ID |
Date |
Author |
Subject |
55
|
Mon May 27 07:18:26 2019 |
A. Algora | AIDA Plastik |
Nishimura san found the module used for the AIDA plastik in the last BRIKEN run. We put it instead of the TFA. It was kept in the same
setting conditions as in the earlier experiment.
AIDA plastic module configuration (Ortec 671 Amplifier)
Coarse gain: 20
Fine gain: ?
Shaping time (us): 0.5
Type: Gaus
Signal pos
The configuration of the digitizer might need to be updated (5) |
56
|
Mon May 27 07:34:58 2019 |
A. Algora | HV status |
EMPTY00 │ 0.0 V 1500.0 uA 0.5 V 0.5 uA Off │00.0000
DET01 │ 599.0 V 1500.0 uA 598.5 V 90.0 uA On │00.0001
DET02 │ 548.5 V 1500.0 uA 548.0 V 82.5 uA On │00.0002
DET03 │ 530.5 V 1500.0 uA 530.0 V 80.0 uA On │00.0003
DET04 │ 697.5 V 1500.0 uA 697.5 V 105.0 uA On Admi4
DET05 │ 451.5 V 1500.0 uA 451.0 V 67.5 uA On 05
DET06 │ 820.0 V 1500.0 uA 820.5 V 123.0 uA On Status Ch06
DET07 │ 560.0──────────────560.0 V 84.0 uA On ────────────────┬─0.0007
DET08 │ 486.0 V 1500.0 uA 486.0 V 73.0 uA On │00.0008
EMPTY09 │ 0.0 V 1500.0 uA 1.0 V 0.5 uA Off │00.0009
EMPTY10 │ 0.0 V 1500.0 uA 0.5 V 0.0 uA Off │00.0010
EMPTY11 │ 0.0 V 1500.0 uA 1.0 V 0.5 uA Off │00.0011
EMPTY12 │ 0.0 V 1500.0 uA 0.0 V 0.0 uA Off │00.0000
DET11 │ 579.0 V 300.0 uA 579.5 V 86.5 uA On │00.0001
DET12 │ 506.0 V 300.0 uA 506.5 V 76.0 uA On │00.0002
DET13 │ 623.5 V 300.0 uA 623.5 V 92.5 uA On │00.0003
DET14 │ 649.5 V 300.0 uA 649.5 V 96.5 uA On │00.0004
DET15 │ 656.5 V 300.0 uA 656.5 V 97.5 uA On │00.0005
DET16 │ 530.5 V 300.0 uA 530.0 V 78.5 uA On │00.0006
EMPTY12 │ 0.0 V 1500.0 uA 0.0 V 0.0 uA Off │02.0000
DET11 │ 578.0 V 300.0 uA 578.0 V 86.5 uA On │02.0001
DET12 │505.5 V 300.0 uA 506.0 V 76.0 uA On │02.0002
DET13 │ 624.5 V 300.0 uA 624.5 V 92.5 uA On │02.0003
DET14 │ 650.0 V 300.0 uA 650.0 V 96.0 uA On │02.0004
DET15 │ 657.0 V 300.0 uA 657.0 V 98.0 uA On │02.0
DET16 │ 530.5 V 300.0 uA 530.0 V 78.5 uA On │00.0006
DET17 │ 462.0 V 300.0 uA 461.5 V 68.0 uA On │00.0007
DET18 │ 518.5 V 300.0 uA 518.5 V 77.5 uA On │00.0008
EMPTY19 │ 0.0 V 300.0 uA 0.0 V 0.0 uA Off │00.0009
EMPTY20 │ 0.0 V 300.0 uA 0.0 V 0.0 uA Off │00.0010
REF01 │ 800.0 V 300.0 uA 800.0 V 174.0 uA On │00.0011
EMPTY12 │ 0.0 V 1500.0 uA 0.0 V 0.0 uA Off │02.0000
|
57
|
Mon May 27 08:26:20 2019 |
A. Algora | Co source on top of Aida |
Start: 16:28
Stop: 17:03
Rates: 14.4 kHz (Dinodes), 12.8 kHz (Anodes)
Lmd: 60Co_source_AIDA_201905271627_196.dlt .... 60Co_source_AIDA_201905271627_209.dlt
Root: 201905271703_60Co_AIDA_in.root
Comments: AIDA in, sligthly less deeper than before. It is 17 mm less deeper. So from the other side we measure 17.7 mm to the nose of AIDa.
The source is on top of AIDA with a help of a plastic ruler, that it is lying on top of it. The other extreme of the ruler is attached with tape to the edge
of the frame on the beam side. The source is positioned on top of the detectors 3, and 4, in practical terms in the geometrical center, but on top of
AIDA. 60Co source from source box (key 3). No sources around, the sources are kept inside BELEN. |
Attachment 1: 201905271703_60Co_AIDA_in.root
|
58
|
Mon May 27 09:12:22 2019 |
A. Algora | Background, with AIDA |
Start: 17:13
Stop: 18:03
Rates: 1.8 kHz (Dynodes), 1.8 kHz Anodes
Lmd: Background_AIDA_201905271711_210.dlt ... Background_AIDA_201905271711_226.dlt
Root: 201905271803_background_AIDA_in.root
Comments: AIDA in, in the same conditions as the earlier entry. Some asymetry in the spectra is seen. See EDy05 vs EDy15 and similarly EDy03 vs EDy13 |
Attachment 1: Screenshot_from_2019-05-27_17-50-45.png
|
|
Attachment 2: Screenshot_from_2019-05-27_17-51-02.png
|
|
Attachment 3: 201905271803_background_AIDA_in.root
|
59
|
Mon May 27 12:54:48 2019 |
A. Algora | Reference detector |
The reference detector was very noisy. The threshold has been changed from 2000 to 3200. We should look at this more carefully.
The position of the fiber was also changed. It moved too close to the Cs peak. Probably it has to do with the flexibility of the plastik screw that fixes it.
|
Attachment 1: Screenshot_from_2019-05-27_20-54-21.png
|
|
60
|
Mon May 27 13:50:54 2019 |
A. Algora | preparation directory moved |
The preparation directory was moved to /data/DRikenPreparation (from its first place /home/dacq/DRikenPreparation) because the lack of space. |
61
|
Mon May 27 13:58:03 2019 |
A. Algora | Background, no lmd, AIDA in, no sources around |
Start: 21:59
Stop: 13:15 (next day, 28/05)
Rates: 1.9-2.3 (dynodes), 1.8-1.9 kHz (anodes)
Root: 201905281315_background_AIDA_in.root
Comments: Aida inside. No sources around |
Attachment 1: 201905281315_background_AIDA_in.root
|
62
|
Tue May 28 05:30:03 2019 |
A. Algora | Reference detector threshold changed |
Again the Reference detector was firing mainly in the noise. Threshold increased to 5000 !!!!
New config file was saved, rate of Ref detector now: aprox. 400 kHz |
Attachment 1: Screenshot_from_2019-05-28_13-29-46.png
|
|
63
|
Tue May 28 05:43:41 2019 |
A. Algora | 133Ba source. (IR052 from the third floor) |
Start: 13:43
Stop: 14:05
Comments: The source looks also too strong. Rates: 39 kHz (Dynodes), 40 kHz (anodes).
Source ID: IR052 from the third floor, small plastic rectangular form
Rates:
RefAn: 180 Hz
Dy01: 3.7 kHz
Dy02: 0.4 kHz
Dy11: 0.5 kHz
Dy12: 0.13 KHz
AIDA in. The source is put at the extreme of the source holder, at 16.5 cm from the edge of the detectors. Almost touching AIDA.
Lmd: 133Ba_source_AIDA_in_201905281343_227.dlt ..133Ba_source_AIDA_in_201905281343_230.dlt
Root: 201905281405_133Ba_AIDA_in.root |
64
|
Tue May 28 06:13:27 2019 |
A. Algora | Background, Aida in |
Start: 14:13
Stop: 15:47
Lmd: Background_AIDA_in_201905281413_231.dlt .. Background_AIDA_in_201905281413_233.dlt
Root: 201905281547_background_AIDA_in.root
Rates: 2.1 kHz (Dyn), 1.8 kHz (Anodes) |
65
|
Tue May 28 07:49:51 2019 |
A. Algora | Background, no lmd |
Start: 15:50
Stop: 10:09 (next day)
Root: 201905291009_background_AIDA_in.root
Comments: background, Aida in |
66
|
Wed May 29 02:07:20 2019 |
A. Algora | Irradiation for 24Na |
Irradiation of two pieces of aluminium of 80 mm x 40 mm x 10 mm will be performed at the second BigRips dipole.
The pieces were placed with the help of Fukuda and Nishimura san. |
67
|
Wed May 29 07:04:19 2019 |
A. Algora | F11 plastic connected |
F11 plastic has been connected to our dacq as it was in the earlier BRIKEN setting.
The detector cables were followed and connections checked.
The outputs of the detector (L and R) go through an Ortec TFA 474 with the following settings:
Coarse gain: x4
Fine gain: looks minimal
Integrate (ns): 200
Diff (ns): 100
Non-inv setting
After the TFA, the Right signal is plugged in channel 1 of our card, and an attenuated signal (50 Ohm) is
plugged in channel 3. Similarly the Left signal after the TFA is plugged in channel 2 and the attenuated
signal is plugged in channel 4.
Parameters of the channels of the dacq have to be checked.
|
68
|
Wed May 29 08:29:26 2019 |
A. Algora | latest conf file |
Latest configuration file from 28 of May |
Attachment 1: 20190528_Conf.xlsx
|
69
|
Wed May 29 09:08:11 2019 |
A. Algora | AIDA plastic setting changed |
AIDA plastic changed.
HV was at 1000 V, it was changed to 1300 V
Coarse gain was changed to x5, Fg remains as it was.
I had only two minutes to enter (it was a rush). Better to verify it again |
70
|
Wed May 29 09:49:25 2019 |
A. Algora | New configuration file |
New configuration file after a few small changes in thresholds. |
Attachment 1: 20190529_Conf.xlsx
|
71
|
Wed May 29 10:01:10 2019 |
A. Algora | Background, no lmd, AIDA in |
Start: 18:02
Stop: 18:33, partial file (201905291833_background_AIDA_in.root), Stop: 09:53 saved also next day (30/05) 201905300953_background_AIDA_in.root
Comments: AIDA in, Bigrips is doing settings (105Te to my best knowledge). There is a lead wall before our setting. No apparent change in the rate of background
Root: 201905300953_background_AIDA_in.root
|
Attachment 1: 201905291833_background_AIDA_in.root
|
Attachment 2: 201905300953_background_AIDA_in.root
|
72
|
Thu May 30 02:03:33 2019 |
A. Algora | Ref detector position changed, AIDA plastik HV changed |
The position of the reference detector in the ancilliary card was changed from channel 10 to channel 9. Remember that this is important
in relation to all files saved until now.
The HV of AIDA plastic was changed from 1300 V back to 1000 V.
A new configuration file provided by Alvaro and Jose, with some small changes will be loaded.
cp 20190529_Conf_ATJLT.xlsx 20190530_Conf.xlsx
The new configuration file has been uploaded. Reference detector appears in channel V2A8C9. It was necessary to change the range of the histogram, and to recalibrate it.
With the new configuration parameters the pulser appears at 1360 keV in the calibrated spectra of the reference detector.
The threshold of the Ref detector has also been changed. From 4400 to 2800. |
Attachment 1: 20190530_Conf.xlsx
|
73
|
Thu May 30 02:44:56 2019 |
A. Algora | Background, just a test |
Start: 11:01
Stop: 13:37
Root: 201905301337_background_AIDA_in.root
Comments: Some of the TAC spectra have desapeared with the new configuration file. To be explored.
I have tried with both configuration files that were sent. |
Attachment 1: 20190530_Conf.xlsx
|
74
|
Thu May 30 06:41:18 2019 |
A. Algora | Background with the beam |
There was an small increase in the counting rate of the detectors after the beam was delivered. The rate at
F7 was of the order of 5 kHz. But remember that there is a lead wall protecting the detector. Root file saved. Look at TSumA for example
Root: 201905301442_background_AIDA_in.root |