Tue Mar 19 10:17:30 2024, NH, Dry Run 2024 - 19th March 24    
|
AIDA has 8 FEEs and 1 DSSSD
Aida08 (HDMI#12) had no WR again, I moved it to a different MACB and now it gets WR
|
Mon Oct 13 15:04:53 2025, JPB, NH, GB, MP, Dismount AIDA and realign the cables 
|
We have dismounted the AIDA stack and bPlast and realigned the cables so that they can fit into the adaptors.
The plan for the next couple days is the following:
|
Wed Oct 30 11:51:08 2019, CA, TD, NH, Detector biasing  
|
12.51 - detector bias and leakage currents - detector 3 has no current
- other detectors ok
|
Wed Jun 5 08:34:47 2024, JB. AM, MP, MY, Detector bias tests 14x
|
We are trying to bias each system on its own and in sequence biasing each other detector system to see the interdependence.
HV OK. TEMP OK. Attachment 1 &2. |
Thu Dec 6 15:31:40 2018, TD, CA, NH, December 6th 2018 11x
|
15.32 1.8.W spectra for aida01, aida09, aida10, aida11, aida12 (attachments 1, 2, 3)
2 and 3 expanded in Y and X direction respectively
|
Fri Dec 6 20:34:56 2019, CA, DK, NH, Dec 6 Night Shift 15x
|
21.35 AIDA DAQ currently running, TapeService on, not writing to storage
System wide checks ok
|
Thu Dec 5 22:00:11 2019, CA, DK, NH, Dec 5 Night Shift 30x
|
15:30
TD found the merger was running at ~43 Mb/s
The odd-numbered FEE64s all had significant (>200k) disk rates
|
Fri Nov 1 09:30:17 2019, NH, Database error stopping run?
|
Strange error about Database options appearing during run stop.
Does not seem to affect stopping/starting/data? |
Wed Oct 30 08:06:49 2019, CA, TD, NH, DSSSD stack 6x
|
Ribbon cables fitted with kapton PCB couplers -> 3 isolated + 1 non-isolated for each cable length/DSSSD
Ribbon cable lengths:
|
Wed Mar 13 13:02:19 2024, JB, NH, HA, MA, DSSD 1 biasing tests
|
Dismounted Snout and biased DSSSD1 channels
leftmost waifer working
middle waifer reaches 90V then current ramps up |
Mon Oct 18 13:03:23 2021, OH, DHCPd.conf backup
|
| |
Sun Apr 28 15:49:58 2024, TD, DAQ restart short(er) cut
|
16.45 aida04 stopped producing data
|
Wed Mar 2 09:08:44 2022, OH, DAQ Status 2nd March 2022    
|
DAQ was found yesterday with none of the FEEs reporting any good events on any FEE - attachment 1
Temperature menu is still responsive and all FEEs report a temperature
Looking at the white rabbit timestamps aida09 has a gibberish timestamp - attachment 2
|
Wed Mar 2 10:05:13 2022, OH, DAQ Status 2nd March 2022
|
> DAQ was found yesterday with none of the FEEs reporting any good events on any FEE - attachment 1
> Temperature menu is still responsive and all FEEs report a temperature
> Looking at the white rabbit timestamps aida09 has a gibberish timestamp - attachment 2
|
Thu May 9 09:08:26 2024, TD, Current NewMerger install - aida-gsi
|
[npg@aidas-gsi MIDAS]$ ls -l
total 524
drwxrwxr-x. 2 npg npg 156 Mar 17 2021 Aida
|
Wed Apr 14 12:14:47 2021, PJCS, Corrections and changes
|
14/4/21 @12:00 UK time
Corrected the failure of FADC re-calibration All modules. The problem was a missing > in the .tml file
Edited the sys.tcl file in /MIDAS/TclHttpd/Html/RunControl to comment out where the status of the waveform enable was able to stop the creation |
Tue Apr 9 13:22:41 2024, TD, JB, CC, Continued of 9th of April  
|
Disconnected test - pulser from aida08. Noise conditions improved slightly, probably less pulser events. (Attachments 1 & 2).
We recabled the dew point sensor and flow sensor for the water supply interlock to the FEE64s. When we tried to power up the interlock box, the
PSU (charging block) LED was blinking indicating an internal power instability. We then plugged each of the sensors out and into the interlock box and |
Thu May 13 19:52:36 2021, Philippos and Marc, Connection problem
|
At 20:15, We lost connection with AnyDesk.
Liliana restarted Anydesk on the AIDA DAQ PC. All seemed fine on AIDA DAQ but there was a more generic issue on the main DAQ, possibly the timesorter.
Helena restarted all after ~30mn and it's all working again now. |
Mon May 10 11:45:15 2021, NH, Confusing behaviour of rates
|
Testing the behaviour of powering down the fast ADCs in aida01 (the hot FEE)
Set local control 2 to 0xff in aida01
|
Thu Apr 29 16:56:42 2021, OH, Config changes for 16FEE
|
New set of ASIC settings produced 2021Apr29-13-16-00
- Note no longer odd/even for polarity of signal
|