Wed Apr 21 11:27:55 2021, OH, Wed 21st April 12:00-16:00 6x
|
12:30 MBS has crashed
This time AIDA is still running fine.
|
Wed May 12 06:59:55 2021, OH, RDP, Wed 12 May 08:00-16:00   
|
08:00 While the rates appeared to have quietened around 7am they are back to their previous rate now.
They still seem to be working on the setting. Implant rate in AIDA about 8 per spill. Thick degrader is still in place
|
Wed May 12 02:07:59 2021, MA, BA, Wed 12 May 00:00-08:00 9x
|
03:12 System Check
attachment 1 : Current
attachment 2 : Temperature |
Wed May 12 18:53:38 2021, JS, Wed 12 20:00-00:00    
|
19:51 Shift change
Stats ok
|
Wed May 12 15:03:10 2021, CB, OH, Wed 12 16:00-20:00 15x
|
16:00 Shift change
Statistics OK (attach 1)
|
Tue Nov 24 13:10:42 2020, NH, Waveforms with frame grounded   
|
AIDA Frame is now grounded to platform (will add photo later)
Waveforms improved from April (c.f. https://elog.ph.ed.ac.uk/DESPEC/125) but comparison not truly fair (other systems may be offline now) |
Fri Nov 22 08:55:31 2019, NH, Waveforms   
|
| AIDA Waveforms immediately after beam at 21.11.19 8am or so. |
Mon Nov 25 13:25:51 2019, NH, Waveforms 
|
> AIDA Waveforms immediately after beam at 21.11.19 8am or so.
Update 25.11.19 with all FEE64 waveforms added
|
Thu Apr 18 13:59:55 2024, TD, JB, WR timestamp OK, readout time not OK    
|
WR timestamps appear to be OK but readout times fort aida03, aida04 and aida09 appear to be out of sequence. Attachment 1-3.
|
Thu Feb 20 17:06:42 2025, JB, CC, TD,MP, WR error for aida 13,14,15,16
|
| WR timestamp errors resolved after reseating the HDMI cables to the MACB |
Thu Oct 31 14:27:06 2019, NH, WR Timestamps
|
All 12 FEEs have valid WR Timestamps
Had to powercycle aida09 once as before raw readout was displaying upper 12 bits of WR timestamp as 0. Unsure of other method.
|
Thu Oct 31 19:02:42 2019, Patrick, WR Timestamps
|
> All 12 FEEs have valid WR Timestamps
> Had to powercycle aida09 once as before raw readout was displaying upper 12 bits of WR timestamp as 0. Unsure of other method.
>
|
Sun Feb 3 16:01:49 2019, TD, Updated analyses GSIJan19/R1_0 - R10_0 7x
|
Summary statistics for WR test data files R1_0, R2_0, R3_0, R4_0, R7_0 and R10_0 (attachments 1-6)
|
Wed Oct 29 13:01:08 2025, NH, Update of AnyDesk
|
AnyDesk was updated to the latest version that supports CentOS 7 (EOL OS)
First the YUM repositories were migrated, as the original address was offline due to EOL
|
Tue Mar 26 18:56:03 2024, TD, USB-controlled ac mains relay interlock box - wiring
|
Sensor ( 4 pins )
|
Mon Mar 8 22:51:47 2021, OH, Tuesday 9th March 10x
|
18:00 ASIC settings 2019Dec19-16.19.51
DSSSD#1 slow comparator 0xa
DSSSD#2 slow comparator 0xa
|
Tue Mar 9 20:04:20 2021, TD, Tuesday 9 March 7x
|
21.04 DAQ continues OK - file R20_379
ASIC settings 2019Dec19-16.19.51
|
Tue Apr 9 08:37:47 2024, TD, Tuesday 9 April 16x
|
09.36 Cooling water temperature and flow OK
FEE64 power ON
|
Tue Mar 8 10:26:56 2022, TD NH, Tuesday 8 March 2022 40x
|
11.25 System wide checks OK except aida06 clock fail status 6 - attachments 1-13
|
Tue Jun 8 10:51:55 2021, TD, Tuesday 8 June 9x
|
11.52 DAQ continues file S496/R32_340
High data ratefrom c. 18.00 yesterday evening to c. 06.00 this morning
|