Entry  Wed Nov 14 07:43:35 2018, TD, Wednesday 14 November 16:00 - 00:00 21x
16.44 DAQ continues OK - AIDA file RIBF148/R5_432
      165Pm BigRIPS setting + 2.8mm Al degrader

      ASIC settings 2018Oct03-13.26.58
       from c. 09.00 13.11.18
       n+n FEE64s Ibias LF feedback 0xb -> 0xd
       HEC fast comparator threshold 0x20 -> 0xff

      BNC PB-4 pulser 
      Amplitude 90,000, x5 attenuator IN
      t_r 50ns tau_d 1ms
      Frequency min
      Delay min
      
16.45 BigRIPS and BRIKEN stop
      AIDA on File R5_432

16.48 BigRIPS starts run 1047
      BRIKEN starts run 62
      AIDA on file R5_433

16.57 online MIDASsort

      m_p versus m_n (LEC) - attachment 1
      x versus y (LEC) - attachment 2
      E_p versus E_n (LEC) - attachment 3
      m_p versus m_n (HEC) - attachment 4
      x versus y (LEC) - attachment 5
      E_p versus E_n (LEC) - attachment 6

      all MIDASsort spectra zero'd

17.17 file RIBF128/R5_436 ADC data synchronised - attachment 7
      analysis of RIBF148/R5_436 - attachment 8
       max dead time nnaida19 ~0.25%

17.48 BigRIPS and BRIKEN stop
      AIDA on File R5_443

17.48 BigRIPS starts run 1048
      BRIKEN starts run 63
      AIDA on file R5_443

18.45 BigRIPS and BRIKEN stop
      AIDA on File R5_452

18.47 BigRIPS starts run 1049
      BRIKEN starts run 64
      AIDA on file R5_453

19.45 BigRIPS and BRIKEN stop
      AIDA on File R5_463

19.46 BigRIPS starts run 1050
      BRIKEN starts run 65
      AIDA on file R5_463

20.49 BigRIPS and BRIKEN stop
      AIDA on File R5_473

20.49 BigRIPS starts run 1051
      BRIKEN starts run 66
      AIDA on file R5_473

21.49 BigRIPS and BRIKEN stop
      AIDA on File R5_483

      ASIC check load

21.50 BigRIPS starts run 1052
      BRIKEN starts run 67
      AIDA on file R5_484

22.00 all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 9 & 10

      FEE64 temperatures OK - see attachment 11
      good event statistics OK - see attachments 12

      Merger OK - see attachment 13
      TapeServer OK - see attachment 14
      1x merger error message reported - trying to add items to Q index 12 (Link 12) but queue alreday full

      per FEE64 Rate spectra - attachment 15

      1.8.L spectra - attachments 16 & 17
       nnaida1 pulser peak width 73 ch FWHM
      1.8.H spectra - attachments 18 & 19
      1.8.W spectra 20ms FSR - attachments 20 & 21


22.48 BigRIPS and BRIKEN stop
      AIDA on File R5_493

22.50 BigRIPS starts run 1053
      BRIKEN starts run 68
      AIDA on file R5_493

23.24 BRIKEN DAQ found stopped - no sync check data - BigRIPS DAQ stopped and restarted
      AIDA on File R5_499

23.26 BigRIPS starts run 1054
      BRIKEN starts run 69
      AIDA on file R5_499


      
Entry  Tue Nov 13 22:58:49 2018, OH , Wednesday 14 November 08:00 - 16:00 30x
08:01 BigRips and BRIKEN stop
      AIDA on file R5_346

08:02 BigRIPS starts run 1038
      BRIKEN starts run 181114_0800_165Pm_053.dlt
      AIDA on file R5_346

      System wide checks ok * except nnaida 7 fails adc*
      FEE Temps ok - attachment 1
      Stats ok - attachment 2 
      Leakage currents ok - attachment 3

09:15 BigRIPS and BRIKEN stop
      AIDA on file R5_358

09:16 BigRIPS starts run 1039
      BRIKEN starts run 181114_0915_165Pm_054.dlt
      AIDA on file R5_358

10:15 BigRIPS and BRIKEN stop
      AIDA on file R5_368

10:15 BigRIPS starts run 1040
      BRIKEN starts run 181114_1015_165Pm_055.dlt
      AIDA on file R5_368

10:36 Analyser output of R5_368 - attachment 4

11:15 BugRIPS and BRIKEN stop
      AIDA on File R5_378

11:16 BigRIPS starts run 1041
      BRIKEN starts run
      AIDA on file R5_378



08.03 DAQ continues - AIDA file RIBF148/R5_384
      165Pm BigRIPS setting + 2.8mm Al degrader

      ASIC settings 2018Oct03-13.26.58
       from c. 09.00 13.11.18
       n+n FEE64s Ibias LF feedback 0xb -> 0xd
       HEC fast comparator threshold 0x20 -> 0xff
      
11.50 all system wide checks OK *except* nnaida7 fails ADC calibration

      detector biases & leakage currents OK - see attachments 5 & 6

      FEE64 temperatures OK - see attachment 7
      good event & AIDA discriminator statistics OK - see attachments 8

      Merger OK - see attachment 9
      TapeServer OK - see attachment 10
      no errors or warnings shown by server terminal sessions 

      per FEE64 Rate spectra - attachment 11
      per FEE64 HitRate spectraq - attachment 12
      1.8.L spectra - attachments 13-16
        some evidence of pulser peak shifts, notably nnaida5, 11 & 15
        nnaida1 pulser peak width 78 ch FWHM
      1.8.H spectra - attachments 17 & 18
 
      1.8.W spectra 20us FSR - attachments 19 & 23
      1.8.W spectra 200us FSR - attachments 20 & 24
      1.8.W spectra 2ms FSR - attachments 21 & 25
      1.8.W spectra 20ms FSR - attachments 22 & 26

12.15 BugRIPS and BRIKEN stop
      AIDA on File R5_388

12.16 BigRIPS starts run 1042
      BRIKEN starts run 57
      AIDA on file R5_388

12.22 *all* online and MIDASsort spectra zero'd

13:15 BigRIPS and BRIKEN stop
      AIDA on file R5_398

13:16 BigRIPS starts run 1043
      BRIKEN starts run58
      AIDA on file R5_398

13:28 BigRIPS and BRIKEN stop
      AIDA on file
      beam halted to give access to F11 area

13.44 beam started
      BigRIPS and BRIKEN start
      AIDA file RIBF148/R5_402
      BRIKEN run 59
      BigRIPS run 1043

14.45 BRIKEN and BigRIPS stop
      AIDA file RIBF148/R5_412

14.47 BRIKEN and BigRIPS start
      AIDA file RIBF148/R5_413
      BRIKEN run 60
      BigRIPS run 1044

15.43 BRIKEN and BigRIPS stop
      AIDA file RIBF148/R5_422

      BRIKEN Resets the timestamp

15:44 BRIKEN and BigRIPS start
      AIDA file RIBF148/R5_422
      BRIKEN run 61
      BigRIPS run 1045

16:02 System wide checks all ok *Except nnaida7 fails adc calibration*
      Leakage currents ok - attachment 27
      Stats ok - attachment 28
      FEE Temps ok - attachment 29

16:27 First 200ms of decay correlations in DSSD 3 and 4 for the 3 degrader settings - attachment 30
      DSSD 3 along the top row and DSSD 4 along the bottom.
      Dgrader settings are 3mm, 2.3mm and 2.8mm
      1 hour data in each
Entry  Tue Nov 13 14:27:38 2018, LS, Wednesday 14 November 00:00 to 08:00 9x
00:00 BRIKEN stop AIDA file RIBF148/R5_269 

---------------------------------------------------------------- 

00.02 BRIKEN start 
      BRIKEN file 45 
      BigRIPS file 1030 
      AIDA file RIBF148/R5_270 

00:07 system wide checks ok, except nnaida7 ADC calibration failure 
      FEE64 temperatures ok (attachment 1) 
      good event statistics ok (attachment 2) 
      detector bias/leakage currents ok (attachment 3) 

01:01 BRIKEN stop 
      AIDA file RIBF148/R5_279 

--------------------------------------------------------------- 

01.02 BRIKEN start 
      BRIKEN file 46 
      BigRIPS file 1031 
      AIDA file RIBF148/R5_279 

02:00 BRIKEN stop 
      AIDA file RIBF148/R5_289 

---------------------------------------------------------------- 

02.01 BRIKEN start 
      BRIKEN file 47 
      BigRIPS file 1032 
      AIDA file RIBF148/R5_289 

03:00 BRIKEN stop 
      AIDA file RIBF148/R5_298 

---------------------------------------------------------------- 

03.01 BRIKEN start 
      BRIKEN file 48 
      BigRIPS file 1033 
      AIDA file RIBF148/R5_298 

03:04 system wide checks ok, except nnaida7 ADC calibration failure 
      FEE64 temperatures ok (attachment 4) 
      good event statistics ok (attachment 5) 
      detector bias/leakage currents ok (attachment 6) 

04:00 BRIKEN stop 
      AIDA file RIBF148/R5_308 

----------------------------------------------------------------

04.03 BRIKEN start 
      BRIKEN file 49 
      BigRIPS file 1034 
      AIDA file RIBF148/R5_308 

05:00 BRIKEN stop 
      AIDA file RIBF148/R5_317 

---------------------------------------------------------------- 

05.01 BRIKEN start 
      BRIKEN file 50 
      BigRIPS file 1035 
      AIDA file RIBF148/R5_318 

06:00 BRIKEN stop 
      AIDA file RIBF148/R5_327 

---------------------------------------------------------------- 

06.02 BRIKEN start 
      BRIKEN file 51 
      BigRIPS file 1036
      AIDA file RIBF148/R5_327 

06:07 system wide checks ok, except nnaida7 ADC calibration failure 
      FEE64 temperatures ok (attachment 7) 
      good event statistics ok (attachment 8) 
      detector bias/leakage currents ok (attachment 9)

07:00 BRIKEN stop 
      AIDA file RIBF148/R5_337 

---------------------------------------------------------------- 

07.01 BRIKEN start 
      BRIKEN file 52 
      BigRIPS file 1037
      AIDA file RIBF148/R5_327 
Entry  Wed Jun 14 07:01:34 2017, TD, Wednesday 14 June 21x
15.01 DAQ continues
      AIDA file RIBF148/R8_22

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency 25Hz range 2.6Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Mar27-16.12.39
      *except* 
      slow comparator 0x64
      LEC/MEC fast comparator 0x14

15.02 Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      good events statistics stable & OK - see attachment 4 
      No warnings/errors reported by merger since start of RIBF148/R3  - see attachment 5
      *all* system wide checks OK (zero fails)

15.12 online analysis 
      ADC & disc data synchronised - note the effect of the higher slow comparator threshold on the after-pulse
- see attachment 6
      per DSSSD m_p versus m_n (LEC) - see attachments 7-8
      per DSSSD x versus y (LEC) - see attachment 9
      per DSSSD E_p versus E_n (LEC) - 512x512 spectra - nominal 20keV per channel - see attachment 10

15.30 per FEE64 1.8.* spectra - see attachments 11-18
      note all waveforms x-scale 1000ch = 2ms
      note absence of saw tooth transients/noise p+n waveforms - attachments 15-16
      Brief discussion with Nishimura-san and Gabor - clover detectors have been removed from BRIKEN
      otherwise no major changes in last 24h
      see also - attachments 17-18

15.42 Ambient temperature +24.7 deg C, d.p. +9.4 deg C, RH 37.8%
      Julabo FL11006 recirculating chiller set point +20.0deg C , actual +20.0 deg C, water level c. 75% 

15.54 analysis file RIBF148/R8_22 - zero deadtime, 1x MBS & zero ts timewarps - see attachment 19
Entry  Wed Oct 12 02:58:19 2016, TD & CG, Wednesday 12 October 1.png2.png3.png4.png5.png
10.50 Merger & TapeServer found stopped
      connect errors - see attachment 1

      DSSSD bias and leakage currents OK - see attachments 2 & 3

      Most recent data file R17_440 last accessed 08.22 12.10.16

      R17_440 ADC data synchronised

aidas1> ./analyser2 /TapeData/Sept2016/R17_440
 *** GREAT format 3.2.0 analyser - TD - May 2014
 *** ERROR: READ I/O error:       5002
                   blocks:      30546
          ADC data format:  147630891 ( 1189229.1 Hz)
        Other data format:  101593923 (  818381.9 Hz)
 Sample trace data format:          0 (       0.0 Hz)
         Undefined format:          0 (       0.0 Hz)
   Other data format type:      PAUSE:        101 (       0.8 Hz)
                               RESUME:        101 (       0.8 Hz)
                              SYNC100:    1133804 (    9133.3 Hz)
                           FEE64 disc:  100456290 (  809217.8 Hz)
                             MBS info:       3627 (      29.2 Hz)
                           Other info:          0 (       0.0 Hz)

   ADC data range bit set:        151 (       1.2 Hz)

                Timewarps:        ADC:          0 (       0.0 Hz)
                                PAUSE:          0 (       0.0 Hz)
                               RESUME:          0 (       0.0 Hz)
                              SYNC100:          0 (       0.0 Hz)
                           FEE64 disc:          0 (       0.0 Hz)
                             MBS info:          0 (       0.0 Hz)
                            Undefined:          0 (       0.0 Hz)
                         Sample trace:          0 (       0.0 Hz)

 Timestamp elapsed time:      124.140 s
 FEE module #:   1 elapsed dead time       0.728 s
 FEE module #:   2 elapsed dead time       0.184 s
 FEE module #:   3 elapsed dead time       0.000 s
 FEE module #:   4 elapsed dead time       0.000 s
 FEE module #:   5 elapsed dead time       1.034 s
 FEE module #:   6 elapsed dead time       3.462 s
 FEE module #:   7 elapsed dead time       0.000 s
 FEE module #:   8 elapsed dead time       0.000 s
 FEE module #:   9 elapsed dead time       0.064 s
 FEE module #:  10 elapsed dead time       0.003 s
 FEE module #:  11 elapsed dead time       0.000 s
 FEE module #:  12 elapsed dead time       0.000 s
 FEE module #:  13 elapsed dead time       0.000 s
 FEE module #:  14 elapsed dead time       0.062 s
 FEE module #:  15 elapsed dead time       0.018 s
 FEE module #:  16 elapsed dead time       0.005 s
 FEE module #:  17 elapsed dead time       0.817 s
 FEE module #:  18 elapsed dead time       0.035 s
 FEE module #:  19 elapsed dead time       0.048 s
 FEE module #:  20 elapsed dead time       0.000 s
 FEE module #:  21 elapsed dead time       0.000 s
 FEE module #:  22 elapsed dead time       0.103 s
 FEE module #:  23 elapsed dead time       0.216 s
 FEE module #:  24 elapsed dead time       0.000 s
 FEE module #:  25 elapsed dead time       0.000 s
 FEE module #:  26 elapsed dead time       0.000 s
 FEE module #:  27 elapsed dead time       0.000 s
 FEE module #:  28 elapsed dead time       0.000 s
 FEE module #:  29 elapsed dead time       0.000 s
 FEE module #:  30 elapsed dead time       0.000 s
 FEE module #:  31 elapsed dead time       0.000 s
 FEE module #:  32 elapsed dead time       0.000 s

 *** Program elapsed time:   14.129s ( 2161.951 blocks/s, 135.122 Mb/s)

Conclude DAQ ran without problem for c. 16.5h which is a significant improvement
assuming that it can be consistently reproduced.

11.20 Merger restart did not clear connect/timeout issues 

      Re-route BNC - BNC RG58 cable with 50MHz clock from BRIKEN to
      AIDA Master MACB external clock input (rear panel) so that it will
      not be stepped upon. 

      Power cycle AIDA FEE64s

      N.B. I selected 'Switch ALL On' rather than 'Sequence ALL On' by mistake and one
      of the USB ac mains controller output 250VAC/16A fuses blew - use 'Sequence ALL On'

      TS_SYNC_PHASE 0

      FEE64 temperatures OK
      *All* system-wide checks OK

      B2F/F11 temperature +24.6 deg C, d.p. +8.9 deg C,RH 36.6%

      ASIC settings 2016Sep02-18.17.01
      slow comparator 0x10 -> 0x0a
      LEC/MEC fast comparators # 1, 55 & 58 disabled for all FEE64s

      Master MACB serial #13 (see https://elog.ph.ed.ac.uk/AIDA/367)
      Master MACB mode 5

14.05 DAQ starts (file Sept2016/R21)

15.30 Sequence of three electrical power dips over 30 minute period. RIKEN public address
      system reports that TEPCO (Tokyo Electric Power Company) - the site supplier - is
      having region-wide problems.

      The third power dip stopped DAQ.

      Unable to recover by restarting merger

      Power cycle FEE64s

16.20 DAQ restarts (file Sept2016/R22)

19.05 DAQ stops
      
      Because of risk of power outages due to TEPCO distribution problems we have powered
      down AIDA and Julabo FL1006 chiller and disconnected from ac mains.
Entry  Wed Jun 12 01:54:04 2019, OH, TD, CG, CB, CA, Wednesday 12 June 11x
09:54 DAQ Continues ok with merger running at ~ 560k data items per second

      ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range, variable min. c. 2.5Hz
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      System wide checks all ok - except nnaida7 fails adc calibration

      Bias and leakage currents ok - attachment 1

      FEE Temps ok - attachment 2

      Good event statistics good - attachment 3

      Merger and tape service good - attachment 4 and 5


16:33  DAQ continues OK with merger running at ~ 600k data items per second

     ASIC settings 2019May19-17.54.03
      slow comparator 0xa

      BNC PB-4 Pulser 
      Amplitude 90,000
      x5 attenuator IN
      Frequency 25Hz range, variable min. c. 2.5Hz
      t_r 100ns, tau_d 50us
      - polarity
      Delay MIN, tail pulse

      All system-wide checks OK except nnaida7 fails ADC calibration

      Bias and leakage currents OK - attachment 6

      FEE temperatures OK - attach 7

      Good event statistics OK - attach 8
      check nnaida10 & nnaida20 counts vs. rate

      Merger and tape service OK - attachment 9 & 10

17.00 Average FEE64 rates since last DAQ stats zero @ 17.05 yesterday - data from attachment 11

      DSSSD#1   count      rate  
                           24h   most recent
      nnaida17 	1914894752 22.1k 23.2k
      nnaida20  3061905664 35.4k 35.6k
      nnaida22  1872820000 21.7k 21.8k
      nnaida23  2217883584 25.7k 24.8k  

      DSSSD#2

      nnaida18  1760374528 20.3k 21.1k
      nnaida19  1402771232 16.2k 66.6k => 32 bit overflow => 4.3E+09 + 1.4E+09 = 5.6E+09 => 65.9k
      nnaida21  2038432092 23.6k 23.5k
      nnaida24  2844103808 32.9k 31.3k

      Conclude rates have bee stable overnight - no evidence of noise transients for DSSSD#1 to 150-250k good
events/s

17.15 DAQ stats zero'd
      all spectra zero'd
Entry  Wed Oct 11 16:53:20 2023, TD, Wednesday 11 October contd. 14x
      CAEN N1419ET ch 2 - all outputs non-floating ( internal LK fitted )
      Standard coaxial HV cable chained to aida12, aida03, aida11, aida09, aida01 & aida10
      aida04 LK1 ( local gnd ) fitted

      ground cable from aida04 adaptor PCB ground to aida12 to aida02

      aida03 LK3 ( PCB gnd ) fitted

      BNC PB-5 
      amplitude 10.0V
      attenuation x10
      decay time 1ms
      polarity +
      frequency 25Hz


      test + direct from PB-5 to aida12 *only*

      slow comparator 0xf

      DSSSD bias -150V -7.280uA

      aida12 1.8.L pulser peak width 55 ch FWHM

      aida12 *.*.L spectra - attachments 1-4
      All channels have similar pulser peak widths to 1.8.L *except* 1.1.L, 1.4.L, 4.7.L & 4.13.L ( hot channels ) and 2.2.L, 2.9.L & 2.12.L no pulser peak



      test + direct from PB-5 to aida01 *only*

      aida01 1.8.L pulser peak width 70 ch FWHM

      aida01 *.*.L spectra - attachments 5-8
      All channels have similar pulser peak widths to 1.8.L *except* 1.1.L, 1.4.L, 4.7.L & 4.13.L ( hot channels )


      per FEE64 1.8.W spectra - 20us FSR - attachments 9-12
      per FEE64 1.8.W spectra - 200us FSR - attachments 13 & 14
Entry  Wed Oct 11 09:37:30 2023, TD, Wednesday 11 October 48x
10.00 Water flow & temp OK
      Power ON

      All system wide checks OK

      CAEN N1419ET ch 2 - all outputs non-floating ( internal LK fitted )
      Standard coaxial HV cable chained to aida12, aida03 & aida11
      aida04 LK1 ( local gnd ) fitted

      ground cable from aida04 adaptor PCB ground to aida12 to aida02

      aida03 LK3 ( PCB gnd ) fitted

      BNC PB-5 
      amplitude 10.0V
      attenuation x10
      decay time 1ms
      polarity -
      frequency 25Hz

      test + cable and test + chain removed
      test - direct from PB-5 to aida02 *only*

      DSSSD bias -120V -6.105uA - attachment 1

      FEE64 tempweratures OK - attachment 2

      slow comparator 0x64

      ADC data item stats - attachment 3
      Note 
       aida04 asic #1 u/s
        ASIC mezzanine to be replaced
       aida10 asic #4 v. high rates observed
        see https://elog.ph.ed.ac.uk/AIDA/901 attachment 45
        ASIC/cable/wafer issue ?

      rate ~ 1600Hz observed for aida02 - pulser
      All other rates 0Hz

      per FEE64 Rate spectra - attachment 4

      aida02 1.8.L pulser peak width 137 ch FWHM


      1.8.W spetcra 20us, 200us & 2ms FSR - attachments 5-10


      slow comparator 0x64 -> 0xf 

      ADC data item stats - attachment 11
      per FEE64 Rate spectra - attachments 12 & 13
       p+n rates entirely dominated by hot channels, otherwise rates ~0Hz
       n+n rates ~ 200-300Hz/channel

      slow comparator 0xf -> 0xa

      ADC data item stats - attachment 14
      per FEE64 Rate spectra - attachments 15-17
       p+n rates aida12, aida03 & aida11 entirely dominated by hot channels, otherwise rates ~0Hz
       p+n rates aida10, aida01 & aida09 ~500-1000Hz/channel
       n+n rates ~ 2000-3000Hz/channel

11.25 Extend HV chain from aida11 to aida09

      ADC data item stats - attachment 18
       significant decrease in aida09 rate

      1.8.W spectra 2ms FSR - attachments 19 & 20
       significant improvement in aida09 1.8.W noise

      aida02 1.8.L pulser peak width 165 ch FWHM ( slow comparator 0xa ), 127 ch FWHM ( 0xf )


12.00 Extend HV chain from aida09 to aida01 & aida10

      slow comparator 0xf

      ADC data item stats - attachment 21

      per FEE64 Rate spectra - attachment 22

      1.8.W spectra 2ms FSR - attachments 23-26
       significant improvement in aida01 & aida10 1.8.W noise

      aida02 1.8.L pulser peak width 127 ch FWHM ( 0xf )


12.35 slow comparator 0xf
      
      ADC data item stats - attachment 27
      per FEE64 Rate spectra - attachment 28




      slow comparator 0xa
      
      ADC data item stats - attachment 29
      per FEE64 Rate spectra - attachment 30



      slow comparator 0x9
      
      ADC data item stats - attachment 31
      per FEE64 Rate spectra - attachment 32 & 33



      slow comparator 0x8
      
      ADC data item stats - attachment 34
      per FEE64 Rate spectra - attachment 35



      slow comparator 0x7
      
      ADC data item stats - attachment 36
      per FEE64 Rate spectra - attachment 37



      slow comparator 0x6
      
      ADC data item stats - attachment 38
      per FEE64 Rate spectra - attachment 39



      slow comparator 0x5
      
      ADC data item stats - attachment 40
      per FEE64 Rate spectra - attachment 41



16.55 Background alpha 3*L & 3*W ( 20us FSR ) spectra - attachments 42 & 43
      Slow comparator 0xf
      Waveform threshold 7000 ( - polarity ), 9000 ( + polarity )

17.00 DSSSD bias -120V -7.125uA - attachment 44

      aida02 1.8.L ( n+n Ohmic side strip ) pulser peak width as function of DSSSD bias 

      Bias I_l    FWHM  Centroid
      (V)  (uA)   (ch)  (ch)
      -120 -7.125 124.8 43090
      -140 -7.230 124.5 43092
      -160 -7.370 120.2 43093
      -180 -7.570 120.0 43095
      -100 -7.020 143.8 43089
       -80 -6.820 continuum/no peak
      -150 -7.300 122.6 43093


      DSSSD bias -150V -7.280uA

      aida02 *.*.L spectra - attachments 45-48
      All channels have similar pulser peak widths to 1.8.L
 








      
Entry  Wed May 11 06:37:32 2016, AE, VP, BM, Wednesday 11 May Eurica_timestamp.pngAIDA_timestamp.pngTimestamp_correlation_AIDA_EURICA.png
*2:30 pm*: changed power source of USB-controlled power supply for FEEs. Now one of the power cables is connected
to the 'B' power supply of F11, and the second to the 'C' power supply (these are in the unit on top of WASBI
crates). The reason for change is that before both were connected to 'C', and drained current was very close to
limit  of 10 A (and power cord getting noticeable hot). We'll check resolution after this switch.

Detectors 1, 2, 3, and 4 biased to 100V, detectors 5, 6, and veto biased to 70V.

n-side is more noisy than p-side, but resolution comparable to previous configuration:

NNAIDA#   FWHM [ch]
    1     91
    2     61
    3     84
    4     -
    5    160
    6     -
    7     79
    8     84
    9     71
   10     76
   11    101
   12     81
   13     69
   14     78
   15    199
   16    107
   17     82
   18     62
   19     -
   20     89
   21    168
   22     70
   23     80
   24     73
   25     81
   26     13
   27    109
   28     13
   29     15
   30    126
   31     13
   32    100



*5 pm*: Set-up correlation scalers for AIDA<->EURICA<->BigRIPS (detail diagram coming soon, now only in paper
logbook). 


*6:30 pm*: Started EURICA DAQ with Pieter and Gabor.

A folder is created to record data of June campaign: /raid/11/mbs/2016_ucampaign/

Tested correlation of AIDA and EURICA with a pulser run, triggering EURICA and AIDA correlation scaler with 250
Hz. The corresponding files are

/raid/11/mbs/2016_ucampaign/test0003.lmd
/TapeData/may2016/R12_0.gz

Phong did a quick analysis and the pattern of correlation scalers in both files is similar (attached figures). It
is also consistent with the number of reset requests we sent during the data taking (three requests closely spaced
in the middle of run, and two more in the end of the run).
Entry  Wed May 10 02:55:13 2017, TD, OH, Wednesday 10 May 8x
10.55 DAQ continues OK 
      Current file May2017/R6_44

      ASIC settings 2017Mar27-16.12.39 *except*
      slow comparator 0x20
      LEC/MEC fast comparator 0xff

      Detector biases & leakage currents OK - see attachments 1 & 2
      FEE64 Temperatures OK - see attachment 3
      'good events' statistics OK - see attachment 4
      No warnings/errors reported by merger since DAQ GO - see attachment 5

      *all* system wide checks OK

10.48 E17/F11 ambient temperature +24.7 deg C, RH 35.3%, d.p. +8.3 deg C
      Julabo FL11006 set point +20 deg C, actual +20.0 deg C, water level OK (c. 65%)

18.32 DAQ continues OK 
      Current file May2017/R6_67

      Detector biases & leakage currents OK - see attachments 6 & 7
      FEE64 Temperatures OK - see attachment 8
      'good events' statistics OK - see attachment 9

      *all* system wide checks OK
Entry  Wed Nov 1 02:24:20 2017, OH, Wednesday 1 November 011117_1120_bias.png011117_1122_feetemps.png011117_1123_stats.png
11:24 DAQ running OK - data transfer not enabled

      BNC PB-4
      Amplitude 90,000
      x5 attenuator IN
      t_r 100ns tau_d 50us
      - polarity
      frequency range 25Hz
      delay MIN, tail pulse, INT ref

      ASIC settings 2017Oct23-14.38.03
      *all* FEE64s slow comparators 0xa *except* nnaida13 0xe, nnaida19-22 0xc

      Detector biases & leakage currents OK - see attachments 1
      FEE64 Temperatures OK - see attachment 2
      DAQ statistics OK - see attachment 3
      *all* system wide checks OK *except* nnaida21 fails ADC calibration
Entry  Wed Jun 1 01:33:58 2016, TD, Wednesday 1 June 2016 9x
09.30 Detector biases and leakage currents - ambient temp +24.9 deg C - attachments 1-2

10.08 Connection lost to nnaida27 and nnaida28 
      Most recent save of settings DB key 2016May31-23.32.46

12.00 Replaced nnaida17 & nnaida18 with FEE module previously used for nnaida31 & nnaida32

      Pulser walkthrough

      BNC PB-4 settings 

      90,000 - 10,000 @ 10,000 step
      x5 attenuator (was x10 previously)
      t_r = 50ns tau_d = 50us
      polarity -
      rate ~ 150Hz
      INT, delay MIN

      polarity + by EG&G Ortec 433A

      N.B. - polarity terminated at end of AIDA adaptor chain, + polarity not terminated

      Spectra saved 12.32 ( to directories tagged around June 1, 2016 05.29 )

      BNC PB4 pulser rate set to ~5Hz

      207Bi source at front of snout and 2x 60Co sources above DSSD3

14.00 Rebooted aidas1 to cleanup merge 
      
      Changed config to 24x FEE64s 

      Clock issues with nnaida23 - clock not locked status 0x6 - see attachments 3-9

18.00 Replace current nnaida19 (no histogramming, reSYNCs OK but timestamp not sync'd wrt timestamps of
      other FEE64s).
      Replace current nnaida23

      Following powerup obtained clock not locked - status 0x6 for nnaida5 (master)

      Replacement nnaida19 also does not histogram, reSYNCs OK but timestamp not sync'd wrt timestamps
      of other FEE64s)

      With help from PJCS investigate above issues and change timestamp master from nnaida5 to nnaida6
      ( /MIDAS/config/HttpdTcl/aidas1/startup.tcl and /MIDAS/config/HttpdTcl/nnaida5/startup.tcl ) to 
      change definition of timestamp master.

      Able to operate DAQ and obtain AIDA SYNCs and WAVE SYNCs with TS_SYNC 0x0 _only_ 
      PJCS will also investigate using system in T9 at STFC DL

      Restored settings 2016June1-17.27.?
    Reply  Wed Jun 1 09:00:13 2016, TD - Reply PJCS, Wednesday 1 June 2016 
> 09.30 Detector biases and leakage currents - ambient temp +24.9 deg C - attachments 1-2
> 
> 10.08 Connection lost to nnaida27 and nnaida28 
>       Most recent save of settings DB key 2016May31-23.32.46
> 
> 12.00 Replaced nnaida17 & nnaida18 with FEE module previously used for nnaida31 & nnaida32
> 
>       Pulser walkthrough
> 
>       BNC PB-4 settings 
> 
>       90,000 - 10,000 @ 10,000 step
>       x5 attenuator (was x10 previously)
>       t_r = 50ns tau_d = 50us
>       polarity -
>       rate ~ 150Hz
>       INT, delay MIN
> 
>       polarity + by EG&G Ortec 433A
> 
>       N.B. - polarity terminated at end of AIDA adaptor chain, + polarity not terminated
> 
>       Spectra saved 12.32 ( to directories tagged around June 1, 2016 05.29 )
> 
>       BNC PB4 pulser rate set to ~5Hz
> 
>       207Bi source at front of snout and 2x 60Co sources above DSSD3
> 
> 14.00 Rebooted aidas1 to cleanup merge 
>       
>       Changed config to 24 FEE64s 
> 
>       Clock issues with nnaida23 - clock not locked status 0x6 - see attachments 3-9

Check the Local Controls for the setting of the LMK3200 control rgister.
If you recall from the problem with nnaida5 the value should be 0xB.
If the clock still doesn't lock then maybe it's an HDMI cable problem like nnaida5.
Since the other FEE64 are ok then it is likely to be just the MACB to nnaida23.
    Reply  Wed Jun 1 13:58:48 2016, TD - Reply PJCS, Wednesday 1 June 2016 500.png501.png503.png505.png
> Check the Local Controls for the setting of the LMK3200 control rgister.
> If you recall from the problem with nnaida5 the value should be 0xB.
> If the clock still doesn't lock then maybe it's an HDMI cable problem like nnaida5.
> Since the other FEE64 are ok then it is likely to be just the MACB to nnaida23.

I attach SYNC stats for WAVE and AIDA - you will notice that most (but not all) FEE64s
produce WAVE SYNCS

Plus (load balanced) statistics for disc, good events
Entry  Tue Jun 18 23:58:46 2019, CB, Wed Jun 19 8:00-16:00  19x
08:00 Shift change
      No change in settings

      DAQ continues running OK
      Good events rate OK - see attach 1
      FEE64 temperature OK - see attach 2
      Merger & Tape server OK - see 3 & 4
      Leakage currents OK - see attach 5
      
      Online HEC xy plot OK, most implants in DSSD2 - see attach 6
      Zeroed online histograms

09:29 DTAS DAQ stops on run 135
      AIDA on run R2_787

09:32 DTAS DAQ restarts on run 136
      AIDA on run R2_787
      BigRIPS on run 2053

09:33 DAQ continues OK
      Good events rate OK - see attach 7
      Merger ~ 700k items/s
      Tape ~ 5400 Kbytes/s
      Leakage currents OK - see attach 8

      Online HEC xy plot OK - see attach 9

10:31 Good event rate in nnaida 19, 24 steadily creeping up. See attach 10.
      Currently 160k and 120k
      Merger ~ 710 k items/s
      Tape ~ 5689 kbytes/s
      All system-wide checks OK, except nnaida7 fails ADC calibration as usual

11:20 Good event rate stable - see attach 11
      FEE64 temperatures OK - see attach 12
      Merger ~700k items/s
      Tape server ~ 5500 Kbytes/s
      Leakage currents OK - see attach 13 & 14

11:30  DTAS stops DAQ on run 139
       BigRIPS on run 2053
       AIDA on run R2_806
       
11:32  DTAS restarts on run 140
       BigRIPS 2053
       AIDA R2_807

13:14  DAQ continues OK
       Good statistics events stable - see attach 15
       Merger ~ 730k items/s
       Tape server ~ 5500 KBytes/s
       Leakage current OK - see attach 16

13:34  DTAS stops DAQ on AIDA R2_826

13:37  DTAS restarts DAQ on run 144
       BigRIPS run 2055
       AIDA run 827

15:10  DTAS DAQ reports 20% drop in F11 counting rate
       AIDA plastic unchanged, beam current unchanged. Focusing issue?

       Reset online histograms to check we are still implanting. We are.

       Acquisition continues uninterrupted.

15:17  Good statistics OK - see attach 17
       FEE64 temperature OK - see attach 18
       Merger ~ 700k items/s
       TapeServer ~ 5500 Kbytes/s
       Detector leakage current OK - see attach 19

15:30  DTAS stops DAQ on AIDA run R2_845

15:33  DTAS restarts DAQ on run 147
       BigRIPS run 2056
       AIDA run R2_846 
       
Entry  Wed Jun 19 08:30:36 2019, CG, Wed Jun 19 16.00 - end of run AIDAimplants.png
16.10 DAQ continues as before.
      Merger rate ~700k items/s and ~5.3Kb/s
      Individual FEE rates remain roughly constant
      Temps and bias ok.

17.33 DTAS run 149 stopped

      DTAS run 150 started
      BigRIPS run 2057 
      AIDA run ~1.3GB in R2_865

19.30 DTAS run 152 stopped

      DTAS run 153 started
      BigRIPS run 2058
      AIDA run ~1.8GB in R2_884

      Temps and bias ok.
      Rates in merger, tape server and individual FEEs as before.
      
19.50 In looking at previous data, AIDA is seeing ~30% fewer implants that what is observed at F11Plastic
      Beam spot is not central on AIDA and some ions will be lost over sides of DSSD, but does not look
      like it would account for a ~30% reduction
      AIDA implant profile shown in Attachment 1 - from file
merged_a2_687_708_br190618_2315_100Sn118_121bgrun2049_sw_Tsi.root

21.01 Beam stopped.
      DTAS run 155 stopped.
      AIDA file R2_899.

      DTAS and AIDA will continue with background measurements.

21.04 DTAS run 156 started
      AIDA run R2_899
      
      Chiller is full.
      Lead bricks upstream of moderator.

      That's all folks.
Entry  Tue Jun 6 15:59:19 2017, DK, Wed 7 June night shift 14x
7 June 2017   

00:05  All system-wide checks are passed
       Sync pulses nera 44 x 10^6
       Memory 35k ~ 38k
       Biases uploaded as 33.png, 34.png - attachments 1 & 2
          Entered into spreadsheet
          SSD 3 bias is going up slowly since this isotope setting began
       Around "2 to 4" Hz in SSDs 3 and 4 -> 50 to 80 pps implant rate
          See attachment 3, 35.png
       All rates under 100k except nnaida 13, 21
          See attachment 4, 36.png
       FEE temperatures shown as 37.png, attachment #5

00:30  BRIKEN number 32, AIDA R6_299, BigRIPS 3032

01:01  Timewarps look okay (only 2 MBS seen in FEE 6)
          See R6_307 - attachment #6

01:30  BRIKEN and BigRIPS runs stopped.  AIDA R6_317 at that time.

01:31  AIDA R6_318, BRIKEN run #33, BigRIPS #3033

       All system-wide checks passed.
       SYNC pulses 46 x 10^6
       Memory 35 ~ 39k

02:08  Biases shown as 38.png, attachment #7

02:34  AIDA run R6_337 when BRIKEN and BigRIPS runs are stopped.

AIDA passes all system-wide checks.

02:37  AIDA just begins R6_338l.  BRIKEN run #34, BigRIPS #3034

Timewarps look okay, see attachment #8, R6_336
Statistics consistent, see attachment #9, 39.png
Implant rate around 75 cps

03:31  AIDA run R6_354 at the stop of BRIKEN and BigRIPS runs


03:34  AIDA R6_355, BRIKEN run #35, BigRIPS #3035

AIDA passed all system checks.  Merger shows low rate for nnaida8, but this is consistent for tonight.

4:03 Biases shown as 40.png, attachment #10.

4:08 Stats shown as 41.png, attachment #11
     As before, only nnaida13 and nnaida21 show over 100k, and nnaida8 is quite low.

4:30 AIDA run R6_372 at the close of BRIKEN and BigRIPS runs.

4:32 AIDA R6_373 with new BRIKEN run #36 and BigRIPS #3036.
     All system-wide checks okay.

5:38 AIDA R6_393 at the stop of BRIKEN and BigRIPS runs
     All system-wide checks okay.

5:39 AIDA still R6_393, BRIKEN run #37 and BigRIPS run #3037

6:00 Leak currents shown as 42.png, attachment #12

6:16 Beam stopped so we can check the signal of one Ge detector that became noisy
     R6_404 at that time
     Added some Al foil to the Ge signal cables

6:28 Beam is back R6_408

6:29 BRIKEN run #41, BigRIPS #38

7:28 AIDA R6_426 at stop of other runs
     System-wide checks are okay.

7:30 AIDA R6_427.  BRIKEN #42, BigRIPS #39
     Implant rates 60 ~ 70 cps in AIDA
     One MBS timewarp in R6_426, as attachment #13

7:57 Leak currents shown as attachment #14, 43.png
Entry  Wed Feb 18 05:49:29 2015, CG, AE, TD, PJC-S, Wed 18th Feb progress 15x

18/02/15

1300: Error seen when trying to check ASIC settings. See Attachment #1. Power cycle removed error.

           nnaida3 still performing much better than other three cards connected to detector (nnaida1,6,8). Why?

           Noticed nnaida3 was connected to only adapter PCB which did not feature a metal shielding plate on the front?

           Noise in other cards caused by possible short or capacitance between PCB pins and shielding plate?

           -> Removed all shielding plates and no difference seen. Pulser peak widths all still approximately the same.

           nnaida3+6 PCB grounds connected by thin braid, as well as nnaida1+8. These were all removed. No change.

 

1830: Still seeing errors when data is sent through them Merger, see attachments #2-6.

            Attached heavy duty copper braiding in loop, connecting all nnaida PCBs in use. Improved FWHM of nnaida1 by factor of 2. No change in other nnaida.

            Moved kapton and bias from nnaida3 to nnaida11 (same crate, next module along) and saw no appreciable difference in the spectra.

            Changed voltages in PSUs for first 16 nnaida.

           +5V -> +5.5V, -6V -> -6.5V, +7V -> +8V.

           All +5.5V rails at 5.5V, however in PSU#1 (supplying inner FEE modules) -6.5V and +8V rails show variations of +- ~12mV and +- ~4mV respectively, between outlets.

           Installed PSUs with updated voltages.

           nnaida8 FWHM improved by factor of 2 but no difference in other nnaida.

          Current configuration nnaida1,6,8,11. nnaida1+8 connected to p+n side, nnaida6+11 connected to n+n side.

          Current FWHM: nnaida1 ~500ch, nnaida6 ~1500ch, nnaida8 ~600ch, nnaida11 ~120ch.

 Update response from Patrick:-

Please see posting number 39 for ideas to improve noise and possible reasons why some FEE64 are quieter than others.

I think the SYNC problem is likely due to the version of VHDL perhaps not including the flush of buffers and with the widely different data rates. 224k against 3k. 

This should improve with the VHDL update as I think the flush system is more efficient.

 

 

Entry  Fri Jun 9 04:53:24 2017, TD, Waveforms 28x
x-scale 1000ch = 2ms
y-scale expanded

p+n junction strips

DSSSD #1 .. #6 attachments 1-12

n+n ohmic strips

DSSSD #1 .. #6 attachments 13-24

DSSSD #4 pulser off, attachments 25-28
Entry  Sat Oct 28 06:25:07 2017, TD, Waveforms 10x
For comparison with waveforms shown in https://elog.ph.ed.ac.uk/AIDA/665 and https://elog.ph.ed.ac.uk/AIDA/668

Note; nnaida21 ADC did not calibrate during setup therefore no waveforms for this FEE64

Current configuration

p+n junction strips FEE64 adaptor PCB rev D -> FEE64 adaptor PCB rev E
ac coupling 10n/100V/X7R -> 33n/100V/X7R
detector bias RC filter 3.3k/10n added
ASIC setting Ibias LF feedback 0x8 -> 0xf *except* nnaida2 ->0xc

n+n ohmic strips FEE64 adaptor PCB rev C *no change*
ASIC setting Ibias LF feedback 0xb *no change*


Attachments 1-5 waveforms for n+n ohmic strips

Attachment 1 - per FEE64 1.8.W, 20us range
Attachment 2 - per FEE64 1.8.W, 20us range, expanded y=scale
Attachment 3 - per FEE64 1.8.W, 200us range, expanded y=scale
Attachment 4 - per FEE64 1.8.W, 2ms range, expanded y=scale
Attachment 5 - per FEE64 1.8.W, 20ms range, expanded y=scale

Attachments 6-10 waveforms for p+n junction strips

Attachment 6 - per FEE64 1.8.W, 20us range
Attachment 7 - per FEE64 1.8.W, 20us range, expanded y=scale
Attachment 8 - per FEE64 1.8.W, 200us range, expanded y=scale
Attachment 9 - per FEE64 1.8.W, 2ms range, expanded y=scale
Attachment 10 - per FEE64 1.8.W, 20ms range, expanded y=scale