783
|
Wed Nov 14 07:43:35 2018 |
TD | Wednesday 14 November 16:00 - 00:00 | 16.44 DAQ continues OK - AIDA file RIBF148/R5_432
165Pm BigRIPS setting + 2.8mm Al degrader
ASIC settings 2018Oct03-13.26.58
from c. 09.00 13.11.18
n+n FEE64s Ibias LF feedback 0xb -> 0xd
HEC fast comparator threshold 0x20 -> 0xff
BNC PB-4 pulser
Amplitude 90,000, x5 attenuator IN
t_r 50ns tau_d 1ms
Frequency min
Delay min
16.45 BigRIPS and BRIKEN stop
AIDA on File R5_432
16.48 BigRIPS starts run 1047
BRIKEN starts run 62
AIDA on file R5_433
16.57 online MIDASsort
m_p versus m_n (LEC) - attachment 1
x versus y (LEC) - attachment 2
E_p versus E_n (LEC) - attachment 3
m_p versus m_n (HEC) - attachment 4
x versus y (LEC) - attachment 5
E_p versus E_n (LEC) - attachment 6
all MIDASsort spectra zero'd
17.17 file RIBF128/R5_436 ADC data synchronised - attachment 7
analysis of RIBF148/R5_436 - attachment 8
max dead time nnaida19 ~0.25%
17.48 BigRIPS and BRIKEN stop
AIDA on File R5_443
17.48 BigRIPS starts run 1048
BRIKEN starts run 63
AIDA on file R5_443
18.45 BigRIPS and BRIKEN stop
AIDA on File R5_452
18.47 BigRIPS starts run 1049
BRIKEN starts run 64
AIDA on file R5_453
19.45 BigRIPS and BRIKEN stop
AIDA on File R5_463
19.46 BigRIPS starts run 1050
BRIKEN starts run 65
AIDA on file R5_463
20.49 BigRIPS and BRIKEN stop
AIDA on File R5_473
20.49 BigRIPS starts run 1051
BRIKEN starts run 66
AIDA on file R5_473
21.49 BigRIPS and BRIKEN stop
AIDA on File R5_483
ASIC check load
21.50 BigRIPS starts run 1052
BRIKEN starts run 67
AIDA on file R5_484
22.00 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 9 & 10
FEE64 temperatures OK - see attachment 11
good event statistics OK - see attachments 12
Merger OK - see attachment 13
TapeServer OK - see attachment 14
1x merger error message reported - trying to add items to Q index 12 (Link 12) but queue alreday full
per FEE64 Rate spectra - attachment 15
1.8.L spectra - attachments 16 & 17
nnaida1 pulser peak width 73 ch FWHM
1.8.H spectra - attachments 18 & 19
1.8.W spectra 20ms FSR - attachments 20 & 21
22.48 BigRIPS and BRIKEN stop
AIDA on File R5_493
22.50 BigRIPS starts run 1053
BRIKEN starts run 68
AIDA on file R5_493
23.24 BRIKEN DAQ found stopped - no sync check data - BigRIPS DAQ stopped and restarted
AIDA on File R5_499
23.26 BigRIPS starts run 1054
BRIKEN starts run 69
AIDA on file R5_499
|
|
782
|
Tue Nov 13 22:58:49 2018 |
OH | Wednesday 14 November 08:00 - 16:00 | 08:01 BigRips and BRIKEN stop
AIDA on file R5_346
08:02 BigRIPS starts run 1038
BRIKEN starts run 181114_0800_165Pm_053.dlt
AIDA on file R5_346
System wide checks ok * except nnaida 7 fails adc*
FEE Temps ok - attachment 1
Stats ok - attachment 2
Leakage currents ok - attachment 3
09:15 BigRIPS and BRIKEN stop
AIDA on file R5_358
09:16 BigRIPS starts run 1039
BRIKEN starts run 181114_0915_165Pm_054.dlt
AIDA on file R5_358
10:15 BigRIPS and BRIKEN stop
AIDA on file R5_368
10:15 BigRIPS starts run 1040
BRIKEN starts run 181114_1015_165Pm_055.dlt
AIDA on file R5_368
10:36 Analyser output of R5_368 - attachment 4
11:15 BugRIPS and BRIKEN stop
AIDA on File R5_378
11:16 BigRIPS starts run 1041
BRIKEN starts run
AIDA on file R5_378
08.03 DAQ continues - AIDA file RIBF148/R5_384
165Pm BigRIPS setting + 2.8mm Al degrader
ASIC settings 2018Oct03-13.26.58
from c. 09.00 13.11.18
n+n FEE64s Ibias LF feedback 0xb -> 0xd
HEC fast comparator threshold 0x20 -> 0xff
11.50 all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 5 & 6
FEE64 temperatures OK - see attachment 7
good event & AIDA discriminator statistics OK - see attachments 8
Merger OK - see attachment 9
TapeServer OK - see attachment 10
no errors or warnings shown by server terminal sessions
per FEE64 Rate spectra - attachment 11
per FEE64 HitRate spectraq - attachment 12
1.8.L spectra - attachments 13-16
some evidence of pulser peak shifts, notably nnaida5, 11 & 15
nnaida1 pulser peak width 78 ch FWHM
1.8.H spectra - attachments 17 & 18
1.8.W spectra 20us FSR - attachments 19 & 23
1.8.W spectra 200us FSR - attachments 20 & 24
1.8.W spectra 2ms FSR - attachments 21 & 25
1.8.W spectra 20ms FSR - attachments 22 & 26
12.15 BugRIPS and BRIKEN stop
AIDA on File R5_388
12.16 BigRIPS starts run 1042
BRIKEN starts run 57
AIDA on file R5_388
12.22 *all* online and MIDASsort spectra zero'd
13:15 BigRIPS and BRIKEN stop
AIDA on file R5_398
13:16 BigRIPS starts run 1043
BRIKEN starts run58
AIDA on file R5_398
13:28 BigRIPS and BRIKEN stop
AIDA on file
beam halted to give access to F11 area
13.44 beam started
BigRIPS and BRIKEN start
AIDA file RIBF148/R5_402
BRIKEN run 59
BigRIPS run 1043
14.45 BRIKEN and BigRIPS stop
AIDA file RIBF148/R5_412
14.47 BRIKEN and BigRIPS start
AIDA file RIBF148/R5_413
BRIKEN run 60
BigRIPS run 1044
15.43 BRIKEN and BigRIPS stop
AIDA file RIBF148/R5_422
BRIKEN Resets the timestamp
15:44 BRIKEN and BigRIPS start
AIDA file RIBF148/R5_422
BRIKEN run 61
BigRIPS run 1045
16:02 System wide checks all ok *Except nnaida7 fails adc calibration*
Leakage currents ok - attachment 27
Stats ok - attachment 28
FEE Temps ok - attachment 29
16:27 First 200ms of decay correlations in DSSD 3 and 4 for the 3 degrader settings - attachment 30
DSSD 3 along the top row and DSSD 4 along the bottom.
Dgrader settings are 3mm, 2.3mm and 2.8mm
1 hour data in each |
|
781
|
Tue Nov 13 14:27:38 2018 |
LS | Wednesday 14 November 00:00 to 08:00 |
00:00 BRIKEN stop AIDA file RIBF148/R5_269
----------------------------------------------------------------
00.02 BRIKEN start
BRIKEN file 45
BigRIPS file 1030
AIDA file RIBF148/R5_270
00:07 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 1)
good event statistics ok (attachment 2)
detector bias/leakage currents ok (attachment 3)
01:01 BRIKEN stop
AIDA file RIBF148/R5_279
---------------------------------------------------------------
01.02 BRIKEN start
BRIKEN file 46
BigRIPS file 1031
AIDA file RIBF148/R5_279
02:00 BRIKEN stop
AIDA file RIBF148/R5_289
----------------------------------------------------------------
02.01 BRIKEN start
BRIKEN file 47
BigRIPS file 1032
AIDA file RIBF148/R5_289
03:00 BRIKEN stop
AIDA file RIBF148/R5_298
----------------------------------------------------------------
03.01 BRIKEN start
BRIKEN file 48
BigRIPS file 1033
AIDA file RIBF148/R5_298
03:04 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 4)
good event statistics ok (attachment 5)
detector bias/leakage currents ok (attachment 6)
04:00 BRIKEN stop
AIDA file RIBF148/R5_308
----------------------------------------------------------------
04.03 BRIKEN start
BRIKEN file 49
BigRIPS file 1034
AIDA file RIBF148/R5_308
05:00 BRIKEN stop
AIDA file RIBF148/R5_317
----------------------------------------------------------------
05.01 BRIKEN start
BRIKEN file 50
BigRIPS file 1035
AIDA file RIBF148/R5_318
06:00 BRIKEN stop
AIDA file RIBF148/R5_327
----------------------------------------------------------------
06.02 BRIKEN start
BRIKEN file 51
BigRIPS file 1036
AIDA file RIBF148/R5_327
06:07 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 7)
good event statistics ok (attachment 8)
detector bias/leakage currents ok (attachment 9)
07:00 BRIKEN stop
AIDA file RIBF148/R5_337
----------------------------------------------------------------
07.01 BRIKEN start
BRIKEN file 52
BigRIPS file 1037
AIDA file RIBF148/R5_327 |
|
661
|
Wed Jun 14 07:01:34 2017 |
TD | Wednesday 14 June | 15.01 DAQ continues
AIDA file RIBF148/R8_22
BNC PB-4
Amplitude 90,000
x5 attenuator IN
t_r 100ns tau_d 50us
- polarity
frequency 25Hz range 2.6Hz
delay MIN, tail pulse, INT ref
ASIC settings 2017Mar27-16.12.39
*except*
slow comparator 0x64
LEC/MEC fast comparator 0x14
15.02 Detector biases & leakage currents OK - see attachments 1 & 2
FEE64 Temperatures OK - see attachment 3
good events statistics stable & OK - see attachment 4
No warnings/errors reported by merger since start of RIBF148/R3 - see attachment 5
*all* system wide checks OK (zero fails)
15.12 online analysis
ADC & disc data synchronised - note the effect of the higher slow comparator threshold on the after-pulse
- see attachment 6
per DSSSD m_p versus m_n (LEC) - see attachments 7-8
per DSSSD x versus y (LEC) - see attachment 9
per DSSSD E_p versus E_n (LEC) - 512x512 spectra - nominal 20keV per channel - see attachment 10
15.30 per FEE64 1.8.* spectra - see attachments 11-18
note all waveforms x-scale 1000ch = 2ms
note absence of saw tooth transients/noise p+n waveforms - attachments 15-16
Brief discussion with Nishimura-san and Gabor - clover detectors have been removed from BRIKEN
otherwise no major changes in last 24h
see also - attachments 17-18
15.42 Ambient temperature +24.7 deg C, d.p. +9.4 deg C, RH 37.8%
Julabo FL11006 recirculating chiller set point +20.0deg C , actual +20.0 deg C, water level c. 75%
15.54 analysis file RIBF148/R8_22 - zero deadtime, 1x MBS & zero ts timewarps - see attachment 19 |
|
383
|
Wed Oct 12 02:58:19 2016 |
TD & CG | Wednesday 12 October | 10.50 Merger & TapeServer found stopped
connect errors - see attachment 1
DSSSD bias and leakage currents OK - see attachments 2 & 3
Most recent data file R17_440 last accessed 08.22 12.10.16
R17_440 ADC data synchronised
aidas1> ./analyser2 /TapeData/Sept2016/R17_440
*** GREAT format 3.2.0 analyser - TD - May 2014
*** ERROR: READ I/O error: 5002
blocks: 30546
ADC data format: 147630891 ( 1189229.1 Hz)
Other data format: 101593923 ( 818381.9 Hz)
Sample trace data format: 0 ( 0.0 Hz)
Undefined format: 0 ( 0.0 Hz)
Other data format type: PAUSE: 101 ( 0.8 Hz)
RESUME: 101 ( 0.8 Hz)
SYNC100: 1133804 ( 9133.3 Hz)
FEE64 disc: 100456290 ( 809217.8 Hz)
MBS info: 3627 ( 29.2 Hz)
Other info: 0 ( 0.0 Hz)
ADC data range bit set: 151 ( 1.2 Hz)
Timewarps: ADC: 0 ( 0.0 Hz)
PAUSE: 0 ( 0.0 Hz)
RESUME: 0 ( 0.0 Hz)
SYNC100: 0 ( 0.0 Hz)
FEE64 disc: 0 ( 0.0 Hz)
MBS info: 0 ( 0.0 Hz)
Undefined: 0 ( 0.0 Hz)
Sample trace: 0 ( 0.0 Hz)
Timestamp elapsed time: 124.140 s
FEE module #: 1 elapsed dead time 0.728 s
FEE module #: 2 elapsed dead time 0.184 s
FEE module #: 3 elapsed dead time 0.000 s
FEE module #: 4 elapsed dead time 0.000 s
FEE module #: 5 elapsed dead time 1.034 s
FEE module #: 6 elapsed dead time 3.462 s
FEE module #: 7 elapsed dead time 0.000 s
FEE module #: 8 elapsed dead time 0.000 s
FEE module #: 9 elapsed dead time 0.064 s
FEE module #: 10 elapsed dead time 0.003 s
FEE module #: 11 elapsed dead time 0.000 s
FEE module #: 12 elapsed dead time 0.000 s
FEE module #: 13 elapsed dead time 0.000 s
FEE module #: 14 elapsed dead time 0.062 s
FEE module #: 15 elapsed dead time 0.018 s
FEE module #: 16 elapsed dead time 0.005 s
FEE module #: 17 elapsed dead time 0.817 s
FEE module #: 18 elapsed dead time 0.035 s
FEE module #: 19 elapsed dead time 0.048 s
FEE module #: 20 elapsed dead time 0.000 s
FEE module #: 21 elapsed dead time 0.000 s
FEE module #: 22 elapsed dead time 0.103 s
FEE module #: 23 elapsed dead time 0.216 s
FEE module #: 24 elapsed dead time 0.000 s
FEE module #: 25 elapsed dead time 0.000 s
FEE module #: 26 elapsed dead time 0.000 s
FEE module #: 27 elapsed dead time 0.000 s
FEE module #: 28 elapsed dead time 0.000 s
FEE module #: 29 elapsed dead time 0.000 s
FEE module #: 30 elapsed dead time 0.000 s
FEE module #: 31 elapsed dead time 0.000 s
FEE module #: 32 elapsed dead time 0.000 s
*** Program elapsed time: 14.129s ( 2161.951 blocks/s, 135.122 Mb/s)
Conclude DAQ ran without problem for c. 16.5h which is a significant improvement
assuming that it can be consistently reproduced.
11.20 Merger restart did not clear connect/timeout issues
Re-route BNC - BNC RG58 cable with 50MHz clock from BRIKEN to
AIDA Master MACB external clock input (rear panel) so that it will
not be stepped upon.
Power cycle AIDA FEE64s
N.B. I selected 'Switch ALL On' rather than 'Sequence ALL On' by mistake and one
of the USB ac mains controller output 250VAC/16A fuses blew - use 'Sequence ALL On'
TS_SYNC_PHASE 0
FEE64 temperatures OK
*All* system-wide checks OK
B2F/F11 temperature +24.6 deg C, d.p. +8.9 deg C,RH 36.6%
ASIC settings 2016Sep02-18.17.01
slow comparator 0x10 -> 0x0a
LEC/MEC fast comparators # 1, 55 & 58 disabled for all FEE64s
Master MACB serial #13 (see https://elog.ph.ed.ac.uk/AIDA/367)
Master MACB mode 5
14.05 DAQ starts (file Sept2016/R21)
15.30 Sequence of three electrical power dips over 30 minute period. RIKEN public address
system reports that TEPCO (Tokyo Electric Power Company) - the site supplier - is
having region-wide problems.
The third power dip stopped DAQ.
Unable to recover by restarting merger
Power cycle FEE64s
16.20 DAQ restarts (file Sept2016/R22)
19.05 DAQ stops
Because of risk of power outages due to TEPCO distribution problems we have powered
down AIDA and Julabo FL1006 chiller and disconnected from ac mains. |
|
838
|
Wed Jun 12 01:54:04 2019 |
OH, TD, CG, CB, CA | Wednesday 12 June | 09:54 DAQ Continues ok with merger running at ~ 560k data items per second
ASIC settings 2019May19-17.54.03
slow comparator 0xa
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range, variable min. c. 2.5Hz
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
System wide checks all ok - except nnaida7 fails adc calibration
Bias and leakage currents ok - attachment 1
FEE Temps ok - attachment 2
Good event statistics good - attachment 3
Merger and tape service good - attachment 4 and 5
16:33 DAQ continues OK with merger running at ~ 600k data items per second
ASIC settings 2019May19-17.54.03
slow comparator 0xa
BNC PB-4 Pulser
Amplitude 90,000
x5 attenuator IN
Frequency 25Hz range, variable min. c. 2.5Hz
t_r 100ns, tau_d 50us
- polarity
Delay MIN, tail pulse
All system-wide checks OK except nnaida7 fails ADC calibration
Bias and leakage currents OK - attachment 6
FEE temperatures OK - attach 7
Good event statistics OK - attach 8
check nnaida10 & nnaida20 counts vs. rate
Merger and tape service OK - attachment 9 & 10
17.00 Average FEE64 rates since last DAQ stats zero @ 17.05 yesterday - data from attachment 11
DSSSD#1 count rate
24h most recent
nnaida17 1914894752 22.1k 23.2k
nnaida20 3061905664 35.4k 35.6k
nnaida22 1872820000 21.7k 21.8k
nnaida23 2217883584 25.7k 24.8k
DSSSD#2
nnaida18 1760374528 20.3k 21.1k
nnaida19 1402771232 16.2k 66.6k => 32 bit overflow => 4.3E+09 + 1.4E+09 = 5.6E+09 => 65.9k
nnaida21 2038432092 23.6k 23.5k
nnaida24 2844103808 32.9k 31.3k
Conclude rates have bee stable overnight - no evidence of noise transients for DSSSD#1 to 150-250k good
events/s
17.15 DAQ stats zero'd
all spectra zero'd |
|
905
|
Wed Oct 11 16:53:20 2023 |
TD | Wednesday 11 October contd. | CAEN N1419ET ch 2 - all outputs non-floating ( internal LK fitted )
Standard coaxial HV cable chained to aida12, aida03, aida11, aida09, aida01 & aida10
aida04 LK1 ( local gnd ) fitted
ground cable from aida04 adaptor PCB ground to aida12 to aida02
aida03 LK3 ( PCB gnd ) fitted
BNC PB-5
amplitude 10.0V
attenuation x10
decay time 1ms
polarity +
frequency 25Hz
test + direct from PB-5 to aida12 *only*
slow comparator 0xf
DSSSD bias -150V -7.280uA
aida12 1.8.L pulser peak width 55 ch FWHM
aida12 *.*.L spectra - attachments 1-4
All channels have similar pulser peak widths to 1.8.L *except* 1.1.L, 1.4.L, 4.7.L & 4.13.L ( hot channels ) and 2.2.L, 2.9.L & 2.12.L no pulser peak
test + direct from PB-5 to aida01 *only*
aida01 1.8.L pulser peak width 70 ch FWHM
aida01 *.*.L spectra - attachments 5-8
All channels have similar pulser peak widths to 1.8.L *except* 1.1.L, 1.4.L, 4.7.L & 4.13.L ( hot channels )
per FEE64 1.8.W spectra - 20us FSR - attachments 9-12
per FEE64 1.8.W spectra - 200us FSR - attachments 13 & 14 |
|
904
|
Wed Oct 11 09:37:30 2023 |
TD | Wednesday 11 October |
10.00 Water flow & temp OK
Power ON
All system wide checks OK
CAEN N1419ET ch 2 - all outputs non-floating ( internal LK fitted )
Standard coaxial HV cable chained to aida12, aida03 & aida11
aida04 LK1 ( local gnd ) fitted
ground cable from aida04 adaptor PCB ground to aida12 to aida02
aida03 LK3 ( PCB gnd ) fitted
BNC PB-5
amplitude 10.0V
attenuation x10
decay time 1ms
polarity -
frequency 25Hz
test + cable and test + chain removed
test - direct from PB-5 to aida02 *only*
DSSSD bias -120V -6.105uA - attachment 1
FEE64 tempweratures OK - attachment 2
slow comparator 0x64
ADC data item stats - attachment 3
Note
aida04 asic #1 u/s
ASIC mezzanine to be replaced
aida10 asic #4 v. high rates observed
see https://elog.ph.ed.ac.uk/AIDA/901 attachment 45
ASIC/cable/wafer issue ?
rate ~ 1600Hz observed for aida02 - pulser
All other rates 0Hz
per FEE64 Rate spectra - attachment 4
aida02 1.8.L pulser peak width 137 ch FWHM
1.8.W spetcra 20us, 200us & 2ms FSR - attachments 5-10
slow comparator 0x64 -> 0xf
ADC data item stats - attachment 11
per FEE64 Rate spectra - attachments 12 & 13
p+n rates entirely dominated by hot channels, otherwise rates ~0Hz
n+n rates ~ 200-300Hz/channel
slow comparator 0xf -> 0xa
ADC data item stats - attachment 14
per FEE64 Rate spectra - attachments 15-17
p+n rates aida12, aida03 & aida11 entirely dominated by hot channels, otherwise rates ~0Hz
p+n rates aida10, aida01 & aida09 ~500-1000Hz/channel
n+n rates ~ 2000-3000Hz/channel
11.25 Extend HV chain from aida11 to aida09
ADC data item stats - attachment 18
significant decrease in aida09 rate
1.8.W spectra 2ms FSR - attachments 19 & 20
significant improvement in aida09 1.8.W noise
aida02 1.8.L pulser peak width 165 ch FWHM ( slow comparator 0xa ), 127 ch FWHM ( 0xf )
12.00 Extend HV chain from aida09 to aida01 & aida10
slow comparator 0xf
ADC data item stats - attachment 21
per FEE64 Rate spectra - attachment 22
1.8.W spectra 2ms FSR - attachments 23-26
significant improvement in aida01 & aida10 1.8.W noise
aida02 1.8.L pulser peak width 127 ch FWHM ( 0xf )
12.35 slow comparator 0xf
ADC data item stats - attachment 27
per FEE64 Rate spectra - attachment 28
slow comparator 0xa
ADC data item stats - attachment 29
per FEE64 Rate spectra - attachment 30
slow comparator 0x9
ADC data item stats - attachment 31
per FEE64 Rate spectra - attachment 32 & 33
slow comparator 0x8
ADC data item stats - attachment 34
per FEE64 Rate spectra - attachment 35
slow comparator 0x7
ADC data item stats - attachment 36
per FEE64 Rate spectra - attachment 37
slow comparator 0x6
ADC data item stats - attachment 38
per FEE64 Rate spectra - attachment 39
slow comparator 0x5
ADC data item stats - attachment 40
per FEE64 Rate spectra - attachment 41
16.55 Background alpha 3*L & 3*W ( 20us FSR ) spectra - attachments 42 & 43
Slow comparator 0xf
Waveform threshold 7000 ( - polarity ), 9000 ( + polarity )
17.00 DSSSD bias -120V -7.125uA - attachment 44
aida02 1.8.L ( n+n Ohmic side strip ) pulser peak width as function of DSSSD bias
Bias I_l FWHM Centroid
(V) (uA) (ch) (ch)
-120 -7.125 124.8 43090
-140 -7.230 124.5 43092
-160 -7.370 120.2 43093
-180 -7.570 120.0 43095
-100 -7.020 143.8 43089
-80 -6.820 continuum/no peak
-150 -7.300 122.6 43093
DSSSD bias -150V -7.280uA
aida02 *.*.L spectra - attachments 45-48
All channels have similar pulser peak widths to 1.8.L
|
|
220
|
Wed May 11 06:37:32 2016 |
AE, VP, BM | Wednesday 11 May | *2:30 pm*: changed power source of USB-controlled power supply for FEEs. Now one of the power cables is connected
to the 'B' power supply of F11, and the second to the 'C' power supply (these are in the unit on top of WASBI
crates). The reason for change is that before both were connected to 'C', and drained current was very close to
limit of 10 A (and power cord getting noticeable hot). We'll check resolution after this switch.
Detectors 1, 2, 3, and 4 biased to 100V, detectors 5, 6, and veto biased to 70V.
n-side is more noisy than p-side, but resolution comparable to previous configuration:
NNAIDA# FWHM [ch]
1 91
2 61
3 84
4 -
5 160
6 -
7 79
8 84
9 71
10 76
11 101
12 81
13 69
14 78
15 199
16 107
17 82
18 62
19 -
20 89
21 168
22 70
23 80
24 73
25 81
26 13
27 109
28 13
29 15
30 126
31 13
32 100
*5 pm*: Set-up correlation scalers for AIDA<->EURICA<->BigRIPS (detail diagram coming soon, now only in paper
logbook).
*6:30 pm*: Started EURICA DAQ with Pieter and Gabor.
A folder is created to record data of June campaign: /raid/11/mbs/2016_ucampaign/
Tested correlation of AIDA and EURICA with a pulser run, triggering EURICA and AIDA correlation scaler with 250
Hz. The corresponding files are
/raid/11/mbs/2016_ucampaign/test0003.lmd
/TapeData/may2016/R12_0.gz
Phong did a quick analysis and the pattern of correlation scalers in both files is similar (attached figures). It
is also consistent with the number of reset requests we sent during the data taking (three requests closely spaced
in the middle of run, and two more in the end of the run). |
|
561
|
Wed May 10 02:55:13 2017 |
TD, OH | Wednesday 10 May | 10.55 DAQ continues OK
Current file May2017/R6_44
ASIC settings 2017Mar27-16.12.39 *except*
slow comparator 0x20
LEC/MEC fast comparator 0xff
Detector biases & leakage currents OK - see attachments 1 & 2
FEE64 Temperatures OK - see attachment 3
'good events' statistics OK - see attachment 4
No warnings/errors reported by merger since DAQ GO - see attachment 5
*all* system wide checks OK
10.48 E17/F11 ambient temperature +24.7 deg C, RH 35.3%, d.p. +8.3 deg C
Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 65%)
18.32 DAQ continues OK
Current file May2017/R6_67
Detector biases & leakage currents OK - see attachments 6 & 7
FEE64 Temperatures OK - see attachment 8
'good events' statistics OK - see attachment 9
*all* system wide checks OK |
|
694
|
Wed Nov 1 02:24:20 2017 |
OH | Wednesday 1 November | 11:24 DAQ running OK - data transfer not enabled
BNC PB-4
Amplitude 90,000
x5 attenuator IN
t_r 100ns tau_d 50us
- polarity
frequency range 25Hz
delay MIN, tail pulse, INT ref
ASIC settings 2017Oct23-14.38.03
*all* FEE64s slow comparators 0xa *except* nnaida13 0xe, nnaida19-22 0xc
Detector biases & leakage currents OK - see attachments 1
FEE64 Temperatures OK - see attachment 2
DAQ statistics OK - see attachment 3
*all* system wide checks OK *except* nnaida21 fails ADC calibration |
|
265
|
Wed Jun 1 01:33:58 2016 |
TD | Wednesday 1 June 2016 | 09.30 Detector biases and leakage currents - ambient temp +24.9 deg C - attachments 1-2
10.08 Connection lost to nnaida27 and nnaida28
Most recent save of settings DB key 2016May31-23.32.46
12.00 Replaced nnaida17 & nnaida18 with FEE module previously used for nnaida31 & nnaida32
Pulser walkthrough
BNC PB-4 settings
90,000 - 10,000 @ 10,000 step
x5 attenuator (was x10 previously)
t_r = 50ns tau_d = 50us
polarity -
rate ~ 150Hz
INT, delay MIN
polarity + by EG&G Ortec 433A
N.B. - polarity terminated at end of AIDA adaptor chain, + polarity not terminated
Spectra saved 12.32 ( to directories tagged around June 1, 2016 05.29 )
BNC PB4 pulser rate set to ~5Hz
207Bi source at front of snout and 2x 60Co sources above DSSD3
14.00 Rebooted aidas1 to cleanup merge
Changed config to 24x FEE64s
Clock issues with nnaida23 - clock not locked status 0x6 - see attachments 3-9
18.00 Replace current nnaida19 (no histogramming, reSYNCs OK but timestamp not sync'd wrt timestamps of
other FEE64s).
Replace current nnaida23
Following powerup obtained clock not locked - status 0x6 for nnaida5 (master)
Replacement nnaida19 also does not histogram, reSYNCs OK but timestamp not sync'd wrt timestamps
of other FEE64s)
With help from PJCS investigate above issues and change timestamp master from nnaida5 to nnaida6
( /MIDAS/config/HttpdTcl/aidas1/startup.tcl and /MIDAS/config/HttpdTcl/nnaida5/startup.tcl ) to
change definition of timestamp master.
Able to operate DAQ and obtain AIDA SYNCs and WAVE SYNCs with TS_SYNC 0x0 _only_
PJCS will also investigate using system in T9 at STFC DL
Restored settings 2016June1-17.27.? |
|
267
|
Wed Jun 1 09:00:13 2016 |
TD - Reply PJCS | Wednesday 1 June 2016 | > 09.30 Detector biases and leakage currents - ambient temp +24.9 deg C - attachments 1-2
>
> 10.08 Connection lost to nnaida27 and nnaida28
> Most recent save of settings DB key 2016May31-23.32.46
>
> 12.00 Replaced nnaida17 & nnaida18 with FEE module previously used for nnaida31 & nnaida32
>
> Pulser walkthrough
>
> BNC PB-4 settings
>
> 90,000 - 10,000 @ 10,000 step
> x5 attenuator (was x10 previously)
> t_r = 50ns tau_d = 50us
> polarity -
> rate ~ 150Hz
> INT, delay MIN
>
> polarity + by EG&G Ortec 433A
>
> N.B. - polarity terminated at end of AIDA adaptor chain, + polarity not terminated
>
> Spectra saved 12.32 ( to directories tagged around June 1, 2016 05.29 )
>
> BNC PB4 pulser rate set to ~5Hz
>
> 207Bi source at front of snout and 2x 60Co sources above DSSD3
>
> 14.00 Rebooted aidas1 to cleanup merge
>
> Changed config to 24 FEE64s
>
> Clock issues with nnaida23 - clock not locked status 0x6 - see attachments 3-9
Check the Local Controls for the setting of the LMK3200 control rgister.
If you recall from the problem with nnaida5 the value should be 0xB.
If the clock still doesn't lock then maybe it's an HDMI cable problem like nnaida5.
Since the other FEE64 are ok then it is likely to be just the MACB to nnaida23. |
|
269
|
Wed Jun 1 13:58:48 2016 |
TD - Reply PJCS | Wednesday 1 June 2016 |
> Check the Local Controls for the setting of the LMK3200 control rgister.
> If you recall from the problem with nnaida5 the value should be 0xB.
> If the clock still doesn't lock then maybe it's an HDMI cable problem like nnaida5.
> Since the other FEE64 are ok then it is likely to be just the MACB to nnaida23.
I attach SYNC stats for WAVE and AIDA - you will notice that most (but not all) FEE64s
produce WAVE SYNCS
Plus (load balanced) statistics for disc, good events |
|
854
|
Tue Jun 18 23:58:46 2019 |
CB | Wed Jun 19 8:00-16:00 | 08:00 Shift change
No change in settings
DAQ continues running OK
Good events rate OK - see attach 1
FEE64 temperature OK - see attach 2
Merger & Tape server OK - see 3 & 4
Leakage currents OK - see attach 5
Online HEC xy plot OK, most implants in DSSD2 - see attach 6
Zeroed online histograms
09:29 DTAS DAQ stops on run 135
AIDA on run R2_787
09:32 DTAS DAQ restarts on run 136
AIDA on run R2_787
BigRIPS on run 2053
09:33 DAQ continues OK
Good events rate OK - see attach 7
Merger ~ 700k items/s
Tape ~ 5400 Kbytes/s
Leakage currents OK - see attach 8
Online HEC xy plot OK - see attach 9
10:31 Good event rate in nnaida 19, 24 steadily creeping up. See attach 10.
Currently 160k and 120k
Merger ~ 710 k items/s
Tape ~ 5689 kbytes/s
All system-wide checks OK, except nnaida7 fails ADC calibration as usual
11:20 Good event rate stable - see attach 11
FEE64 temperatures OK - see attach 12
Merger ~700k items/s
Tape server ~ 5500 Kbytes/s
Leakage currents OK - see attach 13 & 14
11:30 DTAS stops DAQ on run 139
BigRIPS on run 2053
AIDA on run R2_806
11:32 DTAS restarts on run 140
BigRIPS 2053
AIDA R2_807
13:14 DAQ continues OK
Good statistics events stable - see attach 15
Merger ~ 730k items/s
Tape server ~ 5500 KBytes/s
Leakage current OK - see attach 16
13:34 DTAS stops DAQ on AIDA R2_826
13:37 DTAS restarts DAQ on run 144
BigRIPS run 2055
AIDA run 827
15:10 DTAS DAQ reports 20% drop in F11 counting rate
AIDA plastic unchanged, beam current unchanged. Focusing issue?
Reset online histograms to check we are still implanting. We are.
Acquisition continues uninterrupted.
15:17 Good statistics OK - see attach 17
FEE64 temperature OK - see attach 18
Merger ~ 700k items/s
TapeServer ~ 5500 Kbytes/s
Detector leakage current OK - see attach 19
15:30 DTAS stops DAQ on AIDA run R2_845
15:33 DTAS restarts DAQ on run 147
BigRIPS run 2056
AIDA run R2_846
|
|
855
|
Wed Jun 19 08:30:36 2019 |
CG | Wed Jun 19 16.00 - end of run | 16.10 DAQ continues as before.
Merger rate ~700k items/s and ~5.3Kb/s
Individual FEE rates remain roughly constant
Temps and bias ok.
17.33 DTAS run 149 stopped
DTAS run 150 started
BigRIPS run 2057
AIDA run ~1.3GB in R2_865
19.30 DTAS run 152 stopped
DTAS run 153 started
BigRIPS run 2058
AIDA run ~1.8GB in R2_884
Temps and bias ok.
Rates in merger, tape server and individual FEEs as before.
19.50 In looking at previous data, AIDA is seeing ~30% fewer implants that what is observed at F11Plastic
Beam spot is not central on AIDA and some ions will be lost over sides of DSSD, but does not look
like it would account for a ~30% reduction
AIDA implant profile shown in Attachment 1 - from file
merged_a2_687_708_br190618_2315_100Sn118_121bgrun2049_sw_Tsi.root
21.01 Beam stopped.
DTAS run 155 stopped.
AIDA file R2_899.
DTAS and AIDA will continue with background measurements.
21.04 DTAS run 156 started
AIDA run R2_899
Chiller is full.
Lead bricks upstream of moderator.
That's all folks. |
|
640
|
Tue Jun 6 15:59:19 2017 |
DK | Wed 7 June night shift | 7 June 2017
00:05 All system-wide checks are passed
Sync pulses nera 44 x 10^6
Memory 35k ~ 38k
Biases uploaded as 33.png, 34.png - attachments 1 & 2
Entered into spreadsheet
SSD 3 bias is going up slowly since this isotope setting began
Around "2 to 4" Hz in SSDs 3 and 4 -> 50 to 80 pps implant rate
See attachment 3, 35.png
All rates under 100k except nnaida 13, 21
See attachment 4, 36.png
FEE temperatures shown as 37.png, attachment #5
00:30 BRIKEN number 32, AIDA R6_299, BigRIPS 3032
01:01 Timewarps look okay (only 2 MBS seen in FEE 6)
See R6_307 - attachment #6
01:30 BRIKEN and BigRIPS runs stopped. AIDA R6_317 at that time.
01:31 AIDA R6_318, BRIKEN run #33, BigRIPS #3033
All system-wide checks passed.
SYNC pulses 46 x 10^6
Memory 35 ~ 39k
02:08 Biases shown as 38.png, attachment #7
02:34 AIDA run R6_337 when BRIKEN and BigRIPS runs are stopped.
AIDA passes all system-wide checks.
02:37 AIDA just begins R6_338l. BRIKEN run #34, BigRIPS #3034
Timewarps look okay, see attachment #8, R6_336
Statistics consistent, see attachment #9, 39.png
Implant rate around 75 cps
03:31 AIDA run R6_354 at the stop of BRIKEN and BigRIPS runs
03:34 AIDA R6_355, BRIKEN run #35, BigRIPS #3035
AIDA passed all system checks. Merger shows low rate for nnaida8, but this is consistent for tonight.
4:03 Biases shown as 40.png, attachment #10.
4:08 Stats shown as 41.png, attachment #11
As before, only nnaida13 and nnaida21 show over 100k, and nnaida8 is quite low.
4:30 AIDA run R6_372 at the close of BRIKEN and BigRIPS runs.
4:32 AIDA R6_373 with new BRIKEN run #36 and BigRIPS #3036.
All system-wide checks okay.
5:38 AIDA R6_393 at the stop of BRIKEN and BigRIPS runs
All system-wide checks okay.
5:39 AIDA still R6_393, BRIKEN run #37 and BigRIPS run #3037
6:00 Leak currents shown as 42.png, attachment #12
6:16 Beam stopped so we can check the signal of one Ge detector that became noisy
R6_404 at that time
Added some Al foil to the Ge signal cables
6:28 Beam is back R6_408
6:29 BRIKEN run #41, BigRIPS #38
7:28 AIDA R6_426 at stop of other runs
System-wide checks are okay.
7:30 AIDA R6_427. BRIKEN #42, BigRIPS #39
Implant rates 60 ~ 70 cps in AIDA
One MBS timewarp in R6_426, as attachment #13
7:57 Leak currents shown as attachment #14, 43.png |
|
38
|
Wed Feb 18 05:49:29 2015 |
CG, AE, TD, PJC-S | Wed 18th Feb progress | 18/02/15
1300: Error seen when trying to check ASIC settings. See Attachment #1. Power cycle removed error.
nnaida3 still performing much better than other three cards connected to detector (nnaida1,6,8). Why?
Noticed nnaida3 was connected to only adapter PCB which did not feature a metal shielding plate on the front?
Noise in other cards caused by possible short or capacitance between PCB pins and shielding plate?
-> Removed all shielding plates and no difference seen. Pulser peak widths all still approximately the same.
nnaida3+6 PCB grounds connected by thin braid, as well as nnaida1+8. These were all removed. No change.
1830: Still seeing errors when data is sent through them Merger, see attachments #2-6.
Attached heavy duty copper braiding in loop, connecting all nnaida PCBs in use. Improved FWHM of nnaida1 by factor of 2. No change in other nnaida.
Moved kapton and bias from nnaida3 to nnaida11 (same crate, next module along) and saw no appreciable difference in the spectra.
Changed voltages in PSUs for first 16 nnaida.
+5V -> +5.5V, -6V -> -6.5V, +7V -> +8V.
All +5.5V rails at 5.5V, however in PSU#1 (supplying inner FEE modules) -6.5V and +8V rails show variations of +- ~12mV and +- ~4mV respectively, between outlets.
Installed PSUs with updated voltages.
nnaida8 FWHM improved by factor of 2 but no difference in other nnaida.
Current configuration nnaida1,6,8,11. nnaida1+8 connected to p+n side, nnaida6+11 connected to n+n side.
Current FWHM: nnaida1 ~500ch, nnaida6 ~1500ch, nnaida8 ~600ch, nnaida11 ~120ch.
Update response from Patrick:-
Please see posting number 39 for ideas to improve noise and possible reasons why some FEE64 are quieter than others.
I think the SYNC problem is likely due to the version of VHDL perhaps not including the flush of buffers and with the widely different data rates. 224k against 3k.
This should improve with the VHDL update as I think the flush system is more efficient.
|
|
648
|
Fri Jun 9 04:53:24 2017 |
TD | Waveforms |
x-scale 1000ch = 2ms
y-scale expanded
p+n junction strips
DSSSD #1 .. #6 attachments 1-12
n+n ohmic strips
DSSSD #1 .. #6 attachments 13-24
DSSSD #4 pulser off, attachments 25-28 |
|
689
|
Sat Oct 28 06:25:07 2017 |
TD | Waveforms |
For comparison with waveforms shown in https://elog.ph.ed.ac.uk/AIDA/665 and https://elog.ph.ed.ac.uk/AIDA/668
Note; nnaida21 ADC did not calibrate during setup therefore no waveforms for this FEE64
Current configuration
p+n junction strips FEE64 adaptor PCB rev D -> FEE64 adaptor PCB rev E
ac coupling 10n/100V/X7R -> 33n/100V/X7R
detector bias RC filter 3.3k/10n added
ASIC setting Ibias LF feedback 0x8 -> 0xf *except* nnaida2 ->0xc
n+n ohmic strips FEE64 adaptor PCB rev C *no change*
ASIC setting Ibias LF feedback 0xb *no change*
Attachments 1-5 waveforms for n+n ohmic strips
Attachment 1 - per FEE64 1.8.W, 20us range
Attachment 2 - per FEE64 1.8.W, 20us range, expanded y=scale
Attachment 3 - per FEE64 1.8.W, 200us range, expanded y=scale
Attachment 4 - per FEE64 1.8.W, 2ms range, expanded y=scale
Attachment 5 - per FEE64 1.8.W, 20ms range, expanded y=scale
Attachments 6-10 waveforms for p+n junction strips
Attachment 6 - per FEE64 1.8.W, 20us range
Attachment 7 - per FEE64 1.8.W, 20us range, expanded y=scale
Attachment 8 - per FEE64 1.8.W, 200us range, expanded y=scale
Attachment 9 - per FEE64 1.8.W, 2ms range, expanded y=scale
Attachment 10 - per FEE64 1.8.W, 20ms range, expanded y=scale |
|