Fri Mar 22 13:32:55 2019, NH, AIDA Startup @ 22.03.2019 
|
FEEs have been on this morning, temperatures attached
Leakage current attached
the DSSD is attached to the following AIDA modules: |
Thu Mar 21 15:45:48 2019, NH, AS, SA, MP, Snout Mounting & DSSD Bias
|
The AIDA snout (with single DSSD and plastic) has been mounted in S4 cave again. DSSD was connected to the outer 4 FEEs again.
-160 V Bias applied and DSSD holds voltage with leakage current of around 1.3 uA (was dropping as only biased briefly)
Bias off and FEEs offline again now as day nearly done. MBS testing will be performed shortly, as well as resolution checks. |
Thu Mar 21 14:10:10 2019, NH, FEE Temperatures
|
All 12 FEEs powered up, and temperatures are OK.
Passed ASIC checks and system wide checks - now checking cabling and to mount snout. |
Thu Mar 21 13:59:08 2019, NH, AIDA Water
|
The AIDA water system has been repaired and not leaking.
The interlock did not light immediately, but power cable reconnected to box and now it works.
|
Wed Mar 20 15:35:10 2019, NH, AIDA Setup @ GSI
|
Currently
- 1 DSSD and Plastic is in the AIDA snout, which is not mounted
- The FEEs are offline due to a water supply issue, this will be repaired early 21/03/2019 |
Sun Feb 3 16:01:49 2019, TD, Updated analyses GSIJan19/R1_0 - R10_0 7x
|
Summary statistics for WR test data files R1_0, R2_0, R3_0, R4_0, R7_0 and R10_0 (attachments 1-6)
|
Fri Jan 25 12:33:03 2019, NH, aida-gsi Access
|
If you have a GSI account it should be possible to connect to aida-gsi remotely via the following
SSH into lx-pool.gsi.de with GSI Linux username/password (ssh username@lx-pool.gsi.de)
Then SSH into aida-gsi via ssh npg@aida-3 |
Fri Jan 25 12:26:17 2019, NH, Raspberry Pi Startup & Information
|
Two raspberry pis:
nnrpi1 - FEE64 system consoles
nnrpi2 - AC Mains Relay & CAEN HV |
Fri Jan 25 09:54:12 2019, CA, TD, NH, VP, 25th January 2019 8x
|
09.30: AIDA still running, not writing to file
Still running ok
No water leaks
|
Thu Jan 24 17:16:32 2019, CA, TD, NH, VP, MACB time switch settings 
|
AIDA@DESPEC MACB time switch settings before and after WR are as follows.
Attachment 1 - MACB 1-4 from left to right
|
Thu Jan 24 14:53:26 2019, TD, Analyses GSIJan19/R2_0 - R4_0   
|
Attachment 1 - summary analysis R2_0
|
Thu Jan 24 13:29:40 2019, TD, Analysis of GSIJan19/R1_0   
|
Analysis of file GSIJan19/R1_0 (white rabbit data)
|
Thu Jan 24 10:26:09 2019, VP, TD, CA, NH, Start up Merger
|
1) Click on item "Merger MK2" in top panel
This kills any existing Merger existence and loads a new copy.
|
Thu Jan 24 10:21:26 2019, CA, TD, NH, VP, January 24th 2019 7x
|
10.30 AIDA setup and DAQ start complete
writing to file TapeData/NULL/R1
|
Wed Jan 23 17:03:08 2019, CA, TD, NH, VP, January 23rd 2019 6x
|
Post White Rabbit update
Attachment 1 - 1.8.W pulser waveform spectra for all FEE64
|
Wed Jan 23 09:25:55 2019, CA, TD, NH, VP, MACB and setup pics 23.01.19 11x
|
Pictures of leaking cooling pipe
MACB switch settings
|
Wed Jan 23 08:12:01 2019, CA, TD, NH, VP, January 22nd 2019   
|
Attachment 1 - 1.8.W waveform spectra with pulser signal
Attachment 2 - 1.8.L pulser peak spectra
|
Mon Dec 10 11:09:53 2018, CA, TD, Scintillator pictures/system checks for 07.12.18 14x
|
11.12 System wide checks ok (attachment 1)
FEE temperatures ok (attachment 2)
bias/leakage currents ok (attachment 3)
|
Sat Dec 8 15:22:52 2018, TD, Raspberry Pi startup
|
nnrpi1 - FEE64 system consoles
nnrpi2 - USB-controlled ac mains relay, CAEN N1419ET
|
Thu Dec 6 20:45:47 2018, TD, Analysis of R2_0 
|
Analysis of data file R2_0
|