ID |
Date |
Author |
Subject |
Text |
 |
887
|
Sat Aug 19 14:52:42 2023 |
TD | BNC PB-4 & BNC PB-5 pulser test - JCMB 18.8.23 | BNC PB-4 - see attachment 1
Amplitude 35,000
|
21x |
886
|
Wed Aug 9 14:40:28 2023 |
Carole, Nic, Jeroen | AIDA HV SUPPLY & PULSER TEST | Tests with the HV supply (CAEN N1419ET)
and Pulse Generator ( BNC PB-5 ) in the detector
lab on the crate behind the DEGAS detectors, |
    |
885
|
Fri Aug 4 10:04:58 2023 |
TD | Friday 4 August contd. |
10.45 Power cycle and reboot
|
17x |
884
|
Fri Aug 4 09:07:08 2023 |
TD | Friday 4 August |
07.00 DAQ stopped
Overnight run for background alphas
|
23x |
883
|
Thu Aug 3 14:16:12 2023 |
TD | Thursday 3 August contd. |
15.16 DSSSD bias +100V leakage current +8.415uA
|
21x |
882
|
Thu Aug 3 09:21:26 2023 |
TD | Thursday 3 August |
10.20 DSSSD bias +100V leakage current +6.565uA
- attachment 1
|
47x |
881
|
Wed Aug 2 15:39:52 2023 |
TD | Wednesday 2 August contd. | 16.22 Add LK1 to aida01
DSSSD bias -100V leakage current -7.950uA
|
11x |
880
|
Wed Aug 2 08:44:00 2023 |
TD | Wednesday 2 August |
09.40 Al snout electrically connected ( <
1 Ohm ) to Al gasket and Al base plate
|
48x |
879
|
Tue Aug 1 08:50:36 2023 |
TD | Tuesday 1 August |
09.50 DSSSD bias +100V leakage current +9.255uA
- attachment 2
|
30x |
878
|
Mon Jul 31 09:43:37 2023 |
TD | Monday 31 July |
10.43 DSSSD bias +100V leakage current +7.550uA
|
44x |
877
|
Mon Jul 24 08:54:48 2023 |
TD | Monday 24 July |
09.54 DSSSD#1 bias +100V leakage current
+9.065uA
|
17x |
876
|
Mon Jul 24 08:39:39 2023 |
TD | aida10 sys console output & aida10 options | Options and system cosnole disagree about
which version of the DataAcq program is in
use? |
 |
875
|
Sun Jul 23 08:50:48 2023 |
TD | Sunday 23 July | 09.50 Cooknell HV2
2x + polarity, grounded HV channels
|
19x |
874
|
Sat Jul 22 08:17:48 2023 |
TD | Saturday 22 July | 09.20 CAEN N1419ET
Floating outputs
Channel #3 + polarity, Channels #0-2 |
26x |
873
|
Sun Jul 16 14:20:31 2023 |
TD | Sunday 16 July contd. | 15.20 All ERNI pins of aida01 adaptor PCB
OK => FEE64 #1 ASIC #2 channel #5 issue faulty
test C? faulty ASIC channel?
|
27x |
872
|
Sun Jul 16 09:36:40 2023 |
TD | 15-16 July 2023 |
Saturday 15 July
|
40x |
871
|
Fri Jul 7 11:59:06 2023 |
TD | Test of AIDA FEE64 - summary | Test of AIDA FEE64s
ISOL test station STFC DL, T9
|
11x |
870
|
Fri Jul 7 10:52:34 2023 |
TD, PJCS | Test of AIDA FEE64 d8:80:39:42:0d:0b ( aida03 ) - ASIC mezzanine replaced |
Test of AIDA FEE64 d8:80:39:42:0d:0b ( aida03
) - ASIC mezzanine replaced
|
16x |
869
|
Fri Jul 7 10:15:09 2023 |
TD | Test of AIDA FEE64 d8:80:39:41:f6:ee ( aida01 ) - HDMI connector replaced |
Test of AIDA FEE64 d8:80:39:41:f6:ee ( aida01
) - HDMI connector replaced. Cooling plate |
17x |
868
|
Thu Jul 6 10:20:25 2023 |
TD | Unable to start Rly16 server |
Unable to start Rly16 server - see attachment
1
|
|