AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC, Page 24 of 35  ELOG logo
ID Date Author Subject Text Attachments
  237   Sun Apr 18 00:57:36 2021 BA, MASanday 18 April 00.00-08.0002:01 Beam has stopped at 01:37 and returend
at 01:43 for few min and then stopped again
and not knowen how it will take until it
 12x 
  236   Sat Apr 17 19:39:56 2021 DJ, TDSaturday 17 April 20.00-00.00
19.37 per FEE64 rate spectra - attachments
1 & 2
 13x 
  235   Sat Apr 17 15:29:27 2021 MSSaturday 17 April 16:00-20:00 202116:21

 6x 
  234   Sat Apr 17 11:27:38 2021 JS, TDSaturday 17th April 12:00-16:00  Base  Current  Difference
aida05 fault   0x1a52 :   0x1a53 :   1  
aida06 fault   0x4f3e :   0x4f3f :   1  
 10x 
  233   Sat Apr 17 07:28:02 2021 LPG, TDSaturday 08:00 - 12:0008:30 CEST

HV and leakage currents: elog:233/1
 9x 
  232   Fri Apr 16 23:13:57 2021 CA, MA, BASaturday 16th 00:00 - 08:00 00:05 DESPEC tuning beam, changing spill length

00:21 DAQ continues ok
 14x 
  231   Fri Apr 16 15:52:08 2021 DSJ 16 April 16.00 shiftClock status test result: Passed 12, Failed
0
 12x 
  230   Fri Apr 16 11:42:45 2021 OHShifter checklistEvery 30 minutes:

- Reload statistics (Web browser tab screen
  
  229   Fri Apr 16 07:19:00 2021 OH, MAFriday 16th April 08:00-16:0008:19 System wide checks all ok
      Baselines reset on WR and FPGA checks
for the start of the day
 14x 
  228   Fri Apr 16 05:13:09 2021 MLsystem wide checkssystem wide checks:

Done around 4am.

#################################
 16x 
  227   Fri Apr 16 02:15:12 2021 NHImplants/PunchthroughRate with beam < 1000 Hz still

Not Uranium but some fragments possibly produced
 53_AM.png 
  225   Fri Apr 16 00:47:36 2021 OH-MLAIDA DAQ Reset1:18 (CET) No rate in AIDA 07.

Oscar did a AIDA DAQ reset and
a quick wide check.
  
  224   Thu Apr 15 22:38:18 2021 OHThursday 15th April - 16th April23:38 They are testing the finger detector
      It is a lot of U they are putting into
it which is passing through to AIDA ~3500Hz
  
  223   Thu Apr 15 13:28:07 2021 TDThursday 15 April
13.28 System wide checks
 19x 
  222   Thu Apr 15 12:23:22 2021 NHFast Discriminator MasksIt is possible to get good rates with the
fast discriminators with a few hot channels
mask in aida03 and aida10
 Image_Pasted_at_2021-4-14_18-57.png 
  221   Wed Apr 14 15:14:55 2021 PJCSNote WellPlease note that for the forseeable future
the system function in the menu on the System
Wide Checks browser page labelled as 
  
  220   Wed Apr 14 12:14:47 2021 PJCSCorrections and changes14/4/21 @12:00 UK time

Corrected the failure of FADC re-calibration
All modules. The problem was a missing >
  
  219   Wed Apr 14 09:55:12 2021 TDTuesday 13 April
Two further comments
  
  218   Tue Apr 13 16:49:41 2021 TDTuesday 13 April
Rename Options directory
 8x 
  217   Tue Apr 13 14:24:48 2021 TDS460 informationaida-gsi Anydesk 897170655

DESPEC remote working - https://sf.gsi.de/d/a4fb2134e06a450ca777/
  
ELOG V3.1.4-unknown