Tue Mar 12 16:16:49 2024, NH, TD, JB, HA, Summary of DSSSD Biasing 12.03
|
DSSD#1 undergoes a breakdown at 90V, two of the three wafers show this
- The adapter PCBs themselves have no breakdown at 100V, indicating the issue is internal to the snout
|
Mon Mar 11 15:19:44 2024, JB, NH, Priyanka, Michael Armstrong, Helena Albers, To Do: AIDA PCB tests
|
To do:
1) Disconnect ribbon cables from p+n junction FEE64 adaptor PCBs of upstream DSSSD, apply c. 100V bias and check leakage current is zero
i.e. eliminate shorts in PCBs |
Sun Mar 10 17:08:12 2024, NH, AIDA FEE Layout + Cabling Plan for S100   
|
Proposed FEE numbering and wiring plan for upcoming experiment S100 (2x Wide DSSSDs)
Image designed in draw.io, source attached
|
Fri Mar 8 16:05:57 2024, JB, CC, TD, NH, Friday 8 March  
|
Bias tests of AIDA on individual wafers and in parallel. Spreadsheet can be found in attachment 1, and graphic results can be seen in Attachment 2. Summary:
Upstream detector cannot be biased in parallel nor individually. Downstream detector can be biased with positively and each wafer can individually be biased
negatively. In parallel, it was not possible to bias the downstream detector negatively as indicated by the tests below. |
Thu Mar 7 15:15:29 2024, JB, NH, CC, AIDA HV Bias Test IV Curve
|
AIDA I-V Test
Voltage (V)
Current
(uA)
10
1.43
20
2.065
30
2.415
40
2.64
50
2.825
60
2.99
70
3.185
80
3.58
90
5.01
100
9.4
-
Voltage-Current test of newly installed AIDA snout. Breakdown observed at around ~92 V. Probably caused my light leakage into the snout, will investigate |
Thu Mar 7 11:09:05 2024, HA, JB, CC, TD, MG, Thursday 7 March
|
Snout assembly
|
Tue Mar 5 09:10:01 2024, MA, JB, TD, Tuesday 5 March  
|
Old ribbon cable assemblies - attachments 1 2
|
Mon Mar 4 11:35:30 2024, TD, Monday 4 March 11x
|
12.30 FEE64 41:d7:cd ASIC mezzanine u/s, no data, ASIC temperature c. 20 deg C ( ambient ) low
Replaced ASIC mezzanine with new ASIC mezzanine
Installed as aida08
|
Sun Mar 3 16:06:11 2024, TD, Sunday 3 March 24x
|
17.00 FEE64 41:a0:71 ASIC #2 u/s
Replaced ASIC mezzanine with new ASIC mezzanine
|
Mon Feb 26 13:17:51 2024, TD, Offline analysis of DEC23/R9_6 9x
|
First pass analysis of data file /TapeData/DEC23/R9_6
|
Wed Jan 31 15:27:07 2024, TD, Photos of water flow & dew point sensor interlock box   
|
Attachment 1 - 4 socket input connectors ( from sensor to interlock box )
Attachment 2 & 2 - 24V dc relays and internal wiring
|
Mon Nov 13 11:46:54 2023, TD, Summary of AIDA 24cm x 8cm 'triple' DSSSD tests 
|
|
Thu Oct 12 14:01:43 2023, TD, MSL type BB18 24cm x 8cm DSSSD test - update 
|
DSSSD MSL type BB18 24cm x 8cm 3208-3/3208-21/3208-22
|
Sun Oct 1 11:47:07 2023, TD, S4 cooling water 
|
The photographs show the cooling water controls and temperature/pressure gauges outside S4 and the connections used by AIDA within S4. |
Thu Aug 31 15:24:56 2023, NH, New AIDA MBS PC
|
The AIDA MBS FDR will be x86l-119 from now on, not x86l-94
the MBS relay and startup scripts will be changed for this |
Mon Aug 28 12:47:56 2023, NH, Power Failure 24.08.2023
|
There was a power failure in the morning of 24.08.2023 in the Rhein-Main area affecting GSI
The Aida workstation (aida-3) has been restarted, it is unknown if the Pis in S4 rebooted as well (there is a UPS)
|
Sun Jul 16 09:32:57 2023, TD, Firefox browser proxy setting change
|
Firefox browser proxy setting changed to 'Auto-detect proxy settings for this network' ( Firefox -> Edit -> Settings -> Network Settings -> Settings ) |
Wed Jan 18 13:40:33 2023, PJCS TD, MACB settings with either Emulator or VITAR   
|
When using the VETAR connected to the HDMI port of the root MACB the settings for all the MACB in the system 0x3.
This setting takes the Clock and Data line from the input HDMI and outputs it via teh HDMI output ports.
When using the Emulator connected via the SMA connectors on the back of the root MACB then the setting for the root MACB |
Fri Dec 16 14:02:12 2022, NH, AIDA System off for christmas break
|
The AIDA NIM crate, pis and workstation have been powered off for the Christmas break and will not be accessible |
Thu Oct 6 16:51:49 2022, NH, Oscilloscope analysis 24x
|
Investigating AIDA noise with a TA041 differential probe and oscilloscope
AC Mains (DESPEC platform AC, L-N)
|