Wed Oct 11 09:37:30 2023, TD, Wednesday 11 October 48x
|
10.00 Water flow & temp OK
Power ON
|
Tue Oct 10 08:38:37 2023, TD, Tuesday 10 October 40x
|
09.38 Water flow & temp OK
Power ON
|
Mon Oct 9 14:29:12 2023, TD, Monday 9 October contd.
|
15.20 aida10 LK4 ( front f/p -> gnd ) fitted - investigate impact aida10 asic #4 edge channels
DSSSSD HV bias +100V >+50uA
|
Mon Oct 9 09:03:10 2023, TD, Monday 9 October 50x
|
09.55 Water pressure and temp OK
Power on
|
Sun Oct 8 14:51:59 2023, TD, Sunday 8 October contd. 28x
|
Slow comparator 0xf
|
Sun Oct 8 11:44:35 2023, TD, Sunday 8 October 50x
|
12.22 USB-controlled ac mains relay - interlock overidden
Water flow OK. Water temp 19 deg C OK.
Power on.
|
Tue Oct 3 10:16:06 2023, TD, nnrpi2 scripts
|
pi@raspberrypi:~ $ cat ./caen_screen.sh
if ! screen -list | grep -q "caenhv"; then
screen -f -S caenhv /dev/ttyACM0 9600
|
Tue Oct 3 09:14:59 2023, TD, Tuesday 3 October 
|
09.45 Power up
|
Mon Oct 2 14:03:42 2023, TD, Monday 2 October  
|
15.00 DSSSD bias +60V +6.44uA - breakdown above c. 60V, bias +100V leakage current +20-30uA
|
Sun Oct 1 12:38:56 2023, TD, Sunday 1 October 13x
|
12.50 Power up
All system wide checks OK
|
Sat Sep 30 16:37:06 2023, TD, Honeywell HSS-DPS fault
|
The Honeywell HSS-DPS dew point sensor indicator ( red LED ) is flashing c. 0.5Hz indicating humiidity > 90+/-3 %
|
Sat Sep 30 12:14:52 2023, TD, Satuurday 30 September 10x
|
12.54 Power up
All system wide checks OK
|
Thu Sep 28 16:03:40 2023, TD, Thursday 28 September 10x
|
17.00 USB-controlled ac mains relay - no water interlock ( yellow ) LED
Water temperature ( station outside S4 entrance ) c. 19 deg C and pressure OK
|
Thu Aug 31 15:01:11 2023, TD, Test of CAEN N1419B - T9, STFC DL 7x
|
Noise test of CAEN N1419B in T9, STFC DL.
|
Wed Aug 30 15:30:50 2023, TD, Test of AIDA ASIC mezzanines with FEE64s - summary 13x
|
Test of ( previously bench-tested OK ) AIDA ASIC mezzanines using FEE64 nnlycca5 ( MAC d8:80:39:41:d7:c3 )
AIDA test station STFC DL, T9 - see attachments 12-13
|
Mon Aug 21 13:17:54 2023, TD, Silena 7710 HV test - JCMB 21.8.23 11x
|
Silena 7710 - see attachment 1
Bias -60V
|
Sun Aug 20 16:21:45 2023, TD, Silena 7710 HV test - JCMB 20.8.23 6x
|
Silena 7710 - see attachment 1
Bias -60V
|
Sat Aug 19 14:52:42 2023, TD, BNC PB-4 & BNC PB-5 pulser test - JCMB 18.8.23 21x
|
BNC PB-4 - see attachment 1
Amplitude 35,000
|
Wed Aug 9 14:40:28 2023, Carole, Nic, Jeroen, AIDA HV SUPPLY & PULSER TEST    
|
Tests with the HV supply (CAEN N1419ET) and Pulse Generator ( BNC PB-5 ) in the detector lab on the crate behind the DEGAS detectors, carried out the
9th August 2023
Both modules were plugged |
Fri Aug 4 10:04:58 2023, TD, Friday 4 August contd. 17x
|
10.45 Power cycle and reboot
|