Entry  Fri Nov 1 09:30:17 2019, NH, Database error stopping run? ErrorStopDatabase.png
Strange error about Database options appearing during run stop.
Does not seem to affect stopping/starting/data?
Entry  Mon Nov 4 08:48:18 2019, NH, Alpha Status 04.11.2019 alphas-bias.pngalphas-temp.pngalpha-aida10.pngstats-0511.pngspec-0511.png
Alpha run has been running over the weekend. Stats looking positive with notable exception that aida10 is not sending much data. Will attempt a powercycle.

Temperatures and Leakage Currents very good to excellent

Noted that the slow rate of aida10 seems to be slowing down the rate data is sent to tape - perhaps being buffered until an ADC event arrives?
Entry  Mon Nov 4 13:17:12 2019, NH, Report: aida10 database corruption db_working.txtdb_corrupt.txt
Often aida10 database seems to get corrupted (ASIC values are all 0xad)
Unsure why it's only aida10, might be related to weird alpha rate behaviour at the moment.

 
Entry  Mon Nov 11 10:34:19 2019, NH, Alpha Analysis (11.11.19) 9x
Analysis of approx 20GB of data since Thursday night (not continuous - estimate of hours incominG)

Figures 1-3: Energy Front vs Energy Back (No F-B energy gate)

Figures 4-6: 2D hit pattern
Entry  Wed Nov 13 15:38:20 2019, NH, [HowTo] MBS for AIDA (Nov 19) 
For MBS integration AIDA forwards its data to an MBS Foreign Data Receiver (FDR), currently the PC assigned for this task is x86l-94.

To start the MBS system do the following:
Entry  Fri Nov 22 08:55:31 2019, NH, Waveforms waves_21119_even.png21119_waves_odd.pngwaves_21119_even_z.pngwaves_21119_odd_z.png
AIDA Waveforms immediately after beam at 21.11.19 8am or so.
    Reply  Mon Nov 25 13:25:51 2019, NH, Waveforms wave_odd.pngwave_even1.png
> AIDA Waveforms immediately after beam at 21.11.19 8am or so.

Update 25.11.19 with all FEE64 waveforms added
Entry  Wed Nov 27 16:33:50 2019, NH, New Bias & Waveforms waves_new_odd.pngwaves_new_even.png
HV core (-160 V) now attached to bottom 3 FEEs (with grounded kapton)

DSSDs bias OK, leakages unchanged

No appreciable change in waveforms
Entry  Wed Dec 11 10:58:52 2019, NH, Initital Dec. Analysis 11x
Fig 1: 2D hit pattern of each AIDA DSSD for MBS run 24 (40Ar primary beam, centred)

Low statistics in DSSD2 partly due to low energy deposition 
Entry  Wed Dec 18 02:44:27 2019, NH, 17.12.19, 18.12.19  7x
Beam: 238U (850 MeV/u)
Fragment: 168Re
Entry  Fri Dec 20 10:19:29 2019, NH, Alpha Analysis (Pre Run) Alphas.pngAlphas_E.png
======================================                                                                                                                 
                                                                                                                      
AIDA Unpacker Analysis                                                                                                                                
Entry  Wed Jan 15 14:53:41 2020, NH, AIDA Noise 14.01.2019 6x
Checking noise

DESPEC crate entirely offline except:
    Reply  Mon Jan 20 14:01:23 2020, NH, AIDA Noise 20.01.2019 6x
> Checking noise

> DESPEC crate entirely offline except:
Entry  Wed Jan 29 12:22:13 2020, NH, aida10 troubleshooting aida08-noise.pngaida10-noitems.pngStats.png
aida10 has no ASIC events when running without pulser

DSSD ribbon cable was replaced, no change
    Reply  Wed Jan 29 19:09:51 2020, NH, aida10 troubleshooting 
> aida10 has no ASIC events when running without pulser

> DSSD ribbon cable was replaced, no change
Entry  Fri Feb 14 12:50:29 2020, NH, AIDA Noise 14.02.2020 6x
Electricians changed the grounding for the DESPEC platform to a large pillar that goes deep into soil

Change in grounding was apparently successful at reducing the noise complaints in Cave C (R3B)
Entry  Mon Feb 17 09:48:34 2020, NH, Report: aida09 Kernel Panic & Lost WR  ttyUSB12GSI_WR.pngGSI_WR2.pngGSI_WR4.pngGSI_WR3.png
aida09 crashed over the weekend and automatically rebooted.
After the reboot the WR timestamp sent to the merger is in the future and hence incorrect
Entry  Fri Feb 21 09:21:10 2020, NH, (Early) Alpha Analysis  DSSD_1d_HP.pngDSSD_E.pngFEE3_43_Hot.pngfee8_hot.pngFee10_hot.png
Overnight alpha data

======================================
Entry  Tue Feb 25 09:55:59 2020, NH, HV Scope Traces 10x
Looking at the HV bias in a scope to see if it is the source of noise especially on the p+n side

Channel 1: HV Core channel 4
Entry  Sat Feb 29 10:56:38 2020, NH, Alpha Hit Pattern alphas_1d.pdf
Circa 700 GB of data

A few channels in DSSD 1 missing? Otherwise seems OK