804
|
Sat May 18 04:47:22 2019 |
TD | Saturday 18 May | 12.47 DAQ continues OK to file May19/R7_3
alpha background
ASIC settings 2018Oct03-13.26.58
*except* slow comparator 0xa -> 0x64
BNC PB-4 Pulser OFF
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachments 4
Merger OK - see attachment 5
TapeServer OK - see attachment 6
12.51 Check ASIC Control *all* ASICs & FEE64s
1.8.L spectra - see attachments 7 & 8
p+n FEE64s 1.8.W spectra 20us, 200us, 2ms, 20ms FSR - attachments 9-12
n+n FEE64s 1.8.W spectra 20us, 200us, 2ms, 20ms FSR - attachments 13-16
13.13 analysis of May19/R7_3 - see attachment 17
deadtime OK
ADC data for FEE64s nnaida5 & nnaida6 hot channels (Stat spectrum channel # 104)
Also observed at a much lower rate for nnaida1 & nnaida2
13.27 Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C - water topped up to H level marker
Temperature +24.6 deg C dew point +7.1 deg C RH 32.6% |
|
803
|
Fri May 17 08:10:05 2019 |
TD | AIDA+DTAS @ BigRIPS F11 | |
|
802
|
Fri May 17 02:15:56 2019 |
CA, TD, OH | Friday 17 May | 10.18 DAQ continues OK to file May19/R5_1
alpha background
ASIC settings 2018Oct03-13.26.58
*except* slow comparator 0xa -> 0x64
BNC PB-4 Pulser OFF
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachments 4
Merger OK - see attachment 5
TapeServer OK - see attachment 6
1.8.W spectra 20us FSR - attachments 7 & 8
1.8.W spectra 20ms FSR - attachments 9 & 10
Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C - water topped up
Temperature +24.8 deg C dew point +8.3 deg C RH 35.3%
10.42 DAQ stop May19/R5_2
slow comparator 0x64 -> 0xa
Check ASIC Control *all* ASICs & FEE64s
10:52 AIDA inserted ~4cm into DTAS
60Co (~40kBq) + 137Cs (~8kBq) sources located approximately 4cm in front of the front face of the snout
10.54 DAQ start May19/R6
good event statistics OK - see attachments 11
*all* FEE64s <100k, 9x < 20k
data rate 4.9Mb/s 0.64M data items/s
11.04 nnaida17 good events c. 350k returns to 13k following Check ASIC Control (nnaida1 OK - see
https://elog.ph.ed.ac.uk/AIDA/801 )
AIDA disc statistics indicate that fast discriminators were source of higher rate - next time this
happens we should check HitRate to see whether this is a specific channel or channels and possibly
mask
11.17 DTAS DAQ issues correlation scaler reset
AIDA data file May19/R6_3
use files *from* R6_4
11.39 DAQ stop file May19/R6_8
nnrpi1 is offline - we suspect wrench or power tool placed too close to RPi when we moved AIDA
into DTAS and we jostled one or other of the cables connected to nnrpi1
nnrpi1 power cycled to reboot
*all* FEE64s OK
USB ac mains relay controls recovered OK
CAEN N1419 contrs recovered OK
11.48 AIDA withdrawn from DTAS for installation of upstream (CH2)n shield and additional Pb wall
11.52 DAQ starts to file May19/R7
alpha background resumes
ASIC settings 2018Oct03-13.26.58
*except* slow comparator 0xa -> 0x64
BNC PB-4 Pulser OFF
18.06 DAQ continues ok, on file May19/R7_0
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachment 13
FEE64 temperatures OK - see attachment 14
good event statistics OK - see attachment 15
Merger and TapeServer ok
Temperature +25.0 deg C dew point +4.4 deg C RH 27.0%
|
|
801
|
Thu May 16 02:13:10 2019 |
CA, OH, TD | Thursday 16th May | 10:13 all system wide checks OK *except* nnaida7 fails ADC calibration
FEE64 temperatures OK - see attachment 1
good event statistics OK - see attachments 2
detector biases & leakage currents OK - see attachment 3
Temperature + deg 24.7C dew point +8.6 deg C RH 35.0%
Merger and TapeServer ok
10:30 DAQ, merger, TapeServer stop
Alpha run stopped at AIDA file TapeData/May19/R2_5
slow comparator threshold lowered to 10 keV (0xa)
DAQ start, not writing to file
good event statistics after lowering threshold, ok (attachment 4)
10:49 1.8.W spectra 20ms FSR - attachments 5 & 6
11:25 started writing to file R3 (sync test)
low threshold run at 5mb/s
11:33 DAQ stop (R3)
slow comparator threshold raised to 100 kev (0x64)
DAQ start, alpha run
Writing to file TapeData/May19/R4
18.49 alpha background
ASIC settings 2018Oct03-13.26.58
*except* slow comparator 0xa -> 0x64
BNC PB-4 Pulser OFF
FEE64 temperatures OK - see attachment 7
good event statistics - see attachments 8 - nnaida1 405k & nnaida17 368k, all other FEE64s c. 380Hz as
expected
ASIC check for *all* ASICs & FEE64s restored all FEE64 rates to expected c. 380Hz - see attachment 13
detector biases & leakage currents OK - see attachments 9 & 10
Merger OK - no error messages - see attachment 11
TapeServer OK - see attachment 12
all system wide checks OK *except* nnaida7 fails ADC calibration *and* Check Sync errors not counted Since
Baseline - see attachment 14
DAQ stopped May19/R4_3
aidas1> ls -l /TapeData/May19
total 19808048
-rw-rw-rw-. 1 npg npgstaff 1908736000 May 14 18:28 R1_0
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 15 02:30 R2_0
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 15 10:20 R2_1
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 15 18:08 R2_2
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 16 01:57 R2_3
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 16 09:45 R2_4
-rw-rw-rw-. 1 npg npgstaff 190447616 May 16 10:28 R2_5
-rw-rw-rw-. 1 npg npgstaff 1109327872 May 16 11:28 R3_0
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 16 18:39 R4_0
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 16 18:45 R4_1
-rw-rw-rw-. 1 npg npgstaff 2097152000 May 16 18:50 R4_2
-rw-rw-rw-. 1 npg npgstaff 225312768 May 16 18:54 R4_3
-rw-rw-rw-. 1 npg npgstaff 52494336 May 16 19:10 R5_0
Wahtever the issue was it probably occurred before 18.39
DAQ RESET
DAQ SETUP
ASIC check for *all* ASICs & FEE64s
all system wide checks OK *except* nnaida7 fails ADC calibration
Sync all ASIC clocks using Re-SYNC
DAQ started May19/R5
19.23 1.8.W spectra - see attachments 15 & 16 |
|
800
|
Wed May 15 02:17:23 2019 |
CA, OH, TD | Wednesday 15 May | 10.18 DAQ continues OK to file May19/R2
alpha background
ASIC settings 2018Oct03-13.26.58
*except* slow comparator 0xa -> 0x64
BNC PB-4 Pulser OFF
all system wide checks OK *except* nnaida7 fails ADC calibration
detector biases & leakage currents OK - see attachments 1 & 2
FEE64 temperatures OK - see attachment 3
good event statistics OK - see attachments 4
Merger OK - see attachment 5
TapeServer OK - see attachment 6
1.8.L spectra - attachments 7 & 8
1.8.W spectra 20us FSR - attachments 10 & 11
1.8.W spectra 20ms FSR - attachments 10 & 11
Julabo FL11006 chiller set point +20 deg C actual +20.0 deg C - water topped up
Temperature + deg 24.7C dew point +7.5 deg C RH 33.3%
FEE64 nnaida1-4 temperature logs - see attachments 12-15
all FEE64 temperatures appear to be stable
11.05 check ASIC control *all* FEE64s and ASICs
11.13 analysis of file May19/R2_1 - see attachment 16
ADC rate c. 72Hz dominated by same, single hot channel (Stat spectrum ch# 104) in each of FEE64s nnaida1,
2, 5 & 6 otherwise looks OK
sync pulser rate 10Hz confirmed by JA
17.25 all system wide checks OK *except* nnaida7 fails ADC calibration
FEE64 temperatures OK - see attachment 18
good event statistics OK - see attachments 19
detector biases & leakage currents OK - see attachment 20
Temperature + deg 24.5C dew point +8.2 deg C RH 34.1%
Merger and TapeServer ok |
|
799
|
Tue May 14 02:19:20 2019 |
CA, OH, TD | Tuesday 14th May 2019 | 10.20 System wide checks ok (except naida07 ADC calibration failure)
FEE temperatures ok (attachment 1)
good event statistics ok (attachment 2)
detector bias and leakage currents ok (attachment 3)
DAQ stopped
10.40 pulser switched off
slow comparator threshold raised to 0x64
DAQ, merger, TapeServer started - all ok (attachments 4 and 5)
writing to file R1 in TapeData/May19 at 70 kb/s
good event statistics (attachment 6)
16:11 Tapeserver and merger still ok
DAQ still running
System wide checks all ok *except nnaida7 fails ADC calibration*
Bias and leakage currents ok - attachment 7
FEE Temperatures ok - attachment 8
Statistics still ok
Spectrum are showing alpha events (Layout 4) - attachment 9
18:25 Tektronix clock accidentally unplugged. DAQ has stalled as a result.
Clock reset to 50MHz 5V TTL signal
18:29 System wide check of clock status shows error message in attachment 10
DAQ Reset
With slow comp at oxa statistics are still good - attachment 11
System wide checks all ok *except nnaida7 fails ADC calibration*
Resync to the ASIC clock
All histograms zeroed
18:37 Slow comp threshold raised to 0x64 for alpha run
Writing to file:
May19/R2_0
|
|
798
|
Mon May 13 10:13:26 2019 |
TD | AIDA with 25MHz external clock does not work ... ! |
OH setup AIDA external clock using Tektronix AFG3102 - settings as follows
CH1 ON - amplitude +5V, phase 0n deg, offset 0V, frequency 50.000000MHz
JA setup DTAS/BigRIPS clock using Tektronix AFG3102 - settings as follows
CH2 ON - amplitude -0.8V, phase 0n deg, offset 0V, frequency 25.000000MHz
Unfortunately the AGF3102 was set to CH1 frequency = CH2 frequency and this changed the AIDA external
clock frequency from 50MHz to 25MHz.
This stopped the DAQ. Interestingly it was possible to Resync successfully and *all* system wide
checks were OK *except* Check Clock Status which produced errors - see attachment 1.
CH1 & CH2 now set to 50MHz and 25MHz respectively.
AIDA DAQ now running OK. |
|
797
|
Mon May 13 10:00:01 2019 |
TD | Report - low - SETUP/Perform Timestamp Resync produces error message |
AIDA@RIKEN
Selecting SETUP (Run Control tab) or Perform Timestamp Resync (AIDA Control tab) produces error message - see
attachment 1
Selecting Perform Resync (AIDA Master Timestamp Control) works OK |
|
796
|
Mon May 13 03:43:08 2019 |
CA, TD, OH | Sunday 12 May |
19.12 all system wide checks OK *except* nnaida7 fails ADC calibration
FEE64 temperatures OK - see attachment 9
detector biases & leakage currents OK - see attachments 10 & 11
good event statistics OK - see attachment 12
all FEE64s <100k, 10x < 20k
1.8.W spectra 20us FSR - see attachments 13 & 14 |
|
795
|
Mon May 13 03:23:58 2019 |
CA, TD, OH | Monday 13 May |
09.55 all system wide checks OK *except* nnaida7 fails ADC calibration
FEE64 temperatures OK - see attachment 1
good event statistics OK - see attachment 2
all FEE64s <100k, 11x < 20k
1.8.L spectra - see attachments 3 & 4
nnaida1 70ch FWHM, nnaida3 80 ch FWHM
detector biases & leakage currents OK - see attachments 5 & 6
1.8.W spectra 20us FSR - see attachments 7 & 8
nnaida1 Options, Local Controls, Data Redaout, Discriminator Controls - see attachment 9-13
Lost Activity Monitor -see attachment 14
18.01 DAQ found stopped
This was caused by a chnage to the AIDA external clock frequency - see https://elog.ph.ed.ac.uk/AIDA/798
Following change of AIDA external clock frequency to 50MHz
DAQ RESET
DAQ SETUP
Sync all ASIC clocks using Resync
all system wide checks OK *except* nnaida7 fails ADC calibration
zero all histograms
DAQ GO
18.31 FEE64 temperatures OK - see attachment 15
good event statistics OK - see attachment 16
all FEE64s <100k, 10x < 20k
detector biases & leakage currents OK - see attachments 17 & 18 |
|
794
|
Sat May 11 02:50:14 2019 |
CA, OH, TD | Saturday 11th May 2019 | 10:52 system wide checks ok, except naida07 fails ADC calibration
good event statistics, rates in naida19,20,23,24 high (attachment 1)
FEE temperatures ok (attachment 2)
detector bias/leakage currents stable (attachment 3) |
|
793
|
Fri May 10 05:58:01 2019 |
CA, OH, TD | Friday 10 May 2019 | 11.00 - chiller refilled and switched on
set to 20 degrees celsius
AIDA stand moved and secured in F11
clock connected -> 50 MHz square wave at 5.0 V
AIDA connected to power box
network cables attached
NIM modules switched on
Raspberry Pis switched on
Detector bias switched on for detectors 1-6
13.58 - AIDA power relay 1,2,5,6 sequenced on
bias/leakage currents ok (attachment 1)
DAQ setup
System wide checks ok, except naida07 fails ADC calibration, and clock status check returns error in
nnaida06, 5.
Above issue with clock no longer present
FEE temperatures ok (attachment 2)
good event statistics (attachment 3)
16.30 1.8.W spectra n+n strips
FSR 20us, 200us, 2ms, 20ms attachments 4-7
1.8.W spectra p+n strips
FSR 20us, 200us, 2ms, 20ms attachments 8-11
compare with https://elog.ph.ed.ac.uk/AIDA/746
naida19, naida20, naida23 show low frequency noise structure, worse than in above elog entry
17.22 1.8.L spectra p+n strips
peaks:
aida01 72
aida02 112
aida03 80
aida04 78
aida05 74
aida06 95
aida07 78
aida08 71
aida09 82
aida10 85
aida11 93
aida12 96
aida13 101
aida14 70
aida15 87
aida16 87
aida17 69
aida18 83
aida19 127
aida20 116
aida21 84
aida22 74
aida23 108
aida24 117
|
|
792
|
Tue Mar 26 14:02:59 2019 |
OH | Inventory to be shipped back | Inventory is currently packaged up under the AIDA desk in B3F
2 FEE64 Modules
Module 1 - MAC Addresses
2B:33:15
2B:22:6E <- Loose HDMI
Module 2 - MAC Addresses
2B:09:DA <- ASIC #1 HEC Hot Channel, LEC noisy
2B:11:C5 <- Missing ASIC
MACB A-MP 202394 -2 72-38
MACB RevB Ser38 AMP 202394-2
MACB RevB AMP 202394-2 81-49 |
|
791
|
Sun Nov 18 02:02:48 2018 |
TD | Latest detector bias & leakage current spreadsheet (URL and csv & xlsx files) | https://docs.google.com/spreadsheets/d/12hgbrywB10hGsKt5uc2HnLfZymh8_uvM6cEASbbqnBE/edit#gid=1347508769 |
|
790
|
Fri Nov 16 06:01:47 2018 |
CA | High Energy Channel Ex vs Ey for different degrader thicknesses | 15.04 Plots of X Energy vs Y Energy for each detector, and for 4 different Al degrader thicknesses, as indicated
on each graph. (Attachments 1-4) |
|
789
|
Fri Nov 16 04:47:39 2018 |
CA, TD, OH | Correlation analysis at different degrader Al thickness | 13.53 Correlation analysis plot of all detectors with 3.3mm Al degrader thickness (attachment 1)
13.54 Correlation analysis plot of detectors 1 and 2 with 3mm, 2.3mm and 2.8mm degrader thickness (attachment 2)
Correlation analysis plot of detectors 3 and 4 with 3mm, 2.3mm and 2.8mm degrader thickness (attachment 3)
Correlation analysis plot of detectors 5 and 6 with 3mm, 2.3mm and 2.8mm degrader thickness (attachment 4)
14.02 Plots made over first 200ms of data |
|
788
|
Thu Nov 15 22:55:38 2018 |
CA, TD, OH, LS | Friday 16th November 08.00 - 16.00 | 07.56 DAQ continues - AIDA file RIBF148/R5_822
165Pm BigRIPS setting + 2.8mm Al degrader
08.13 zeroed xy hit pattern on online analysis and MIDASsort
08.30 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 1)
good event statistics ok (attachment 2)
detector bias/leakage currents ok (attachment 3)
08.34 Merger and TapeServer ok, no error messages in terminal (attachments 4/5)
08.40 zeroed all spectra on online analysis and MIDASsort
08.44 BRIKEN and BigRIPS stop
AIDA file RIBF148/R5_830
08.45 BigRIPS starts run 1087
BRIKEN starts run 102
AIDA on file RIBF148/R5_830
09.13 n vs p multiplicity (attachment 6)
xy hit pattern (*deleted, took wrong screenshot, sorry! -CA@12.16*)
Energy X vs Energy Y plot (*deleted, took wrong screenshot, sorry! -CA@12.16*)
12.17 see elog entry for 15th November 2018 00.00-08.00 (attachments 9 and 10) for screenshots of the above before
degrader change
09.17 BRIKEN and BigRIPS stop
AIDA on file RIBF148/R5_836
beam stop to change Al degrader thickness to 3.3mm
09.23 chiller refilled while access available
humidity, temperature, dew point measurements obtained
09.25 detector bias/ leakage currents ok (attachment 7)
09.30 zeroed all spectra in online analysis and MIDASsort
09.31 beam back on
BRIKEN and BigRIPS start
AIDA file RIBF148/R5_838
BRIKEN run 103
BigRIPS run 1088
10.32 BRIKEN and BigRIPS stop
AIDA file RIBF148/R5_848
10.33 BRIKEN and BigRIPS start
AIDA file RIBF148/R5_848
BRIKEN run 104
BigRIPS run 1089
10.45 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 8)
good event statistics ok (attachment 10)
detector bias/leakage currents ok (attachment 9)
11.20 ran analyser on file R5_847 - attachment 11
no timewarps observed
dead time in nnaida19 at ~1s out of 355.283s (~0.3%)
rest at zero, except nnaida20 at ~0.6s and nnaida21 at ~0.1s
11.33 BRIKEN and BigRIPS stop
AIDA file RIBF148/R5_859
11.34 BRIKEN and BigRIPS start
AIDA file RIBF148/R5_859
BRIKEN run 105
BigRIPS run 1090
12.05 n vs p multiplicity (attachment 12)
xy hit pattern (attachment 13)
Energy X vs Energy Y plot (attachment 14)
this is after change of Al degrader thickness to 3.3mm
12.36 BRIKEN and BigRIPS stop
AIDA file RIBF148/R5_869
12.36 BRIKEN and BigRIPS start
AIDA file RIBF148/R5_869
BRIKEN run 106
BigRIPS run 1091
12.40 online histograms
per FEE64 Rate - attachment 15
per FEE64 HitRate - attachment 16
1.8.L spectra - attachments 17-20
nnaida1 pulser peak width 70ch FWHM
1.8.H spectra - attachments 21-22
1.8.W spectra - attachments 23-24
12.47 beam off
AIDA file RIBF148/R5_870
12.54 BRIKEN and BigRIPS stop
AIDA file RIBF148/R5_871
13.10 MIDASsort analysis RIBF148/R5_871 - all ADC data synchronised - attachments 25-26
13.16 correlation analysis decay plots for each detector with the Al degrader thickness at 3.3mm (attachment 23)
see elog entry November 15th 2018 00:00-08:00 (attachments 16-18) for same plots with previous degrader
thickness.
13.26 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 24)
good event statistics ok (attachment 25)
detector bias/leakage currents ok (attachment 26)
15.04 see other elog entries for correlation analysis and high energy Ex vs Ey plots compared between different Al
degrader thicknesses.
15.38 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 28)
good event statistics ok (attachment 27)
detector bias/leakage currents ok (attachment 29)
15.55 BRIKEN start background measurement
BRIKEN not writing to file
AIDA file RIBF148/R5_896
Online histograms available
16.00 BRIKEN start writing to file
AIDA file RIBF/R5_897
BRIKEN run 107
17.18 DAQ stop AIDA file RIBF148/R5_907
End of run RIBF148 due to accelerator problems requiring ~3-4 days to fix
'low' mass BigRIPS setting postponed
ASIC check load
all onine histograms zero'd
17.22 DAQ start AIDA file RIBF148/R6
Pulser walkthrough
BNC PB-4 amplitudes 90,000 - 10,000 @ 10,000 step
x5 attenuator IN
Frequency 25Hz range
Delay MIN
Polarity -
Tail pulse
17.37 DAQ trop AIDA file RIBF148/R6_2
1.8.L spectra - see attachments 30 & 31
17.40 DAQ start AIDA file RIBF148/R7
Pulser walkthrough
BNC PB-4 amplitudes 1,200 - 1,000 @ 1,000 step
x5 attenuator IN
Frequency 25Hz range
Delay MIN
Polarity -
Tail pulse
17.50 DAQ stop AIDA file RIBF148/R7_1
1.8.L spectra - see attachments 32 & 33
17.54 DAQ start AIDA file RIBF148/R8
background
BNC PB-4 pulser OFF
18.00 detector biases & leakage currents OK - see attachments 34 & 35
FEE64 temperatures OK - see attachment 36
good event statistics OK - see attachments 37
Merger OK - see attachment 38
TapeServer OK - see attachment 39
no error or warning messages terminal server sesssions reported
per FEE64 Rate spectra - attachment 40
1.8.W spectra 20ms FSR - attachments 41 & 42
18.10 DAQ stop AIDA file RIBF148/R8_2
Detector bias OFF
FEE64 power OFF
18.12 Ambient temperature +25.5 deg C, d.p. +2.2 deg C, RH 21.5%
18.15 Withdraw AIDA from BRIKEN - done
2x NIM bin OFF
1x NIM bin fan tray OFF
1x USB ac mains relay OFF
Julabo FL11006 chiller OFF front panel/rear panel/wall-mounted 3ph outlet
2x network cables disconnected and coiled/cable-tied to stairwell railings
disconnect coorelation scaler trigger & reset inputs to root MACB front panel
disconnect 50MHz TTL clock root MACB rear panel
disconnect AIDA plastic veto
discopnnect 2x ac mains cabnles from ac mains distribution box
remove AIDA support assembly nose support
19.00 raise AIDA support assembly feet/lower AIDA support assembly wheels from/to F11 floor
|
|
787
|
Thu Nov 15 15:05:33 2018 |
LS | Friday 16 November 00.00 to 08.00 |
00.25 System wide checks ok * except nnaida 7 fails adc*
FEE Temps ok - attachment 1
Stats ok - attachment 2
Leakage currents ok - attachment 3
tape server and merger ok
00:57 BigRIPS and BRIKEN stop
AIDA on file R5_753
00.59 BigRIPS starts run 1079
BRIKEN starts run 94
AIDA on file R5_753
01:56 BigRIPS and BRIKEN stop
AIDA on file R5_762
01.57 BigRIPS starts run 1080
BRIKEN starts run 95
AIDA on file R5_762
02:47 BigRIPS and BRIKEN stop
AIDA on file R5_770
02.48 BigRIPS starts run 1081
BRIKEN starts run 96
AIDA on file R5_770
03:45 BigRIPS and BRIKEN stop
AIDA on file R5_780
03.47 BigRIPS starts run 1082
BRIKEN starts run 97
AIDA on file R5_780
04:45 BigRIPS and BRIKEN stop
AIDA on file R5_790
04.47 BigRIPS starts run 1083
BRIKEN starts run 98
AIDA on file R5_790
04.50 System wide checks ok * except nnaida 7 fails adc*
FEE Temps ok - attachment 4
Stats ok - attachment 5
Leakage currents ok - attachment 6
tape server and merger ok
05.46 BigRIPS and BRIKEN stop
AIDA on file R5_800
05.47 BigRIPS starts run 1084
BRIKEN starts run 99
AIDA on file R5_800
06.45 BigRIPS and BRIKEN stop
AIDA on file R5_810
06.46 BigRIPS starts run 1085
BRIKEN starts run 100
AIDA on file R5_810
07.25 System wide checks ok * except nnaida 7 fails adc*
FEE Temps ok - attachment 7
Stats ok - attachment 8
Leakage currents ok - attachment 9
tape server and merger ok
07.44 BigRIPS and BRIKEN stop
AIDA on file R5_820
07.45 BigRIPS starts run 1086
BRIKEN starts run 101
AIDA on file R5_820 |
|
786
|
Thu Nov 15 06:52:09 2018 |
OH | Thursday 15th November 16:00 - 24:00 | 15:51 DAQ continues - AIDA file RIBF148/R5_664
165Pm BigRIPS setting + 2.8mm Al degrader
ASIC settings 2018Oct03-13.26.58
from c. 09.00 13.11.18
n+n FEE64s Ibias LF feedback 0xb -> 0xd
HEC fast comparator threshold 0x20 -> 0xff
15:52 Beam out due to power fault somewhere.
15:53 Beam back
BRIKEN and BigRIPS start
BRIKEN file 181115_1553_165Pm_085.dlt
BigRIPS run 1070
AIDA on run R5_664
16:00 System wide checks all ok *Except nnaida7 fails adc calibration*
Leakage currents and bias ok - attachment 1
FEE Temps ok - attachment 2
Statistics ok - attachment 3
16:54 BRIKEN and BigRIPS stop
AIDA on file R5_675
16:55 BRIKEN starts run 181115_1654_165Pm_086.dlt
BigRIPS starts run 1071
AIDA on file R5_675
17:30 BRIKEBN and BigRIPS stop for BigRIPS team to get access
13:46 BRIKEN sends sync pulse resetting timestamp
AIDA on file R5_683
17:49 BRIKEN starts run 181115_1748_165Pm_087.dlt
bigRIPS starts run 1072
AIDA on file R5_683
18.00 System wide checks ok, except nnaida7 ADC calibration failure
FEE64 Temperatures ok (attachment 4)
Good event statistics ok (attachment 5)
Detector bias/Leakage currents ok (attachment 6)
18.04 Merger and TapeServer run ok - no error messages in terminals (attachments 7 and 8)
18:48 BigRIPS and BRIKEN stop
AIDA on file R5_693
18:49 BigRIPS starts run 1073
BRIKEN starts run 181115_1849_165Pm_088.dlt
AIDA on file R5_693
19:47 BgiRIPS and BRIKEN stop
AIDA on file R5_703
19:48 BigRIPS starts run 1074
BRIKEN starts run 181115_1947_165Pm_089.dlt
AIDA on file R5_703
20:54 bigRIPS and BRIKEN stop
AIDA on file R5_714
20:55 bigRIPS starts run 1075
BRIKEN starts run 181115_2053_165Pm_090.dlt
AIDA on file R5_714
21:25 System wide checks all ok *Except nnaida 7 fails adc calibration*
Bias and leakage currents ok - attachment 9
FEE Temps all ok - attachment 10
Statistics ok - attachment 11
21:55 BRIKEN and BigRIPS stop
AIDA on file R5_724
21:56 BigRIPS starts on run 1076
BRIKEN starts run
AIDA on file R5_724
22:55 BigRIPS and BRIKEN stop
AIDA on file R5_733
22:56 BRIKEN Starts run
BigRIPS starts run 1077
AIDA on file R5_733
23.53 Briken stops
AIDA on file R5_743
23.55 BRIKEN starts
bigRIPS run 1078
BRIKEN starts run 93
AIDA on file R5_743 |
|
785
|
Wed Nov 14 22:54:11 2018 |
LS | Thursday 15th November 08:00 - 16:00 |
08.10 BRIKEN stop
AIDA file RIBF148/R5_587
08.11 BRIKEN start
AIDA file RIBF148/R5_588
BRIKEN run 77
BigRIPS run 1062
08.16 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 1)
good event statistics ok (attachment 2)
detector bias/leakage currents ok (attachment 3)
Merger and TapeServer still ok
09.11 BRIKEN stop
AIDA file RIBF148/R5_598
09.13 BRIKEN start
AIDA file RIBF148/R5_598
BRIKEN run 78
BigRIPS run 1063
10.10 BRIKEN stop
AIDA file RIBF148/R5_608
10.12 BRIKEN start
AIDA file RIBF148/R5_608
BRIKEN run 79
BigRIPS run 1064
11.10 BRIKEN stop
AIDA file RIBF148/R5_618
11.12 BRIKEN start
AIDA file RIBF148/R5_618
BRIKEN run 80
BigRIPS run 1065
11.17 system wide checks ok, except nnaida7 ADC calibration failure
FEE64 temperatures ok (attachment 4)
good event statistics ok (attachment 5)
detector bias/leakage currents ok (attachment 6)
Merger and TapeServer still ok
12:10 BigRIPS and BRIKEN stop
AIDA on file R5_628
12:11 BRIKEN start
AIDA on file R5_628
BigRIPS run 1066
BRIKEN run 81 181115_1211_165Pm_081.dlt
13.10 BRIKEN stop
AIDA file RIBF148/R5_638
13.12 BRIKEN start
AIDA file RIBF148/R5_638
BRIKEN run 82
BigRIPS run 1067
13.32 AIDA file RIBF148/R5_642
BRIKEN stopped
Turning beam off to access area
13.50 AIDA file RIBF148/R5_644
beam back on
BRIKEN run 83
BigRIPS run 1068
14.26 MIDASsort analysis file RIBF148/R5_648 - all ADC data synchronised - attachment 7
analysis of file RIBF148/R5_648 - attachment 8
nnaida19 dead time ~0.5% - all other deadtimes lower or 0% - no timewarps
14.35 1.8.W spectra layout ID 8 (attachment 9)
14.50 BRIKEN stop
AIDA file RIBF148/R5_654
14.51 BRIKEN start
AIDA file RIBF148/R5_654
BRIKEN run 83
BigRIPS run 1068
15.10 1.8.H high energy channel spectra (attachment 10)
15.35 beam off
AIDA file RIBF148/R5_662
BRIKEN and BigRIPS stopped run |
|