92
|
Tue Nov 12 13:22:19 2019 |
OH, NH, CA | AIDA09 Replacement |
ASIC 1 on FEE9 was not producing signals.
A new HDMI cable was installed but still no signals.
AIDA09 was replaced with a new FEE.
Current list of FEE to serial numbers installed attachment 1
MAC address of new FEE was obtained.
Backup of DHCPD.conf was made dhcpd.confBACKUP191112
dhcpd.conf was updated with new MAC address.
All FEEs powered on and 09 was seen to mount and is seen by AIDAServer.
AIDA09 flashed to newest version of firmware. (0x18430701) - attachment 2 |
Attachment 1: FEE-SerialNumber.txt
|
FEE# Messanie Serial No. MAC Address
4 026-1690101-0016 d8-80-39-41-f6-b7
8 026-1690101-0013 d8-80-39-41-ba-2b
12 026-1690101-0030 d8-80-39-41-cf-ac
2 026-1690101-0011 d8-80-39-41-ba-22
6 026-1690101-0035 d8-80-39-41-ee-72
10 026-1690101-0032 d8-80-39-41-ee-10
1 026-1690101-0027 d8-80-39-41-ba-8a
5 026-1690101-0008 d8-80-39-41-d7-cc
9 026-1690101-0015 d8-80-39-41-f6-ee
7 Not checked d8-80-39-41-b4-0c
11 026-1690101-0006 d8-80-39-41-f6-5a
3 026-1690101-0019 d8-80-39-41-d8-21
Dead FEE Messanie Serial No. MAC Address
ASIC 1 Dead 026-1690101-0005 d8-80-39-42-0d-0c
|
|
91
|
Mon Nov 11 10:34:19 2019 |
NH | Alpha Analysis (11.11.19) |
Analysis of approx 20GB of data since Thursday night (not continuous - estimate of hours incominG)
Figures 1-3: Energy Front vs Energy Back (No F-B energy gate)
Figures 4-6: 2D hit pattern
Figures 7-9: 1D hit pattern
aida10 does not seem to be reading out still, unsure of reason? Looks fine with pulser.
Slow comparator not setting or unusual gain?
Otherwise good |
Attachment 1: DSSD1-Energy.png
|
|
Attachment 2: DSSD2-Energy.png
|
|
Attachment 3: DSSD3-Energy.png
|
|
Attachment 4: DSSD1-2DHP.png
|
|
Attachment 5: DSSD2-2DHP.png
|
|
Attachment 6: DSSD3-2DHP.png
|
|
Attachment 7: DSSD1-1DHP.png
|
|
Attachment 8: DSSD2-1DHP.png
|
|
Attachment 9: DSSD3-1DHP.png
|
|
|
90
|
Thu Nov 7 10:22:26 2019 |
Nic, Patrick | Reply: WR Timestamps |
> > > All 12 FEEs have valid WR Timestamps
> > > Had to powercycle aida09 once as before raw readout was displaying upper 12 bits of WR timestamp as 0. Unsure of other method.
> > >
> > > HDMI cables in aida09 checked and good.
> >
> > The problem would be the cable , one end or the other.
> > I think ( if I recall ) a setup would restart the WR decoder.
> >
> > I notice you have set the WR info word rate to be quite high , 6123/sec typ, is this intentional ?
>
> Cable will be replaced once a spare is available.
>
> Setup did not restart the WR decoder when this problem occurred beforehand - Reset/Setup tried.
>
> WR rate was chosen by Vic I believe, I am unsure of reasoning myself.
An update/clarification:
Although the upper bits are zero I believe actually it is a total failure to synchronise to WR:
aida09
WR Time Item 0x80500000 0x0fbd8000; Time (48:63)=0x0; Time (28:47)=0x249; Time (0:27)=0x0fbd8000
WR Time Item 0x80400249 0x0fbd8000; Time (28:47)=0x249; Time (0:27)=0x0fbd8000
WR Time Item 0x80500000 0x0fbdc000; Time (48:63)=0x0; Time (28:47)=0x249; Time (0:27)=0x0fbdc000
WR Time Item 0x80400249 0x0fbdc000; Time (28:47)=0x249; Time (0:27)=0x0fbdc000
WR Time Item 0x80500000 0x0fbe0000; Time (48:63)=0x0; Time (28:47)=0x249; Time (0:27)=0x0fbe0000
WR Time Item 0x80400249 0x0fbe0000; Time (28:47)=0x249; Time (0:27)=0x0fbe0000
aida10
WR Time Item 0x8050022e 0x0bde8000; Time (48:63)=0x22e; Time (28:47)=0xe29d5; Time (0:27)=0x0bde8000
WR Time Item 0x804e29d5 0x0bde8000; Time (28:47)=0xe29d5; Time (0:27)=0x0bde8000
WR Time Item 0x8050022e 0x0bdec000; Time (48:63)=0x22e; Time (28:47)=0xe29d5; Time (0:27)=0x0bdec000
WR Time Item 0x804e29d5 0x0bdec000; Time (28:47)=0xe29d5; Time (0:27)=0x0bdec000
WR Time Item 0x8050022e 0x0bdf0000; Time (48:63)=0x22e; Time (28:47)=0xe29d5; Time (0:27)=0x0bdf0000
WR Time Item 0x804e29d5 0x0bdf0000; Time (28:47)=0xe29d5; Time (0:27)=0x0bdf0000 |
|
89
|
Tue Nov 5 11:15:41 2019 |
NH PJCS | Report: aida10 database corruption |
If ypou now check the actual values in the aida10 Options screen are the correct ?
It can happen that corruptions once loaded are propagated when Options are generally saved.
Looking at the dates the files were modified does that make sense to you. It doesn't look as if the corrupted database was changed except when they all were ? Can you check ? Have you checked the other Options files ? I seem to recall that RIKEN AIDA had a program to to this ?
Quote: |
Hi, see the attached files in the original comment. I think the BACKUPCorrupt is quite old now.
The Options data gets corrupted and for example the ASIC folder gets set to 'undefined' which I think is why the ASIC data loads incorrectly.
Quote: |
This looks like the Options not the ASIC data. That will be in the CONTENTS file of the FEE64 named directory within the dated EXPERIMENT directory. Looks like the latest is at /MIDAS/DB/EXPERIMENTS/AIDA/2019Oct31-13.24.23/aida10 ?
A quick look at this file and it looks normal , not all 0xad ?
Looked at the Options file in the /MIDAS/DB/EXPERIMENTS/AIDA/Options.BACKUPCorrupt/aida10 and the worst bit is the name of the Data Acquistion program. That could welll cause you problems with readout as it won't understand some of the new data.
Quote: |
Often aida10 database seems to get corrupted (ASIC values are all 0xad)
Unsure why it's only aida10, might be related to weird alpha rate behaviour at the moment.
Attached is corrupted database and correct database
|
|
|
|
|
88
|
Tue Nov 5 10:32:21 2019 |
NH PJCS | Report: aida10 database corruption |
Hi, see the attached files in the original comment. I think the BACKUPCorrupt is quite old now.
The Options data gets corrupted and for example the ASIC folder gets set to 'undefined' which I think is why the ASIC data loads incorrectly.
Quote: |
This looks like the Options not the ASIC data. That will be in the CONTENTS file of the FEE64 named directory within the dated EXPERIMENT directory. Looks like the latest is at /MIDAS/DB/EXPERIMENTS/AIDA/2019Oct31-13.24.23/aida10 ?
A quick look at this file and it looks normal , not all 0xad ?
Looked at the Options file in the /MIDAS/DB/EXPERIMENTS/AIDA/Options.BACKUPCorrupt/aida10 and the worst bit is the name of the Data Acquistion program. That could welll cause you problems with readout as it won't understand some of the new data.
Quote: |
Often aida10 database seems to get corrupted (ASIC values are all 0xad)
Unsure why it's only aida10, might be related to weird alpha rate behaviour at the moment.
Attached is corrupted database and correct database
|
|
|
|
87
|
Tue Nov 5 10:02:45 2019 |
NH PJCS | Report: aida10 database corruption |
This looks like the Options not the ASIC data. That will be in the CONTENTS file of the FEE64 named directory within the dated EXPERIMENT directory. Looks like the latest is at /MIDAS/DB/EXPERIMENTS/AIDA/2019Oct31-13.24.23/aida10 ?
A quick look at this file and it looks normal , not all 0xad ?
Looked at the Options file in the /MIDAS/DB/EXPERIMENTS/AIDA/Options.BACKUPCorrupt/aida10 and the worst bit is the name of the Data Acquistion program. That could well cause you problems with readout as it won't understand some of the new data.
Hope this helps.
Quote: |
Often aida10 database seems to get corrupted (ASIC values are all 0xad)
Unsure why it's only aida10, might be related to weird alpha rate behaviour at the moment.
Attached is corrupted database and correct database
|
|
|
86
|
Mon Nov 4 13:17:12 2019 |
NH | Report: aida10 database corruption |
Often aida10 database seems to get corrupted (ASIC values are all 0xad)
Unsure why it's only aida10, might be related to weird alpha rate behaviour at the moment.
Attached is corrupted database and correct database |
Attachment 1: db_working.txt
|
Index string Stat.offset&&Aida_Hist_W_Enable&&TS_SYNC_PHASE&&ExtClk&&Aida.shift&&MACB_TRIG_MODE&&Aida.offset&&RunNumber&&Aida_GroupBase&&Rate.channels&&Stat.channels&&Aida.Vchannels&&ASIC.settings&&Aida_Hist_D_Enable&&WAVE_DMA_HWM&&Stat.shift&&Aida.Wchannels&&Aida_Hist_V_Enable&&Include.Aida&&Aida.channels&&DataAcqPgm&&Aida_Hist_H_Enable&&DataFormat&&ASIC_DMA_HWM&&Aida_Hist_L_Enable
Stat.offset string 64
Aida_Hist_W_Enable string 1
Aida.shift string 0
TS_SYNC_PHASE string 0x0
ExtClk string 1
0x0006dead string 0x0000
Aida.offset string 0
MACB_TRIG_MODE string 4
Aida_GroupBase string 1
Stat.channels string 512
RunNumber string 12
Rate.channels string 512
WAVE_DMA_HWM string 0x0007ffff
Aida.Vchannels string 256
ASIC.settings string 2019Oct31-13.24.23
Aida_Hist_D_Enable string 1
Stat.shift string 6
Aida.Wchannels string 1020
Aida.channels string 65536
Aida_Hist_V_Enable string 0
Include.Aida string 0
Aida_Hist_H_Enable string 1
DataAcqPgm string AidaExecV9
DataFormat string 0x0000
ASIC_DMA_HWM string 0x000fffff
Aida_Hist_L_Enable string 1
|
Attachment 2: db_corrupt.txt
|
Index string Stat.offset&&Aida_Hist_W_Enable&&TS_SYNC_PHASE&&ExtClk&&Aida.shift&&MACB_TRIG_MODE&&Aida.offset&&0x0006dead&&Aida_GroupBase&&Rate.channels&&Stat.channels&&Aida_Hist_D_Enable&&Aida.Vchannels&&ASIC.settings&&WAVE_DMA_HWM&&Stat.shift&&Aida.Wchannels&&Aida_Hist_V_Enable&&Include.Aida&&Aida.channels&&Aida_Hist_H_Enable&&DataAcqPgm&&DataFormat&&ASIC_DMA_HWM&&Aida_Hist_L_Enable
Stat.offset string 64
Aida_Hist_W_Enable string 0
Aida.shift string 1
TS_SYNC_PHASE string 1
ExtClk string 1
Aida.offset string 0
MACB_TRIG_MODE string 14
0x0006dead string 0x0000
Stat.channels string 150
Aida_GroupBase string 1
Rate.channels string 150
RunNumber string 2
Aida_Hist_D_Enable string 1
ASIC.settings string undefined
WAVE_DMA_HWM string 0x0007ffff
Aida.Vchannels string 256
Stat.shift string 6
Aida.Wchannels string 1020
Include.Aida string 1
Aida.channels string 32768
Aida_Hist_V_Enable string 0
Aida_Hist_H_Enable string 0
DataFormat string AIDA
DataAcqPgm string AidaExecV9
ASIC_DMA_HWM string 0x000fffff
Aida_Hist_L_Enable string 1
|
|
85
|
Mon Nov 4 08:48:18 2019 |
NH | Alpha Status 04.11.2019 |
Alpha run has been running over the weekend. Stats looking positive with notable exception that aida10 is not sending much data. Will attempt a powercycle.
Temperatures and Leakage Currents very good to excellent
Noted that the slow rate of aida10 seems to be slowing down the rate data is sent to tape - perhaps being buffered until an ADC event arrives?
Update 05.11.2019
Most rates OK, aida10 only recorded 1 event over night, aida09 spectra had a number of weird channels.
Performed ASIC check/load, aida10 found another 201 events but mostly in the HEC channel (as did all FEEs)
All other FEEs look very nice however
15:44 CET: Check/load performed as aida07 rate went very high (in HEC it seems) - now back to 0
Update 07.11.2019
Run stopped to move DESPEC platform, overnight the USB relay disconnected/reconnected powering down the FEEs (probably someone knocked the interlock wire)
|
Attachment 1: alphas-bias.png
|
|
Attachment 2: alphas-temp.png
|
|
Attachment 3: alpha-aida10.png
|
|
Attachment 4: stats-0511.png
|
|
Attachment 5: spec-0511.png
|
|
|
84
|
Fri Nov 1 18:09:03 2019 |
CA, TD, NH | Summary - 29.10-1.11.19 |
> >
> >
> > - evidence of c. 1MHz extrinsic noise for most FEE64s
>
>
> Is the 1MHz noise actually 1.58Mhz? as this is the frequency of observed noise at LYCCA.
Judge for yourself
https://elog.ph.ed.ac.uk/DESPEC/191031_125102/1350_18W.png
I would estimate c. 2.5 cycles in 2us => c. 1.2MHz … ?
Note that all of the waveforms are shown on an expanded scale c. 7000-9000 of 0-16383
so the amplitude is significantly less than that observed at LYCCA.
Tom |
|
83
|
Fri Nov 1 15:45:04 2019 |
CA, TD, NH | Summary - 29.10-1.11.19 |
>
>
> - evidence of c. 1MHz extrinsic noise for most FEE64s
Is the 1MHz noise actually 1.58Mhz? as this is the frequency of observed noise at LYCCA. |
|
82
|
Fri Nov 1 15:24:40 2019 |
CA, TD, NH | Summary - 29.10-1.11.19 |
3x MSL type BB18(DS)-1000 installed
detector bias -160V, leakage current c. 1uA @ +21 deg C for all 3x DSSSDs
all FEE64 good event rates (slow comparator 0xa, LEC fast comparator 0xff) c. 120k, or less, typically 30-50k, overall merge rate c. 1.6M data item/s
See https://elog.ph.ed.ac.uk/DESPEC/70
Outstanding issues
- occasional loss of bits 48-63 of WR timestamp for aida09 - can be recovered by power cycle - replace HDMI cable?
- aida09 asic 1 not producing ADC or disc data - replace aida09?
ASIC Check works OK
aida09 asic temp c. 30deg C < other asic temps
- evidence of c. 1MHz extrinsic noise for most FEE64s
- acquisition of waveform data not robust - most ASIC channels do not produce data or quickly stop producing data - PJCS to review firmware rev for Jan/Feb 2020
- require spare DSSSDs, FEE64s, HDMI cables etc |
|
81
|
Fri Nov 1 14:17:15 2019 |
Nic, Patrick | Reply: WR Timestamps |
> > All 12 FEEs have valid WR Timestamps
> > Had to powercycle aida09 once as before raw readout was displaying upper 12 bits of WR timestamp as 0. Unsure of other method.
> >
> > HDMI cables in aida09 checked and good.
>
> The problem would be the cable , one end or the other.
> I think ( if I recall ) a setup would restart the WR decoder.
>
> I notice you have set the WR info word rate to be quite high , 6123/sec typ, is this intentional ?
Cable will be replaced once a spare is available.
Setup did not restart the WR decoder when this problem occurred beforehand - Reset/Setup tried.
WR rate was chosen by Vic I believe, I am unsure of reasoning myself. |
|
80
|
Fri Nov 1 10:46:13 2019 |
CA, TD, NH | Friday 1st November 2019 |
11.46 - attachments 1,2,3: bias/leakage currents, good event statistics and fee temperatures (respectively)
before removing aluminium foil upstream of AIDA
note - slow comparator threshold 0x64 (alpha background), LEC fast comparator threshold 0xff, pulser OFF
11.52 - attachments 4,5,6: bias/leakage currents, good event statistics and fee temperatures (respectively)
after removing aluminium foil upstream of AIDA
- base of AIDA snout has Mylar shielding |
Attachment 1: biaswfoil.png
|
|
Attachment 2: statswfoil.png
|
|
Attachment 3: tempwfoil.png
|
|
Attachment 4: biaswofoil.png
|
|
Attachment 5: statswofoil.png
|
|
Attachment 6: tempswofoil.png
|
|
|
79
|
Fri Nov 1 09:30:17 2019 |
NH | Database error stopping run? |
Strange error about Database options appearing during run stop.
Does not seem to affect stopping/starting/data? |
Attachment 1: ErrorStopDatabase.png
|
|
|
78
|
Thu Oct 31 19:02:42 2019 |
Patrick | WR Timestamps |
> All 12 FEEs have valid WR Timestamps
> Had to powercycle aida09 once as before raw readout was displaying upper 12 bits of WR timestamp as 0. Unsure of other method.
>
> HDMI cables in aida09 checked and good.
The problem would be the cable , one end or the other.
I think ( if I recall ) a setup would restart the WR decoder.
I notice you have set the WR info word rate to be quite high , 6123/sec typ, is this intentional ? |
|
77
|
Thu Oct 31 17:13:54 2019 |
NH | Alpha Run |
Beginning Alpha run
Thresholds set to 100 (LEC/Slow) 255 (LEC/Fast) 2 (HEC)
Pulser is OFF
Recording to MIDAS: /TapeData/31Oct19/R1
Recording to MBS: /lustre/gamma/nhubbard/AIDA/Alphas311019_
Disk rate approximately 0 as expected.
01.11.19 19:11 CET: FEE10 Stat histogram was empty - performed a check/load and now rare events come in - Merger also emitted 10 blocks to storage quickly (presumably waiting on FEE10 data)
All FEEs now slowly showing alpha data |
|
76
|
Thu Oct 31 16:43:03 2019 |
NH CA TD | Merger & MBS Performance |
Testing merger with 3 DSSDs connected, with thresholds of 10 (slow comparator), 2 (HEC) and 255 (fast discriminator)
Merger handling rate of 1.6 million events per second comfortably. Forwarding to MBS fine.
Network usage: 130 Mbps (AIDA network) and 20 Mbps (MBS network) |
Attachment 1: MergerStats.png
|
|
Attachment 2: MBSRates.png
|
|
Attachment 3: Mergernetworking.png
|
|
Attachment 4: MergerStatistics.png
|
|
|
75
|
Thu Oct 31 15:24:35 2019 |
TD | waveform spectra issues? |
|
Attachment 1: 20.png
|
|
Attachment 2: 21.png
|
|
Attachment 3: 22.png
|
|
Attachment 4: 23.png
|
|
Attachment 5: 30.png
|
|
Attachment 6: 31.png
|
|
Attachment 7: 32.png
|
|
Attachment 8: 33.png
|
|
Attachment 9: 40.png
|
|
Attachment 10: 41.png
|
|
Attachment 11: 42.png
|
|
Attachment 12: 43.png
|
|
|
74
|
Thu Oct 31 14:27:06 2019 |
NH | WR Timestamps |
All 12 FEEs have valid WR Timestamps
Had to powercycle aida09 once as before raw readout was displaying upper 12 bits of WR timestamp as 0. Unsure of other method.
HDMI cables in aida09 checked and good. |
Attachment 1: WRTimes.png
|
|
|
73
|
Thu Oct 31 14:23:37 2019 |
TD | aida09 asic 1 no data readout? |
|
Attachment 1: 10.png
|
|
Attachment 2: 11.png
|
|
Attachment 3: 12.png
|
|
Attachment 4: 13.png
|
|