Fri Dec 20 12:04:03 2019, NH, TD, Friday 20 December 2019 7x
|
Fig 1: DSSD Leakage currents ok
Fig 2: FEE Temps ok
|
Fri Nov 1 10:46:13 2019, CA, TD, NH, Friday 1st November 2019 6x
|
11.46 - attachments 1,2,3: bias/leakage currents, good event statistics and fee temperatures (respectively)
before removing aluminium foil upstream of AIDA
|
Fri Mar 19 17:16:45 2021, TD, Friday 19 March 8x
|
Screenshots of successive displays of Options for FEE64s aida01, 02, 03 and 04. |
Fri Apr 19 13:52:34 2024, JB, CC, TD, Friday 19 April contd. 23x
|
12.00 Replaced AIDA ASIC mezzanine of aida01 to fix issue with asic #3
|
Fri Apr 19 02:35:14 2024, TD, Friday 19 April 38x
|
03.30
|
Sat Jun 18 09:46:57 2022, OH, NH, Friday 17th June 23x
|
Summary of the day:
Small improvements were made throughout the day but by far the biggest change to the rate in the FEEs came from disabling the waveforms
This was done by setting ADC con
|
Fri Apr 16 07:19:00 2021, OH, MA, Friday 16th April 08:00-16:00 14x
|
08:19 System wide checks all ok
Baselines reset on WR and FPGA checks for the start of the day
|
Fri May 14 19:01:09 2021, JS, Friday 14th May 20:00-00:00 27x
|
20:00 Taking over from CB
20:33 Doing full checks.
|
Fri May 14 14:57:32 2021, CB + OH, TD, NH, Friday 14th May 16:00-20:00 33x
|
16:00 Shift taken over from CA
16:00
|
Fri May 14 07:05:08 2021, CA, Friday 14th May 08:00 - 16:00 shift 24x
|
08:00 CA takes over
08:28 Problem at UNILAC - no beam
|
Thu May 13 23:58:05 2021, OH, Friday 14th May 00:00 - 08:00 21x
|
00:00 OH Takes over.
PULSER SETTINGS
---------------------------
|
Fri Mar 13 08:01:53 2020, DK, Friday 13th 8:00 to 16:00 shift 8x
|
8:00 - See previous shift's elog entry. All good.
9:00
|
Thu Mar 12 23:04:24 2020, LS, DK, Friday 13th 00:00 to 08:00 shift 20x
|
Shift file starting R9_639
system wide checks okay
|
Fri May 13 07:36:18 2022, OH, TD, Friday 13 May 35x
|
08:36 DAQ Still running smoothly
Currently on run R1_328
At current data rates we have just over 4 days of space left on the current disk
|
Fri Jun 13 12:42:39 2025, CC, TD, Friday 13 June 2025 10x
|
13.42 ASIC settings 2025Jun12-13.53.04
*all* FEE64s slow comparator 0x64
|
Fri Mar 12 10:09:19 2021, LS, CA, Friday 12th March 11.00- 17x
|
11.00(Germany) System wide checks okay except:
FEE64 module aida09 global clocks failed, 6
|
Fri Mar 12 08:32:46 2021, TD, Friday 12 March 08.00-    
|
08.31 DAQ continues OK - file R33_263
ASIC settings 2019Dec19-16.19.51
|
Fri Mar 12 20:12:10 2021, TD, Friday 12 March    
|
21.12 DAQ continues OK - file R33_617
ASIC settings 2019Dec19-16.19.51
|
Fri Feb 11 12:09:24 2022, NH, Fri 11 Feb AIDA Work
|
Use beam break between FRS experiments to finalise getting AIDA testable for PJCS etc.
USB hubs reconnected - all 16 ttyUSBs seem to be visible now
|
Fri Jun 10 10:32:46 2022, NH, Fri 10 Jun 8x
|
Probe PSU output with 'scope again and 10X probe
Use a AC mains scope as was available, means referenced to ground of AC, but should be ground of platform (and hence Fee64)
|