AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  AIDA, Page 29 of 46  ELOG logo
Entry  Wed Nov 9 22:16:11 2016, CG, DK, TD, ML, NP1412 R24 R25 34x
07.15 R24 started. BRIKEN run 33.
      76~78Ni beam setting.
      F5 C target out. F7 CH2 in. F11 C out.
Entry  Wed Nov 9 11:07:14 2016, TD, CG, DK, NP1412 R22 R23 6x
DAQ start (R22) failed - merger remained in 'want first sync' state, no output to TapeServer -> soft reset OK

start 20.07
Entry  Wed Nov 9 09:50:50 2016, TD, NP1412 R20, R21 1.png2.png3.png4.png

DAQ start (R20) failed - nnaida5 dropped data transfer #1 link
Entry  Tue Nov 8 13:37:06 2016, TD, OH, DK, NP1412 R2 6x
10:36 Run 2 Start
      Degraders (5.4mm Al) in at start and all removed once started.
      Other settings same as R1.
Entry  Wed Nov 9 08:47:03 2016, OH, TD, NP1412 R16, R17 and R18 r16bias1.pngr16bias2.png
start 17:44
stop 17:49
AIDA filename R16
Entry  Wed Nov 9 07:16:59 2016, OH, TD, NP1412 R15 6x
Run 15

start 16:14
Entry  Wed Nov 9 02:14:08 2016, OH, TD, NP1412 R14 bias1.pngbias2.pngR14bias1.pngR14bias2.png
11.13 Still waiting to get beam back got access so took temp measurement.
      T=25.6 DP=0.0 H=18.1
      Biases and leakages good. Attachments 1 and 2
Entry  Wed Nov 9 00:49:03 2016, TD, OH, NP1412 R13 r13bias1.pngr13bias2.png
09.50 Run 13 Start
      Thick C target (6mm) removed from F11. All variable Al degraders put in place
      AIDA settings file: 2016Nov07-23.03.06
Entry  Tue Nov 8 23:36:56 2016, DK, TD, NP1412 R12 10x
Takechi team changes the condition: F5 carbon; F7 empty; F11 carbon

  F11 Plastic is 1.1kHz

BRIKEN: 161109_0828_747576Ni_013.dlt
Entry  Tue Nov 8 23:16:15 2016, DK, TD, NP1412 R11 R11_USB1.pngR11_USB2.png
Recovered AIDA after power cycle (unknown cause of crash from the end of R10)

Corresponding run in BRIKEN: 161109_0814_747576Ni_012

 
Entry  Tue Nov 8 22:04:21 2016, DK, NP1412 R10 R10_USB1.pngR10_USB2.png
Put the variable Al degraders all back in for safety (was 4.4 mm, now 5.4 mm)

Takechi team switches the condition: F5 empty; F7 CH2; F11 empty

 
Entry  Wed Feb 1 16:54:40 2017, CG, NP1306 run summary Run_Summary_NP1306.xlsx
 
Entry  Mon May 23 09:17:55 2016, Patrick Coleman-Smith, Multiplicity Trigger Specification Specification_for_the_multiplicity_Trigger_for_AIDA.pdf
Please read and comment on the attached specification.
Entry  Mon Oct 9 14:29:12 2023, TD, Monday 9 October contd. Screenshot_from_2023-10-09_15-40-36.png

15.20 aida10 LK4 ( front f/p -> gnd ) fitted - investigate impact aida10 asic #4 edge channels
      DSSSSD HV bias +100V >+50uA
Entry  Mon Oct 9 09:03:10 2023, TD, Monday 9 October 50x
09.55 Water pressure and temp OK
      Power on
Entry  Mon May 8 02:11:37 2017, OH, TD, Monday 8 May 11x
10.11 DAQ continues OK 
      Current file May2017/R1_58
Entry  Mon Nov 7 02:13:05 2016, CG, OH, TD, DK, Monday 7 November 10x
11.10 DAQ found stopped. No storage mode so can't tell when, but presumably at 06.30 during LN2 fill cycle.
      Unable to regain control of (see errors after merger restart in attachment 1) system so power cycling FEEs.
Entry  Sun Jun 5 17:09:49 2016, CG, DK, AE, Monday 6th June 20x
00.45 Run R1100 started on n-rich setting (EURICA run_1100, BigRIPS run 1165)
            Degraders as noted in previous entry
           Check clock status failed for nnaida5. ADC calib failed for nnaida8. nnaida19 failed
Entry  Mon Sep 5 06:54:02 2016, TD, Monday 5 September 16x
14.20 R17 continues

      DSSSD bias/leakage current OK - see attachments 1-2
Entry  Mon Nov 5 01:40:10 2018, CA, CB, TD, OH, PJW, Monday 5 November 16x
10.40 Lowered slow comparator threshold from 100 to 10
      Good event statistics ok (attachment 1)
      System wide checks ok, except for nnaida7 ADC calibration failure
ELOG V3.1.3-7933898