214
|
Sat Apr 10 12:15:54 2021 |
OH, TD | Saturday 10 April | 12:37 DAQ found crashed. Had crashed at some time following 7:50am following previous statistics update
Unable to recover the FEEs so forced to power cycle
Following power cycle statistics much improved from yesterday.
In the waveforms the lower frequency component that was prevalent yesterday is no longer visible.
The waveforms are not perfect by any means though
Also note that following an ASIC synchronisation all of the FEEs lose ADC calibration and must be manually recalibrated.
16.42 UTC+1
per FEE64 1.8.W spectra
20us FSR attachments 7 & 8
200us FSR attachments 9 & 10
2ms FSR attachments 11 & 12
20ms FSR attachments 15 & 16
rate spectra attachment 13
good events statistics attachment 14 |
|
291
|
Tue May 4 13:53:22 2021 |
OH, TD | Tuesday 4 May |
NH has re-checked alignment of AIDA adaptor PCBs cf. Elog:290
14.40 Rate spectra - attachment 1
adc data items - attachment 2
slow comparator 0xa
stats for p+n much improved, n+n remain poor
1.8.W spectra - attachments 3 & 4
1.8.L spectra - attachments 5 & 6
pulser peak widths aida01 69 ch FWHM, aida02 247 ch FWHM
17:31 Bias lowered to 100V on DSSD2 after noticing the leakage current has increased 2uA over the previous 3 hours. - attachment 7 |
|
341
|
Fri May 21 07:03:50 2021 |
OH, TD | Friday 21 May 08:00-16:00 | 08:00 OH and TD take over
Current settings p+n sides at 0xc
DSSD1 n+n sides 0x20 on ASICs 1-3 and 0x64 on ASIC 4
DSSD2 n+n sides 0x1b on ASICs 1-3 and 0x64 on ASIC 4
PULSER SETTINGS
---------------------------
Pulse is ON
Positive Tail Pulse
Trigger Source is Internal Clock
Trigger Threshold is 3.5
Amplitude : 2.0 Volts
Rep Rate : 2.0 hZ
Delay : 250.0 ns
Fall Time : 1 ms
Attenuation : 1
Display is : Volts
Equivalent keV is : 200.0
Ramp Start at 0.01 Volts
Ramp Stop at 9.99 Volts
Ramp Start at 1.0 keV
Ramp Stop at 999.0 keV
Ramp Time is 60 seconds
# Ramp Cycles is 1
08:29 System wide checks ok - WR difference at 111
Statistics ok (Still high following the sharp rise at 18:30 yesterday) - attachment 1
Temperature ok - attachment 2
Bias and leakage current ok - attachment 3
09:58 System wide checks ok - WR difference at 111
Statistics ok - attachment 4
Temperature ok - attachment 5
Bias and leakage current ok - attachment 6
Max deadtime this morning has been around 5.5% on FEE8
12:32 System wide checks ok - WR difference at 111
Statistics ok - attachment 7
Temperature ok - attachment 9
Bias and leakage current ok - attachment 8 |
|
353
|
Wed Jun 2 14:53:05 2021 |
OH, TD | Merger/Tapeserver issues after reboot of aida-3 | On 01/06/21 aida-3 was rebooted as the graphical user interface had frozen.
We were also unable to connected to MIDAS via ssh port forwarding.
Since the restart we have been unable to write to file.
The tapeserver will allocate/mount/open a file but nothing will be written to it.
The merger will be receiving data but never shows as having any "Current links with data".
/MIDAS currently points to lrwxrwxrwx. 1 root root 45 Jan 23 2019 /MIDAS -> /home/npg/MIDAS_Releases/23Jan19/MIDAS_200119
The current disk structure is:
[npg@aidas-gsi npg]$ lsblk
NAME MAJ:MIN RM SIZE RO TYPE MOUNTPOINT
sr0 11:0 1 1024M 0 rom
sdb 8:16 0 7.3T 0 disk
├─sdb1 8:17 0 200M 0 part
├─sdb2 8:18 0 500M 0 part
└─sdb3 8:19 0 7.3T 0 part
├─vg_aidas2-lv_root (dm-2)
253:2 0 50G 0 lvm
├─vg_aidas2-lv_home (dm-3)
253:3 0 7.2T 0 lvm /media/1e121361-83d3-4825-b6ae-8700b07e0ca7
└─vg_aidas2-lv_swap (dm-4)
253:4 0 7.8G 0 lvm
sdc 8:32 0 7.3T 0 disk
└─sdc1 8:33 0 7.3T 0 part /media/ThirdDrive
sdd 8:48 0 7.3T 0 disk
└─sdd1 8:49 0 7.3T 0 part /media/SecondDrive
sda 8:0 0 465.8G 0 disk
├─sda1 8:1 0 200M 0 part /boot/efi
├─sda2 8:2 0 500M 0 part /boot
└─sda3 8:3 0 465.1G 0 part
├─vg_aidasgsi-lv_root (dm-0)
253:0 0 50G 0 lvm /
├─vg_aidasgsi-lv_swap (dm-1)
253:1 0 7.8G 0 lvm [SWAP]
└─vg_aidasgsi-lv_home (dm-5)
253:5 0 407.3G 0 lvm /home
Terminal outputs for the HTTPd, Merger and TapeServer terminals - attachments 1-3
TapeServer config file - attachment 4
Merger options - attachment 5
Note in the merger window the top bar is different to what we have seen previously - Attachment 6 and https://elog.ph.ed.ac.uk/DESPEC/210418_124830/51.png |
|
442
|
Fri May 13 07:36:18 2022 |
OH, TD | Friday 13 May | 08:36 DAQ Still running smoothly
Currently on run R1_328
At current data rates we have just over 4 days of space left on the current disk
Analysis of R1_327 - attachment 1
14.15 all histograms zero'd
baseline system wide checks counters
DAQ continues file S450/R4_33
netint pushenable 6 - was 60 - to investigate FEE64 throughput issues
netint flushenable 60 - not changed
ASIC settings
slow comparator aida02 aida04 0x16, aida06 aida08 0x23, aida12 0xd, all others 0xc
BNC PB-5 settings
amplitude 1.000V
attenuation x1
frequency 2Hz
tau_d 1ms (tail pulse)
+ polarity
Attachments 2 & 3 - ucesb
Pb beam intensity c. 4e+8/spill - lower than expected 1e+9/spill . Low implant rates (~Hz/spill) are expected for setting.
Attachment 4 & 5 - grafana DSSSD bias, leakage current & temp - OK
Attachments 6-8 - NewMerger stats blocks 0, 1, & 6
Attachment 9 - NewMerger link stats
Attachments 10-11 - TapeServer/NewMerger
Attachment 12 - iptraf aida-gsi network interfaces
Attachments 13-14 - per FEE64 1.8.L spectra
pulser peak width aida01 123 ch FWHM, aida04 430 ch FWHM
Attachments 15-18 - per FEE64 rates and stat spectra
Attachments 19-22 - system wide checks
aida09 clock status 6, WR decoder status errors
Attachments 23-26 - adc, pause, resume correlation sclaer data item stats
Attachment 27 - FEE64 temps OK
Attachment 28 - DSSSSD bias and leakage currents OK
14.30 check ASIC load
14.47 attachment 29 - analysis S450/R4_38
15.11 Attachments 30-35 - data push, flush stats for all FEE64s and aida01 (low ADC data item rate) and aida06 (high ADC data item rate) |
|
486
|
Mon Jun 20 09:26:23 2022 |
OH, TD | Monday 20 June | 10:26 MACB HDMI for AIDA03 and AIDA04 reseated.
Merger and DAQ restarted - All happily sending data
Has been running for 10 minutes with no WR errors.
Rates from yesterday achieved again.
DTAS will move back into position at some point.
Once this is done, will perform pulser walkthrough.
11:14 DTAS now back in position for experiment
Slight increase in noise noticed across a few FEEs
Over time this has decreased on all but AIDA07 which is now running at ~80k - attachment 1
AIDA03 is now however gaining white rabbit errors at a rate of 1 every few minutes.
Less frequently gaining FPGA errors
12:20 AIDA power cycle to reseat the HDMI for AIDA03 - It would eventually crash if not done
Upon powering up again rates in AIDA08 slightly worse - attachment 2
13:26 No WR errors since the last power cycle.
21:16 Baseline system wide checks, zero histograms & stats
All system wide checks OK
per p+n FEE64 1.8.L spectra - attachment 3
aida01 pulser peak width 82 channels FWHM
per FEE64 rate spectra - attachment 4
adc, pause, resume and correlation scaler data item stats - attachments 5-8
FEE64 temps OK - attachment 9
DSSSD bias & leakage currents OK - attachments 10 & 11 |
|
307
|
Wed May 12 06:59:55 2021 |
OH, RDP | Wed 12 May 08:00-16:00 | 08:00 While the rates appeared to have quietened around 7am they are back to their previous rate now.
They still seem to be working on the setting. Implant rate in AIDA about 8 per spill. Thick degrader is still in place
They have collected 25 minutes of implantation data. They are analysing this data to confirm the PID
08:38 The PID confirmation was apparently run number 4. During this run they were implanting Cd in AIDA DSSD2
09:09 They are now taking data implanting somewhere in the snout. I have turned off no storage.
Unfortunately we are writing to NOTAPE R3. I was unable to change it as they were already taking data
Data rate 16.3MB/s
ucesb shows similar rates in SC42 and AIDA01. We may be implanting Te in AIDA then.
09:35 They stopped the DAQ as their is no beam. Switched the tapeserver over to R1 in directory S496. Setting to no storage
09:47 Beam back turned off no storage. On file R1_10
11:04 Now that we have implants on the detector we get a better look at the strips
From the 1d pattern it is obvious that there is an adapter card misaligned in DSSD1- attachments 1 and 2
This adaptor board corresponds to FEE11
12:17 Leakage currents screenshot 512.png
13:30 Beam off. Expected back at 15:00...
14:09 Making use of the beam off time NH and HA entered S4 to investigate FEE1 and FEE11.
They noticed FEE11 was off and have realigned that. All channels are now observed in that FEE
They couldn't see anything wrong with the connector for FEE1. We are still missing a block of channels there.
Rates are as before. No better, no worse.
15:31 Leakage currents screenshot 513.png. They are continuing to rise following the intervention in S4. It could just be because the temperature is rising... |
|
323
|
Tue May 18 06:50:53 2021 |
OH, PP | Tuesday 18th May 08:00 - 16:00 | 08:00 OH Takes over
Statistics still look good from yesterday. Rate is still down.
Dead time in FEEs 2 and 4 is 2.04%
System wide checks clock ok
Base Current Difference
aida05 fault 0x4da : 0x500 : 38
White Rabbit error counter test result: Passed 15, Failed 1
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
*Baseline reset*
FPGA check ok
Stats - attachment 1
Temp - attachment 2
Bias - attachment 3
08:11 With the decreased noise I am investigating whether we can lower the thresholds on DSSD2 y strips.
I have set them to 0x19 on ASICS 1-3 and 0x64 on ASIC4
The results of the test appear to indicate that it is the network link that we are saturating causing dead time
*** Before After
FEE DT (s) Dead time % DT (s) Dead time %
0 0.013 0.004184451 0.026 0.010025681
1 6.352 2.04458693 4.097 1.57981599
2 0.295 0.09495484 0.292 0.112596112
3 6.349 2.043621288 8.255 3.183153771
4 0.021 0.006759497 0.03 0.011568094
5 0.285 0.091736032 3.388 1.306423377
6 0.02 0.006437616 0.149 0.057454865
7 1.219 0.392372712 2.133 0.822491459
8 1.137 0.365978485 0.457 0.176220627
9 0 0 0 0
10 1.633 0.525631369 1.545 0.595756823
11 0.061 0.01963473 0.166 0.064010118
12 0.059 0.018990968 0.085 0.032776265
13 0 0 0 0
14 0.016 0.005150093 0.023 0.008868872
15 0.282 0.09077039 0.168 0.064781324
Note that the dead time for FEE6 and 8 remains fairly low but the dead time for FEE4 has increased
09:00 They are stopping the run to decrease the bias on the bPlast detector
09:25 No noticable change in AIDA
They will now begin to change the spill structure.
09:46 While they are optimising the 1s spill structure raised the thresholds of FEE6 and 8 back to 0x20. Will start from here when optimising the thresholds post spill change
10:21 Beam is back on AIDA
10:39 System wide checks.
- ADC calibration check failed (this is anticipated because waveforms are off).
- 1 Failed test in White Rabbit:
Base Current Difference
aida05 fault 0x500 : 0x506 : 6
White Rabbit error counter test result: Passed 15, Failed 1
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
- All other systems passed
11:00
Stats - attachment 4
Temp - attachment 5
Bias - attachment 6
12:46 System wide checks
- ADC calibration check failed (this is anticipated because waveforms are off).
- 1 Failed test in White Rabbit:
Base Current Difference
aida05 fault 0x500 : 0x50b : 11
White Rabbit error counter test result: Passed 15, Failed 1
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
- All other systems passed
14:36 System wide checks
- ADC calibration check failed (this is anticipated because waveforms are off).
- 1 Failed test in White Rabbit:
Base Current Difference
aida05 fault 0x500 : 0x50f : 15
White Rabbit error counter test result: Passed 15, Failed 1
Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR
- All other systems passed |
|
412
|
Wed Apr 6 10:09:21 2022 |
OH, NH, TD | Wednesday 6 April | DSSSD bias OFF. DAQ STOPped
Attachment 6 FEE64 temps OK
Attachment 1-5 system wide checks OK |
|
413
|
Thu Apr 7 12:42:14 2022 |
OH, NH, TD | Thursday 7 April |
Re-assembled triple snout re-installed on the AIDA stand - DSSSD ribbon cables and grounds re-connected.
FEE64 adaptor PCB LKs per https://elog.ph.ed.ac.uk/DESPEC/409
Attachments 1 & 2 1.8.L spectra
p+n FEE64 aida10 pulser peak width 97 ch FWHM
n+n FEE64 aida04 pulser peak width 550 ch FWHM
Attachment 3 rate spectra
missing channels aida01, aida07 & aida11 - misaligned ERNI and/or Samtec connectors?
Attachments 4 & 5 1.8.W spectra
Attachment 6 ADC data item rates
aida10 OK, all others c. 300-400k
Attachment 7 FEE64 temperatures OK
Attachment 8 DSSSD bias & leakage currents OK |
|
428
|
Wed May 11 07:04:13 2022 |
OH, NH, TD | Wednesday 11th May | 08:00 Overnight at some point it appears that aida14 has stopped producing any ASIC data
It is currently sending data pause requests and data push requests but nothing else - attachment 1
telnet into aida14 and run top to see what the current state of the FEE is. AIDAExec currently using 0% CPU usage.
FEE appears to have hung. - attachment 2
Still passes all system wide checks.
When performing a check load can see the tclsh process take up CPU usage
09:02 AIDA out of time sorter. Going to perform a power cycle to see if it recovers FEE14
AIDA14 recovered but ASIC2 still missing
Current working hypothesis is ASIC2 on AIDA14 is dead and that the problem with AIDA05 is either DSSD, cable or adaptor board.
Given access would change the adapter board on AIDA05.
Could consider installing the old AIDA05 FEE above AIDA14 and moving the cables up 1.
Would retain the option of a speedybacktrack
Alternatively the HEC events in AIDA05 are just above threshold. Can raise the HEC threshold
A threshold of 0x4 reduced the rate in the HEC channels considerably
A threshold of 0x6 removes entirely
Note when the rate of HEC channels decreases can see the pulser in all channels of AIDA05 again. Showing that the adapter board is properly aligned
09:59 Had access so replaced the adapter card on AIDA05
Did note that while replacing the card the right hand edge of the right hand ribbon cable was lifted slightly.
This could also have been the cause
10:09 Restarted DAQ and there is no change to the behavior of AIDA05
Still see events in HEC
Again raising threshold to 0x5 in ASIC1 largely removes them.
13.10 Attachments 3-5
DAQ histogramming only enabled, waveforms/data transfer disabled, sampling ADCs OFF
ASIC check load
ASIC slow comparator all FEE64s all ASICS 0xa
ADC data item stats
13.17 Attachments 6-9
DAQ histogramming & data transfer enabled, waveforms disabled, sampling ADCs OFF
ADC data item stats
NewMerger/Tape Server
18:00 As part of tests in the afternoon it was investigated powering only one DSSD worth of FEEs at a time
It was found that reasonable rates could be found for DSSD1 but poor rates only for DSSD2
To try and help isolate the two detectors the jumpers 2 and 4 were removed from FEEs 13, 5 and 14 (Top, DSSD2)
After powering on it was found there was no improvement and possible worse noise than before
We were able to recover the rates in the 1DSSD setup so instead have removed FEE power from FEE5 and FEE14 |
|
92
|
Tue Nov 12 13:22:19 2019 |
OH, NH, CA | AIDA09 Replacement | ASIC 1 on FEE9 was not producing signals.
A new HDMI cable was installed but still no signals.
AIDA09 was replaced with a new FEE.
Current list of FEE to serial numbers installed attachment 1
MAC address of new FEE was obtained.
Backup of DHCPD.conf was made dhcpd.confBACKUP191112
dhcpd.conf was updated with new MAC address.
All FEEs powered on and 09 was seen to mount and is seen by AIDAServer.
AIDA09 flashed to newest version of firmware. (0x18430701) - attachment 2 |
|
390
|
Wed Oct 27 15:12:50 2021 |
OH, NH | Noise tests - 27th October | Began by installing the triple.
Only the upstream detector was connected to the FEEs the downstream detector cables were left loose. The two detectors are isotolated though.
Initial cabling configuration
No pulser connections to any of the FEEs connected to the DSSD.
Bias core connected to FEE 13 - FEE 3 and FEE 11 with connections made by T pieces
Bias braid connected to FEE 4 - FEE 2 with connection made by T piece
Adaptor boards grounded to FEE cooling plate via lemo
DSSD cable drain wires also connected to LEMOs
LK1 not on any adaptor boards.
LK2, 3 and 4 on all
System started and rates very high in all of the FEEs connected to DSSD.
N.B FEE 8 connected to pulser but not to cable and adaptor board not grounded to FEE cooling plate
FEE14 connected to pulser and adaptor board grounded to FEE cooling plate
Statistics - Attachment 1
It was noted one of the ribbon cable connectors for FEE13 had come off
Statistics fixed - Attachment 2
Representative waveforms showing the noise - Attachments 3 and 4
Bias turned off and HV filters connected
Issue at first as the solder connection in one filter had broken. Filter was replaced and the DSSD bias
Statistics are unchanged
Waveforms with the HV filters installed - Attachments 5 and 6 |
|
397
|
Tue Mar 1 08:29:19 2022 |
OH, NH | Findings from the upgrade to centos7 from SL6 | Required packages not included as default
- csh (For running MIDAS scripts)
- xinetd (Runs the rdate server for the FEEs)
- Telnet (Communication with the FEEs)
- dhcp
Configuration changes
Optional extras
- For top bar icons similar to SL6 can use the gnome extension "Frippery panel favorites"
- Then just need to create desktop icons for each of the startup scripts within /usr/share/applications - Attachment 1
- Can also add screenshot to this - Attachment 2
- Gnome tweaks also allows the number of workspaces to be changed
- Also recommend setting workspaces to span displays
|
|
400
|
Wed Mar 2 16:00:01 2022 |
OH, NH | Noise tests - 2nd March | An interesting find with the new sum inverter is that with the pulser at default settings the pulser peak on the n+n side is above the range of the ADC.
We have checked signals from the sum inverter and pulser and both are equal in magnitude.
We therefor determine it is over range as the charge is spread over 4 FEEs vs 12 on the p+n side.
This possibly wasn't an issue with the previous inverter if it produced a lower amplitude.
For tests going forward p+n voltage will be 2V. n+n voltage will be 0.5V - Attachment 1
Going to look into a 20dB attenuator
Seems there were issues with the lemo cables in the n+n pulser chain |
|
401
|
Thu Mar 3 08:45:03 2022 |
OH, NH | AIDA Noise and grounding tests 3rd March | Pulser spectra
- Changes to the nn pulser circuit yesterday have improved the peak widths on the p+n side
- p+n peak widths averact around 100 channels
- n+n around 1000 at 0xa threshold
- n+n 1.8.L spectra - attachment 1
- p+n 1.8.L spectra attachment 2
p+n peak widths
| FEE |
Width |
| 9 |
105 |
| 1 |
82 |
| 10 |
71 |
| 11 |
79 |
| 3 |
96 |
| 12 |
72 |
| 13 |
191 |
| 5 |
101 |
| 14 |
129 |
| 15 |
118 |
| 7 |
110 |
| 16 |
100
|
n+n peak widths
| FEE |
Width |
| 2 |
1200 |
| 4 |
1179 |
| 6 |
- |
| 8 |
1813 |
Daily checks
- Fee temperatures ok - attachment 3
- Bias and leakage currents ok - attachment 4
- Stats spectra with current grounding - attachment 5
- Current grounding is just a cable going between each adaptor board and its respective FEE
- Interfee grounding done by power supply and also pulser circuit
- Rates for each of the asics - attachment 6
- p+n waveforms - attachment 7
- n+n waveforms - attachment 8
- Lots of noise evident on them
Plans for the day
- Connect FEEs up to the copper earth bars installed
- Powercycle and check noise levels
- If n+n remains significantly higher investigate the n+n pulser circuit further
- Investigate the effects of different jumper connetions At present n+n have all LK jumpers on and p+n have lk2-4
|
|
402
|
Fri Mar 4 08:40:36 2022 |
OH, NH | AIDA Noise and grounding tests 3rd March - Results | Grounding implemented
- Each FEE given direct connection to heavy duty earth bar via adaptor board lemo ground
- Earth bars are then connected to platform ground on AIDA frame
- Photo of grounding - attachment 1
- Improvements not as large as expected from previous experience though
- Particularly in the n+n side where widths still appear awful
- While implementing the grounding it was noticed the ground lemo socket on FEE7 was damaged and no longer gripped onto a lemo cable
- It was noted that it would need changing at some point but continued with it for the time being
- Jumper configuration was n+n LK1-4 all on, p+n LK2-4 all on
p+n widths
| FEE |
Width |
| 9 |
147 |
| 1 |
116 |
| 10 |
92 |
| 11 |
77 |
| 3 |
128 |
| 12 |
72 |
| 13 |
75 |
| 5 |
59 |
| 14 |
75 |
| 15 |
80 |
| 7 |
150 |
| 16 |
75 |
- Compared to the width with no grounding the FEEs for the first DSSD (First 6) range from no change to slightly worse
- 1.8.L spectra - attachment 3
n+n widths
- Widths still very large
- Width for 8 is around 500
- 1.8.L attachment 2
Additional info
- Statistics - attachment 4
Jumper optimisation
- Went to our previously found optimal jumper configuration for the trips
- n+n LK1 only
- p+n LK 2 and 4 on all FEE
- p+n LK3 on non isolated cables (PCB ground on lower middle cable and n+n side bias on top middle cable)
- While changing the jumpers we also swapped out the damaged adaptor board on aida07 for a new one
- Saw an improvement for most FEEs
| FEE |
Width |
| 9 |
134 |
| 1 |
83 |
| 10 |
80 |
| 11 |
68 |
| 3 |
60 |
| 12 |
60 |
| 13 |
71 |
| 5 |
60 |
| 14 |
75 |
| 15 |
82 |
| 7 |
97 |
| 16 |
77 |
- Unfortunately saw no improvement for the n+n
- We event tested having the pulser only going into a single FEE (aida08)
- Still have a width of 622 for that FEE
- To compare FEEs we have layout 1 - attachment 5
- Layouot 1 on a common timescale - attachment 6
- Note that the rate is typically dominated by a single strip or a couple of strips within a FEE
- Many channels are just showing the pulser rate 25Hz
- p+n 1.8.L attachment 7
- p+n waveforms attachment 8
- n+n waveforms attachment 9
- Testing the noise on the n+n strips we observe noise with slow comparator set to 200 (2MeV)
- Meanwhile the p+n strips are silent at 0x64 for the most part.
- DAQ left running overnight with 0x64 on ASIC settings to see if we observe any alphas
Plan for the day
- Test daisy chaining the n+n across to the other n+n strips
- Currently the n+n braid is only plugged into either aida06 or aida08 (depending on DSSD)
- All of the jumpers are however grounded
- Will still test
- May investigate also trying this on LK3 on adaptors for aida01 and aida03
- Observe the output of the new sum inverter on the scope to check for variation in amplitude
|
|
403
|
Sat Mar 5 10:43:45 2022 |
OH, NH | Grounding and noise tests 4th-5th March Summary | Tests performed
- After leaving overnight noticed the pulser was not fully inserted into aida09
- Little change in rates observed in 9 but otherwise nothing else
- Connected bias of aida02 (n+n ground) to aida06 (n+n ground) and likewise 4-8
- No change observed in rates
- Also tested covering the snout in light tight material
- Unplugged the system monitor from 6 and 8, no change
- Connected the snout to ground with copper tape
- Rates became very unpredictable in the FEEs unsure if this was snout grounding related though
- Removed the ground and powercycled and rates restored to where previously
- Changed the bias supply jumper inside to have the bias line connected to the rack ground
- At the same time noticed that the ground cable connecting the adaptor board of aida06 to the FEE had become disconnected from the ring terminal
- Recrimped and inserted again
- Saw an improvment in the rates of the p+n strips and also the n+n strips - attachment 1
- Pulser widths however larger than yesterday
p+n width
| FEE |
Width |
| 9 |
118 |
| 1 |
88 |
| 10 |
75 |
| 11 |
77 |
| 3 |
67 |
| 12 |
66 |
| 13 |
96 |
| 5 |
89 |
| 14 |
107 |
| 15 |
93 |
| 7 |
118 |
| 16 |
87 |
n+n width
- Connected up entire pulser circuit to n+n FEEs
| FEE |
Width |
| 2 |
428 |
| 4 |
N/A |
| 6 |
428 |
| 8 |
343 |
Switching bias supply back
- because changes were made to the n+n ground, and the pulser circuit it was decided to go back and check the previous bias configuration with the module set in a floating configuration
- DAQ was then left overnight as had to leave to catch the bus
- Rates are observed to be back at their higher values - attachment 2
- n+n 1.8.L spectra - attachment 3
- p+n 1.8.L spectra - attachment 4
p+n widths
| FEE |
Wdith |
| 9 |
130 |
| 1 |
128 |
| 10 |
75 |
| 11 |
78 |
| 3 |
71 |
| 12 |
71 |
| 13 |
96 |
| 5 |
85 |
| 14 |
110 |
| 15 |
99 |
| 7 |
123 Dbl |
| 16 |
92 |
n+n widths
| FEE |
width |
| 2 |
56 |
| 4 |
983 |
| 6 |
523 |
| 8 |
358 |
- The results show no improvment to slightly worse performance so will go back to having the bias module grounded internally and check the previous conclusions hold before making further changes
Return to grounded bias module
Widths
p+n
| FEE |
Width |
| 9 |
106 |
| 1 |
79 |
| 10 |
72 |
| 11 |
78 |
| 3 |
70 |
| 12 |
73 |
| 13 |
102 |
| 5 |
112 |
| 14 |
105 |
| 15 |
110 |
| 7 |
131 |
| 16 |
104 |
n+n
| FEE |
Width |
| 2 |
- |
| 4 |
298 |
| 6- |
- |
| 8 |
278 |
- Don't see a pulser peak in 4 or 2
- Will replace cable between 4 and 2 and 2 and 6
- Statistics at 0x64 with pulser on - atachment 5
- Now see peaks - attachment 6
- Statistics at 0xa - attachment 7
| FEE |
Width |
| 2 |
284 |
| 4 |
1089 (Triple) |
| 6 |
371 |
| 8 |
318 |
With the grounded bias in place in the module move the braid cable to the frame rather than being connected to bias line
- Still observe a triple peak in 4 and have worse perfromance in 2 and no improvement in p+n
- 0x64 stats - attachment 7
- 0xa statistics - attachment 8
- p+n 1.8.L - attachment 9
- p+n waveform - attachment 10
- n+n waveform - attachment 11
- Same high frequency component in both
- 35 channel spacing corresponds to a frequency of ~1.4MHz
| FEE p+n |
Width |
| 9 |
109 |
| 1 |
118 |
| 10 |
78 |
| 11 |
80 |
| 3 |
70 |
| 12 |
73 |
| 13 |
92 |
| 5 |
85 |
| 14 |
99 |
| 15 |
100 |
| 7 |
123 |
| 16 |
92 |
| FEE n+n |
Width |
| 2 |
531 |
| 4 |
1185 (Triple) |
| 6 |
536 |
| 8 |
376 |
Next tests
- Returned braid to FEE 4 and 8
- Recovered earlier rates and performance
- Re-added the ground cable to the snout
- Much worse noise performance across all FEEs
- p+n rates in 200k
- Removed the ground from the snout
- Got really good rates originally - attachment 12
- Noticed 1 and 5 failed ADC calibration
- Calibrated them and there rates increased massively - attachment 13
- Notice that the pulser peak in FEE2 and 6 keeps dropping out
- As the cables have been changed multiple times thought it could be that an adaptor board was broken
- 4 seemed the most likely suspect with its high rates and triple peaks
- Replaced the adaptor board with one missing an outer ERNI pin (PIN doesn't interface with the FEE so didn't see any issues)
- Pulser peaks were still muissing
- Issue could be then with adaptor board 2 possibly
Conclusions
- Rates on the FEEs are in a better place now than they were at the start of the week
- p+n in particular have shown a large amount of improvement
- There are still however issues with the n+n FEEs
- I think there is likely an issue with one of the adaptor boards
- Possible aida02 at the pulser connection point
- Could explain the triple peaking in aida04 and the pulser issues with aida02 and aida06 dropping out.
- Would recomend swapping out aida02 as a test next week
- Other possible issues coiuld be the clingfilm around the cables has torn and we are grounding the cables to the inside of the snout
- Nic would like to use something other than clingfilm in future
- One suggestion would be mylar which could be held in place with strips of double sided tape
- This would make the cables less likely to stick to each other which is an issue with the tightness of the snout
- Could also try removing LK3 from FEE3 and 7 leaving the PCB ground floating
- I am at a slight loss of what else to try
- I have left the daq running at 0x64 overnight to check for alphas
- Histograms and statistics zeroed
|
|
414
|
Fri Apr 8 08:59:42 2022 |
OH, NH | Friday 8th April | 09:59 Noticed that the right hand ribbon cable for aida07 was offset vertically by 1. i.e. Only 1 of the 2 rows of pins in. Unbiassed detectors and reconnected.
Rebiassed detectors - leakage current on DSSD2 increased to 9.96uA - Attachment 1
Power cycling FEEs
Following restart the FEEs show some improvement in rates - attachment 2
Rate spectra - attachment 3
Still issues with aida11 in the rate spectra (Likely a misaligned ERNI connector)
Pulser on rate spectra - attachment 4
Pulser off rate spectra - attachment 5
We can see that with the pulser on there are events in the channels in aida01 and aida11
With the pulser off these channels disappear.
ERNI connector is then likely aligned the issue is more likely the DSSD ribbon cable is misaligned.
Taking bias off to investigate.
10:39 Could not observe anything out of alignement about either the connectors or the adaptor boards
Note that after pushing on the connector AIDA01 had largely recovered
13:30 Top of snout taken off while mounted in S4. It was observed that one of the ribbon cables for aida11 was off by 1 row.
This was corrected in situ and the snout was put back together and made light tight again.
Oddly leakage current still the same - DSSD2 has come down though - attachment 6
FEEs powercycled for new tests
Half the channels in aida11 had appeared again
Took apart once more and found the other header also off by one row which has been corrected
FEEs powercycled for new test
All of aida11 now showing - attachment 7
Statistics at 0xa shown in attachment 8
All p+n for DSSD1 are good (aida01, 03, 09, 10, 11 and 12)
p+n for DSSD2 are so so (aida05, 07, 13, 14, 15 and 16)
All n+n 2, 4, 6 and 8 are poor still
14:50 p+n waveforms - attachment 9
n+n waveforms - attachment 10
Current jumper configuration
LK1 (Bias ground) on FEEs 6 and 8
LK2&4 on all p+n
LK3 on aida03 and aida07 (Bottom middle)
15:00 Investigated whether there was a light leak. Placed black material over snout.
Leakage on DSSD2 dropped to 5.6 - attachment 11
There was a light leak
Stats have decreased slightly - attachment 12
16:00 Move raspberry pis atop 5mm alminimum plate to keep away from FEE64 PSUs
No change
16:30 Remove pulser network from all FEEs
Rates get noticably worse in many FEEs... why?
Stats - attachment 13
Possible ideas include braid touching things it shouldn't (Adapter PCBs?) |
|
482
|
Sat Jun 18 09:46:57 2022 |
OH, NH | Friday 17th June | Summary of the day:
Small improvements were made throughout the day but by far the biggest change to the rate in the FEEs came from disabling the waveforms
This was done by setting ADC con
Initial jumper configuration:
FEE LK
1 2 4
2 1 2 4
3 2 3 4
4 2 4
5 2 4
6 1 2 4
7 2 3 4
8 1 2 4
13:22 LK2, 3 and 4 on all p+n - attachment 1
14:20 AC mains relay connected to AC voltage stabilisers + on separate mains socket from DESPEC - attachment 2
Same configuration but waveforms disabled (p+n improvment) - attachment 3
Layout 1 - attachment 4
14:32 all FEE power cables moved to PSU 2 (Lower one in rack, upper one was slow to start for 1 FEE pair, likely the 5V slow to start)
Previously all FEEs on their own pair within a psu
New order 1,3 2,4 5,7 6,8
Waveforms on - all bad - attachment 5
Waveforms off - all p+n better - attachment 6
14:54 interlock removed from AC mains relay
Also noticed adaptor board on 2 not in fully -> Now in
p+n generally better - attachment 7
Without waveforms - p+n better again - attachment 8
16:04 Platform moved into position
Interlock power cable removed from extension lead
w/ waveform - p+n slightly worse
w/o waveform - p+n same as beffore
16:14 LK3 removed from 1 and 5
w/ waveform -> p+n slightly worse - attachment 9
w/o waveform -> p+n best yet - attachment 10
16:33 LK1 wasn't on all n+n was only on 2, 6 and 8
Removed from 6 -> only one LK1 per DSSD
w/ waveforms - all bad - attachment 11
w/o waveforms - no change - attachment 12
16:59 LK1 removed from 2->4 This is the FEE with the BIAS braid
w/ waveform p+n better - attachment 13
w/0 waveform p+n better, n+n worse? - attachment 14
17:21 Bias filters added to n+n
w/ waveform - no change - attachment 15
w/o wwaveofrm - P=n worse - attachment 16
Stats 0x21 - attachment 17
Stats 0x14 - attachment 18
Stats 0x12 - attachment 19
17:42 Bias filter removed and bias floated (LK1 removed)
w/ waveform all worse - attachment 20
w/o waveform -> p+n worse, n+n similar
18:29 LK1 back on 4+8
w/ waveform -> poor
w/0 waveform - recovered previous - attachment 20
18:32 LK 2 and 4 removed from 2 +6
Also noticed ground lemo on aida02 not fully inserted which has been corrected
w/ waveform poor - attachment 21
Threshold 0xF - attachment 22
w/o waveform - attachment 23 |
|