Entry  Tue May 4 14:05:48 2021, NH, MACB Cables macb_front.jpgmacb_side.jpg
During the installation of the 4 new HDMI cables for aidas13-16 the HDMI cables between the MACB Root and the MACB leaves were looped through the cable
guide at the top to support them more
This may improve their connection and reduce the chance of WR errors seen in S460/S452
Entry  Tue May 4 13:53:22 2021, OH, TD, Tuesday 4 May 7x

NH has re-checked alignment of AIDA adaptor PCBs cf. Elog:290
Entry  Tue May 4 12:11:39 2021, OH TD, Bias test of triple 210504_1325_Rate.png210504_1326_Layout7.png210504_1327_Layout08.png210504_1334_Stats.pngVICurve.png
13:11 System wide checks all ok except *aida02 fails ADC calibration*
 Collecting the file size of each FEE64 Options CONTENTS file to check they are all the same
Entry  Sat May 1 08:45:19 2021, TD, AIDA 24cm x 8cm DSSSD + triPlast assembly for S496 12x

Photos provided by Nic Hubbard and Helena Albers from the assembly of 2x triPlast
plus 2x MSL type BB18(DS)-1000 24cm x 8cm detector assembly - attachments 1-3.
Entry  Fri Apr 30 08:38:12 2021, TD, Friday 30 April 6x
09.40 All system wide checks OK *except*

FEE64 module aida07 failed
Entry  Thu Apr 29 16:56:42 2021, OH, Config changes for 16FEE 
New set of ASIC settings produced 2021Apr29-13-16-00
- Note no longer odd/even for polarity of signal
Entry  Wed Apr 28 17:39:35 2021, TD, Preparations for setup of aida13-aida16 dhcpd.confdhcpd.conf.BAK-280421startup.tclstartup.tcl.BAK-280421

Updated /etc/dhcpd.conf
Entry  Wed Apr 28 10:06:11 2021, NH, Wednesday 28 April elog_temps.pngelog_gswr.pngelog_tse.pngelog_hv.pngAIDA_FEE.png.jpeg
11.05 All system wide checks OK *except*
WR
Entry  Tue Apr 27 08:26:01 2021, TD, Tuesday 27 April - aida09 system console reports 'received undefined information code: 1' and 'code: 9' 6x

09.25 All system wide checks OK *except*
Entry  Mon Apr 26 09:41:21 2021, TD, Monday 26 April 302.pngScreenshot_2021-04-26_Temperature_and_status_scan_aidas-gsi.pngScreenshot_2021-04-26_Statistics_aidas-gsi.pngScreenshot_2021-04-26_NewMerger_Statistics_aidas-gsi.pngScreenshot_2021-04-26_AIDA_Alerting_-_Grafana.png
10.42 All system wide checks OK *except*

 Base  Current  Difference
Entry  Sun Apr 25 10:22:16 2021, TD, Sunday 25 April 9x
11.25 All system wide checks OK *except*

 Base  Current  Difference
Entry  Sat Apr 24 15:48:38 2021, TD, aida12 crash - 08:33:03 Monday 19 April  

from ttyUSB5 system console log
Entry  Sat Apr 24 15:28:17 2021, TD, aida05 crash - 11:08:07 Tuesday 20 April 

From ttyUSB7 system console log
Entry  Sat Apr 24 15:13:33 2021, TD, aida09 crash - 08:10:52 Tuesday 20 April 

From ttyUSB6 system console log
Entry  Sat Apr 24 13:33:44 2021, TD, aida01 restart - 22.07UTC+2 Tuesday 20 April 

From aida01 /var/log/messages
Entry  Sat Apr 24 13:25:23 2021, NH, TD, AIDA FEE64 crash log for S460 

https://docs.google.com/spreadsheets/d/1CyL5L-WLsh9fsS7CN989GC3ktqiPQOkG8YIDJrrGxYA/edit#gid=0
Entry  Sat Apr 24 10:21:06 2021, NH, Saturday 24.04.2021 AnyDesk_2021-04-24_11-27-49.pngAnyDesk_2021-04-24_11-28-50.png
I have restarted some more of the consoles that disappeared since the aidas-gsi issue last night
Grafana is working again
Entry  Sat Apr 24 05:17:51 2021, ML, End of beam 
Beam has now stopped (end of expriment) and background measurements starts.
Entry  Sat Apr 24 04:50:01 2021, ML, Spectra 8x
Some screenshot of spectra attached
Entry  Sat Apr 24 04:07:23 2021, ML, ucesb scalers 24-04-2021_5am_ucesb.png
ucesb is back - see attachment