Wed Nov 9 22:16:11 2016, CG, DK, TD, ML, NP1412 R24 R25 34x
|
07.15 R24 started. BRIKEN run 33.
76~78Ni beam setting.
F5 C target out. F7 CH2 in. F11 C out.
F11 plastic rate ~500cps.
1mm Al vairable degrader removed after ~3mins. Now have 4.4mm in.
07.26 R24 stopped for unknown reason - fill cycle had previously happened, similar scenario to yesterday.
Soft reset performed successfully.
07.36 R25 started.
No changes to beam settings above. F11 plastic rate ~500cps.
Offline implant rates:
*** DSSSD # 1 count: 894 old count: 682 dt: 1.02 s HEC rate: 207.66 Hz
*** DSSSD # 2 count: 781 old count: 590 dt: 1.02 s HEC rate: 187.09 Hz
*** DSSSD # 3 count: 900 old count: 682 dt: 1.02 s HEC rate: 213.54 Hz
*** DSSSD # 4 count: 790 old count: 594 dt: 1.02 s HEC rate: 191.99 Hz
*** DSSSD # 5 count: 421 old count: 317 dt: 1.02 s HEC rate: 101.87 Hz
*** DSSSD # 6 count: 162 old count: 128 dt: 1.02 s HEC rate: 33.30 Hz
08.13 Beam stopped.
corresponds to file R25_9.
08.15 Leakage currents attachments 1+2.
08.20 Takechi group changing settings: F5 C now in, F7 now empty, F11 C (6mm) now in.
Beam returned to AIDA.
Corresponding to 1.0Gb through file R25_10.
4.4mm of Al variable degrader currently in place. Will gradually remove.
08.23 All variable degrader removed.
Corresponds to start of file R25_11.
Rates from offline implant finder:
*** DSSSD # 1 count: 474 old count: 310 dt: 1.02 s HEC rate: 160.08 Hz
*** DSSSD # 2 count: 407 old count: 269 dt: 1.02 s HEC rate: 134.70 Hz
*** DSSSD # 3 count: 473 old count: 306 dt: 1.02 s HEC rate: 163.01 Hz
*** DSSSD # 4 count: 463 old count: 303 dt: 1.02 s HEC rate: 156.18 Hz
*** DSSSD # 5 count: 331 old count: 220 dt: 1.02 s HEC rate: 108.35 Hz
*** DSSSD # 6 count: 64 old count: 43 dt: 1.02 s HEC rate: 20.50 Hz
09.18 Beam stopped
Corresponds to file R25_23
09.20 Takechi et al.change settings - F5, F7, F11 out
Variable degrader 5.4mm Al in
First AIDA data file R25_24, BRIKEN 161110_0923_767778Ni_036
from file R25_26 (offline):
*** DSSSD # 1 count: 13306 old count: 13159 dt: 1.01 s HEC rate: 145.63 Hz
*** DSSSD # 2 count: 3578 old count: 3536 dt: 1.01 s HEC rate: 41.61 Hz
*** DSSSD # 3 count: 7699 old count: 7609 dt: 1.01 s HEC rate: 89.16 Hz
*** DSSSD # 4 count: 6191 old count: 6121 dt: 1.01 s HEC rate: 69.35 Hz
*** DSSSD # 5 count: 1891 old count: 1872 dt: 1.01 s HEC rate: 18.82 Hz
*** DSSSD # 6 count: 1377 old count: 1364 dt: 1.01 s HEC rate: 12.88 Hz
10.08 All system wide checks OK *except* nnaida8 & 16 fail clock status, nnaida23 fails ADC calibration
10.14 Beam stopped
Corresponds to file R25_36
10.19 Takechi et al.change settings - F5 in, F7 out, F11 in
Variable degrader all out
First AIDA data file R25_37, BRIKEN 161110_1016_767778Ni_37
10.26 Rate, Hitrate, 1.8.L and 1.8.W spectra - see attachments 3-12
All spectra cleared
10.30 Merge statistics - see attachments 13-17
10.32 Noted that nnaida19 histogram link has dropped - non-critical - continue DAQ
11.09 Change of target upstream: F5 empty, F7 CH2, F11 C 6mm
Corresponds to AIDA data file R25_48
11:14 Change completed
Corresponds to AIDA data file R25_49
11:21 All degraders out
Corresponds to AIDA data file R25_51
From AIDA R25_55:
*** DSSSD # 1 count: 241 old count: 0 dt: 1.02 s HEC rate: 235.75 Hz
*** DSSSD # 2 count: 210 old count: 0 dt: 1.02 s HEC rate: 205.42 Hz
*** DSSSD # 3 count: 242 old count: 0 dt: 1.02 s HEC rate: 236.72 Hz
*** DSSSD # 4 count: 209 old count: 0 dt: 1.02 s HEC rate: 204.44 Hz
*** DSSSD # 5 count: 118 old count: 0 dt: 1.02 s HEC rate: 115.43 Hz
*** DSSSD # 6 count: 46 old count: 0 dt: 1.02 s HEC rate: 45.00 Hz
11:40 lost beam
Current AIDA data R25_56
11:45 beam back
Current AIDA data R25_57
12.19 Takechi et al. change settings - F5 in, F7 out, F11 C 6mm
Corresponds to AIDA data file R25_65
All degraders 5.4mm Al in
12.24 All degraders out
Corresponds to AIDA data file R25_67
From offline analysis of R25_67
*** DSSSD # 1 count: 584 old count: 472 dt: 3.91 s HEC rate: 149.31 Hz
*** DSSSD # 2 count: 516 old count: 410 dt: 3.91 s HEC rate: 131.92 Hz
*** DSSSD # 3 count: 607 old count: 488 dt: 3.91 s HEC rate: 155.19 Hz
*** DSSSD # 4 count: 595 old count: 475 dt: 3.91 s HEC rate: 152.12 Hz
*** DSSSD # 5 count: 439 old count: 355 dt: 3.91 s HEC rate: 112.24 Hz
*** DSSSD # 6 count: 87 old count: 69 dt: 3.91 s HEC rate: 22.24 Hz
ADC & disc data synchronised - see attachments 17-18
13.38 R25_84 Bias and leakage currents ok - see attachments 19 and 20
13.57 FEE settings checked by saving current settings to file and comparing against the last saved settings. No change was found in the ASIC settings.
14.33 BRIKEN now on run 40 file name 161110_1419. DSSD Rates
*** DSSSD # 1 count: 138 old count: 0 dt: 1.02 s HEC rate: 135.54 Hz
*** DSSSD # 2 count: 119 old count: 0 dt: 1.02 s HEC rate: 116.88 Hz
*** DSSSD # 3 count: 139 old count: 0 dt: 1.02 s HEC rate: 136.52 Hz
*** DSSSD # 4 count: 134 old count: 0 dt: 1.02 s HEC rate: 131.61 Hz
*** DSSSD # 5 count: 101 old count: 0 dt: 1.02 s HEC rate: 99.20 Hz
*** DSSSD # 6 count: 20 old count: 0 dt: 1.02 s HEC rate: 19.64 Hz
15:14 BigRIPS now on Run 439 (CH2 Target in at F7)
AIDA file R25_104
15.17 BRIKEN now on run 41 file name 161110_1516
*** DSSSD # 1 count: 380 old count: 169 dt: 1.05 s HEC rate: 200.67 Hz
*** DSSSD # 2 count: 326 old count: 144 dt: 1.05 s HEC rate: 173.09 Hz
*** DSSSD # 3 count: 381 old count: 172 dt: 1.05 s HEC rate: 198.77 Hz
*** DSSSD # 4 count: 336 old count: 150 dt: 1.05 s HEC rate: 176.90 Hz
*** DSSSD # 5 count: 191 old count: 86 dt: 1.05 s HEC rate: 99.86 Hz
*** DSSSD # 6 count: 73 old count: 35 dt: 1.05 s HEC rate: 36.14 Hz
15.26 bias and leakage ok see attachment 21 and 22
16.04 BigRIPS Setting change now on 440 F5 In F7 Out F11 In
16.08 BRIKEN Now on run 42 File 161110_1608
16.09 Bias and leakage currents ok attachments 23 and 24
From offline r25_119
*** DSSSD # 1 count: 255 old count: 117 dt: 1.04 s HEC rate: 132.79 Hz
*** DSSSD # 2 count: 234 old count: 111 dt: 1.04 s HEC rate: 118.35 Hz
*** DSSSD # 3 count: 259 old count: 120 dt: 1.04 s HEC rate: 133.75 Hz
*** DSSSD # 4 count: 250 old count: 115 dt: 1.04 s HEC rate: 129.90 Hz
*** DSSSD # 5 count: 181 old count: 84 dt: 1.04 s HEC rate: 93.34 Hz
*** DSSSD # 6 count: 33 old count: 18 dt: 1.04 s HEC rate: 14.43 Hz
16.54 BRIKEN AND BigRIPS change runs. Corresponding to AIDA file R25_129
BigRIPS Now F5 In, F7 Out F11 Out
All variable Al degraders in 5.5mm
16.48 New BRIKEN run
Run number 043
File name: 161110_1658_767778Ni
17.02 Bias and leakage currents ok see attachment 25 and 26
DSSD Rates from offline file R25_131
*** DSSSD # 1 count: 269 old count: 127 dt: 1.04 s HEC rate: 136.32 Hz
*** DSSSD # 2 count: 224 old count: 107 dt: 1.04 s HEC rate: 112.32 Hz
*** DSSSD # 3 count: 257 old count: 118 dt: 1.04 s HEC rate: 133.44 Hz
*** DSSSD # 4 count: 160 old count: 71 dt: 1.04 s HEC rate: 85.44 Hz
*** DSSSD # 5 count: 29 old count: 13 dt: 1.04 s HEC rate: 15.36 Hz
*** DSSSD # 6 count: 3 old count: 2 dt: 1.04 s HEC rate: 0.96 Hz
17.48 BigRIPS change setting putting back the thick Carbon at F11 6mm. Corresponds to AIDA file R25_141
F5 In, F7 Out, F11 In C 6mm
All variable Al degraders removed.
BigRIPS run 442
17.52 BRIKEN starts run 44
File name 161110_1752767778Ni
AIDA file R25_142
bias and leakage currents ok attachments 27 and 28
DSSD rates offline from R25_142
*** DSSSD # 1 count: 275 old count: 136 dt: 1.03 s HEC rate: 134.63 Hz
*** DSSSD # 2 count: 240 old count: 114 dt: 1.03 s HEC rate: 122.04 Hz
*** DSSSD # 3 count: 275 old count: 134 dt: 1.03 s HEC rate: 136.57 Hz
*** DSSSD # 4 count: 266 old count: 130 dt: 1.03 s HEC rate: 131.73 Hz
*** DSSSD # 5 count: 180 old count: 86 dt: 1.03 s HEC rate: 91.05 Hz
*** DSSSD # 6 count: 46 old count: 24 dt: 1.03 s HEC rate: 21.31 Hz
18.16 System-wide checks OK *except* clock status (some/all of nnaida5,7,8,15&16 - see attachment 29) and ADC calibration (nnaida23)
18.24 DAQ stops OK
AIDA data filename R25_149
18.28 System wide checks show SYNC pulse and SYNC errors - see attachments 30-32
18.31 Stop & close merger & TapeServer
DAQ reset
DAQ setup - not complete - use Master Timestamp tab to ReSYNC
but note unusual(?) ReSYNC LSW value before second ReSYNC attempt- see attachments 33 & 34 before and after ReSYNC
|
Wed Nov 9 11:07:14 2016, TD, CG, DK, NP1412 R22 R23 6x
|
DAQ start (R22) failed - merger remained in 'want first sync' state, no output to TapeServer -> soft reset OK
start 20.07
stop
AIDA filename R23
BRIKEN filename 161109_12012_767778Ni_23
BigRIPS filename
ASIC settings 2016Nov07_23.03.06
DSSSD bias & leakage currents OK see attachments 1 & 2
Degrader 5.4mm Al from start of run
F11 rate 450Hz
BigRIPS setting 76-78Ni
N.B. there are a number of correlation resets at the start of R23
Offline analysis of R23_2
*** DSSSD # 1 count: 297 old count: 146 dt: 1.03 s HEC rate: 146.07 Hz
*** DSSSD # 2 count: 259 old count: 126 dt: 1.03 s HEC rate: 128.66 Hz
*** DSSSD # 3 count: 281 old count: 141 dt: 1.03 s HEC rate: 135.43 Hz
*** DSSSD # 4 count: 178 old count: 86 dt: 1.03 s HEC rate: 89.00 Hz
*** DSSSD # 5 count: 30 old count: 13 dt: 1.03 s HEC rate: 16.45 Hz
*** DSSSD # 6 count: 3 old count: 2 dt: 1.03 s HEC rate: 0.97 Hz
00.04 R23 continuing.
F5 C wedge in. F7 empty. F11 now empty. All variable degraders in place.
F11 plastic rate ~400-500Hz.
Change made around start of R23_70.
Offline implant rates shown below for R23_72:
*** DSSSD # 1 count: 562 old count: 414 dt: 1.06 s HEC rate: 139.98 Hz
*** DSSSD # 2 count: 507 old count: 372 dt: 1.06 s HEC rate: 127.68 Hz
*** DSSSD # 3 count: 552 old count: 404 dt: 1.06 s HEC rate: 139.98 Hz
*** DSSSD # 4 count: 350 old count: 262 dt: 1.06 s HEC rate: 83.23 Hz
*** DSSSD # 5 count: 46 old count: 37 dt: 1.06 s HEC rate: 8.51 Hz
*** DSSSD # 6 count: 4 old count: 3 dt: 1.06 s HEC rate: 0.95 Hz
00:34 (corresponds to the beginning of R23_77) Remove 1 mm of Al variable degrader at F11 (now total Al is
2.2+2.2 = 4.4 mm)
*** DSSSD # 1 count: 715 old count: 579 dt: 5.00 s HEC rate: 142.96 Hz
*** DSSSD # 2 count: 619 old count: 511 dt: 5.00 s HEC rate: 123.77 Hz
*** DSSSD # 3 count: 717 old count: 579 dt: 5.00 s HEC rate: 143.36 Hz
*** DSSSD # 4 count: 714 old count: 574 dt: 5.00 s HEC rate: 142.76 Hz
*** DSSSD # 5 count: 493 old count: 398 dt: 5.00 s HEC rate: 98.57 Hz
*** DSSSD # 6 count: 93 old count: 81 dt: 5.00 s HEC rate: 18.59 Hz
00:43 R23_80 biases are shown as attachments 3 and 4
01.08 F11 C target installed.
Corresponds to start of R23_90.
01.09 Final 4.4mm of variable Al degrader removed by start of R23_91.
Implant rates sensible.
02.12 F5 C target removed. F7 CH2 target put in. F11 C target remains in place.
Rate reduction during R23_115 as beam components changed.
implant rate >0 by start of R23_116 and BRIKEN run 30.
F11 plastic rate ~500Hz.
3:15 Takechi group changes the condition. This is beginning R23_132
F5 carbon, F7 empty.
F11 pllastic rate is 450 cps
F11 carbon remains in.
06.20 R23 stopped in anticipation of LN2 fill cycle.
Takechi group changed setting > F5 C out, F7 CH2 out, FF C in.
All variable degraders in.
06.30 Leakage currents attachments 5+6. |
Wed Nov 9 09:50:50 2016, TD, NP1412 R20, R21   
|
DAQ start (R20) failed - nnaida5 dropped data transfer #1 link
start 18.51
stop 19.30
AIDA filename R21
BRIKEN filename 161109_1858_767778Ni_21
BigRIPS filename 410
ASIC settings 2016Nov07_23.03.06
DSSSD bias & leakage currents OK see attachments 1 & 2
Degrader 1mm Al from start of run
F11 rate 450Hz
BigRIPS setting 76-78Ni
19.35 Offline analysis of R21_12
*** DSSSD # 1 count: 289 old count: 141 dt: 1.01 s HEC rate: 146.23 Hz
*** DSSSD # 2 count: 253 old count: 127 dt: 1.01 s HEC rate: 124.49 Hz
*** DSSSD # 3 count: 292 old count: 143 dt: 1.01 s HEC rate: 147.22 Hz
*** DSSSD # 4 count: 284 old count: 138 dt: 1.01 s HEC rate: 144.25 Hz
*** DSSSD # 5 count: 210 old count: 98 dt: 1.01 s HEC rate: 110.66 Hz
*** DSSSD # 6 count: 46 old count: 25 dt: 1.01 s HEC rate: 20.75 Hz
Scalers - see attachment 4
1 F11
2 Upstream plastic top
3 Upstream plastic bottom
4 Si top
5 Si bottom
6 AIDA plastic veto
7 F11 and (upstream plastic top or upstream plastic bottom)
8 F11 and AIDA plastic veto
19.42 Online analysis confirms ADC data synchronised |
Tue Nov 8 13:37:06 2016, TD, OH, DK, NP1412 R2 6x
|
10:36 Run 2 Start
Degraders (5.4mm Al) in at start and all removed once started.
Other settings same as R1.
22.45 DAQ stop OK
All variable degraders (5.4mm Al) in
From analysis of R2_3
S O R T C O M P L E T E ..... Tue Nov 8 22:59:02 2016
*** ENTRY finish
*** data items: 146156010 ( 541115.06 Hz)
*** ADC events: 7815683 ( 28936.09 Hz)
*** time warps: 0 ( 0.00 Hz)
*** DSSSD # 1 count: 51901 old count: 51732 dt: 84.55 s HEC rate: 613.87 Hz
*** DSSSD # 2 count: 42966 old count: 42831 dt: 84.55 s HEC rate: 508.19 Hz
*** DSSSD # 3 count: 37092 old count: 36983 dt: 84.55 s HEC rate: 438.71 Hz
*** DSSSD # 4 count: 12570 old count: 12532 dt: 84.55 s HEC rate: 148.67 Hz
*** DSSSD # 5 count: 1927 old count: 1921 dt: 84.55 s HEC rate: 22.79 Hz
*** DSSSD # 6 count: 659 old count: 658 dt: 84.55 s HEC rate: 7.79 Hz
*** ENTRY finish ends
S O R T S T O P P E D ..... Tue Nov 8 22:59:16 2016
s2000-s2005 m_p versus m_n DSSSD #1-6
s2010-s2015 x versus y DSSSD #1-6
s2020-s2025 E_p versus E_n DSSSD #1-6
23.05 DSSSD bias & leakage currents - see attachments 5 & 6
FEE64 temperatures - see attachment 7 |
Wed Nov 9 08:47:03 2016, OH, TD, NP1412 R16, R17 and R18 
|
start 17:44
stop 17:49
AIDA filename R16
BRIKEN filename 161105_1507
BigRIPS filename run405
ASIC settings 2016Nov07_23.03.06
DSSSD bias & leakage currents ok attachments 1 and 2
Degrader All variable in at start (5.5mm Al)
F11 rate 430Hz
BigRIPS setting 76-78Ni
start 17:52
stop 17.55
AIDA filename R17
BRIKEN filename
BigRIPS filename
ASIC settings 2016Nov07_23.03.06
Degrader All variable in at start (5.5mm Al)
F11 rate 430Hz
BigRIPS setting 76-78Ni
start 17:56
stop 18:02
AIDA filename R18
BRIKEN filename 161105_1507
BigRIPS filename run405
ASIC settings 2016Nov07_23.03.06
DSSSD bias & leakage currents ok
Degrader All variable in at start (5.5mm Al)
F11 rate 430Hz
BigRIPS setting 76-78Ni
start 18:04
stop 18.28
AIDA filename R19
BRIKEN filename
BigRIPS filename
ASIC settings 2016Nov07_23.03.06
DSSSD bias & leakage currents ok
Degrader All variable in at start (5.5mm Al)
18.07 2x2.2mmAl degrader removed now running with 1.1mm.
F11 rate 430Hz
BigRIPS setting 76-78Ni
*** DSSSD # 1 count: 474 old count: 224 dt: 1.03 s HEC rate: 242.26 Hz
*** DSSSD # 2 count: 414 old count: 194 dt: 1.03 s HEC rate: 213.18 Hz
*** DSSSD # 3 count: 480 old count: 225 dt: 1.03 s HEC rate: 247.10 Hz
*** DSSSD # 4 count: 414 old count: 196 dt: 1.03 s HEC rate: 211.25 Hz
*** DSSSD # 5 count: 232 old count: 111 dt: 1.03 s HEC rate: 117.25 Hz
*** DSSSD # 6 count: 95 old count: 42 dt: 1.03 s HEC rate: 51.36 Hz
18.28 DAQ stopped - presumably ORNL HPGe clover LN2 refill cycle commences |
Wed Nov 9 07:16:59 2016, OH, TD, NP1412 R15 6x
|
Run 15
start 16:14
stop
AIDA filename
BRIKEN filename 161109_1614
BigRIPS filename run406
ASIC settings 2016Nov07_23.03.06
DSSSD bias & leakage currents ok attachments 1 and 2
Degrader All variable in at start (5.5mm Al) at
F11 rate 430Hz
BigRIPS setting 76-78Ni
16:16 removed all variable degrader
Attachments 3, 4, 5 and 6 are plots for FEE 24 ASICs 1, 2, 3, 4
17.08 BRIKEN issues correlation scaler reset
17.32 Run stop |
Wed Nov 9 02:14:08 2016, OH, TD, NP1412 R14   
|
11.13 Still waiting to get beam back got access so took temp measurement.
T=25.6 DP=0.0 H=18.1
Biases and leakages good. Attachments 1 and 2
11.58 *All* system wide checks OK
14.29 *All* system wide checks OK
start 15:09
stop 16.04
AIDA filename R14
BRIKEN filename 161105_1507
BigRIPS filename run405
ASIC settings 2016Nov07_23.03.06
DSSSD bias & leakage currents ok attachments 3 and 4
Degrader All variable in at start (5.5mm Al)
F11 rate 430Hz
BigRIPS setting 76-78Ni
*** DSSSD # 1 count: 316 old count: 146 dt: 1.02 s HEC rate: 166.21 Hz
*** DSSSD # 2 count: 294 old count: 143 dt: 1.02 s HEC rate: 147.64 Hz
*** DSSSD # 3 count: 396 old count: 233 dt: 1.02 s HEC rate: 159.37 Hz
*** DSSSD # 4 count: 257 old count: 147 dt: 1.02 s HEC rate: 107.55 Hz
*** DSSSD # 5 count: 43 old count: 28 dt: 1.02 s HEC rate: 14.67 Hz
*** DSSSD # 6 count: 3 old count: 1 dt: 1.02 s HEC rate: 1.96 Hz
15.24 Briefly reduce Al degrader from 5.4mm Al to 3.2mm Al
Observe increased rate in DSSSDD #6, similar rates in DSSSD #1 and increased
rates in AIDA plastic veto
15.26 Return to 5.4mm Al degrader
16.06 *All* system wide checks OK |
Wed Nov 9 00:49:03 2016, TD, OH, NP1412 R13 
|
09.50 Run 13 Start
Thick C target (6mm) removed from F11. All variable Al degraders put in place
AIDA settings file: 2016Nov07-23.03.06
Keep alive script updated to current directory.
Output from Tom's Implants.f script
*** DSSSD # 1 count: 6316 old count: 6290 dt: 1.19 s HEC rate: 21.89 Hz
*** DSSSD # 2 count: 5435 old count: 5410 dt: 1.19 s HEC rate: 21.05 Hz
*** DSSSD # 3 count: 6314 old count: 6286 dt: 1.19 s HEC rate: 23.57 Hz
*** DSSSD # 4 count: 5211 old count: 5186 dt: 1.19 s HEC rate: 21.05 Hz
*** DSSSD # 5 count: 1383 old count: 1375 dt: 1.19 s HEC rate: 6.73 Hz
*** DSSSD # 6 count: 67 old count: 66 dt: 1.19 s HEC rate: 0.84 Hz
Measured rates from file:
DSSD1: 309
DSSD2: 270
DSSD3: 308
DSSD4: 249
DSSD5: 67
DSSD6: 3.3
10.03 biases and leakage currents attachments 1 and 2
10.26 Run stopped as Takechi group optimise upstream lowering our rate. |
Tue Nov 8 23:36:56 2016, DK, TD, NP1412 R12 10x
|
Takechi team changes the condition: F5 carbon; F7 empty; F11 carbon
F11 Plastic is 1.1kHz
BRIKEN: 161109_0828_747576Ni_013.dlt
AIDA settings file: 2016Nov07-23.03.06
Initially we have all the Al variable degraders in, but then after the run began and we saw nothing, we reduce it to 1 mm of Aluminium.
Biases shown as attachments 1 and 2
Offline data rates (from R12_2)
*** DSSSD # 1 count: 1022 old count: 709 dt: 2.91 s HEC rate: 351.69 Hz
*** DSSSD # 2 count: 863 old count: 597 dt: 2.91 s HEC rate: 296.98 Hz
*** DSSSD # 3 count: 995 old count: 693 dt: 2.91 s HEC rate: 342.40 Hz
*** DSSSD # 4 count: 827 old count: 575 dt: 2.91 s HEC rate: 284.59 Hz
*** DSSSD # 5 count: 246 old count: 164 dt: 2.91 s HEC rate: 84.65 Hz
*** DSSSD # 6 count: 11 old count: 10 dt: 2.91 s HEC rate: 3.79 Hz
FEE64 temperatures OK - see attachment 3
08.55 Online analysis shows ADC data synchronised - see attachment 4
09.05 Rate, HitRate, 1.8.L and 1.8.H energy spectra - see attachmemnts 5-10
Run stop 9:46
|
Tue Nov 8 23:16:15 2016, DK, TD, NP1412 R11 
|
Recovered AIDA after power cycle (unknown cause of crash from the end of R10)
Corresponding run in BRIKEN: 161109_0814_747576Ni_012
Run condition for Takechi has changed since our crash
F5 empty; F7 CH2; F11 carbon
At the beginning of our run, we had all Aluminium degraders in (1 mm was added back after our crash for safety)
We finally removed all but 1 mm Al on the variable degrader
SSD biases shown as attachments 1 and 2
8:20 They inform us that they want to take the F11 carbon out (in the middle of elog entry) |
Tue Nov 8 22:04:21 2016, DK, NP1412 R10 
|
Put the variable Al degraders all back in for safety (was 4.4 mm, now 5.4 mm)
Takechi team switches the condition: F5 empty; F7 CH2; F11 empty
7:00 Reset Run Control, run Setup, everything looks good. Start the run
Data rate around 10 MB/s
7:04 Checking the online data rate with MIDAS, the event rate is nearly zero...but it comes back (presume Takechi team was playing with the beam condition)
Events come back...
F11 Plastic rate is 1.2kHz
Here is a sample of the offline analysis of event rates with all Al degraders in:
*** DSSSD # 1 count: 640 old count: 321 dt: 1.03 s HEC rate: 309.51 Hz
*** DSSSD # 2 count: 551 old count: 277 dt: 1.03 s HEC rate: 265.85 Hz
*** DSSSD # 3 count: 773 old count: 452 dt: 1.03 s HEC rate: 311.45 Hz
*** DSSSD # 4 count: 633 old count: 371 dt: 1.03 s HEC rate: 254.20 Hz
*** DSSSD # 5 count: 211 old count: 140 dt: 1.03 s HEC rate: 68.89 Hz
*** DSSSD # 6 count: 6 old count: 4 dt: 1.03 s HEC rate: 1.94 Hz
7:11 Remove 1 mm of Al variable degrader (back to 4.4 mm total before AIDA)
Offline rates from R10_3:
*** DSSSD # 1 count: 2596 old count: 2447 dt: 6.52 s HEC rate: 398.43 Hz
*** DSSSD # 2 count: 2193 old count: 2067 dt: 6.52 s HEC rate: 336.58 Hz
*** DSSSD # 3 count: 2641 old count: 2485 dt: 6.52 s HEC rate: 405.34 Hz
*** DSSSD # 4 count: 2626 old count: 2470 dt: 6.52 s HEC rate: 403.04 Hz
*** DSSSD # 5 count: 1940 old count: 1825 dt: 6.52 s HEC rate: 297.75 Hz
*** DSSSD # 6 count: 852 old count: 801 dt: 6.52 s HEC rate: 130.76 Hz
7:18 Leak currents shown as attachments 1 and 2
7:20 Notice that the data rate has gone to zero...
Follow along with https://elog.ph.ed.ac.uk/AIDA/303 to fix it. This looks good and the merger seems to come back successfully.
Trying to STOP on Run Control gave:
STATE for nnaida16 returned with an error
error: SOAP http transport timed out after 20000 ms
NONE
error: SOAP http transport timed out after 20000 ms
while executing
"$transport $procVarName $url $req"
(procedure "::SOAP::invoke" line 18)
invoked from within
"::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState"
("eval" body line 1)
invoked from within
"eval ::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState $args"
(procedure "DataAcquisitionControlClient__GetState" line 1)
invoked from within
"DataAcquisitionControlClient__GetState"
Try a Reset on Run Control (this is the only option at present)
nnaida16 showed "going" but after Reset shows "reset" like all the rest.
Now we get another error
JavaScript error occurred!
Error description: SyntaxError: missing ) after argument list
Page address: http://localhost:8015/DataAcquisitionControl/DataAcquisitionControl.tml
Line number: 203
nnaida16 shows error
STATE for nnaida16 returned with an error
connect failed
NONE
connect failed
while executing
"::http::geturl http://nnaida16:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 20000 -query {<?xml version="1.0" encoding="UTF-8..."
("eval" body line 1)
invoked from within
"eval [list ::http::geturl $url] $args"
(procedure "::http::geturl_followRedirects" line 4)
invoked from within
"::http::geturl_followRedirects http://nnaida16:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 20000 -query {<?xml version="1.0"..."
("eval" body line 1)
invoked from within
"eval ::http::geturl_followRedirects [list $url] -headers [list $local_headers] -type text/xml -timeout $timeout -query [list $request] $local_pro..."
(procedure "::SOAP::Transport::http::xfer" line 61)
invoked from within
"$transport $procVarName $url $req"
(procedure "::SOAP::invoke" line 18)
invoked from within
"::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState"
("eval" body line 1)
invoked from within
"eval ::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetState $args"
(procedure "DataAcquisitionControlClient__GetState" line 1)
invoked from within
"DataAcquisitionControlClient__GetState"
GET OPTION for nnaida16 returned with an error
connect failed
NONE
connect failed
while executing
"::http::geturl http://nnaida16:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 60000 -query {<?xml version="1.0" encoding="UTF-8..."
("eval" body line 1)
invoked from within
"eval [list ::http::geturl $url] $args"
(procedure "::http::geturl_followRedirects" line 4)
invoked from within
"::http::geturl_followRedirects http://nnaida16:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 60000 -query {<?xml version="1.0"..."
("eval" body line 1)
invoked from within
"eval ::http::geturl_followRedirects [list $url] -headers [list $local_headers] -type text/xml -timeout $timeout -query [list $request] $local_pro..."
(procedure "::SOAP::Transport::http::xfer" line 61)
invoked from within
"$transport $procVarName $url $req"
(procedure "::SOAP::invoke" line 18)
invoked from within
"::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetOption HistEnable"
("eval" body line 1)
invoked from within
"eval ::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetOption $args"
(procedure "DataAcquisitionControlClient__GetOption" line 1)
invoked from within
"DataAcquisitionControlClient__GetOption "$AcqGetOption""
GET OPTION for nnaida16 returned with an error
connect failed
NONE
connect failed
while executing
"::http::geturl http://nnaida16:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 60000 -query {<?xml version="1.0" encoding="UTF-8..."
("eval" body line 1)
invoked from within
"eval [list ::http::geturl $url] $args"
(procedure "::http::geturl_followRedirects" line 4)
invoked from within
"::http::geturl_followRedirects http://nnaida16:8015/DataAcquisitionControlServer -headers {} -type text/xml -timeout 60000 -query {<?xml version="1.0"..."
("eval" body line 1)
invoked from within
"eval ::http::geturl_followRedirects [list $url] -headers [list $local_headers] -type text/xml -timeout $timeout -query [list $request] $local_pro..."
(procedure "::SOAP::Transport::http::xfer" line 61)
invoked from within
"$transport $procVarName $url $req"
(procedure "::SOAP::invoke" line 18)
invoked from within
"::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetOption Xfer2Enable"
("eval" body line 1)
invoked from within
"eval ::SOAP::invoke ::SOAP::_DataAcquisitionControlClient__GetOption $args"
(procedure "DataAcquisitionControlClient__GetOption" line 1)
invoked from within
"DataAcquisitionControlClient__GetOption "$AcqGetOption""
TapeServer was still running, so maybe that was a mistake...?
Power cycle... |
Wed Feb 1 16:54:40 2017, CG, NP1306 run summary
|
| |
Mon May 23 09:17:55 2016, Patrick Coleman-Smith, Multiplicity Trigger Specification
|
Please read and comment on the attached specification. |
Mon Oct 9 14:29:12 2023, TD, Monday 9 October contd.
|
15.20 aida10 LK4 ( front f/p -> gnd ) fitted - investigate impact aida10 asic #4 edge channels
DSSSSD HV bias +100V >+50uA
breakdown MGR & Front F/P?
wafer?
cabling?
aida10 LK4 removed
DSSSD HV +120V +6.760uA stable - attachment 1 |
Mon Oct 9 09:03:10 2023, TD, Monday 9 October 50x
|
09.55 Water pressure and temp OK
Power on
All system wide checks OK
DSSSD HV +120V +5.600uA - attachment 1
CAEN N1419ET floating output
standard coaxial cable to aida04, HV daisy chain aida12, aida03 & aida11
aida03 LK1 & LK3 fitted
test + & - daisy chains removed
FEE64 temps OK - attachment 2
slow comparator 0x64
ADC data item stats - attachment 3
per FEE64 Rate spectra - attachments 4-6
1.8.W spectra - 20us & 200us FSR - attachments 7-10
11.00 CAEN N1419ET floating -> gnd output ( via internal LK )
standard coaxial cable to aida04
aida03 LK1 & LK3 fitted
test + daisy chain removed, test - daisy chain aida02 - aida04
DSSSD HV +120V +5.955uA - attachment 11
slow comparator 0x64
ADC data item stats - attachment 12
per FEE64 Rate spectra - attachments 13 & 14
1.8.W spectra - 200us FSR - attachments 15 & 16
Overall, significant improvement in extrinsic noise for both p+n and n+n sides
slow comparator 0xf
ADC data item stats - attachment 17
per FEE64 Rate spectra - attachment 18
11.50 LK1 fitted to aida12, aida03 & aida11
DSSSD HV +120V +6.030uA - attachment 19
slow comparator 0xf
ADC data item stats - attachment 20
1.8.W spectra - 200us FSR - attachments 21 & 22
Incremental improvement in extrinsic noise for both p+n and n+n sides
12.20 LK1 fitted to aida12, aida03 & aida11, HV daisy chain aida12, aida03 & aida11 removed
DSSSD HV +120V +6.125uA - attachment 23
slow comparator 0xf
ADC data item stats - attachment 24
1.8.W spectra - 200us FSR - attachments 25 & 26
Little observable difference
12.40 test - daisy chain removed
slow comparator 0xf
ADC data item stats - attachment 29
per FEE64 Rate spectra - attachments 30 & 31
1.8.W spectra - 200us FSR - attachments 32 & 33
Further improvement in extrinsic noise observed for both p+n and n+n sides
14.00 additional ThorLab inline filter added to HV coaxial to aida04
DSSSD HV +120V +6.505uA - attachment 34
slow comparator 0xf
ADC data item stats - attachment 35
per FEE64 Rate spectra - attachments 36
1.8.W spectra - 200us FSR - attachments 37 & 38
Little observable difference
14.15 DSSSD HV +140V +6.685uA - attachment 39
1.8.W spectra - 200us FSR - attachment 40
DSSSD HV +160V +6.765uA - attachment 41
1.8.W spectra - 200us FSR - attachment 42
additional detector HV to dep+30V ( 160V ) mkaes little difference to n+n noise
14.45 per FEE64 4.*.W spectra - 200us FSR
n+n aida02 aida04 - attachments 43 & 44
p+n aida10, aida01, aida09 - attachments 45-47
p+n aida12, aida03, aida11 - attachments 48-50
aida09 missing strips? cable?
aida10 large transients for edge strips |
Mon May 8 02:11:37 2017, OH, TD, Monday 8 May 11x
|
10.11 DAQ continues OK
Current file May2017/R1_58
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator 0x64
LEC/MEC fast comparator 0xff
BNC PB-4 pulser OFF
Detector biases & leakage currents OK - see attachments 1 & 2
FEE64 Temperatures OK - see attachment 3
'good events' statistics OK - see attachment 4
No warnings/errors reported by merger since DAQ GO - see attachment 5
*all* system wide checks OK
10.19 1.8.L ADC spectra - see attachments 6 & 7
1.8.W waveforms - see attachments 8-11
10.35 E17/F11 ambient temperature +24.2 deg C, RH 33.2%, d.p. +7.0 deg C
Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 65%) |
Mon Nov 7 02:13:05 2016, CG, OH, TD, DK, Monday 7 November 10x
|
11.10 DAQ found stopped. No storage mode so can't tell when, but presumably at 06.30 during LN2 fill cycle.
Unable to regain control of (see errors after merger restart in attachment 1) system so power cycling FEEs.
11.15 Bias voltages @ ambient temp 25.5oC before DSSD noise investigation (attachment 2+3).
11.40 NB+DK attempting shielding of LN2 solenoid.
Manual fill initiated. DAQ still stopped at 11.41.
Unable to regain control of system. Power cycle FEEs while DK+NB install more shielding.
2:55 Made attempts to solve the noise issue that was plaguing DSSD1.
Attempts at checking the cabling, making sure the detectors were light sealed and switching the bias cables
around were unsuccessful.
A blue cable was found to be laying across the snout of AIDA that runs as a temperature probe for one of the
clover detectors.
Moving this cable appears to have solved this issue (Has been locked in position). See pulser spectrum in
attachments 4 and 5
23.15 DSSSD bias & leakage current Ok - see attachments 6-7
FEE64 temperatures OK - see attachment 8
00.05 DAQ start (R137)
Pulser Walkthrough
BNC PB_4 amplitudes 90,000-10,000 @ 10,000 step
x5 attenuator IN
t_r 50ns tau 50us
Frequency 250Hz range
Polarity -, tail pulse, delay MIN
ASIC settings 2016Nov07-23.03.06
00.21 DAQ start (R138)
Amplitude 14,000-2,000 @ 2,000 step
00.42 Background run R139 started. Pulser set to external trigger.
01.01 Run R139 closed successfully.
01.04 Started R140 on no storage mode to run overnight.
Leakage currents @ 01.06 shown in 9+10.
R140 still running ok @ 01.40. |
Sun Jun 5 17:09:49 2016, CG, DK, AE, Monday 6th June 20x
|
00.45 Run R1100 started on n-rich setting (EURICA run_1100, BigRIPS run 1165)
Degraders as noted in previous entry
Check clock status failed for nnaida5. ADC calib failed for nnaida8. nnaida19 failed Check ASIC clock and Check SYNC errors.
Temps ok. Leakage currents ok.
Settings file 2016Jun06-00.31.34
AIDA fluctuating but all > 0.
Tape rate ~8Mb/s.
Merge was outputting some interesting looking messages to terminal about queue being full and timestamps going backwards (attachment 3).
TapeServer window shows data being written as does repeated ls in /TapeData/NP1306. Writing one file every ~5-7min.
01.45 Merger went into 'Want next SYNC'
AIDA SYNC rate went to zero and TapeService stopped writing data.
Run stopped and FEEs power cycled as contact was lost with ~50%.
Beam operators want to do some tuning so control has been given to them while we start back up.
02.22 Started run R1101 on n-rich setting (EURICA run_1101-1102, BigRIPS run 1168-1169)
Increased fast disc threshold to 5 in all nnaida to prevent merger from running so fast and possibly choking.
Stopped @ 04.20
Realised after, nnaida24 had been on 1GeV range.
04.28 Started run R1102 on n-rich setting (EURICA run_1103, BigRIPS run 1170)
Stopped as 04.37 nnaida24 in 1GeV range.
04.39 Started run R1103 on n-rich setting (EURICA run_1104, BigRIPS run 1170)
Settings file 2016Jun06-04.39.22 -> same as previous settings but nnaida24 changed to 20MeV range.
Stopped at 04.59 at request of beam operators.
05.11 Started run R1104 on n-rich setting (EURICA run_1105, BigRIPS run 1171)
Merger died.
Stopped @ 05.15.
Ctrl-c out of all tape/merger related windows and restarted them. Setup/Configure etc.
05.34 Started run R1105 on n-rich setting ( EURICA run_1106, BigRIPS run 1172)
Settings file 2016Jun06-05.34.56 -> nnaida24 reverted back to 1GeV range. New settings file to make sure nnaida24 is set to 1GeV range. Changed in Save/Restore options.
Temps ok. Leakage currents ok.
Rates in visual scalar ~10% higher.
07.22 Lost contact with nnaida4 and merger stopped outputting data.
Stopped run R1105.
Power cycle FEEs and start new run.
On two power cycles, certain FEEs were not contactable.
Beam stopped by operators for brief period.
08.14 Run 1106 started on n-rich setting (EURICA run_1108, BigRIPS run 1175)
Settings file 2016Jun06-05.34.56
Temps ok. Leakage currents ok.
Stopped 09.06 for setting change.
09.41 Run 1107 started on reference mass setting (EURICA run_1200, BigRIPS run 1177)
Settings file 2016Jun06-05.34.56
11:24 Network link to the outside world is dropped across B3F
Beam is being controlled by the primary operators.
Run 1108 - background run with beam off from 10:29 to 11:31 (EURCA run 1201, no BigRIPS run)
(Was still editing the elog without submitting it. Sorry!)
Settings file was saved: 2016Jun06-10.45.31
Leak currents from paper logbook in consecutive order bias1 ch0-3, bias2 ch0-1 in uA: 3.225, 4.155, 3.515, 3.315, 3.65, 3.205
12.45 start R1110 (AIDA) run_1103 (EURICA)
Continuing reference mass measurement following earlier network outage B3F
ASIC settings 2016Jun06-05.34.56
see attachments 13-20
13.30 Stopped R1110 for B3F entry
'unable to connect' to nnaida1 etc
AIDA FEE64s reset OK but would not ReSync wuth about half of the FEE64s returning -1 status code
power cycle FEE64s, reboot aidas1 & FEE64s
14.03 ASIC seetings DB key 2016Jun06-14.03.48
14.20 System-wide checks
All FEE64s pass all tests *except*
nnaida21 ADC calibration
nnaida19 ASIC clock timestamp & SYNC errors not counting
|
Mon Sep 5 06:54:02 2016, TD, Monday 5 September 16x
|
14.20 R17 continues
DSSSD bias/leakage current OK - see attachments 1-2
FEE64 temperatures OK - see attachment 3
System-wide checks OK *except* nnaida19 fails ASIC timestamp check (no change)
No merger error messages
AIDA SYNC, PAUSE, RESUME, Correlation L, SYNC time warp, Good Events, AIDA disc, nnaida6, nnaida5 &
nnaida20 statistics - see attachments 4-14
Analysis of R17_584 - FEE64 data synaidas1> ./analyser2 /TapeData/Aug2016/R17_584
*** GREAT format 3.2.0 analyser - TD - May 2014
*** ERROR: READ I/O error: 5002
blocks: 32000
ADC data format: 258868072 ( 1733415.6 Hz)
Other data format: 2219928 ( 14864.9 Hz)
Sample trace data format: 0 ( 0.0 Hz)
Undefined format: 0 ( 0.0 Hz)
Other data format type: PAUSE: 94 ( 0.6 Hz)
RESUME: 94 ( 0.6 Hz)
SYNC100: 1365786 ( 9145.5 Hz)
FEE64 disc: 842248 ( 5639.8 Hz)
MBS info: 11706 ( 78.4 Hz)
Other info: 0 ( 0.0 Hz)
ADC data range bit set: 400 ( 2.7 Hz)
Timewarps: ADC: 0 ( 0.0 Hz)
PAUSE: 0 ( 0.0 Hz)
RESUME: 0 ( 0.0 Hz)
SYNC100: 0 ( 0.0 Hz)
FEE64 disc: 0 ( 0.0 Hz)
MBS info: 0 ( 0.0 Hz)
Undefined: 0 ( 0.0 Hz)
Sample trace: 0 ( 0.0 Hz)
Timestamp elapsed time: 149.340 s
FEE module #: 1 elapsed dead time 0.000 s
FEE module #: 2 elapsed dead time 0.026 s
FEE module #: 3 elapsed dead time 0.007 s
FEE module #: 4 elapsed dead time 0.000 s
FEE module #: 5 elapsed dead time 0.439 s
FEE module #: 6 elapsed dead time 0.329 s
FEE module #: 7 elapsed dead time 0.000 s
FEE module #: 8 elapsed dead time 0.000 s
FEE module #: 9 elapsed dead time 0.104 s
FEE module #: 10 elapsed dead time 0.000 s
FEE module #: 11 elapsed dead time 0.000 s
FEE module #: 12 elapsed dead time 0.001 s
FEE module #: 13 elapsed dead time 0.515 s
FEE module #: 14 elapsed dead time 0.290 s
FEE module #: 15 elapsed dead time 0.078 s
FEE module #: 16 elapsed dead time 0.036 s
FEE module #: 17 elapsed dead time 0.379 s
FEE module #: 18 elapsed dead time 0.044 s
FEE module #: 19 elapsed dead time 0.122 s
FEE module #: 20 elapsed dead time 0.063 s
FEE module #: 21 elapsed dead time 0.218 s
FEE module #: 22 elapsed dead time 0.467 s
FEE module #: 23 elapsed dead time 0.232 s
FEE module #: 24 elapsed dead time 0.059 s
FEE module #: 25 elapsed dead time 0.000 s
FEE module #: 26 elapsed dead time 0.000 s
FEE module #: 27 elapsed dead time 0.000 s
FEE module #: 28 elapsed dead time 0.000 s
FEE module #: 29 elapsed dead time 0.000 s
FEE module #: 30 elapsed dead time 0.000 s
FEE module #: 31 elapsed dead time 0.000 s
FEE module #: 32 elapsed dead time 0.000 s
*** Program elapsed time: 26.172s ( 1222.687 blocks/s, 76.418 Mb/s)
chronised - see attachment 16
aidas1> ./analyser2 /TapeData/Aug2016/R17_584
*** GREAT format 3.2.0 analyser - TD - May 2014
*** ERROR: READ I/O error: 5002
blocks: 32000
ADC data format: 258868072 ( 1733415.6 Hz)
Other data format: 2219928 ( 14864.9 Hz)
Sample trace data format: 0 ( 0.0 Hz)
Undefined format: 0 ( 0.0 Hz)
Other data format type: PAUSE: 94 ( 0.6 Hz)
RESUME: 94 ( 0.6 Hz)
SYNC100: 1365786 ( 9145.5 Hz)
FEE64 disc: 842248 ( 5639.8 Hz)
MBS info: 11706 ( 78.4 Hz)
Other info: 0 ( 0.0 Hz)
ADC data range bit set: 400 ( 2.7 Hz)
Timewarps: ADC: 0 ( 0.0 Hz)
PAUSE: 0 ( 0.0 Hz)
RESUME: 0 ( 0.0 Hz)
SYNC100: 0 ( 0.0 Hz)
FEE64 disc: 0 ( 0.0 Hz)
MBS info: 0 ( 0.0 Hz)
Undefined: 0 ( 0.0 Hz)
Sample trace: 0 ( 0.0 Hz)
Timestamp elapsed time: 149.340 s
FEE module #: 1 elapsed dead time 0.000 s
FEE module #: 2 elapsed dead time 0.026 s
FEE module #: 3 elapsed dead time 0.007 s
FEE module #: 4 elapsed dead time 0.000 s
FEE module #: 5 elapsed dead time 0.439 s
FEE module #: 6 elapsed dead time 0.329 s
FEE module #: 7 elapsed dead time 0.000 s
FEE module #: 8 elapsed dead time 0.000 s
FEE module #: 9 elapsed dead time 0.104 s
FEE module #: 10 elapsed dead time 0.000 s
FEE module #: 11 elapsed dead time 0.000 s
FEE module #: 12 elapsed dead time 0.001 s
FEE module #: 13 elapsed dead time 0.515 s
FEE module #: 14 elapsed dead time 0.290 s
FEE module #: 15 elapsed dead time 0.078 s
FEE module #: 16 elapsed dead time 0.036 s
FEE module #: 17 elapsed dead time 0.379 s
FEE module #: 18 elapsed dead time 0.044 s
FEE module #: 19 elapsed dead time 0.122 s
FEE module #: 20 elapsed dead time 0.063 s
FEE module #: 21 elapsed dead time 0.218 s
FEE module #: 22 elapsed dead time 0.467 s
FEE module #: 23 elapsed dead time 0.232 s
FEE module #: 24 elapsed dead time 0.059 s
FEE module #: 25 elapsed dead time 0.000 s
FEE module #: 26 elapsed dead time 0.000 s
FEE module #: 27 elapsed dead time 0.000 s
FEE module #: 28 elapsed dead time 0.000 s
FEE module #: 29 elapsed dead time 0.000 s
FEE module #: 30 elapsed dead time 0.000 s
FEE module #: 31 elapsed dead time 0.000 s
FEE module #: 32 elapsed dead time 0.000 s
*** Program elapsed time: 26.172s ( 1222.687 blocks/s, 76.418 Mb/s)
15.08 R17 stops
Detector bias OFF
FEE64 power OFF
USB-controlled ac mains relay OFF
Julabo FL11006 recirculating chiller OFF
ac mains cables to ac mains distributor disconnected
network cables aida1 and aida2 disconnected, rolled up and secured to stairwell rails
207Bi source removed and returned to B2F/F11 storage location |
Mon Nov 5 01:40:10 2018, CA, CB, TD, OH, PJW, Monday 5 November 16x
|
10.40 Lowered slow comparator threshold from 100 to 10
Good event statistics ok (attachment 1)
System wide checks ok, except for nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 2)
Bias/Leakage currents ok (attachment 3)
1.0.L spectra obtained, look ok (attachment 4)
DAQ stopped (R16)
10.49 Slow comparator threshold raised from 10 to 100
DAQ restarted (R17)
Merger and TapeServer ok, no error messages in terminals
Good event statistics ok (attachment 5)
nnaida2 is now at a good rate, after fixing kapton connection.
15:09 DAQ still running (Oct18/R17_0)
Merger and TapeServer ok
System wide checks ok, except for nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 6)
Bias/Leakage currents ok (attachment 7)
Good event statistics ok (attachment 8)
18.48 good event statistics OK *except* nnaida17 c. 400k
ASIC Control -> Check ASIC Control
*all* FEE64 good events statistics now OK
DAQ stop - current file Oct18/R17_20
18.50 DAQ start - file Oct18/R18
BNC PB-4 OFF
LEC/MEC slow comparator 0x64
alpha background run
18.54 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 9 & 10
FEE64 temperatures OK - see attachment 11
good event statistics OK - see attachment 12
Merger - Link16 errors reported - see attachment 13
TapeServer OK - see attachment 14
1.8.W spectra - attachment 15 & 16
|
|