ID |
Date |
Author |
Subject |
405
|
Fri May 26 09:08:06 2023 |
JM, CB | Detector movement |
Hopefully we will not be opening the chamber for quite some time so find attached images of all four installed detectors and a video of them both moving into the beam axis for future reference |
Attachment 1: 20230525_135756.jpg
|
|
Attachment 2: detector-movement-in.mp4
|
404
|
Fri May 26 08:53:54 2023 |
JM, CB | Strain reliefs (flange side) installed |
Strain reliefs on the flange side of the cable harness have been installed for two detectors (top right and bottom right detectors).
We require the strain relief parts from edinburgh in order to install strain reliefs for the left two detectors.
Find attached images for the strain relief installed and a video of the clamped cables during a movement cycle.
A V-I test was performed for the two detectors which have all strain reliefs installed (see attachment 4). V-I curve appears similar to the previous test (elog 398). Aluminium foil shield is now a bit battered so may have some small light leaks. Detectors were moved using the motors while bias was applied at 150V. Small change in leakage current observed (~0.01-0.02 uA) |
Attachment 1: strain-relief-flange-side.jpg
|
|
Attachment 2: strain-relief1_(1).mp4
|
Attachment 3: 20230526_140923.jpg
|
|
Attachment 4: V-I-plot.png
|
|
403
|
Wed May 24 09:54:56 2023 |
JM, CB | Maximum detector positions |
The detectors were moved into the beam axis using the motors to check the maximum distance they can be moved in without collision with one another.
Movement of both pneumatic motors results in a separation of ~ 4 cm between the detectors.
The servo motors for left and right were each moved in by 1 cm -> results in separation of 2 cm between the detectors
The servo motors for left and right were then each moved in by 5 mm -> results in separation of 1 cm between the detectors
The separation between the jacking plate of one of the detectors and the side of the adjacent detectors is closer than the separation of the detector silicons. The detectors could not be moved in any further than this to avoid collisions from the jacking plate
Attached are photos after movement of the detectors. Parallax may affect the distance on the ruler in the photos.
During beam time -> Maximum distance that can be moved on the servo motors is 30 mm combined for left and right. (15 mm left and right or 10 mm left, 20mm right etc)
|
Attachment 1: pneumatic-in.jpg
|
|
Attachment 2: pneumatic-in2.jpg
|
|
Attachment 3: pneumatic-in-servo-10mm.jpg
|
|
Attachment 4: pneumatic-in-servo-15mm.jpg
|
|
Attachment 5: max-in.jpg
|
|
Attachment 6: max-in2.jpg
|
|
Attachment 7: max-in-zoom.jpg
|
|
Attachment 8: max-in-jacking-plate.jpg
|
|
Attachment 9: max-in-jacking-plate2.jpg
|
|
402
|
Wed May 24 08:35:12 2023 |
JM, CB | Strain reliefs (detector side) installed |
The new MACOR strain reliefs on the (detector side) have been installed. See attached photos
The two pieces of MACOR are not fully closed but we are confident that the kapton wires are sufficiently clamped.
In order for the strain reliefs to be installed in the correct poisition, two nuts on the jacking plate of the cable harness had to be removed. This is as the groove in the strain relief was not deep enough to accomadate the nut.
Attached is also a video of the wires (not) moving with the strain reliefs mounted |
Attachment 1: 20230524_092010.jpg
|
|
Attachment 2: 20230524_092017.jpg
|
|
Attachment 3: 20230524_092101.jpg
|
|
Attachment 4: 20230524_093000.jpg
|
|
Attachment 5: 20230524_092159.mp4
|
401
|
Mon May 15 07:34:22 2023 |
TD | Monday 15 May |
08.35 DSSSD#3 bias -149.95V leakage current -6.418uA temperature +19.6 deg C ( https://web-docs.gsi.de/~lestinsk/tempplot.php )
DSSSD#3 bias OFF |
400
|
Sun May 14 11:55:12 2023 |
TD | Sunday 14 May |
12.44 DSSSD#3 bias -149.95V leakage current -6.448uA temperature +19.7 deg C ( https://web-docs.gsi.de/~lestinsk/tempplot.php )
CRYRING cave lights ON - leakage current measurements @ 60s intervals
6.448
6.449
6.448
6.448
6.449
6.448
6.448
6.448
6.449
6.448
18.24 DSSSD#3 bias -149.95V leakage current -6.442uA temperature + 21.0 deg C ( https://web-docs.gsi.de/~lestinsk/tempplot.php )
CRYRING cave lights ON - leakage current measurements @ 60s intervals
6.441
6.441
6.441
6.441
6.441
6.442
6.442
6.442
6.440
6.441 |
399
|
Sat May 13 11:34:25 2023 |
TD, RSS | Saturday 13 May |
12.34 DSSSD#3 bias -149.95V leakage current -6.763uA temperature +21.2 deg C ( https://web-docs.gsi.de/~lestinsk/tempplot.php )
CRYRING cave lights ON - leakage current measurements @ 60s intervals
6.763
6.764
6.764
6.765
6.767
6.767
6.767
6.767
6.768
6.769
12.48 CRYRING cave lights OFF
13.03 leakage current -6.771uA - no obvious change due to the CRYRING cave lighting
17.31 DSSSD#3 bias -149.95V leakage current -6.658uA temperature +21.0 deg C ( https://web-docs.gsi.de/~lestinsk/tempplot.php )
leakage current measurements @ 60s intervals
6.656
6.655
6.655
6.655
6.654
6.652
6.652
6.651
6.650
6.649
CRYRING cave temperature variations/cycles ~0.1 deg C at +21 deg C ambient temperature can produce leakage current variations ~60nA for a leakage current of ~6.6uA
td@winder:~/f77$ ./jgen
*** Enter reference temperature (K)
294
*** Enter minimum temperature, maximum temperature
*** and temperature increment (K)
290 298 0.1
Temperature (K) Ratio (J(T)/J(294.0))
290.0 0.702
290.1 0.708
290.2 0.714
290.3 0.721
290.4 0.727
290.5 0.734
290.6 0.740
290.7 0.747
290.8 0.754
290.9 0.761
291.0 0.767
291.1 0.774
291.2 0.781
291.3 0.788
291.4 0.795
291.5 0.802
291.6 0.809
291.7 0.817
291.8 0.824
291.9 0.831
292.0 0.839
292.1 0.846
292.2 0.854
292.3 0.861
292.4 0.869
292.5 0.877
292.6 0.884
292.7 0.892
292.8 0.900
292.9 0.908
293.0 0.916
293.1 0.924
293.2 0.932
293.3 0.941
293.4 0.949
293.5 0.957
293.6 0.966
293.7 0.974
293.8 0.983
293.9 0.991
294.0 1.000
294.1 1.009
294.2 1.018
294.3 1.027
294.4 1.036
294.5 1.045
294.6 1.054
294.7 1.063
294.8 1.072
294.9 1.082
295.0 1.091
295.1 1.101
295.2 1.110
295.3 1.120
295.4 1.130
295.5 1.139
295.6 1.149
295.7 1.159
295.8 1.169
295.9 1.179
296.0 1.190
296.1 1.200
296.2 1.210
296.3 1.221
296.4 1.231
296.5 1.242
296.6 1.253
296.7 1.263
296.8 1.274
296.9 1.285
297.0 1.296
297.1 1.308
297.2 1.319
297.3 1.330
297.4 1.342
297.5 1.353
297.6 1.365
297.7 1.376
297.8 1.388
297.9 1.400
STOP *** Program ends |
398
|
Fri May 12 16:24:20 2023 |
RSS, TD | V-I test |
Since yesterday evening, the bottom right detector was at -150 V. In the morning, the current was quite stable ~ 6.77 uA. The bias was switched off, the foils opened, both of the detector arms were moved in and out three times, and the chamber was tightly covered again. A negative bias voltage was applied from 0 to 150 V in steps of 10 V one by one to all four detectors, and the V-I results can be seen in Attachments 1-4. All detectors seem to be fine. The variation of the current with time at a constant bias voltage of -150 V was checked for all detectors and the results are in Attachments 5-8. Also, with an applied bias voltage of -150 V to the bottom right and left detector, the detector arms were moved in and out three times which showed a variation of 1-2 nA and 1 nA in the leakage current. For further conditioning of the bottom right detector, it is again left with a negative bias voltage of 150 V over the weekend. |
Attachment 1: detector_1.png
|
|
Attachment 2: detector_2.png
|
|
Attachment 3: detector_3.png
|
|
Attachment 4: detector_4.png
|
|
Attachment 5: detector_1_variation.png
|
|
Attachment 6: detector_2_variation.png
|
|
Attachment 7: detector_3_variation.png
|
|
Attachment 8: detector_4_variation.png
|
|
397
|
Thu May 11 21:22:54 2023 |
RSS, TD | V-I test |
Today, the foil was opened, the pins on the bottom right detector were checked and looked fine, the right-hand arm was moved a bit in and then out, and the foil was tightly closed again. The negative bias voltage was applied from 0 to 50 V in steps of 10 V. During the measurement, the lights in the cave were switched off to ensure no leakage current from the light*. The results are shown in Attachment 1. To condition the detector, it was left at a negative bias voltage of 50 V for ~ 3 hours with lights on. No increase in current was seen (see Attachment 2), indicating the chamber was nicely covered with foil. Then the negative voltage was increased from 60 V to 150 V in steps of 10 V. No large current readings, like yesterday, were seen today (see Attachment 3). The detector was then left at -150 V for ~ 30 minutes with the current to be nearly stable with fluctuations in the range of 70-100 nA. The foil was then taken off to check if there are any dust particles on the detector surface. As can be seen in Attachment 6, some dust particle was found at the bias/multi-guard ring (MGR) boundary structures of the DSSSD. The area was then nicely cleaned with isopropanol + lab tissue wipe (see Attachment 7) and the chamber was covered with aluminum foil again. The negative bias voltage was applied from 0 to 150 V in steps of 10 V again and the results can be seen in attachment 5, indicating the response of the detector to be normal, unlike yesterday. Before calling the day off, the detector was left at -150 V overnight.
* Every time, before switching off the lights, it was checked that no one else was in the cave. |
Attachment 1: test1.png
|
|
Attachment 2: test2.png
|
|
Attachment 3: test3.png
|
|
Attachment 4: test4.png
|
|
Attachment 5: test5.png
|
|
Attachment 6: IMG_8800.jpg
|
|
Attachment 7: IMG_8804.jpg
|
|
396
|
Wed May 10 19:06:18 2023 |
RSS, TD | V-I test |
To test the detectors, adaptors were installed on the feed through flanges. A negative bias voltage was applied in steps of 10 V from 0 to 150 V. The V-I plots are attached for the four detectors. Three detectors - top right, top left, and bottom left (in the direction of the beam) - are fine. For the bottom right detector (in the direction of the beam), the current was fluctuating and was considerably high (1st test). The adaptors connecting to this detector were removed and checked by applying the bias voltage which showed zero current for voltages ranging from 0 to - 150 V, confirming the adaptors to be fine. They were then exchanged by other adaptors and the V-I test was repeated again (2nd test) with fluctuating and high current values. One of the flanges connecting to this detector was found to be light-tight. These tests were done at an ambient temperature of 21 degrees. Also, we couldn't find the rest of the three SHV to 2x Lemo 00 cables for the detector HV. |
Attachment 1: detector_1.png
|
|
Attachment 2: detector_2.png
|
|
Attachment 3: detector_3.png
|
|
Attachment 4: detector_4.png
|
|
395
|
Tue May 2 15:02:09 2023 |
CB, JM | All detectors and harnesses mounted |
Completed mounting harnesses and detectors.
The strain relief support for the bottom most two detectors on the feedthrough side was not mounted because we could not find enough MACOR parts and mounting flanges. We cannot find them in GSI, possibly they are somewhere not obvious in Edinburgh.
The strain relief on the detectors' side are still being manufactured.
We plan to test the detectors electrically from next week without moving the dolly. DR and OG to find a temporary mounting point for the manifold so that FEEs can be cooled in current configuration. |
Attachment 1: 20230502_153410.jpg
|
|
Attachment 2: 20230502_153503.jpg
|
|
Attachment 3: 20230502_153509.jpg
|
|
Attachment 4: 20230502_153511.jpg
|
|
394
|
Mon May 1 18:21:18 2023 |
CB, JM | All detectors mounted |
Mounted four detectors on Carme ad well as four cable harnesses.
Cables for new, shorter harnesses appear still too long. Likely the model is not correct somehow.
One feedthrough flange was found to be damaged by an adaptor card. Tried to fix in vain. One pin is missing. Replaced with another. |
Attachment 1: 20230501_191533.jpg
|
|
Attachment 2: 20230501_191511.jpg
|
|
Attachment 3: 20230501_145740.jpg
|
|
393
|
Mon May 1 08:26:09 2023 |
CB, JM | Left hand detectors remounted |
(31 April)
Remounted 3335-12 on top and new 3335-1 on bottom of left hand plate. Used new dog legs and old standoffs.
Dismounted all harnesses and feedthroughs from CARME. Dismounted all blind flanges on right hand side. Also had to dismount bottom SAES feedthrough due to a bolt falling inside.
Remounted feedthough. Remounted left hand detector plate.
Prepared new 3335-11 and 3335-13 on top and bottom of right hand plate. Planning to mount tomorrow.
Also planning to start jacking in detectors to new harnesses.
Not possible to test any detector on the bench due to lack of appropriate multimeter probe. Will have to test with harnesses. |
392
|
Sun Apr 30 10:01:39 2023 |
CB, JM | Left hand detectors unmounted |
(29 April)
Inspected CARME after opening. See attached. Some key takeaways.
- No visible damage to the detectors or the Kapton cables. No visible deposits or dust on the detectors either.
- No damage even to the fragile thermocouple wires
- The strain relief on the top and bottom did nothing. Those on the sides may have helped.
- The split funnels on top fell and fortunately missed the detectors. We should not keep them. Unclear if they feel when moving CARME or during the beamtime but either is too dangerous. We need another solution or we'll have to do without.
- Obvious traces of white dust particles over the chamber. Most likely MACOR.
Jacked out harness for bottom left detector that was not accepting bias. MACOR shavings fell off while jacking out. MACOR connector on side with no pins is clearly damaged. See attached. Should not affect performance.
Tested bias pins - they work. No visible damage to the detector except one bias bond wire has disconnected. Possibly one more nearby. TD thinks this is not enough to explain observed behaviour.
Mounting a DSSD on the plate while it hangs in Carme appears possible but very risky. Dismounted whole left support plate, and right support plate.
Unless a clear fault can be identified for either the bottom left detector or its harnesses, both the detector and the harnesses will be replaced. Plan to remount top left detector with the same harnesses.
Plan to mount two more detectors on the right hand plate. |
Attachment 1: 20230428_181020.jpg
|
|
Attachment 2: 20230428_181123.jpg
|
|
Attachment 3: 20230428_181126.jpg
|
|
Attachment 4: 20230428_181129.jpg
|
|
Attachment 5: 20230428_181131.jpg
|
|
Attachment 6: 20230428_181133.jpg
|
|
Attachment 7: 20230428_181139.jpg
|
|
Attachment 8: 20230428_181141.jpg
|
|
Attachment 9: 20230429_150436.jpg
|
|
Attachment 10: 20230429_150717.jpg
|
|
Attachment 11: 20230429_151421.jpg
|
|
Attachment 12: 20230429_151425.jpg
|
|
Attachment 13: 20230429_151428.jpg
|
|
Attachment 14: 20230429_151431.jpg
|
|
Attachment 15: 20230429_151432.jpg
|
|
Attachment 16: 20230429_151435.jpg
|
|
Attachment 17: 20230429_151436.jpg
|
|
Attachment 18: 20230429_151444.jpg
|
|
Attachment 19: 20230429_151449.jpg
|
|
Attachment 20: 20230429_151453.jpg
|
|
Attachment 21: 20230429_152026.jpg
|
|
Attachment 22: 20230429_152045.jpg
|
|
Attachment 23: 20230429_160614.jpg
|
|
Attachment 24: 20230429_162251.jpg
|
|
Attachment 25: 20230429_162257.jpg
|
|
Attachment 26: 20230429_172209.jpg
|
|
391
|
Sat Apr 29 10:12:58 2023 |
CB, JM | CARME opened |
(28 April)
Mounted rails on dolly. The rails don't smoothly connect to those on the Aluminium frame and some thin metal shims were required to get the correct height. A metal support had to be placed above the Delrin pads to support the rails above the gap between frames.
All side feet were mounted but extensive modifications were required to slotted feet due to them not being manufactured to drawing, and to Carme middle section flange width being incorrect (see previous entry).
Removed brakes and started sliding CARME out. The dolly and the aluminium frame came partly separated as the weight of the side feet was shifted on the latter. Strapped them together. No damage.
Kept sliding. Sliding when the carriages passed from one rail to the other was quite hard. The back carriage on the left looking downstream lost a single ball bearing when crossing over. A more sophisticated solution is probably required.
Once Carme was on the dolly, opened it using M8 jacking bolts and lifted away front and middle section.
No obvious damage inside but strain relief did not work as intended. Photos to be uploaded after more detailed inspection on Saturday.
Plan to install right hand side detectors over the weekend. Unclear if bottom left detector is damaged and needs to be replaced. |
Attachment 1: 20230428_094931.jpg
|
|
Attachment 2: 20230428_155649.jpg
|
|
Attachment 3: 20230428_164926.jpg
|
|
Attachment 4: 20230428_155748.jpg
|
|
Attachment 5: 20230428_165448.jpg
|
|
Attachment 6: 20230428_174021.jpg
|
|
Attachment 7: 20230428_181020.jpg
|
|
Attachment 8: 20230428_181919.jpg
|
|
390
|
Thu Apr 27 19:43:23 2023 |
CB, JM | Preparing to open |
Removed turbo pump and set aside in a box. Considering switching to the other turbo to avoid degradation due to lack of use.
Installed bellows support and OG devised support to avoid bellows sagging when Carme is split open.
Attempted to install new CARME feet, but there were issues. Feet with slots produced do not match design drawings by PB. Slot is in the wrong side. Furthermore Carme interaction chamber flange thickness is 36 mm vs. 36.5 mm in drawing. DR can modify feet to make them fit tomorrow.
Now just waiting for rails to open CARME. Will arrive tomorrow or on Tuesday, after weekend and Mayday.
|
Attachment 1: 20230427_123456.jpg
|
|
Attachment 2: 20230427_161602.jpg
|
|
389
|
Wed Apr 26 16:49:28 2023 |
CB, JM | CARME dolly mostly built |
Disconnected cables from CARME motors and turbo. The turbo has to go to allow Carme to slide out. Crane may be required to lift because it's in a very hard spot to access.
Dolly mostly built. See attached. Left side doesn't quite fit right by 1-2 mm. Might be okay anyway but will put some shims to avoid any shaking during sliding.
Waiting for the rails to arrive due to GSI procurement issues.
Plan to mount new feet and bellow support tomorrow.
Also plan to split CARME at the smaller flange towards the interaction chamber, after the bellows and before the bellows towards the magnet on the downstream side. |
Attachment 1: 20230426_152121.jpg
|
|
Attachment 2: 20230426_162619.jpg
|
|
Attachment 3: 20230426_162627.jpg
|
|
388
|
Tue Apr 25 13:22:19 2023 |
JM | CARME venting |
CARME venting using nitrogen gas bottle.
Coolant pipes dismounted from the chamber
Feed-through flange for n+n bias flange for bottom detector has a bent pin image attached. Clearer image of bent pin for ERNI connector used for aida03 also attached. All other feed-throughs and adaptor cards appear undamaged. |
Attachment 1: 20230425_134119.jpg
|
|
Attachment 2: 20230425_132944.jpg
|
|
387
|
Mon Apr 24 21:24:26 2023 |
JM | Preparation for CARME dismount |
CARME currently in vacuum, pressure ~3E-10 mbar. Chamber to be vented tomorrow afternoon by vacuum division. We have active both pirani and hot cathode gauge to aid.
Equipment removed from the chamber in preparation for dismount of CARME from the ring this week. Equipment removed includes:
- Neg pump cabling
- Heating wire
- FEE electronic units and associated cabling (power, hdmi etc)
- Stainless steel FEE holders
- Red and blue cooling hoses
Attached image shows chamber following equipment removal. Adaptor cards still attached to feed-through flanges. ERNI connnectors examined, all fine except for FEE aida03 which has a bent pin (see attached image). Will dismount adaptor cards to examine further tomorrow. Feed-through flanges will also be examined to check for bent pins following adaptor card removal.
New CARME frame is part constructed and is in the CRYRING area. Requires rails and castors which should arrive this week. The height of the frame will require shortening, to be level with the current frame. This will be looked into further tomorrow. |
Attachment 1: 20230424_155635.jpg
|
|
Attachment 2: 20230424_155607.jpg
|
|
Attachment 3: 20230424_150526.jpg
|
|
386
|
Wed Mar 22 16:33:32 2023 |
TD | NIM WR source |
Test of NIM WR source at Edinburgh
Attachment 1 - NIM WR source
NIM module upper SMA - DSO ch 1
NIM module lower SMA - DSO ch 2
Attachment 2
NO SYNC or CLK signals observed.
Noise observed with NIM bin power ON *and* OFF
Attachment 3
NIM WR source in EG&G Ortec NIM bin
... and with the Digilent Zybo board switched ON!
Attacments 4-7
Trigger ch 2
4, 40, 400 & 4000ns/div
Attachment 8
Trigger ch 1
4ns/div |
Attachment 1: 20230322_155735.jpg
|
|
Attachment 2: 20230322_160448.jpg
|
|
Attachment 3: 20230322_160453.jpg
|
|
Attachment 4: 20230322_181015.jpg
|
|
Attachment 5: 20230322_181027.jpg
|
|
Attachment 6: 20230322_181045.jpg
|
|
Attachment 7: 20230322_181054.jpg
|
|
Attachment 8: 20230322_180945.jpg
|
|