Entry  Thu Jul 24 10:28:04 2025, CC, NK, MP, Comparison of channels in the same FEE64 and ASIC 05.1 and with AIDA13.1.8L 8x
BB7 is still connected but not biased (same configuration as the previous elog). The channel of the CAEN power supply #0 is connected to ASIC08 and biased
at 100V. The channel of the CAEN power supply #1 is disconnected but still biased at 100V. AIDA04 (chained to AIDA05) is biased via an unshielded
cable to an iseg NHQ 246L power supplied, biased at 100 V with 15 uA.
Entry  Wed Jul 23 13:08:30 2025, CC, NK, Comparison of channels in the same FEE64 and ASIC 05.1 Screenshot_from_2025-07-23_14-11-36.pngScreenshot_from_2025-07-23_14-13-46.pngScreenshot_from_2025-07-23_14-16-03.pngScreenshot_from_2025-07-23_14-18-27.pngScreenshot_from_2025-07-23_14-20-42.png
Biased at 100, AIDA05 connected, HT cables not shielded CAEN HT

Attachment 1: AIDA05.1.8L --> FWHM 75 channels

Attachment 2: AIDA05.1.1L --> FWHM 70 channels; peak of the pulser shifted to the right (red lines on the left: cursor for the 1.8L)
Entry  Thu Jul 24 15:50:01 2025, CC, MP, NK, Check of BB7 impact Screenshot_from_2025-07-24_16-49-27.pngScreenshot_from_2025-07-24_17-03-44.pngScreenshot_from_2025-07-24_17-22-49.pngScreenshot_from_2025-07-24_17-29-38.png
HT of upstream AIDA from iseg, HT cable shielded, 100V, 17 uA. HT of downstream AIDA from CAEN, AIDA#0. Both channel 0 and 1 of CAEN
are on, even though there is just one channel connected. DEGAS platform closed. The crates of BB7 and bplast have been put to the ground of the AIDA's
structure (cupper sheet)
Entry  Wed Mar 31 15:40:42 2021, PJCS, Check Options files function added 
There is a new operation available in the System Functions menu in the System Wide Checks page. ( The page needs to be reset once when the FEEs are powered
before this function will appear )

Check the Options files are all the same size
Entry  Mon Oct 27 11:55:00 2025, JB, MP, GB, NH, Changing ground setup from 25102025 9x
12:55 We are going to try the follow changes to the AIDA scheme of 241012025.

- Fitting LK1 (aida01, aida14), removing LK1 (aida03, aida07, aida11, aida15) - only 1x LK1 needs to be fitted per Si wafer
    Reply  Mon Oct 27 18:08:11 2025, TD, Changing ground setup from 25102025 
19.06 Found HV OFF, DAQ GOing

      DAQ STOP
Entry  Thu Mar 12 15:25:08 2020, Thursday, March 12 16:00 to midnight shift, CA, DK, LS 12x
16:30
    All system wide checks passed
    Temps shown in attachment #1 
Entry  Tue May 4 12:11:39 2021, OH TD, Bias test of triple 210504_1325_Rate.png210504_1326_Layout7.png210504_1327_Layout08.png210504_1334_Stats.pngVICurve.png
13:11 System wide checks all ok except *aida02 fails ADC calibration*
 Collecting the file size of each FEE64 Options CONTENTS file to check they are all the same
Entry  Tue Nov 4 14:57:19 2025, MP, Bi207 source data, 700keV Screenshot_from_2025-11-04_15-49-20.pngScreenshot_from_2025-11-04_15-33-11.pngScreenshot_from_2025-11-04_15-28-42.pngScreenshot_from_2025-11-04_15-28-35.png
- Bias and basic checks: the pulser FWHM appears to be ~170, which is the same we have obtained at the end of last week.

- Initialised the Merger and Tape server

- loading of settings: EXPERIMENTS/AIDA/2025Nov03-16.21.44
    Reply  Tue Nov 4 16:34:47 2025, MP, NH, Bi207 source data, 700keV 
Some FEES were not acquiring data for some reason we could not understand. We lowered the thresholds to 200 keV, then rose them again to 700 keV. It
seems they are now acquiring. We start data taking again.




Quote:


    Reply  Tue Nov 4 17:29:28 2025, MP, Bi207 source data, 700keV 
Anydesk has crashed. I am not able to set it up again. I switch off bias and fees




Quote:



Some FEES were not acquiring data for some reason we could not understand. We lowered the thresholds to 200 keV,
Entry  Fri Feb 12 17:31:04 2021, NH, Background run 
A background run has been started to check the AIDA channels.
It is in /TapeData/BGFEB21
Entry  Thu Dec 19 14:26:09 2019, TD, BNC PB-5 - Manual pb-5manual.pdf
 
Entry  Tue Nov 12 13:59:07 2019, NH, OH, CA, Awfully noisy waveforms 8x
Rates in all FEEs much higher than before - waveform shows very noisy in all systems. - attachment 1

Good event statistics - attachment 2
Entry  Fri Mar 22 07:34:54 2024, NH, JB, Au Beam 
 
Entry  Tue Apr 20 07:06:20 2021, CA, April 20th 08:00 - 12:00 Screenshot_from_2021-04-20_07-45-09.pngScreenshot_from_2021-04-20_07-45-43.pngScreenshot_from_2021-04-20_07-46-05.pngScreenshot_from_2021-04-20_08-14-58.pngScreenshot_from_2021-04-20_08-15-38.png
08:09 DAQ crashes at start of shift (!)

      again happens at same time as beam being lost
Entry  Tue Apr 20 10:54:41 2021, CB, April 20 12:00 - 16:00 shift 9x
11:55 CA got AIDA back online

12:00 Stats OK (attach 1)
Entry  Fri Mar 22 08:29:55 2024, TD, Anydesk restarted remotely 
Anydesk restarted remotely per https://elog.ph.ed.ac.uk/CARME/489

Anydesk address now restored to 832827869
Entry  Tue Nov 19 11:24:25 2024, TD, Anydesk restarted remotely 

Anydesk restarted remotely per https://elog.ph.ed.ac.uk/CARME/489
Entry  Thu Mar 4 10:13:48 2021, OH, AnyDesk connection instuction Connecting_to_AnyDesk_at_GSI_for_AIDA.pdf
 
ELOG V3.1.3-7933898