  569   Tue Apr 9 13:22:41 2024 TD, JB, CCContinued of 9th of April

Disconnected test - pulser from aida08. Noise conditions improved slightly, probably less pulser events. (Attachments 1 & 2).

We recabled the dew point sensor and flow sensor for the water supply interlock to the FEE64s. When we tried to power up the interlock box, the PSU (charging block) LED was blinking indicating an internal power instability. We then plugged each of the sensors out and into the interlock box and found that the flow sensor was the source of the power supply issue.

The HDMI cable of aida04 appears to be broken or wobbly, but the signals still look the same after powering up again. The adaptor board was reseated.

Pulser peak of test - remeasured the peak in aida02 1.8.L and found the same result as before with 118 channels FWHM c. 95-100 keV FWHM energy resolution. (Attachment 3)

bPlast booster boards connected caused dramatic increase in the noise observed in AIDA. Grounding still needs to take place.

TO-DO for 10.04 - 11.04:

- aida04 asic #1 investigate noise
- aida14 asic #1 & asic #2 u/s? replace ASIC mezzanine?

- noise optimisation of aida-bplast -- (friday if BB7 also powered).

 

  313   Thu May 13 19:52:36 2021 Philippos and MarcConnection problem

At 20:15, We lost connection with AnyDesk.

Liliana restarted Anydesk on the AIDA DAQ PC. All seemed fine on AIDA DAQ but there was a more generic issue on the main DAQ, possibly the timesorter.

Helena restarted all after ~30mn and it's all working again now.

 

 

  304   Mon May 10 11:45:15 2021 NHConfusing behaviour of rates
Testing the behaviour of powering down the fast ADCs in aida01 (the hot FEE)

Set local control 2 to 0xff in aida01
Wait... see temperature drpos by ~3 degrees (PJCS reports much larger drop in T9)
So far so normal

*but*

aida05 rates go very high when this waveform is disabled!!
Set local control to 0x00 again
They drop back...

Recalibrated FADCs and back to normal
  287   Thu Apr 29 16:56:42 2021 OHConfig changes for 16FEE
New set of ASIC settings produced 2021Apr29-13-16-00
- Note no longer odd/even for polarity of signal

New set of Options file for FEEs - Points to the new ASIC settings

New layouts for histograms grouped by side of the detector. Heavily weighted to the front side in terms of number of his

/MIDAS/Linux/startup/NewMerger Updated line from "./master64 -i 12 -l 16 -p 11001  &" to "./master64 -i 16 -l 16 -p 11001  &"

/MIDAS/DB/EXPERIMENTS/MERGE/Options/aidas-gsi/CONTENTS updated the number of links to 16.

System wide checks passed *except aida03 ADC calibration*

System left running to no storage.
  710   Wed Jul 23 09:37:34 2025 CC, NKComparison pulser FWHM with/without HT and detector cables

This elog is the first reliable. Before, the histograms have not been zero'd

Each test are done and compared for AIDA05.1.8L, compared with the pulser (caracteristics, see attachment1)

Test 1: with HT cables shielded

Attachment 2: HT = 100V; cable AIDA05 connected --> FWHM = 77 channels (= 51.3 keV) with a top of the curve at 100 cts (AIDA9.1.8L has also been checked and was at 89 channels)

Attachment 3: HT = 0V; cable AIDA05 connected --> FWHM = 1757 channels with a top of the curve at 15 cts after almost 40 min because the noise had a baseline of 6500 channels

Attachment 4: HT = 0V; cable AIDA05 disconnected --> FWHM = 16 channels with a top of the curve at 110 cts

Attachment 5: HT = 100V; cable AIDA05 disconnected --> FWHM = 17 channels with a top of the curve at 110 cts

Test 2: with HT cables not shielded

Attachment 6: HT = 0V; cable AIDA05 connected --> FWHM = 1452 channels (968 keV) with a top of the curve at 23 cts after 1:15 hours because the noise had a baseline of 7000 channels

Attachment 7: HT = 100V; cable AIDA05 connected --> FWHM = 73 channels with a top of the curve at 100 cts

Attachment 8: HT = 0V; cable AIDA05 disconnected --> FWHM = 15 channels with a top of the curve at 120 cts

Attachment 9: HT = 100V; cable AIDA05 disconnected --> FWHM = 16 channels with a top of the curve at 125 cts

Conclusion: with shielded or unshielded cables, the noise situation seems to be the same, proving that mot of the noise don't come from the cables

Other thing that have been noticed is that the peak of the pulser is shifted when the cables are connected compared to when they are not connected: connected: peak at around channel 21675; disconnected: peak at around channel 21925

  715   Wed Jul 23 17:02:22 2025 CC, NKComparison of channels in the same FEE64 and ASIC 05.1 and with AIDA13.1.8L

BB7 is still connected but not biased (same configuration as the previous elog). The channel of the CAEN power supply #0 is connected to ASIC08 and biased at 100V. The channel of the CAEN power supply #1 is disconnected but still biased at 100V. AIDA04 (chained to AIDA05) is biased via a shielded cable to an iseg NHQ 246L power supplied, biased at 101 V with 16 uA.

All the outside drain wires of bplast are now taped to the snout.

Attachment 1: AIDA05.1.8L --> FWHM 60 channels

Attachment 2: AIDA05.1.1L --> FWHM 51 channels

Attachment 3: AIDA05.1.4L --> FWHM 57 channels

Attachment 4: AIDA05.1.12L --> FWHM 65 channels

Attachment 5: AIDA05.1.15L --> FWHM 69 channels

Attachment 6: AIDA13.1.8L --> FWHM 277 channels, chosen because same place than AIDA05 but for the dowstream detector that is still biased with the CAEN module

Attachment 7: AIDA12.1.8L --> FWHM 88 channels, chosen because same place than AIDA05 but of the bottom of upstream detector that is still biased with the iseg module

Attachment 8: AIDA16.1.8L --> FWHM 133 channels, chosen because same place than AIDA05 of the bottom of dowstream detector that is still biased with the CAEN module. This curve has 2 humps

  716   Thu Jul 24 10:28:04 2025 CC, NK, MPComparison of channels in the same FEE64 and ASIC 05.1 and with AIDA13.1.8L

BB7 is still connected but not biased (same configuration as the previous elog). The channel of the CAEN power supply #0 is connected to ASIC08 and biased at 100V. The channel of the CAEN power supply #1 is disconnected but still biased at 100V. AIDA04 (chained to AIDA05) is biased via an unshielded cable to an iseg NHQ 246L power supplied, biased at 100 V with 15 uA.

All the outside drain wires of bplast are now taped to the snout.

The red markers of the cursors of the previous test have been left for ease of comparison.

Attachment 1: AIDA05.1.8L --> FWHM 67 channels

Attachment 2: AIDA05.1.1L --> FWHM 69 channels

Attachment 3: AIDA05.1.4L --> FWHM 79 channels

Attachment 4: AIDA05.1.12L --> FWHM 79 channels

Attachment 5: AIDA05.1.15L --> FWHM 79 channels

Attachment 6: AIDA13.1.8L --> FWHM 236 channels, chosen because same place than AIDA05 but for the dowstream detector that is still biased with the CAEN module

Attachment 7: AIDA12.1.8L --> FWHM 91 channels, chosen because same place than AIDA05 but of the bottom of upstream detector that is still biased with the iseg module

Attachment 8: AIDA16.1.8L --> FWHM 143 channels, chosen because same place than AIDA05 of the bottom of dowstream detector that is still biased with the CAEN module. This curve has 2 humps

  711   Wed Jul 23 13:08:30 2025 CC, NKComparison of channels in the same FEE64 and ASIC 05.1

Biased at 100, AIDA05 connected, HT cables not shielded CAEN HT

Attachment 1: AIDA05.1.8L --> FWHM 75 channels

Attachment 2: AIDA05.1.1L --> FWHM 70 channels; peak of the pulser shifted to the right (red lines on the left: cursor for the 1.8L)

Attachment 3: AIDA05.1.4L --> FWHM 85 channels; peak of the pulser completely shifted to the left (10 000 on the left)

Attachment 4: AIDA05.1.12L --> FWHM 78 channels

Attachment 5: AIDA05.1.15L --> FWHM 79 channels, peak the most shifted to the right of the 5 curves tested

  718   Thu Jul 24 15:50:01 2025 CC, MP, NKCheck of BB7 impact

HT of upstream AIDA from iseg, HT cable shielded, 100V, 17 uA. HT of downstream AIDA from CAEN, AIDA#0. Both channel 0 and 1 of CAEN are on, even though there is just one channel connected. DEGAS platform closed. The crates of BB7 and bplast have been put to the ground of the AIDA's structure (cupper sheet)

Attachment 1: TEST1: BB7 fully connected = 2 tiles of BB7 have been connected to the mesytec preamps. BB7 is not biased. The mesytec preamps have been fully connected and supplied but 6/8 black cables are not connected on the other side. 4X and 4Y are connected to the 2 FEBEX cards. The cables of BB7 right between the snout and the preamps are shielded.

AIDA05.1.8L --> FWHM 76-80 channels

Attachment 2: TEST2: BB7 fully disconnected

AIDA05.1.8L --> FWHM 71 channels

Attachment 3: TEST3: BB7 fully disconnected, TAMEX crate unbiased, BB7 NIM crate turned off

AIDA05.1.8L --> FWHM 67 channels

Attachment 4: AIDA01.1.0 --> FWHM 171 channels --> same detector, also biased with iseg but bad results

  210   Wed Mar 31 15:40:42 2021 PJCSCheck Options files function added

There is a new operation available in the System Functions menu in the System Wide Checks page. ( The page needs to be reset once when the FEEs are powered before this function will appear )

Check the Options files are all the same size

Using this function will provide a list of the file sizes of the Options/< fee name >/CONTENTS files along with the last time they were accessed as a list.

It's up to the user, at present, to interpret the results but more is possible if required ....

  145   Thu Mar 12 15:25:08 2020 Thursday, March 12 16:00 to midnight shiftCA, DK, LS
16:30
    All system wide checks passed
    Temps shown in attachment #1 
    Histos shown in attachment #2
      NB: Zoomed in, linear scale
    Biases shown in attachment #3
    Stats shown in attachment #4
    Merger rates fine 3 Mega items / sec
    TapeServer looks happy, 25 MB/s

Paddy was asking us to give a rough estimate for the number of implants per spill.
I zoomed in on the rates histogram (attachment #2) and it looks like roughly 5 implants per channel in SSD 1 and SSD2.
So this is 5*128*2=1280.  Note that I don't know the rate window size (e.g., 1 second, 1 spill?)  Probably the window does not catch a full spill.
This might be an over-estimate because an implant in SSD2 may leave a signal in SSD1.
Paddy was getting around 1k or 2k fragments per spill, but to first order it seems consistent.

18:36
   Near R9_372 they slightly opened some FRS slits, as there is a belief we may be blocking some of 94Pd.
   Implant rate seems similar to before, see attachment #5
   Biases did not change more than normal, see attachment #6

20:15
   All system wide checks passed
   Temps shown in attachment #7
   Stats shown in attachment #8
   Biases shown in attchment #9

22:12 
   All system wide checks okay
   Histo #10
   Bias #11
   Stat #12
   Temps #13
  290   Tue May 4 12:11:39 2021 OH TDBias test of triple
13:11 System wide checks all ok except *aida02 fails ADC calibration*
 Collecting the file size of each FEE64 Options CONTENTS file to check they are all the same

 FEE : aida01 =>   Options file size is 1025 	 Last changed Tue May 04 10:57:38 CEST 2021
 FEE : aida02 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:46 CEST 2021
 FEE : aida03 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:50 CEST 2021
 FEE : aida04 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:53 CEST 2021
 FEE : aida05 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:55 CEST 2021
 FEE : aida06 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:43:59 CEST 2021
 FEE : aida07 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:02 CEST 2021
 FEE : aida08 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:05 CEST 2021
 FEE : aida09 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:08 CEST 2021
 FEE : aida10 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida11 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida12 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida13 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida14 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida15 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021
 FEE : aida16 =>   Options file size is 1014 	 Last changed Thu Apr 29 14:44:57 CEST 2021

ASIC clocks have been synchronised and ADC re-calibrated

Rate spectra - attachment 1
Layout7 - attachment 2
Layout8 - attachment 3

Detectors at 130V

Statistics - attachment 4

V-I Curve - attachment 5
Voltage	Ch0	Ch1
10	1.345	1.49
20	2.01	2.285
30	2.405	2.8
40	2.635	3.13
50	2.82	3.37
60	2.95	3.57
70	3.08	3.725
80	3.185	3.865
90	3.26	3.98
100	3.33	4.105
110	3.42	4.22
120	3.525	4.35
130	3.64	4.46
140	3.79	4.37
150	3.98	4.37
150+10min	3.9	4.485
  165   Fri Feb 12 17:31:04 2021 NHBackground run
A background run has been started to check the AIDA channels.
It is in /TapeData/BGFEB21

Due to some noisy channels the data rate is high (1.5 MB/s) even at 0x64

S4 access will be limited. 

I do not know when open if it is worth looking at changing the adapter PCBs of it is more likely DSSD internal.
  115   Thu Dec 19 14:26:09 2019 TDBNC PB-5 - Manual
  93   Tue Nov 12 13:59:07 2019 NH, OH, CAAwfully noisy waveforms
Rates in all FEEs much higher than before - waveform shows very noisy in all systems. - attachment 1

Good event statistics - attachment 2

DISC in 7 and 3 above 300k

ASIC check load performed.

Good event statistics - attachment 3

Thesholds at 0xa for all FEE

Disc rate now 0 across all FEE


Pulser settings: attachment 4

pulser peak widths -   FEE         width(ch)
                              1             128
                              2             145
                              3             84
                              4             74
                              5             207
                              6             125
                              7             172
                              8             80
                              9             193
                              10            183
                              11            238
                              12            138

Attachments 5 & 6: pulser peaks for all FEE64

16.24: DAQ stop

       slow comparator threshold changed to 0x64

       ASIC check performed

       Alpha run/DAQ start (w/ data transfer enabled) - writing to file 31Oct19/R13

       Merger and TapeService ok - attachments 7 & 8
  545   Fri Mar 22 07:34:54 2024 NH, JBAu Beam
  251   Tue Apr 20 07:06:20 2021 CAApril 20th 08:00 - 12:00
08:09 DAQ crashes at start of shift (!)

      again happens at same time as beam being lost

      All FEE64 lost connection

      Beam downtime for next 3-4 hours

      Performed powercycle and reset

08:40 AIDA back up and running, writing to R36

08:42 *all* system wide checks ok

      temperatures ok - attachment 1

      statistics ok - attachment 2

      detector bias/leakage currents ok - attachment 3

08:57 options file size check;


 Collecting the file size of each FEE64 Options CONTENTS file to check they are all the same

 FEE : aida01 =>   Options file size is 1025 	 Last changed Tue Apr 20 08:22:14 CEST 2021
 FEE : aida02 =>   Options file size is 1014 	 Last changed Fri Apr 16 00:56:20 CEST 2021
 FEE : aida03 =>   Options file size is 1014 	 Last changed Wed Apr 14 21:52:04 CEST 2021
 FEE : aida04 =>   Options file size is 1025 	 Last changed Sat Apr 17 06:07:36 CEST 2021
 FEE : aida05 =>   Options file size is 1025 	 Last changed Fri Apr 16 00:53:25 CEST 2021
 FEE : aida06 =>   Options file size is 1014 	 Last changed Wed Apr 14 21:52:04 CEST 2021
 FEE : aida07 =>   Options file size is 1014 	 Last changed Wed Apr 14 21:52:04 CEST 2021
 FEE : aida08 =>   Options file size is 1014 	 Last changed Wed Apr 14 21:52:04 CEST 2021
 FEE : aida09 =>   Options file size is 1014 	 Last changed Wed Apr 14 21:52:05 CEST 2021
 FEE : aida10 =>   Options file size is 1014 	 Last changed Wed Apr 14 21:52:06 CEST 2021
 FEE : aida11 =>   Options file size is 1014 	 Last changed Wed Apr 14 21:52:05 CEST 2021
 FEE : aida12 =>   Options file size is 1025 	 Last changed Mon Apr 19 09:05:25 CEST 2021

08:59 no error messages in database check terminal
      stats ok

09:16 waveforms - attachments 4 & 5

09:30 no error messages in database check terminal
      stats ok

10:07 no error messages in database check terminal
      stats ok

10:30 aida01, aida02, aida03, aida04, aida07 lost connection, DAQ crashes

      powercycle attempted, but FEE modules remain unconnected

      TD performs further power cycle

11.12 reboot aida05

      DAQ reset/setup

      system wide checks

      sync asic clocks OK
      clock status & fpga timestamp errors OK
      
      WR decoder status

		 Base 		Current 	Difference
aida09 fault 	 0xb307 : 	 0xb307 : 	 0  
aida09 : WR status 0x40
 aida10 fault 	 0x3d6c : 	 0x3d6c : 	 0  
aida10 : WR status 0x40
 aida11 fault 	 0xa610 : 	 0xa610 : 	 0  
aida11 : WR status 0x40
 aida12 fault 	 0x7c7 : 	 0x7c7 : 	 0  
aida12 : WR status 0x40
 White Rabbit error counter test result: Passed 8, Failed 4

Understand the status reports as follows:-
Status bit 3 : White Rabbit decoder detected an error in the received data
Status bit 2 : Firmware registered WR error, no reload of Timestamp
Status bit 0 : White Rabbit decoder reports uncertain of Timestamp information from WR

NH had access to area, unplugged and replugged HDMI cables from aida09, aida10, aida11, aida12 root to MACB (at both ends)

earlier on, Akash secured HDMI cable to aida07

11:55 AIDA powercycle and reset, DAQ ok now

      writing to file NULL/R45

12:00 CB takes over
  252   Tue Apr 20 10:54:41 2021 CBApril 20 12:00 - 16:00 shift
11:55 CA got AIDA back online

12:00 Stats OK (attach 1)
No faults found
UCESB OK - no beam atm (attach 2)
Temps OK (running hotter than yesterday, but not worrying yet) - attach 3
Bias OK (attach 4)

System wide checks
Clock - *all* pass
ADC - 06 fails
WR - *all* pass
FPGA *all* pass

Memory info
Returned 0 0 0 0 0 0 0 0 0 0 0 0  
Mem(KB) :	4	8	16	32	64	128	256	512	1k	2k	4k
aida01 :     13      4      1      3      2      1      2      4      2      3      7   : 39876
aida02 :     23      6      1      2      2      2      1      4      2      3      7   : 39772
aida03 :     24      4      4      1      2      3      1      4      2      3      7   : 39904
aida04 :     10      3      1      1      1      3      1      4      2      3      7   : 39728
aida05 :     12      2      3      2      1      4      2      3      1      4      7   : 40688
aida06 :     10      5      4      2      2      3      3      4      2      3      7   : 40400
aida07 :     22      6      1      3      1      3      1      4      2      3      7   : 39864
aida08 :     20      4      4      1      2      3      2      4      2      3      7   : 40144
aida09 :      1      2      2      3      1      2      1      3      3      3      7   : 40148
aida10 :      3      3      1      2      2      2      2      3      2      3      7   : 39412
aida11 :     25      7      0      1      0      3      2      3      2      3      7   : 39484
aida12 :     21      7      3      1      3      2      2      3      2      3      7   : 39580


Waiting for beam.

12:42 Stats OK, UCESB ok, No faults. No beam.
13:21 Stats OK, UCESB ok, No faults. No beam.
13:24 Implants in DSSD#3. ASIC check. Disappeared. No beam.

14:00 No beam.
Stats OK (attach 5)
Temps OK (attach 6)
UCESB OK
Bias  OK (attach 7)

System wide checks
Clock OK
ADC calibration 06 fails
WR OK
FPGA OK

Returned 0 0 0 0 0 0 0 0 0 0 0 0  
Mem(KB) :	4	8	16	32	64	128	256	512	1k	2k	4k
aida01 :     16      9      4      1      3      2      3      3      2      3      7   : 39848
aida02 :     15      8      2      2      2      3      2      3      2      3      7   : 39644
aida03 :      8      1      2      3      4      3      2      3      2      3      7   : 39720
aida04 :     25     12      3      1      1      3      1      4      2      3      7   : 39892
aida05 :     13      4      4      4      0      3      2      3      1      4      7   : 40596
aida06 :     21      8      3      3      0      3      3      4      2      3      7   : 40356
aida07 :     17     11      0      1      1      3      1      4      2      3      7   : 39804
aida08 :     16      8      2      3      2      3      1      4      2      3      7   : 39936
aida09 :     17      6      3      2      1      2      1      4      2      3      7   : 39716
aida10 :      3      3      2      2      4      1      2      2      3      3      7   : 39940
aida11 :      4      4      4      1      1      3      3      2      2      3      7   : 39248
aida12 :      2      4      1      3      1      2      2      3      2      3      7   : 39384

14:30 Stats OK, UCESB ok, No faults. No beam.
15:00 Stats OK, UCESB ok, No faults. No beam.


15:24 Beam is back!
Stats OK (attach 8), UCESB ok (attach 9), No faults.

15:50 Shift handed over to Liam
  547   Fri Mar 22 08:29:55 2024 TDAnydesk restarted remotely
Anydesk restarted remotely per https://elog.ph.ed.ac.uk/CARME/489

Anydesk address now restored to 832827869
  671   Tue Nov 19 11:24:25 2024 TDAnydesk restarted remotely
Anydesk restarted remotely per https://elog.ph.ed.ac.uk/CARME/489