Fri Sep 29 08:54:29 2023, TD, Friday 29 September 11x
|
09.50 CARME IE514 9.0e-11 mbar, ambient temperature 21.3 deg C
|
Fri Apr 26 07:05:01 2024, TD, Friday 26 April 19x
|
08.02 DAQ continues data file R1_14
|
Fri Sep 25 12:58:33 2020, JM, Friday 25/09/20
|
10:00 Vessel P = 3.99e-8 mbar, RGA P = 1.4e-6 mbar
10:54 Vessel P = 3.89e-8 mbar, RGA P = 1.4e-6 mbar
Vessel continues to pump down. Baking Jacket fully installed and testing of baking begins before leaving on automatic mode for the week. Some |
Fri Jan 19 11:17:50 2024, JG, TD, Friday 19 January 18x
|
12.20 CARME IE514 4.1e-11mbar, ambient temperature 16.7 deg C
|
Fri Apr 19 02:47:59 2024, TD, Friday 19 April 12x
|
03.45
|
Fri Oct 15 19:37:12 2021, RSS TD + DR, Friday 15 October 8x
|
Ring seal is out. See picture 1.
Bellow (towards CARME) was installed in the gas jet chamber. Attached are some pictures. |
Fri Jan 14 10:00:18 2022, TD, OH, JM, Friday 14 January
|
11.00 Per https://elog.ph.ed.ac.uk/CARME/170 and https://elog.ph.ed.ac.uk/CARME/171 issues have been identified with the FEE64 carriers and adaptor PCBs
|
Fri Aug 13 07:28:37 2021, RSS TD, Friday 13-17 August 23x
|
Pirani IE514
mbar mbar
|
Wed Oct 13 12:06:09 2021, RSS TD, Flanges cleaned 12x
|
RSS & TD had some success cleaning the COF600 flange surfaces of the staining and residues observed when the downstream COF600 section was opened yesterday
https://elog.ph.ed.ac.uk/CARME/118
|
Fri Nov 5 12:27:11 2021, CB RSS + DR NP ML US, First pump down of YR09    
|
Tightened all flanges.
Added Pirani gauge and third IE514 to interaction chamber.
Stated pump down. Around 100 mbar on Interaction chamber Pirani gauge very loud metallic noise is heard. Found bellows between CARME and interaction |
Wed Sep 8 10:52:54 2021, CB RSS + OG, First feedthrough flange mounted    
|
Assembled feedthrough flange with re-cut
SS support bracket. M3 vented on the
sides, M3x30 partly threaded on top. Top
|
Mon Oct 5 08:21:40 2020, CB (remotely), JM, First bake of CARME with pumps (cont.) 6x
|
5 OCTOBER
08:21 Temperatures stable overnight. See attached. Baking continues.
|
Wed Sep 30 08:10:36 2020, CB (remotely), First bake of CARME with pumps 19x
|
(NOTE: Raspberry Pi internal clock appears to be on GMT, rather than GMT+1 due to daylight saving.)
Baking started on Mon 28 Sep. Due to WiFi connection authorisation expiring, no remote access was possible until
|
Fri Nov 12 08:54:11 2021, RSS ML + CB (remotely) NK AK NP, First CARME baking attempt  
|
07:50 Small leaks (near the motors and upper valve) were detected in the tent which were covered with Al foil.
08:10 ML changed the baking temperature to 120 degrees.
09:40 Since the bottom of CARME is cold, due to uneven distribution of heat, baking is stopped. |
Mon Oct 12 10:49:36 2020, JM, First Bakeout of CARME with pumps (cont)
|
12 October
10:00 Bakeout temperatures stable. Pressure on the backing line is 3e-7mbar which is the lowest we have achieved on the backing line so far.
Vessel pressure can't be measured at the moment as temperature exceeds the max operational temperature of 80C. |
Sat Apr 5 16:15:04 2025, TD, FEE64 power interlock installed 11x
|
Water flow and dew point sensors have been installed on the outlet and inlet cooling water manifolds respectively.
|
Sun Jan 14 12:31:32 2024, JM, CB, FEE positions mac addresses
|
*Updated following FEE changes. Correct as of 31/01/24
FEE modules are labelled according to the diagram in attachment. Note attachment is looking downstream. Each FEE module is also labelled with
its assigned number and whether it is a bias connection on the FEE holders. MAC addresses and FEE descriptions are as follows; |
Sat Jan 15 15:38:25 2022, JM, OH, TD, FEE installation, cable management    
|
All 8 FEEs have been installed in their holders.
The damaged feedthrough does not have the D-connector on the adaptor board inserted into the falnge but the FEE is connected to the adaptor board.
FEE's were carefully installed, with one person inserting FEE and another making sure FEE was straight and on the correct pin positions. |
Wed Jan 12 20:07:20 2022, JM, OH, TD, FEE holder installation 10x
|
4 FEE holders installed on CARME. Mounting procedure devised due to limited working space to install.
Mounting procedure:
>Install FEE holder mounts (image 1) onto feedthrough flange |
Wed Apr 24 13:18:09 2024, TD, Examples of poor CARME cable management/routing 
|
Have observed some instances of poor CARME cable management/routing
Attachment 1 - SAES D-2000 ion pump HV cable (netal braid) transiting FEE64 cabling
|