AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC, Page 9 of 33  ELOG logo
Entry  Tue Mar 12 16:16:49 2024, NH, TD, JB, HA, Summary of DSSSD Biasing 12.03 
DSSD#1 undergoes a breakdown at 90V, two of the three wafers show this
- The adapter PCBs themselves have no breakdown at 100V, indicating the   issue is internal to the snout
Entry  Mon Mar 11 15:19:44 2024, JB, NH, Priyanka, Michael Armstrong, Helena Albers, To Do: AIDA PCB tests 
To do:

1) Disconnect ribbon cables from p+n junction FEE64 adaptor PCBs of upstream DSSSD, apply c. 100V bias and check leakage current is zero
i.e. eliminate shorts in PCBs
Entry  Sun Mar 10 17:08:12 2024, NH, AIDA FEE Layout + Cabling Plan for S100 AIDA_S100_WiringPlan.drawio.pngAIDA_S100_WiringPlan.drawioAIDA_S100_WiringPlan(1).drawioAIDA_S100_WiringPlan.drawio.png
Proposed FEE numbering and wiring plan for upcoming experiment S100 (2x Wide DSSSDs)

Image designed in draw.io, source attached 
Entry  Fri Mar 8 16:05:57 2024, JB, CC, TD, NH, Friday 8 March IV_test_AIDA.xlsxchart.pngDownstream_positive_bias_vs._Current_(uA).png
Bias tests of AIDA on individual wafers and in parallel. Spreadsheet can be found in attachment 1, and graphic results can be seen in Attachment 2. Summary:
Upstream detector cannot be biased in parallel nor individually. Downstream detector can be biased with positively and each wafer can individually be biased
negatively. In parallel, it was not possible to bias the downstream detector negatively as indicated by the tests below.
Entry  Thu Mar 7 15:15:29 2024, JB, NH, CC, AIDA HV Bias Test IV Curve 

AIDA I-V Test


Voltage (V)
Current
(uA)


10
1.43


20
2.065


30
2.415


40
2.64


50
2.825


60
2.99


70
3.185


80
3.58


90
5.01


100
9.4




-
Voltage-Current test of newly installed AIDA snout. Breakdown observed at around ~92 V. Probably caused my light leakage into the snout, will investigate
Entry  Thu Mar 7 11:09:05 2024, HA, JB, CC, TD, MG, Thursday 7 March 

Snout assembly
Entry  Tue Mar 5 09:10:01 2024, MA, JB, TD, Tuesday 5 March 20240305_101140.jpg20240305_101146.jpg20240305_165949.jpg

Old ribbon cable assemblies - attachments 1  2
Entry  Mon Mar 4 11:35:30 2024, TD, Monday 4 March 11x
12.30 FEE64 41:d7:cd ASIC mezzanine u/s, no data, ASIC temperature c. 20 deg C ( ambient ) low
      Replaced ASIC mezzanine with new ASIC mezzanine
      Installed as aida08
Entry  Sun Mar 3 16:06:11 2024, TD, Sunday 3 March 24x

17.00 FEE64 41:a0:71 ASIC #2 u/s
      Replaced ASIC mezzanine with new ASIC mezzanine
Entry  Mon Feb 26 13:17:51 2024, TD, Offline analysis of DEC23/R9_6 9x

First pass analysis of data file /TapeData/DEC23/R9_6
Entry  Wed Jan 31 15:27:07 2024, TD, Photos of water flow & dew point sensor interlock box 20240131_160205.jpg20240131_145249.jpg20240131_145340.jpg20240131_160341.jpg
Attachment 1 - 4 socket input connectors ( from sensor to interlock box )

Attachment 2 & 2 - 24V dc relays and internal wiring
Entry  Mon Nov 13 11:46:54 2023, TD, Summary of AIDA 24cm x 8cm 'triple' DSSSD tests Summary_of_AIDA_tests_2022-3_-_Dec_2023.pdfSummary_of_AIDA_tests_2022-3_-_Dec_2023.pptx
 
Entry  Thu Oct 12 14:01:43 2023, TD, MSL type BB18 24cm x 8cm DSSSD test - update Screenshot_from_2023-10-12_15-13-30.pngScreenshot_from_2023-10-12_15-17-05.png

DSSSD MSL type BB18 24cm x 8cm 3208-3/3208-21/3208-22
Entry  Sun Oct 1 11:47:07 2023, TD, S4 cooling water  1000007356.jpg1000007355.jpg

The photographs show the cooling water controls and temperature/pressure gauges outside S4 and the connections used by AIDA within S4.
Entry  Thu Aug 31 15:24:56 2023, NH, New AIDA MBS PC 
The AIDA MBS FDR will be x86l-119 from now on, not x86l-94

the MBS relay and startup scripts will be changed for this
Entry  Mon Aug 28 12:47:56 2023, NH, Power Failure 24.08.2023 
There was a power failure in the morning of 24.08.2023 in the Rhein-Main area affecting GSI

The Aida workstation (aida-3) has been restarted, it is unknown if the Pis in S4 rebooted as well (there is a UPS)
Entry  Sun Jul 16 09:32:57 2023, TD, Firefox browser proxy setting change 

Firefox browser proxy setting changed to 'Auto-detect proxy settings for this network' ( Firefox -> Edit -> Settings -> Network Settings -> Settings )
Entry  Wed Jan 18 13:40:33 2023, PJCS TD, MACB settings with either Emulator or VITAR macb_apr20.jedmacb_apr20.vhdzybo.jpgMACB.jpg
When using the VETAR connected to the HDMI port of the root MACB the settings for all the MACB in the system 0x3.

This setting takes the Clock and Data line from the input HDMI and outputs it via teh HDMI output ports.

When using the Emulator connected via the SMA connectors on the back of the root MACB then the setting for the root MACB
Entry  Fri Dec 16 14:02:12 2022, NH, AIDA System off for christmas break 
The AIDA NIM crate, pis and workstation have been powered off for the Christmas break and will not be accessible
Entry  Thu Oct 6 16:51:49 2022, NH, Oscilloscope analysis 24x
Investigating AIDA noise with a TA041 differential probe and oscilloscope 

AC Mains (DESPEC platform AC, L-N)
ELOG V3.1.4-unknown