voltage as minimum. This is not a problem with the +4.75v minimum for the TPS51100 requirement setting the voltage for the board.
The supply to the mezzanine is direct from the power connector +5v input. On the mezzanine there is an LT3080 for each ASIC supplying the required 3.3v. These regulators would possibly benefit from a 1uF capacitor at the Control voltage input.
The simplest approach would be to add a capacitor to the bottom layer where the Control voltage enters the mezzanine.
The power cable has a nominal resistance of 13.3ohms/km. The 3 conductors of the cable are supplying 10A when all is in operation. So the expected voltage drop would thus be ( 10 x 13.3 x 0.007 ) /3 => 0.3v each core.
The conclusion would be that the voltage at the power supply should drop to 5.25 v thus ensuring the TPS51100 is supplied as required regardless of the operation of the FEE.
This will be tested at Daresbury. |
|
489
|
Wed Jun 22 07:12:46 2022 |
MS | Wednesday 22 June 8:00-16:00 |
Took over the night shift from Tom.
8:10 : attachments 1-3
10:00 : attachments 4-6
12:00 : attachments 7-9
The beam was off for around 20' starting at around 11:30-11:50 due to ion source issues.
12:19 OH Disabled all discriminators to reduce the data rate.
13:06 Analysis of R2_104 15.66% deadtime on AIDA04- attachment 10
14:00 : attachments 11-13
16:00 : attachments 14-16
|
|
488
|
Tue Jun 21 23:29:16 2022 |
TD | Wednesday 22 June 00:00-08:00 |
00.28 all histograms & stats zero'd
baseline counters
ASIC settings 2021Apr29-13-16-00
*except* slow comparator 0x64 -> 0xa
BNC PB-5 pulser
to p+n FEE64s only
amplitude 1.0V
attenuator x1
decay time 1ms
polarity +
frequency 2Hz
all AD9252 waveform ADCs disabled
Tape Server -> no storage mode
00.36 check ASIC control all FEE64s all ASICs
all system wide checks OK
00.41 BNC PB-5 pulser rate increased from 2Hz to 22Hz per request by HA/NH - to ensure DTAS correlation checker has sufficient rate of identified pulser events
per p+n FEE64 1.8.L spectra - attachment 1
aida01 pulser peak width 88 ch FWHM
per FEE64 rate spectra - attachment 2
adc, pause and correlation scaler data items - push, aida01 and aida02 stats - attachments 3-8
FEE64 temps OK - attachment 9
DSSSD bias & leakage current OK - attachments 10 & 11
02.28 aida05 zero adc data items, unable to halt aida05 - all other FEE64s OK
restart NewMerger -> now able to halt aida05
DAQ go OK, merge rate c. 1.4M data items/s
per FEE64 1.8.H spectra - attachments 12 & 13
approx flat distribution to c. 10GeV
adc data items stats OK - attachments 14
FEE64 temps OK - attachment 15
DSSSD bias & leakage current OK - attachments 16 & 17
02.55 Some problems with GO4 data analysis for PID (FRS PID looks OK). 204Pt setting. Will commence writing data to disk file S505/R2_0
04.12 per FEE64 1.8.H spectra - attachments 18 & 19
c. 9GeV peak in upstream DSSSDs (aida01, 2, 3 & 4) indicating (high energy) heavy fragments are stopping
Low energy events near c. 200MeV lower mass, high energy, low energy loss - fission fragments?
04.20 per FEE64 stats spectra - attachment 20
indicates implant x-y distribution - off centre (more significantly in x than in y) but DSSSDs appear to see most of distribution
remember raw distro will have significant fraction of fission fragments - may be misleading - need to gate on high energy events to see x-y distro of ions of
interest
adc data items stats OK - attachments 21
FEE64 temps OK - attachment 22
DSSSD bias & leakage current OK - attachments 23
06.10 per FEE64 rate spectra - attachment 24
per FEE64 1.8.H spectra - attachments 25 & 26
per FEE64 stats spectra - attachment 27
adc data items stats OK - attachments 28
FEE64 temps OK - attachment 29
system wide checks OK *except* WR & FPGA errors - see attachments 30 & 31
DSSSD bias & leakage current OK - attachments 32
07.52 analysis of file S505/R2_47 - attachment 33
all FEE64 dead times < 1% *except* aida02 c. 7% |
|
487
|
Tue Jun 21 10:19:05 2022 |
OH ML | Tuesday 21st June 08:00-00:00 |
11:19 Stats ok 0xa on all FEE - attachment 1
Current pulser settings - attachment 2
FEE Temperatures ok - attachment 3
Noticed both AIDA03 and AIDA07 had ASICs with missing channels. Did a check ASIC control on them to recover
Layout 1 after recovering channels - attachment 4
New stats (Note increased rate on 3 and 7) attachment 5
Bias and leakage currents ok - attachment 6
S4 now in closed access. Trying to find out if the beam dump has been put in front of AIDA. Will make sure this is checked before beam to S4.
14:05 Have had it confirmed that the flange beam dump is in position in S4
15:43 Stats ok -attachment 7
Temp ok - attachment 8
Bias and Leakage current ok - attachment 9
17:45 Currently the time machine trigger(for correlation) is connected to the DTAS or.
As the beam is tuning they are triggering at 150k. This is 450k worth of scalers in AIDA
During the experiment this will be changed to SC1 trigger to solve this.
19:49: Stats ok - attachment 10
temperatures ok - attachment 11
Bias - Leakage current ok - attachment 12 |
|
486
|
Mon Jun 20 09:26:23 2022 |
OH, TD | Monday 20 June |
10:26 MACB HDMI for AIDA03 and AIDA04 reseated.
Merger and DAQ restarted - All happily sending data
Has been running for 10 minutes with no WR errors.
Rates from yesterday achieved again.
DTAS will move back into position at some point.
Once this is done, will perform pulser walkthrough.
11:14 DTAS now back in position for experiment
Slight increase in noise noticed across a few FEEs
Over time this has decreased on all but AIDA07 which is now running at ~80k - attachment 1
AIDA03 is now however gaining white rabbit errors at a rate of 1 every few minutes.
Less frequently gaining FPGA errors
12:20 AIDA power cycle to reseat the HDMI for AIDA03 - It would eventually crash if not done
Upon powering up again rates in AIDA08 slightly worse - attachment 2
13:26 No WR errors since the last power cycle.
21:16 Baseline system wide checks, zero histograms & stats
All system wide checks OK
per p+n FEE64 1.8.L spectra - attachment 3
aida01 pulser peak width 82 channels FWHM
per FEE64 rate spectra - attachment 4
adc, pause, resume and correlation scaler data item stats - attachments 5-8
FEE64 temps OK - attachment 9
DSSSD bias & leakage currents OK - attachments 10 & 11 |
|
485
|
Sun Jun 19 19:55:20 2022 |
NH | aida03 and aida04 |
aida03 and aida04 seem to stop sending data quickly
stopping DAQ + merger and restarting got them resending data, but quickly stopped again
both are showing large WR errors that increases approx every second.
Assume HDMI cables not well seated and causing all sorts of odd behaviour |
|
484
|
Sun Jun 19 12:36:10 2022 |
OH | Sunday 19th June |
Yesterday the sum/inverter was found to be a considerable source of noise in the system.
Plugging it into an oscilloscope this morning it was clear there was a Vpp 17mv 60kHz sin wave in its signal - attachment 1 yellow trace
Moved the inverter to a crate in the messheute and did not see same behaviour. This time see Vpp 5mV and a ~265Hz wave that looks much less clean. - attachment 2 yellow
Put inverter back in CAEN crate in S4 and confirm similar behavior.
Move CAEN crate to new mains port, same behaviour
Move CAEN crate to AC voltage stabiliser, same behaviour
Remove module one by one from CAEN crate.
Removing BIAS supply - No change
Remove Pulser - No change
Remove right most MACB - No change
Remove second right most MACB - Signal disappears - attachment 3 blue trace
Remove second right most MACB from the rack entirely
Put right most MACB back - Signal comes back - attachment 4 blue trace
Test MACB with and without HDMI connectors connected signal remains
Remove rightmost MACB from the rack - Signal disappears
Put pulser and bias back - Signal does not return
Signal also doesn't return when ramping up bias or running the pulser
Power DAQ up - No change in performance from yesterday with inverter removed.
MACB and Inverter don't share any voltages on NIM:
MACB +-6V and Gnd
Inverter +-24V and Gnd
CAEN Crate does not have any easily accessible monitoring points. Anyone have any ideas?
MACBs and inverter checked on another crate - no signal
Inverter put back in CAEN crate and no noise observed on scope.
With waveforms disabled am able to get <100kHz on all FEEs with the pulser connected
Tested adding bias filters - No change
15:30 Nic realised that one of the mains boxes is labelled measurement network so we connect the AIDA PSU to this via the ac stabiliser
We get ok stats even with waveforms on at 0xa- attachment 5
Waveforms still have the 100kHz oscillation in them though attachments 6 and 7
Without waveforms we get similar performance to before the switch to the new mains - attachment 8
15:56 Move CAEN NIM crate across to this mains network
300+kHz on all FEEs with waveforms on. Can recover waveform off performance though
16:01 Place CAEN NIM crate back onto platform mains but don't recover the ealier rates with waveform. Everything still around 300kHz.
17:43 Current best settings are waveforms off and we can run at 0xa on all FEEs - attachment 9 |
|
483
|
Sun Jun 19 00:22:24 2022 |
TD | Sunday 19 June |
01.20
ASIC settings 2021Apr29-13-16-00
*except* slow comparator 0x64 -> 0xa
All waveform ADCs disabled
BNC PB-5 pulser
to p+n FEE64s only
all system wide checks OK *except* ADC calibration, WR decoder & FPGA timestamp errors - see attachments 1 & 2
DSSSD bias & leakage current OK - see attachments 3 & 4
FEE64 temps OK - see attachment 5
statistics OK - ADC data, pause, resume & correlation scaler data items
data push, data flush, aida01, aida02 - see attachments 6-13
per FEE64 1.8.L spectra - see attachment 14
aida01 pulser peak width 73 ch FWHM |
|
482
|
Sat Jun 18 09:46:57 2022 |
OH, NH | Friday 17th June |
Summary of the day:
Small improvements were made throughout the day but by far the biggest change to the rate in the FEEs came from disabling the waveforms
This was done by setting ADC con
Initial jumper configuration:
FEE LK
1 2 4
2 1 2 4
3 2 3 4
4 2 4
5 2 4
6 1 2 4
7 2 3 4
8 1 2 4
13:22 LK2, 3 and 4 on all p+n - attachment 1
14:20 AC mains relay connected to AC voltage stabilisers + on separate mains socket from DESPEC - attachment 2
Same configuration but waveforms disabled (p+n improvment) - attachment 3
Layout 1 - attachment 4
14:32 all FEE power cables moved to PSU 2 (Lower one in rack, upper one was slow to start for 1 FEE pair, likely the 5V slow to start)
Previously all FEEs on their own pair within a psu
New order 1,3 2,4 5,7 6,8
Waveforms on - all bad - attachment 5
Waveforms off - all p+n better - attachment 6
14:54 interlock removed from AC mains relay
Also noticed adaptor board on 2 not in fully -> Now in
p+n generally better - attachment 7
Without waveforms - p+n better again - attachment 8
16:04 Platform moved into position
Interlock power cable removed from extension lead
w/ waveform - p+n slightly worse
w/o waveform - p+n same as beffore
16:14 LK3 removed from 1 and 5
w/ waveform -> p+n slightly worse - attachment 9
w/o waveform -> p+n best yet - attachment 10
16:33 LK1 wasn't on all n+n was only on 2, 6 and 8
Removed from 6 -> only one LK1 per DSSD
w/ waveforms - all bad - attachment 11
w/o waveforms - no change - attachment 12
16:59 LK1 removed from 2->4 This is the FEE with the BIAS braid
w/ waveform p+n better - attachment 13
w/0 waveform p+n better, n+n worse? - attachment 14
17:21 Bias filters added to n+n
w/ waveform - no change - attachment 15
w/o wwaveofrm - P=n worse - attachment 16
Stats 0x21 - attachment 17
Stats 0x14 - attachment 18
Stats 0x12 - attachment 19
17:42 Bias filter removed and bias floated (LK1 removed)
w/ waveform all worse - attachment 20
w/o waveform -> p+n worse, n+n similar
18:29 LK1 back on 4+8
w/ waveform -> poor
w/0 waveform - recovered previous - attachment 20
18:32 LK 2 and 4 removed from 2 +6
Also noticed ground lemo on aida02 not fully inserted which has been corrected
w/ waveform poor - attachment 21
Threshold 0xF - attachment 22
w/o waveform - attachment 23 |
|
481
|
Tue Jun 14 12:15:33 2022 |
NH | S450 Archived |
All MIDAS files for S450 (/TapeData/S450) have been copied to lustre and archived |
|
480
|
Mon Jun 13 16:38:37 2022 |
NH | Mon 13 Jun |
Remove Caen HV Supply
Open up, C23 was grounded (HV grounded in NIM)
Remove C23 (HV floating at NIM) and reinstall
Rebias HV... No Change
Reboot AIDA... No change |
|
479
|
Fri Jun 10 10:32:46 2022 |
NH | Fri 10 Jun |
Probe PSU output with 'scope again and 10X probe
Use a AC mains scope as was available, means referenced to ground of AC, but should be ground of platform (and hence Fee64)
Figs 1-3: Traces of PSU outputs (5V, -6V, 5V Return) with 10X probe, 1 MOhm impedance AC coupled
Lots of fluctuation and jitter
Figs 4-6: DC 1MOhm output from all 3 power rails
Fig 7-8: Small "loop" using probe and observed scope spectra.
After lunch ideas:
- Try again with 1X probe attenuation, maybe record output to USB
- Remove unconnected MACBs from master
- |
|
478
|
Wed Jun 8 09:55:28 2022 |
NH | Weds 8 May |
10.30 : Turn off AIDA FEE64s
Prepare a ground cable for snout - M10 terminal on one end for copper bar. Will affix with Cu tape for now (no free screws)
Plan for today:
Remove extraneous ground cables
Check ground of all FEE64s esp. 05 and 07 - include on FEE64 side.
14:42 - Done all grounds seem snug, extra ground cables removed. Try power on and check
Try powering just DSSD1 FEE64s
- No change (aida01 125k, aida02,03,04 - 400k)
Try grounding snout
14:59 Connection to snout very bad as tape does not fix well to thick copper in braid. nO difference either way.
Probe PSU output with Battery DSO if available
- See figure 2
Try shuffling FEE64 power allocation
- 15:32 - Now as 1 X 3 X 2 X 4 X
5 X 7 X 6 X 8 X
(Also FEEs on own pair)
No difference to rates
Multimeter voltages for last socket of PSU1 (paired with aida04):
1 : 6.02 V
2: 6.02 V
3: 6.02 V
4: -6.44 V
5: 7.74 V
Similar on PSU2
Voltage between right (GND) pin 5 and 1 (pin 5 is COM) is -0.4 V, similar if pin 4 is COM. Pin1-3 are 0 V when paired
|
|
477
|
Wed Jun 1 08:00:18 2022 |
TD | 2xDSSSDs spectra layouts |
Spectra layouts for 2x DSSSDs can be found in directory ~/LayOut/2xDSSSDs |
|
476
|
Wed Jun 1 07:35:13 2022 |
TD | Wednesday 1 June |
Attachment 1 - DSSSD bias & leakage current - OK
Attachment 2 - ADC data item stats
Attachment 3 - FEE64 temps - OK
Attachments 4-7 - aida01, 02, 03 & 04 2.*.W spectra
Noise observed for asics #1 & #2 of aida02 consistent with no cable attached - confirmed by visual inspection - ribbon cable disconnected from Samtec header of adaptor PCB
To Do list
1) re-connect ribbon cable to aida02 adaptor PCB
2) good quality ground to snout - use screws at inter-stage
3) install heavy duty grounds
4) ...
Attachment 8 - adc data item stats - slow comparator 0x64
Attachment 9 - per FEE64 rate spectra - slow comparator 0x64
Attachment 10 - per FEE64 rate spectra - slow comparator 0xa
-
13:30 - Attach heavy duty ground to AIDA
Attachment 11 - per FEE64 rate spectra - slow comparator 0xa
Try power cycle as system ground changed significantly now
Try to reattach aida04 TTY connector - still wrong
Archive old tty logs (ttyUSB*) to zip and delete them, makes easier to track startup
Attachment 12 - per FEE64 rate spectra - slow comparator 0xa
No major difference, aida01 quieter than before
aida05 has some missing channels - loose ERNI?
Attachements 13 & 14 - Waveforms
Temporarily raise to 0x64 and test merger for DTAS synchro test later... all works good to MBS
Change NETVAR MERGE.LinksAvailable to 8
Notice one copper screw touching aluminium, add some kapton tape behind it to isolate it. No difference
Voltage diff between aida02,aida06 < 1 mV
Voltage diff between copper bars < 0.5 mV (but not 0)
A comment from DTAS people that they observed a large increase in noise recently - perhaps from accelerator
May indicate current power/gronud oddities?
Attachment 15 - Rates at 0x1f (= 310 keV)
DSSSD#1 reasonable (aida04 a bit bad), but mostly just pulser (25 Hz)
DSSSD#2 much worse esp. p+n sides (aida05, aida07)
ERNI connectors pushed in and feel fully in |
|
475
|
Tue May 31 17:09:35 2022 |
TD | Tuesday 31 May |
2x MSL type BB18(DS)-1000 installed
Adaptor PCB LK config
all FEE64s LK2, LK4 fitted
n+n FEE64s LK1 fitted
bottom p+n FEE64s LK3 fitted
AIDA PSU config
top 1-3 2-4
bottom 5-7 6-8
DSSSD bias -> p+n FEE64 core, n+n FEE64 braid
Heavy duty grounds not connected to FEE64 adaptor PCBs
ASIC settings 2021Apr29-13-16-00
slow comparator 0xa
BNC PB-5 settings
amplitude 1.000V
attenuation x1
tau_d 1ms
frequency 25Hz
polarity +
Attachment 1 - DSSSD bias & Leakage current - OK
Attachment 2 - FEE64 temps OK
Attachment 3 - adc data item stats
Attachments 4-9 - system wide checks OK *except* aida07 fails ADC calibration
Attachments 10 & 11 - 1.8.W spectra 20us FSR
Attachments 12 & 13 - 1.8.L spectra - aida03 pulser peak width 155 ch FWHM
Attachment 14 - per FEE64 rate spectra
19.15 Attachments 15-18 - 2*W spectra for aida01, 02, 03 &04
Attachment 19 - per FEE64 rate spectra
Attachment 20 - adc data item stats |