  568   Fri May 31 21:53:20 2024 CB EM, JM,EM ABOvernight beam tests

22:53

Run 15 started

Beam tests. We increased the storage time to 2min and switched to bunched beam so that we can see the beam lifetime over the cycle.

D+ Beam energy is set to 1.5 MeV.

Detectors are positioned in with servos only (left arm = 28.06mm, right arm = 29.23 mm)

No target

We observe a large peak at the start of the cycle at 750 keV. This seems to be molecular hydrogen circulating in the ring. We also see a much smaller peak at 1.5 MeV (D+)

After a period of time the 750 keV peak decreases significantly and we are left with the 1,5 MeV peak only. 

Total energy histogram after running for ~ 1 hour - attachment 1

Total xy plot after ~ 1 hour - attachment 2

We observe the large circular structure on the left hand detectors. Is this related to hydrogen or is it the beam type (bunched instead of coasting)?

In the GSI control plots for the number of ions in the ring we observe a large drop in ions circulating at the start of the cycle, and then remains fairly constant throughout the rest of the cycle. - attachment 3

This seems to confirm our hydrogen hypothesis.

23:54

Run 15 stopped, Run 16 started.

All setting the same and still no target, except beam is now coasting. We will compare to Run 15 after one hour of running.

We do not observe the circular distribution on the detector seen in the previous run. It seems this effect comes from the bunched beam. Reason?

We still observe the 750 keV peak at the start of the cycle.

Total energy histogram after running for ~ 1 hour - attachment 4

Total xy plot after ~ 1 hour - attachment 5

00:59

Run16 stopped. Run 17 started.

All settings the same except we have switched back to bunched beam.

Target now on. (X=-13700, Y=0). Target density ~ 1E11 at/cm2.

02:03

Run 17 continues

Target has increased in the average density to ~4e11 at/cm2, no changes made to nozzle position and fluctuates to 1e11 or 1e12 at/cm2.

Energy histograms and the xy plots appear similar to bunched beam with no target. The rates are slightly lower with the target for the left hand dssds, maybe target interactions destroy molecular hydrogen quicker?

Total energy histogram after running for ~ 1 hour - attachment 6

Total xy plot after ~ 1 hour - attachment 7

We will continue to run for another hour or so in this condition

Target density has decreased over the last minutes. We have tried to move the nozzle to maximise density but fluctuations make this difficult.

03:55

Increase in ADC data items statistics for aida 13 and 14 from ~20 kHz to 90 kHz

ASIC checkload has reduced the rates back to ~ 20 kHz

Total energy histogram after running for ~ 3 hour - attachment 8

Total xy plot after ~ 3 hour - attachment 9

Run 17 stopped. Run 18 started

Beam changed back to coasting. All other settings unchanged. Target on. 

04:55

Run 18 continues.

Energy histograms and the xy plots appear similar to bunched beam with no target. The rates are lower with the target than without.

Total energy histogram after running for ~ 1 hour - attachment 10

Total xy plot after ~ 1 hour - attachment 11

Note we observed fluctuations in the T2 temperature for the target. T2 is set to 80 K but rapid fluctuations were observed up to 100K. This occurred for only 5 minutes and I am not sure if it is a real effect.

 

07:20

Temperatures OK

System wide checks OK

Rates OK

Stats OK

Waveforms attached

Leakage current decreasing slightly.

No changes to beam or target.

  567   Fri May 31 21:03:21 2024 CB, EMRun14

22:03

Before this run we checked quickly  beam with the target on but it does not make sense with the current status of the beam.

We removed target and starte Run14

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

  566   Fri May 31 19:12:01 2024 CB, EMRun 13 - test with beam

20:00

The beam is set at the correct energy, operators cross-checked it again. What we see  at 750 is due to the molecular beam.  The operotors are trying to optimize and get rid of this in the next 30min.

Temps and system checks OK

Leakage current OK

Merger and tape server OK

Potentiometer at 28.06 mm and 29.23 mm

20:30

Keep going with the beam tests. A new peak appears around 950 keV( See attached). Still try to understand what it is. The operators keep going

  565   Fri May 31 06:01:55 2024 JM, TY, PW, EM, CBRun 12 - beam on target

07:00

Run 12 started. D+ beam is incident on the target at 1.5 MeV

Beam intensity in the ring is ~4e7 particles

Target density shows some fluctuations but is ~1e11 atoms/cm2. Target temp 35K

Detectors are positioned fully out (Left arm 18.86mm and right arm 19.8 mm ) and are not moving

Ring pressure looks OK, YR09 E target sensor seems unaffected by the target currently reads ~5e-11 mbar.

Beam is in the ring for ~15s.

Temps and system checks OK

Leakage current OK

Merger and tape server OK

Horizontal beam position 8 mm and vertical beam position 0mm for overlap with target.

Statistics - attachment 1

Energy histogram - attachment 2

XY plot - attachment 3

We see high rate at 1.5 MeV in left dssds but very little in right dssds. XY plots also show counts are concerntrated in a very circular distribution.

07:15

Gleb has just called and requires access to the cave to check on a turbo pump in the source.

Ring vacuum looks good

Beam will be down for 10-15 minutes only hopefully. 

07:45

Beam is back in the ring but injection requires optimisation.

09:00

There are some issues with the source that require a few hours to resolve. Potentially until 7 or 8 pm tonight. We have about an hour before work starts.

We have done a quick test of the motors and moved the left and right arms in using the pnuematic only.

New position left arm = 62.71 mm , right arm = 64.6 mm.

We observed the rates in left detectors went up significantly. In the right detectors we observed little increase.

See attached xy plot - attachment 4

See attached energy histogram - attachment 5

Beam intensity has dropped significantly - below 1e6 particles so even at in positons we see few counts at the moment.

09:25

Source has degraded pretty rapidly so beam will be stopped for now.

Detectors moved back out.

Target nozzle moved from X= -13000, Y=2500 to X= -11000, Y=2500. S1 has dropped to 5e-10 mbar.

Note Gleb mentioned that last night it was likely a large proportion of the ions stored were molecular hydrogen not deuterium.

16:18

Temps and system checks OK

Leakage current OK

Merger and tape server OK

 

18:10

Temps and system checks OK

Leakage current OK

Merger and tape server OK

 

19:38

Operators are trying to optimize some beam. For now the beam is lower than it should be probably because is easy to work with the source like that.

Temps and system checks OK

Leakage current OK

Merger and tape server OK


 

  564   Thu May 30 13:31:01 2024 JM, CB, PW, YLRun 11 - beam on target overlap

14:30

Run 11 started. D+ beam in the ring at 1.5 MeV.

No target at the moment. Valves to the interaction chamber will soon be opened.

Stats and histograms zero'd, monitoring code reset.

All checks and temps OK

Leakage currents OK

ADC data items statistics - attachment 1

Beam in ring ~30s and manipulation ~2s

Beam intensity ~ 5.7E7

 

16:40

System check OK

Temperature OK

Statistics same as before

Merger and server OK

Leakage currents OK

Beam in the ring

Waiting for target , Hydrogen green

 

18:50

System OK

Temperature OK

Statistics same as before

Merger and server OK

Leakage currents OK

Beam ~4x10^7

Target tuning ongoing

 

22:05

System OK

Temperature OK

Statistics same as before

Merger and server OK

Leakage currents OK

Currently no beam -> asked operator to check source

Target tuning ongoing

 

23:22

System OK

Temperature OK

Statistics same as before

Merger and server OK

Leakage currents OK

Currently no beam -> operators are optimizing the intensity of the beam

Optimizing target density as well

 

00:55

The overlaping test started by moving the beam only in the y axis.

 

02:00

Overlap test is over and was found that the optimal position for the horizontal axis is 8mm. Beam on target.

Observation of an increase the 1.5 MeV peak in the CARME monitoring code when overlap with the target is found. Attachment 4.

Drop in the beam lifetime is also observed when overlapped with the target.

System OK

Temperature OK

Statistics same as before

Merger and server OK

Leakage currents OK

 

02:48

CRYRING measurement cycle has been made shorter (from 5 to 3 repetitions) to match it with the decreased beam life time due to optimed overlap with target.

 

04:37

System OK

Temperature OK

Statistics same as before

Merger and server OK

Leakage currents OK

 

 

 

  563   Thu May 30 09:30:33 2024 CB, PW, YL, TDRun 10

10:30

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

10.48 FEE64 temperatures OK - attachment 1

WR timestamps OK - attachment 2

ADC data item stats - attachment 3

per FEE64 Rate spectra - attachment 4

 why do we observe HEC data in aida15 asic #1?

per FEE64 1.8.L spectra - attachments 5-6

 ADC to 20MeV FSR - not clear when these spwectra were last zero;d

 currently 750keV/u D + no target ?

10.53 All histograms & stats zero'd

Merger, Taper Server etc - attachments 7-8

  bind error 98 for NewMerger server - not clear when these messahes were generated - added some returns to monitor whether these, or other, messages are repeated

DSSSD bias & leakage current OK - attachment 9

11:30

We noticed some odd peak structure in the rate energy histograms for the right hand side detectors. We took away the beam but this structure persisted. 

This was not observed in the total energy histograms, the reason for this is the applied multiplicity cut to the total histograms. Removing this cut the peaks are clearly visible (attachment 10). Putting the cut back in we see the peaks removed (attachment 11). These events correspond to some noisy strips on the detectors on the right hand side (see attachment 12). Notice the much larger rate on the RHS. Current ADC data items statistics show FEE modules for the RHS also much larger (attachment 13). We did not observe this with no beam (https://elog.ph.ed.ac.uk/CARME/554) but did with beam last night (https://elog.ph.ed.ac.uk/CARME/558) so there nust be some effect of having beam in the ring increasing the number of events on the RHS. Note the detectors are fully out and the events are localised in these strips. 

Note we have moved the detectorsand removed beam from the ring in this time period to investigate these peaks.

12:50

Run 10 continues. Temps OK. System checks OK. Merger OK, Tape server OK.

Leakage current OK.

Notice the 'beam peak' at 1.5 MeV is larger in the left hand side (attachment 14)

14:26

We have started the new_optimeren.py on pi@gas-stepper to try and align the gas target. Valves to the interaction chamber are closed.

Setting the nozzle X=-13000, Y=1000 we observe an increase in E4 from 4E-10 mbar to 2E-9 mbar. Moving either side we observe a drop in E4. We take this as the gas target being aligned.

Anton will now open the valves to the gas target.

Run 10 stopped.         

  562   Thu May 30 08:33:21 2024 CB, PW, YLRun 8 & 9- detector motion tests

09:34 - Moved to Run 9 in preparation of moving detectors. No target yet. Beam in the ring at 750 keV/A.

Temperatures OK (attach)

System wide checks OK except 16 fails clock and ADC calib

Rates OK (attach)

Writing to disk OK (attach)

Resetting the sorting script. Moving detectors.

Moved detectors in. See attached. Stripe / shadow quite clear. Some oddities on the top-right DSSD. May be due to ExEy cuts and multiplicity cuts.

Moved on tun run 9 to acquire some stats without motion.

 

10:13 Temperatures still OK. Rates OK. Leakage currents stable.

We were cutting the beam.

Increased repetition to x5.

Motors fully out - run 10 started.

  561   Thu May 30 07:25:53 2024 Yuri, Phil, NikosGas target

The cave is closed and beam is back

The d intensity is about half of the intensity before the break

The gas target is prepared but the valvas cannot be opened. Anton is coming in at around 1 PM to open the gate valves.

  560   Thu May 30 06:00:51 2024 Yuri Nikos ...Open the cave

Nikos needs to enter the cave to work on the target cooling.

 

  559   Wed May 29 17:42:06 2024 TDOffline analysis APR24/R2 (pulser walkthrough)
ADC offsets using pulser walkthrough data from data file APR24/R2

ch = channel + ( module * 64 ) + ( range * 2048 )

adc_data( ch ) = INT( RSHIFT( ABS( adc_data( ch ) - 32768 ), 3 ) - offset( ch ) + 0.5 )




per FEE64 ADC spectra - attachments 1-16

Note common x & y-scales - peak height proxy for peak width

aida01 1.8.L (s7) pulser peak width 9 ch FWHM (~50keV FWHM)




ADC offset analysis

/home/td/Calibration/offsets /home/td/CARME/R2 50 2048 20 5

First pass sort variables NAMELIST I/O format ( ADC offsets, FEE64 configuration, LEC & HEC equal energy cuts ) - attachment 17
  558   Wed May 29 16:46:21 2024 NP, AB, PMRun7 - Measurements only with ion beam.

18:19 Start or run 7: In beam has been optimzed up to now by Zoran and Claude. ~7E7 particles at injection. From now on, only the injection beamline is optimized by Operators.

Detectors IN (left arm pot =18.86 mm, right arm pot = 19.80 mm)

No Target

Beam is in the ring for ~6 seconds and total cycle time between injections is ~8 seconds

Injection beamline is still in process of optimization by operators

21:30

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

23:30

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

01:30

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Beam and Electron Coller failure (~00:35); Beam back (~01:08); Cooler back (~01:20)

03:00

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

Beam and Electron Coller failure (~01:35); Beam back (~01:55); Cooler back (~01:55)

04:40

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

06:00

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

07:00

Merger OK, Tape server OK

ADC data item statistics looks as before. OK

FEE temperatures look OK

System checks OK

Leakage current stable

 

 

  557   Wed May 29 09:49:35 2024 JM, CBThreshold changes / D+ beam

09:00

R4 alpha background continues, currently R4_12

Temps OK, system checks unchanged from previous status (OK, aida16 clock and ADC issue)

10:30

R4 stopped and moved to the next run R5.

Lithium 3+ beam is not feasible at sufficient densities so we are switching to D+ at a beam energy of 0.75 MeV/u. A measurement of d+p at this energy would produce 3He at energies of ~900 keV. 

The current threshold (slow comparator) has been 1 MeV (0x64). We lowered the thresholds in all FEE modules to 200 keV (0x14). ADC rates look fine in all FEEs (~tens of Hz) except in aida 15 and aida 16 which was ~several hundred kHz. Data rate to disk was 1.2 Mb/s

We increased the threshold in aida 15 and 16 only to 500 keV (0x32) and rates in aida 15 and 16 have dropped to <20 kHz. Data rate to disk now 220 kB/s.

11:00

R5 stopped. R6 started - background with reduced thresholds. 

Attachment 1 - ADC data items statistics

Attachment 2 - Histogram browser statistics

Attachment 3 - Tape server

Attachment 4 - New Merger

13:12

R6 continues R6_9

Edits to the CARME sort code have been made. EyEx energy cut for online monitoring reduced to 100 keV. Multiplicity cut applies to lowEnergyEy, lowEnergyEx and lowEnergyXY histograms. Multiplicity cut considers only single strip events.

15:00

D+ beam is now in the ring and is being optimised by the beam operators. The CARME detectors are currently fullt out, but we can see some effects of the beam on the online monitoring code.

  556   Wed May 29 08:10:32 2024 JMRun4 - alpha background

09:00

R4 alpha background continues, currently R4_12

Temps OK, system checks unchanged from previous status (OK, aida16 clock and ADC issue)

  555   Tue May 28 15:25:21 2024 TD, JM, CBPulser tests

R3 started

Pulser ON. Approx 2 kHZ.

1.8*w waveforms attachments 1+2

ADC data items statistics - attchment 3

1.8*L spectra - attachments 4+5

Pulser widths are larger than during testing prior to the Feb beamtime. Widths in Channels;

aida01 = 76

aida02 = 82

aida03 = will upate all later tonight..

R3 ended. Pulser OFF

17:15

R4 Started. Alpha background run.

 

  554   Tue May 28 12:38:31 2024 TD, JM, CBRun 2 - Alpha background run
13.35 FEE64 temperatures OK - attachment 1

      ADC data item rates OK - attachment 2
       overall rate dominated by 2x hot channels aida15

     per FEE64 Stat spectra - attachment 3
      spectra consistent with April alpha background test 
      note high HEC data stats for aida12 asic#2 - can probably be fixed by ASIC control check

     per 1.8.W spectra - 20us FSR - attachments 4-5
      note 100kHz ripple observed for some FEE64s cf. https://elog.ph.ed.ac.uk/CARME/544

Run stopped 15:15 
  553   Mon May 27 15:25:08 2024 JMStartup

27/05/24

The carme-gsi computer has been moved from the server room into the shifter control room.

Startup of the DAQ looks OK. Settings loaded 2024Apr24

System wide checks OK. Aida16 clock status fails. Aida16 ADC calibration fails. Rest OK.

All FEEs look OK, we again see high rates in aida15 ~ 12k data items/sec, rate in other FEEs is 0.

Testing of the pulser widths has been done. Pulser widths will be uploaded soon.

Background run started. No beam, no target. File G22-201 and R2. Data being written is ~128 kb/s. Run will be left overnight, no night shift.

28/05/24

DAQ continues with alpha background run.

System wide checks OK. Aida16 clock status fails. Aida16 ADC calibration fails. Rest OK. Temps OK

Data can be seen in the spectrum browser on MIDAS. CARME monitoring code was started incorrectly yesterday. It has now be restarted and a small number of events are seen

  552   Fri May 10 21:23:35 2024 TD, CB, JM20Ne7+ test beam

22.00

Carsten Brandau stopping e-target experiment Unattended beam overnight. No operators watching.

10^5 (nominal) 20Ne7+ at 380keV/u beam available

Carsten reports beam is unstable, often disappearing, and in bunched mode.

4x MSL type TTT14 DSSSDs not moving, locked in out position

Starting CARME MIDAS DAQ ASIC settings 2024Apr24-14.14.03

slow comparator 0x64 E

G&G Ortec 448 Pulser OFF

DSSSD bias & leakage currents - attachment 1

FEE64 temperatures OK - attachment 2

All system wide checks OK, WR timestamps OK - attachment 3

ADC data item stats - attachment 4

Rates dominated by two hot channels aida15 per FEE64 Stat spectra - attachment 5

per FEE64 1.8.L spectra - attachments 6-7 no obvious beam energy peak, continuum of energies from below beam energy to threshold

per FEE64 1.8.W spectra - attachments 8-9

 

22.55 Start data file APR24/R5 Merger, TapeServer etc - attachments 10-12

270k data items/s

 

23:29 Activated CARMESort script. See attachments 13-15.

No obvious beam or signal of any kind. Temperatures OK.

 

00:13 DAQ continues OK. System-wide check OK. Temps OK. Still no clear sign of beam.

 

01:03 Daq continues OK. System wide checks OK, Temps OK.

Only low energies observed in the energy spectrum. Looks a bit like the low energy noise observed in our previous runs with no/very few counts above what would be beam energy (~7.6 MeV)

Note the xy plots are not all orientated correctly (strips are correct in the sort code but root axis need to be flipped - I can;t remember how atm). The part of the plots with counts are those closest to the beam. 

02:00 DAQ continues OK. System wide checks OK. Temps OK.

Spectra still looks the same. I have noticed the right hand side rate is higher than the left. EyEx cut spectrum also looks like much more counts outside of the cut.

03:40

DAQ continues. All checks and temps OK.

I can now see a small peak at 5 MeV in the bottom left dssd. This dssd also has the least number of counts in the low energy continuum.

I have logged back into grafana on x2go to get the number ions in the ring. Attachment 17

04:23 (05:23 EU)

DAQ stopped, merger stopped, tape server stopped. Bias OFF. FEEs turned OFF. Screen shots taken of the monitoring code plots - not yet uploaded.

Windows and servers remain up in case we want to do some more tests in the near future.

Attachment 18 shows beam intensity for total run time from Grafana

  550   Sun Apr 28 23:12:11 2024 TDMonday 28 April
00.09 DAQ continues data file R3_21

      DSSSD bias & leakage current OK - attachment 1

      FEE64 temps OK - attachment 2

      ADC data item stats OK - attachment 3

      per FEE64 Rate spectra - attachment 4

      Merger, TapeServer etc - attachments 5-7

04.50 DAQ continues data file R3_22

      DSSSD bias & leakage current OK - attachment 8

      FEE64 temps OK - attachment 9

      ADC data item stats OK - attachment 10

      per FEE64 Rate spectra - attachment 11

      Merger, TapeServer etc - attachments 12-14


09.41 DAQ continues data file R3_23

      DSSSD bias & leakage current OK - attachment 15

      FEE64 temps OK - attachment 16

      ADC data item stats OK - attachment 17

      per FEE64 Rate spectra - attachment 18

      Merger, TapeServer etc - attachments 19-21


09.43 DAQ stopped
      DSSSD bias OFF
      FEE64 power OFF
  549   Sat Apr 27 23:16:02 2024 TDSunday 28 April
00.12 DAQ continues data file R3_15

      DSSSD bias & leakage current OK - attachment 1

      FEE64 temps OK - attachment 2

      ADC data item stats OK - attachment 3

      per FEE64 Rate spectra - attachment 4

      Merger, TapeServer etc - attachments 5-7


05.32 DAQ continues data file R3_16

      DSSSD bias & leakage current OK - attachment 8

      FEE64 temps OK - attachment 9

      ADC data item stats OK - attachment 10

      per FEE64 Rate spectra - attachment 11

      Merger, TapeServer etc - attachments 12-14

11.27 DAQ continues data file R3_17

      DSSSD bias & leakage current OK - attachment 15

      FEE64 temps OK - attachment 16

      ADC data item stats OK - attachment 17

      per FEE64 Rate spectra - attachment 21

      Merger, TapeServer etc - attachments 18-20

11.33 aida16 asic #2 2x Check ASIC control
      aida16 asic #4 2x Check ASIC control


18.53 DAQ continues data file R3_19

      DSSSD bias & leakage current OK - attachment 22

      FEE64 temps OK - attachment 23

      ADC data item stats OK - attachment 24

      per FEE64 Rate spectra - attachment 25

      Merger, TapeServer etc - attachments 26-28

      per FEE64 1.8.H - attachments 29-30

      per FEE64 1.8.L - attachments 31-32

      per FEE64 1.8.W - 20us FSR - attachments 33-36

19.10 CRYRING ambient temperature per https://web-docs.gsi.de/~lestinsk/tempplot.php - attachment 37
  548   Fri Apr 26 23:41:16 2024 TDSaturday 27 April
00.34 DAQ continues data file R3_9

      DSSSD bias & leakage current OK - attachment 1

      FEE64 temps OK - attachment 2

      ADC data item stats OK - attachment 3

      per FEE64 Rate spectra - attachment 4

      Merger, TapeServer etc - attachments 5-7

06.41 DAQ continues data file R3_10

      DSSSD bias & leakage current OK - attachment 8

      FEE64 temps OK - attachment 9

      ADC data item stats OK - attachment 10

      per FEE64 Rate spectra - attachment 11

      Merger, TapeServer etc - attachments 12-14


10.56 DAQ continues data file R3_11

      DSSSD bias & leakage current OK - attachment 15

      FEE64 temps OK - attachment 16

      ADC data item stats OK - attachment 17

      per FEE64 Rate spectra - attachment 18

      Merger, TapeServer etc - attachments 19-21

14.33 DAQ continues data file R3_12

      DSSSD bias & leakage current OK - attachment 22

      FEE64 temps OK - attachment 23

      ADC data item stats OK - attachment 24

      per FEE64 Rate spectra - attachment 25

      Merger, TapeServer etc - attachments 26-28

18.31 DAQ continues data file R3_13

      DSSSD bias & leakage current OK - attachment 29

      FEE64 temps OK - attachment 30

      ADC data item stats OK - attachment 31

      per FEE64 Rate spectra - attachment 32

      per FEE64 1.8.L spectra - attachments 33-34

      Merger, TapeServer etc - attachments 35-37