  CARME, Page 7 of 34  ELOG logo
  ID Date Author Subject
  547   Fri Apr 26 07:05:01 2024 TDFriday 26 April
08.02 DAQ continues data file R1_14

      DSSSD bias & leakage current OK - attachment 1

      FEE64 temps OK - attachment 2

      ADC data item stats OK - attachment 3

      per FEE64 Rate spectra - attachment 4 

      Merger, TapeServer etc - attachments 5-7

10.40 Close file R1 and open file R2

      Switch on pulser (relay line 50Hz)

      per FEE64 Rate spectra - attachment 8

      Check ASIC load x2 for aida09, aida10, aida12, aida15  aida16

      per FEE64 Rate spectra - attachment 12
       status quo ante

      Pulser walkthrough 

      Pulser settings per https://elog.ph.ed.ac.uk/CARME/544
      Amplituides 70,000-10,000 @ 10,000 step

      per FEE64 1.8.L spectra - attachments 9-10
 
      per FEE64 Rate spectra - attachment 11

11.05 Pulser OFF
      Close file R2, open file R3

      zero all histograms  stats

      background alpha run continues


20.33 DAQ continues data file R3_8

      DSSSD bias & leakage current OK - attachment 13

      FEE64 temps OK - attachment 14

      ADC data item stats OK - attachment 15

      per FEE64 Rate spectra - attachment 16

      Merger, TapeServer etc - attachments 1-19


      CAEN data log stopped 00:21 Friday 26 April 

pi@nnrpi1:~/Programs/caenlogger $ caenlogger --config caenlogger_config_gsi.toml
2024-04-25 15:25:16.194 | INFO     | caenlogger.__main__:main:117 - Config file provided.
Traceback (most recent call last):
  File "/home/pi/.local/bin/caenlogger", line 10, in <module>
    sys.exit(main())
  File "/home/pi/.local/lib/python3.7/site-packages/caenlogger/__main__.py", line 135, in main
    curses.wrapper(process, config_dic)
  File "/usr/lib/python3.7/curses/__init__.py", line 94, in wrapper
    return func(stdscr, *args, **kwds)
  File "/home/pi/.local/lib/python3.7/site-packages/caenlogger/__main__.py", line 57, in process
    Vmon = [ f for f in HV[5].strip().split("  ") if f ]
IndexError: list index out of range

     Old log gzip'd and moved to directory ~/logs

[npg@carme-gsi ~]$ mv caen_current.dat.gz logs/0031260424-caen_current.dat.gz

20.45 Restarted logger program
  546   Thu Apr 25 02:31:25 2024 TDThursday 25 April
03.27 DAQ continues - file APR24/R1_7

      DSSSD bias & leakage current OK - attachment 1
      ambient temperature 16.1 deg C

      FEE64 temperatures OK - attachment 2

      ADC data item stats OK - attachment 3

      per n+n FEE64 1.8.L spectra - attachment 4

      per FEE64 Stat spectra - attachment 5

      Merger, TapeServer etc - attachments 6-8


10.19 DAQ continues - file APR24/R1_8

      DSSSD bias & leakage current OK - attachment 9
      ambient temperature 16.2 deg C

      FEE64 temperatures OK - attachment 10

      ADC data item stats OK - attachment 11

      per FEE64 Stat spectra - attachment 12

      Merger, TapeServer etc - attachments 13-15

      NetVar merge variables - attachment 16


14.44 DAQ continues - file APR24/R1_10

      DSSSD bias & leakage current OK - attachment 17
      ambient temperature 16.5 deg C

      FEE64 temperatures OK - attachment 18

      ADC data item stats OK - attachment 19

      per FEE64 Stat spectra - attachment 20
       some HEC data has appeared

      Merger, TapeServer etc - attachments 21-23

19.12 DAQ continues - file APR24/R1_11

      DSSSD bias & leakage current OK - attachment 24
      ambient temperature 16.5 deg C

      FEE64 temperatures OK - attachment 25

      ADC data item stats OK - attachment 26

      per FEE64 Stat spectra - attachment 27

      per FEE64 1.8.L spectra - attachments 28-29

      Merger, TapeServer etc - attachments 30-32

23.14 DAQ continues - file APR24/R1_12

      DSSSD bias & leakage current OK - attachment 33
      ambient temperature 16.1 deg C

      FEE64 temperatures OK - attachment 34

      ADC data item stats OK - attachment 35

      per FEE64 Stat spectra - attachment 36

      Merger, TapeServer etc - attachments 37-39
  545   Wed Apr 24 13:18:09 2024 TDExamples of poor CARME cable management/routing
Have observed some instances of poor CARME cable management/routing
Attachment 1 - SAES D-2000 ion pump HV cable (netal braid) transiting FEE64 cabling
Attachment 2 - CARME motor controller cable (grey plastic) transiting FEE64 cabling

General check and tidying up required.
  544   Wed Apr 24 12:48:35 2024 TDWednesday 24 April
      Replace AIDA ASIC mezzanine of aida01

13.43 DSSSD bias & leakage current OK - attachment 1
      ambient temperature 16.5 deg C

      FEE64 temperatures OK - attachment 2

      All system wide checks OK 

      WR timestamps OK - attachment 3

      ADC data item stats OK - attachment 4

      per FEE64 Rate spectra - attachment 5
       c. 10 strips (of 1024) not producing data
       most missing strips are from aida05 - misaligned or damaged adaptor PCB?

      per p+n FEE64 1.8.L spectra - attachment 6
       aida01 pulser peak width 75 ch FWHM
       common x/y scale => peak height proxy for peak width

      per n+n FEE64 1.8.L spectra - attachment 7
       aida01 pulser peak width 70 ch FWHM

      per FEE64 1.8.W spectra - 20us FSR - attachments 8-11

Summary

All FEE64s appear to be working (except aida05 which has c. 8 missing channels) with pulser - which 
validates the FEE64, ASICs and adaptor PCB.

Electronic noise c. 50keV FWHM for aida01 & aida03 - other FEE64s better/worse.

Preamplifier noise appears to be similar for all FEE64s which may suggest an issue with the test signal chain

Note
1) e-target is starting up
2) have identified some instances of poor CARME cable management/routing see https://elog.ph.ed.ac.uk/CARME/545



EG&G Ortec 448 Pulser
Amplitude 50,000
Normalise 5.00
x5 attenuator IN
t_r 100ns
tau_d 50us
Polarity +
Frequency relay line (50Hz)

Cooknell SA1 Summing Amplifier
inverted output to test - daisy chain

Current ASIC settings saved to 2024Apr24-14.14.03


14.48 Pulser OFF
 
      All histograms & stats zero'd

      LEC slow comparator 0x64 -> 0xff
      HEC fast comparator 0x2 -> 0x5

      background alpha run to validate signal path from DSSSD to FEE64 adaptor PCB

22.18 Start Merger & TapeServer - attachments 12-14

      DSSSD bias & leakage current OK - attachment 21
      ambient temperature 16.0 deg C


      FEE64 temperatures OK - attachment 15

      WR timestamps OK - attachment 16

      ADC data item stats OK - attachment 17

      per FEE64 Rate spectra - attachment 18

      per p+n FEE64 1.8.L spectra - attachments 19-20
  543   Tue Apr 23 17:44:41 2024 TDTuesday 23 April
18.43 ASIC settings 2024Feb08-15.59.09

aida06 pulser peak width 66 ch FWHM
  542   Fri Apr 19 02:47:59 2024 TDFriday 19 April
03.45




10.01 




10.05 DSSSD bias OFF
      FEE64 power OFF
  541   Thu Apr 18 13:34:06 2024 TD, JGThursday 18 April
14.30 


       per FEE64 1.8.L spectra
        aida11 pulser peak width 52 ch FWHM



14.45 EG&G Ortec 448 Pulser OFF

      all stats & histograms zero'd
 
      alpha background


22.00 DSSSD bias and leakage current OK - attachment 10

      FEE64 temperatures OK - attachment 11

      ADC data item stats - attachment 12

      per FEE64 Stat spectra - attachment 13
       Comparison of attachment 9 (per FEE64 Rate spectra with pulser) indicates significant issues (downstream of adaptor PCB) for DSSSD #1
       Possible issues asics #3 aida01 and possible adaptor PCB misalignment aida05
  540   Wed Mar 6 10:47:11 2024 JG, TDWednesday 6 March - PEXARIA5 WR test
Courtesy Nik Kurz, JG has installed a PEXARIA5 PCIe card adjacent to CARME - BuTIS single-mode fibre input
and HDMI output to root MACB input for distribution to the other MACBs and FEE64s.

The PEXARIA5 card provides equivalent functionality to the VETAR2 VME module that we have used to date. 

Further information https://www.gsi.de/fileadmin/EE/Module/EXPLODER/pexaria5_14.pdf 

Cooling water ON 

Via Anydesk startup MIDAS and FEE64s

Power ON
DSSSD HV OFF ( not required for WR test )

FEE64 temperatures OK - attachment 1

All system wide checks OK

Info code 4 & 5 stats OK - attachments 2 & 3

10:48:46 WR timestamps OK - attachment 4

         Timestamp (ns) converted to (s) corresponds to 09.49.22 GMT 6.3.24 which appears to be correct

11.00 Power OFF
  539   Wed Feb 21 07:38:32 2024 TDWedneday 21 February
10.54 23.2.24 Bias OFF
  538   Tue Feb 20 14:12:32 2024 JMShutdown

CARME DAQ shutdown. FEEs turned OFF and all MIDAS server windows closed.

Detector bias turned OFF.

Final leakage current plot - attachment 1

  537   Tue Feb 20 13:20:44 2024 OF, AG, JG, MLR48 status check

Doing afternoon check. Checked the leakage currents, all ok. Attachment 1

Temperatures ok, attchment 2

System wide check, all correct.

Statistics ok (as far as we can tell) attachment 3

Spectrum browser ok, attachment 4

Checked merger, spinner spinning...

 

Beam current is around 2.2 E7 and the target dencity is 6.3E11 cm2.

  536   Tue Feb 20 09:57:57 2024 OF, AG, MLRun 48 status check

Doing morning check. Checked the leakage currents, all ok. Attachment 1

Temperatures ok, attchment 2

System wide check, all correct.

Statistics ok (as far as we can tell) attachment 3

Spectrum browser ok, attachment 4

Checked merger, spinner spinning...

Attachment 5 shows the BPM scope (i.e. induced signals from the ion buchnes) for the duration of injection, bunching, acceleration, debunching.

 

Beam current is around 2.5E7 and the target dencity is 6.3E11 cm2.

  535   Tue Feb 20 09:51:37 2024 JMRun 48 beam on target

Run 48 started

  534   Tue Feb 20 08:41:08 2024 OF, AGstill no beam CRYRING

We have mosty empty injections into CRYRING. Only sporadically the beam with good intensity is injected.

Michael Lestinsky is currently working on the problem.

  533   Tue Feb 20 07:50:10 2024 JMDAQ restart

DAQ restarted. Run46. All appears OK, aida01 shows high rates across ASIC 4 even with no beam which could not be resolved via ASIC control. May suggest some damge.

Merger is running and tape server is writing data to disk. Data rate is about 1300 kB/sec from 800 kb/sec due to increase in aida01 rate.

FEE temps OK

All system checks OK

Detectors moved to IN position (left arm pot = 36.5 mm, right arm =70.49 mm). Target ON.

Lost beam again.

Run 47 started.

 

  532   Mon Feb 19 21:46:32 2024 EH, EM, CBRun 45 + end of late shift
22:54
We have moved the gas jet target out (y=4000), the density has dropped to zero.
Started a new run (Run45) for background measurement.
Informed HKR that the experiment will be monitored remotely without on-site personnel. They will check on the beam intensity.

02:08
System wide checks OK
Temperatures OK
Leakage current going up - OK
Stats OK
Merger continues OK

Reset the online spectra (had not been done yet for this run).
20s rate looks fine, around 1000 counts/s on top left detector. Maybe we are hitting it with beam? But then we should be see a reduction in beam intensity, we do not.

04:00
System wide checks OK
Temps OK
Stats OK
Merger OK

06:20
Beam intensity quite unstable, but HKR is correcting - attach 7
Systems wide checks OK
Temps OK
Merger OK
Stats OK
Some interesting features can be seen in the online histograms, e.g. no peak at 16 MeV and a clear straight line in the Ex vs x plot.

07:20
DAQ crashed due to multiple merger / tape server / HTTPD server instances being opened on workspace 5
Cannot soft reboot. FEEs off.
  531   Mon Feb 19 08:45:22 2024 JM, JG, SHB, EM, EHRun 44 - cooled beam on target

09:43

Run 44 started cooled beam on target

Detectors IN (left arm pot =36.5 mm, right arm pot = 70.49 mm). Stationary. Same as previous runs

Electron cooler voltage 751.4 V this corresponds to 1.3436 MeV/u. This should be back on the peak of the resonance.

Target density is 5.2e11 at/cm2, ring pressure looks OK YR09 1.72e-10 mbar. Target density is not very stable and requires small corrections in the x position to maintain a target

Horizontal beam position is -7.5 mm, vertical beam position is 0.0 mm. Same as previous run

Beam is in the ring for 6 seconds and total cycle time between injections is 10 seconds. Notice this is different than previous runs. Aim is to have a few different cycle time configurations to see effect on halo

Statistics, merger statistics, histograms all zero'd

10:05

FEE temps OK

System checks OK

Merger OK, tape server OK

Leakage current plot attachment 1

ADC data items stats attachment 2

Monitoring code rates attachment 3

11:08

FEE temps OK

System checks OK

Merger OK, tape server OK

Leakage current OK

ADC data items stats OK

Online spectra OK

12:30

FEE temps OK

System checks OK

Merger OK, tape server OK

Leakage current OK

ADC data items stats OK

Online spectra OK

14:03

FEE temps OK

System checks OK

Merger OK, tape server OK

Leakage current OK

ADC data items stats OK

Online spectra OK

15:36

FEE temps OK, system checks OK, merger OK,

Screenshot 4: temperatures

Screenshot 5: statistics

Screenhot 6 - 9: spectra (there is a persistent peak in aida13 in Layout8)

Screenshot 10: leakage current (current 1 seems to be going up slowly but steadily)

Screenshot 11: spectrum

 

17:18

FEE temps OK, system checks OK, merger OK,

Screenshot 12: temperatures

Screenshot 13: statistics

Screenhot 14 - 17: spectra (we still see a peak in aida13 in Layout8, but not always)

Screenshot 18: leakage current (current 1 seems to be going up slowly but steadily)

Screenshot 19: spectrum

 

19:27

FEE temps OK, system checks OK, merger OK,

Screenshot 20: temperatures

Screenshot 21: statistics

Screenhot 22 - 25: spectra (we still see a peak in aida13 in Layout8, but not always)

Screenshot 26: leakage current (current 1 seems to be going up slowly but steadily)

Screenshot 27: spectrum

 

21:57

FEE temps OK, system checks OK, merger OK,

Screenshot 28: temperatures

Screenshot 29: statistics

Screenhot 30 - 33: spectra (we still see a peak in aida13 in Layout8, but not always)

Screenshot 34: leakage current

Screenshot 35: spectrum

 

 

  530   Mon Feb 19 06:01:17 2024 SHB, JM, JGRun 42 morning shift

07:00

Observed reduction in YR09 presure over last few hours in ML online plot. Beam intensity looks OK but target density was fluctuating quite a lot. Current value was 1.5e11.

Changed the nozzle position slightly which increased target density to 5.5e11. Pressure in YR09 looks OK.

I accidently scrolled slightly with the target code (only by 100 but am not sure). Target looks stable but I will not touch remotely again for fear of losing total target position.

Rates in ADC data items statistics and in the monitoring code look consistent with beam on target. 

08:56

Target density reduced again to 1.5e11. Changed the nozzle position slightly in X and recovered 5.5e11 at/cm2

Beam intensity looks OK

FEE temps OK

System checks OK

Merger and tape server OK

Statistics ADC data items and spectrum browser rates OK

Merger stats attachement 5

leakage current plot attachment 6

Online monitoring code rates OK

09:33

Run stopped to change beam energy

 

  529   Mon Feb 19 05:10:48 2024 TDRun 42 night shift contd.
06.03 DAQ continues file G22-86/R42_32

      DSSSD bias & leakage current - attachments 1-2
       DSSSD #0 trending upwards , DSSSD #1-3 trending downwards with decreasing ambient temperature

      FEE64 temperatures OK - attachments 3-4

      ADC data item stats - attachment 5

      per FEE64 Rate spectra - attachments 6

      target density 5.27e+11/cm2
      beam stable for last 3 hours ( cannot read scale with X2Go )
  528   Sun Feb 18 22:03:56 2024 TDRun 42 night shift
23.03 DAQ continues file G22-86/R42_21

      ML CRYRING ambient temperature and vacuum - attachments 1 & 2

      DSSSD bias & leakage current - attachments 3 & 24
       DSSSD #0 trending upwards , DSSSD #1-3 trending downwards with decreasing ambient temperature

      FEE64 temperatures OK - attachments 4-5

23.07 All system wide checks OK

      WR timestamps OK - attachment 6

      ADC, DISC, PAUSE and correlation stats OK - attachments 7-10

      per FEE64 Rate spectra - attachments 11-12
      per FEE64 Stat spectra - attachments 13-14

      per FEE64 1.8.L, 1.8.H and 1.8.W 20us FSR spectra - attachments 15-20

      Merger, Merger stats and TapeServer OK - attachments 21-23

      online analysis spectra - attachments 25-30

23.28 all histograms, stats, Merger stats zero'd
      online analysis restarted

23.35 target density 5.83e+11/cm2
      beam no change in last 3 hours ( cannot read scale with X2Go )
      see attachments 31-33


00.58 DAQ continues file G22-86/R42_24

      DSSSD bias & leakage current - attachments 34-35
       DSSSD #0 trending upwards , DSSSD #1-3 trending downwards with decreasing ambient temperature

      FEE64 temperatures OK - attachments 36-37

      ADC data item stats - attachment 38

      per FEE64 Rate spectra - attachments 39

      target density 5.75e+11/cm2
      beam no change in last 3 hours ( cannot read scale with X2Go )


03.14 DAQ continues file G22-86/R42_28

      DSSSD bias & leakage current - attachments 40-41
       DSSSD #0 trending upwards , DSSSD #1-3 trending downwards with decreasing ambient temperature

      FEE64 temperatures OK - attachments 42-43

      ADC data item stats - attachment 44

      per FEE64 Rate spectra - attachments 45

      target density 5.91e+11/cm2
      beam decreased in last 3 hours ( cannot read scale with X2Go )
      see attachments 46-47

03:30 beam recovered
      online analysis rates DSSSD#0 recovered to values observed at start of shift