AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  CARME, Page 28 of 33  ELOG logo
  ID Date Author Subject Text Attachments
  117   Tue Oct 12 09:22:37 2021 RSS TD + OG DRMotors problem solvedNeg and ion cables were taken off and were
put in the box.

Top support was taken off.
 IMG_4971.JPGIMG_4970.JPG 
  116   Mon Oct 11 11:24:16 2021 RSS TD MLValve support and cable labelingSplit funnels were given to Oemer for cleaning.
Have received them back after lunch.

Valve support is installed again
 IMG_4962.JPGIMG_4969.JPG.svgIMG_4968.JPG.svg 
  115   Sat Oct 9 09:47:33 2021 CBTo do list for next week0. Give split-funnels to OG or Grossmontage
to clean. Mount back foot support for gate
valves. Critical when we split.
 Capture.PNG 
  114   Sat Oct 9 09:44:36 2021 CB RSSHeaters removed + motors testsRemoved thermocouples, heaters, RE72 modules
and leftover baking tent. Support octagon
removed except topmost part which requires
  
  113   Fri Oct 8 17:56:35 2021 CB RSSMotors testAttempted first motor test in XHV. Wired
and connected limit switches and string potentiometer.

Set up raspberry pi to control
 20211008_184731.jpg20211008_184731.jpg 
  112   Thu Oct 7 08:40:09 2021 CB RSSCARME in XHVCARME reached 9E-12 this morning

Ion pumps on
  
  111   Wed Oct 6 15:36:14 2021 CB RSSNeg activation tests -3Attempted to activate 2xUHV1400 connected
to bottom C4, channel #2 using P 200 W, t
rise 30 min, t hold 90 min.
  
  110   Wed Oct 6 09:47:30 2021 CB RSSNeg activation tests -2Attempted activating 2xUHV1400 connected to
top C4, channel #3 with P 300 W (450 C?),
t rise 30 min, t hold 90
  
  109   Wed Oct 6 09:39:59 2021 PB CBOriginal dirty assembly of CARME internalsPB found pictures of the original dirty assembly,
before screws and components were shipped
to DL.
 7x 
  108   Tue Oct 5 17:30:01 2021 CB RSSNEG activation testsFound D2000 #1 activated and at 250C (25W)

to prevent getting dirty when other NEGs
 20211005_131259.jpg20211005_171403.jpg20211005_161317.jpg 
  107   Tue Oct 5 11:44:11 2021 CB, TD SAES manuals & cable drawingsNEG power
SIP power
UHV1400 operating instructions
 8x 
  106   Mon Oct 4 20:30:46 2021 RSS JM CB (remotely)CARME baking status -21
Current temperature ~95 C (on manual
mode)
See plot and screen attached.

Current
pressure 2.86E-10
See
https://uoe-my.sharepoint.com/:x:/g/personal/cbruno2_ed_ac_uk/EVbsVOubNHRLj0PKiFIvd00BClMJ6WSJ1UDAyLA2F3R4jA?rtime=OK4kVZKB2Ug
 Screenshot_from_2021-10-04_21-28-28.pngScreenshot_from_2021-10-04_21-29-37.png 
  105   Mon Oct 4 09:19:11 2021 JM RSSActivation of NEG elements9:30

Ion pumps de-gassed by switching
on and off multiple times. Large pressure
 IMG_4880.JPGIMG_4881.JPGIMG_4882.JPGIMG_4883.JPG 
  104   Mon Oct 4 06:34:58 2021 RSS JM CB (remotely)CARME baking status -20
Current temperature ~110 C (on manual
mode)
See plot and screen attached.

Current
pressure 3.20E-8
See
https://uoe-my.sharepoint.com/:x:/g/personal/cbruno2_ed_ac_uk/EVbsVOubNHRLj0PKiFIvd00BClMJ6WSJ1UDAyLA2F3R4jA?rtime=OK4kVZKB2Ug
 Screenshot_from_2021-10-04_07-33-14.pngScreenshot_from_2021-10-04_07-34-14.png 
  103   Sun Oct 3 08:36:01 2021 RSS JM CB (remotely)CARME baking status -19
08:36
Current temperature ~132 C (on
manual mode RE72-1: 132 C, RE72-2: 130 C)
See
plot and screen attached.

Current pressure
 Screenshot_from_2021-10-03_09-33-30.pngScreenshot_from_2021-10-03_09-34-24.png 
  102   Sat Oct 2 22:14:14 2021 RSS JM CB (remotely)CARME baking status -18
Current temperature ~132 C (on manual
mode RE72-1: 132 C, RE72-2: 130 C)
See plot
and screen attached.

Current pressure
 Screenshot_from_2021-10-02_23-12-09.pngScreenshot_from_2021-10-02_23-13-16.png 
  101   Sat Oct 2 08:26:49 2021 RSS JM CB (remotely)CARME baking status -17
Current temperature ~132 C (on manual
mode RE72-1: 132 C, RE72-2: 130 C)
See plot
and screen attached.

Current pressure
 Screenshot_from_2021-10-02_09-25-40.pngScreenshot_from_2021-10-02_09-24-44.png 
  100   Fri Oct 1 21:47:51 2021 CB (remotely)CARME baking status -16Current temperature ~132 C (on manual mode
RE72-1: 132 C, RE72-2: 130 C)
See plot and screen attached.
 screen.pngTemps.png 
  99   Fri Oct 1 07:17:46 2021 RSS JMCARME baking status -15
Current temperature ~132 C (on manual
mode RE72-1: 132 C, RE72-2: 130 C)
See plot
and screen attached.

Current pressure
 Screenshot_from_2021-10-01_08-17-02.pngScreenshot_from_2021-10-01_08-16-41.png 
  98   Thu Sep 30 18:16:52 2021 RSS JMCARME baking status -14
Current temperature ~132 C (on manual
mode RE72-1: 132 C, RE72-2: 130 C)
See plot
and screen attached.

Current pressure
 Screenshot_from_2021-09-30_19-18-44.pngScreenshot_from_2021-09-30_19-18-08.png 
ELOG V3.1.4-unknown