AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC, Page 28 of 34  ELOG logo
Entry  Thu Mar 12 23:04:24 2020, LS, DK, Friday 13th 00:00 to 08:00 shift 20x
   Shift file starting R9_639

   system wide checks okay
Entry  Thu Mar 12 15:25:08 2020, Thursday, March 12 16:00 to midnight shift, CA, DK, LS 12x
16:30
    All system wide checks passed
    Temps shown in attachment #1 
Entry  Thu Mar 12 08:04:24 2020, OH, CA, TD, Thursday 12th March 08:00-16:00 14x
ASIC settings 2019Oct31-13.24.23
     slow comparator 0xa
Entry  Wed Mar 11 23:44:27 2020, OH, Thursday 12th March 00:00-08:000 10x
ASIC settings 2019Oct31-13.24.23
     slow comparator 0xa
Entry  Wed Mar 11 17:21:33 2020, CA, TD, OH, Wednesday 11 March 16.00-00.00 23x
18.25 DAQ currently running No Storage mode

ASIC settings 2019Oct31-13.24.23
Entry  Wed Mar 11 16:29:17 2020, CA, LS, TD, slow comparator threshold comparison 100keV_stats.png110keVStats.png120kevstats.png80kevDSSD2.png
17.29 good event statistics for slow comparator threshold at 100 keV (0xa) - attachment 1

17.36 good event statistics for slow comparator threshold at 110 keV (0xb) - attachment 2
Entry  Tue Mar 10 23:17:31 2020, DK, LS, PW, TD, CA, Wednesday 11 March 2020 43x
00:15 Shift change near AIDA R6_475

  We continue with spill cycles around 0.2 Hz or one beginning once each ~5 seconds.
Entry  Tue Mar 10 12:36:44 2020, TD, [How To] Monitor AIDA 
All AIDA operations are controlled from the Dell T410 Workstation aidas1 using web browsers.

AIDA has multiple, independent data acquisition servers connected by an Ethernet network.
Entry  Tue Mar 10 10:16:56 2020, OH, AIDA Implant Range Scan 
A range scan was performed using AIDA.
See the GSI elog: https://elog.gsi.de/despec/S480/19
Entry  Tue Mar 10 08:48:07 2020, TD, OH, NH, LS, CA, DK, PJW, RP, Tuesday 10 March 08.00-16.00 41x

c 09.00 aida03 drops Histogram enable at DAQ GO
        NH determined this was due to Options DB corruption and fixed
Entry  Mon Mar 9 13:50:08 2020, LS, CA, DK, 09/03/2020 system checks 13x
14:52 Attachment 1 shows low energy spectra. AIDA06 shows strange baseline, possible double-hit?

15:06 Systems checks fine.
Entry  Mon Mar 9 12:47:18 2020, NH, Report: Unusual AIDA Timestamp Error  
Noted using my AIDA event builder code, a timewarp was reported and I noticed the ADC event had a different timestamp LSB to the "WR" markers preceding
it, which I didn't think was possible in the merger as it stands
Entry  Sat Feb 29 10:56:38 2020, NH, Alpha Hit Pattern alphas_1d.pdf
Circa 700 GB of data

A few channels in DSSD 1 missing? Otherwise seems OK
Entry  Tue Feb 25 09:55:59 2020, NH, HV Scope Traces 10x
Looking at the HV bias in a scope to see if it is the source of noise especially on the p+n side

Channel 1: HV Core channel 4
Entry  Fri Feb 21 12:16:06 2020, TD, NH, Report - high - unusual FEE64 crash during startup ttyUSB4
 
Entry  Fri Feb 21 09:39:38 2020, TD, NH, Friday 21 February 7x

Attachment 1 - layout 2 - stat spectra
Attachment 2 - detector biases & leakage currents - OK
Entry  Fri Feb 21 09:21:10 2020, NH, (Early) Alpha Analysis  DSSD_1d_HP.pngDSSD_E.pngFEE3_43_Hot.pngfee8_hot.pngFee10_hot.png
Overnight alpha data

======================================
Entry  Thu Feb 20 16:35:55 2020, TD, NH, Thu 20 February 2020 12x
DSSD#3 (most downstream) was replaced

Noted that old DSSD wafer was not sitting properly - detached - in PCB, PCB flexing and/or inadequate epoxy gluing
Entry  Mon Feb 17 09:48:34 2020, NH, Report: aida09 Kernel Panic & Lost WR  ttyUSB12GSI_WR.pngGSI_WR2.pngGSI_WR4.pngGSI_WR3.png
aida09 crashed over the weekend and automatically rebooted.
After the reboot the WR timestamp sent to the merger is in the future and hence incorrect
    Reply  Wed Feb 19 10:58:29 2020, NH, PJCS, Report: aida09 Kernel Panic & Lost WR  
> aida09 crashed over the weekend and automatically rebooted.
> After the reboot the WR timestamp sent to the merger is in the future and hence incorrect
Entry  Fri Feb 14 12:50:29 2020, NH, AIDA Noise 14.02.2020 6x
Electricians changed the grounding for the DESPEC platform to a large pillar that goes deep into soil

Change in grounding was apparently successful at reducing the noise complaints in Cave C (R3B)
ELOG V3.1.4-unknown